Piezoelectric System. Ultrasonichandpiece with cutting tips for abroad range of applications.

QuickReferenceChart

28923-J10400A_28923-J10400A 3/9/11 10:42 AM Page 1

Page 2: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Intended useThe Piezoelectric System distributed by Synthes is an ultrasonic surgicalsystem consisting of handpieces and associated tips for cutting boneand bone substitutes. It can be used for osteotomy, osteoplasty,decorticating, drilling, shaping, and smoothing of bones and teeth in a variety of surgical procedures, including general orthopaedic,otolaryngological, maxillofacial, oral, hand, foot, neurosurgical, spine, and plastic/reconstructive surgery.

Features and benefits

Cruise Control™ System– No loss of power regardless of the environment and/or treatment

performed

– Active only on mineralized tissue, with minimal risk to soft tissue

– Minimum pressure required, resulting in higher precision and lesshand fatigue for the clinician

– Four power setting modes: three for bone cutting based on bonedensity and one for soft tissue detachment

LED handpiece– Up to 60 watts of power for efficient

osteotomies

– No heat generation and vibration free,minimizes the risk of necrosis andsurgeon discomfort

– Six powerful LEDs for enhanced visionat the operative site

– Attached cable

Piezoelectric System Synthes 1

28923-J10400A_28923-J10400A 3/9/11 10:42 AM Page 1

Page 3: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Progressive, multifunction foot pedal

– Real-time adjustment of the ultrasonic power according to the anatomical constraints encountered

– The power of the ultrasonics can be precisely adapted for areas wherenerves, arteries or membranes are present

– Total control of the console through the foot pedal, which allowshands-free operation in the sterile field, without touching the consolescreen

Cutting tips– Irrigation is transmitted through the tip to prevent heat generation

and necrosis

– Narrow kerf for highly precise bone surgery procedures

– Broad range of applications

Console– User-friendly touch-sensitive LCD screen for ease of operation

– Cable connectors for up to two handpieces, eliminating the need tochange cutting tips during procedures

– Two peristaltic pump housings

– Two brackets for irrigation solution

Piezoelectric System Synthes 2

28923-J10400A_28923-J10400A 3/9/11 10:42 AM Page 2

Page 4: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Technical Descriptiona) LCD touch screen The LCD touch screen is used to definethe settings of the Piezoelectric System.The Piezoelectric System settings areadjusted by applying reasonablepressure to the screen keys.

Identification functions

Reduce or increase the irrigation flow rate in ml/min

Flush or prime irrigation

ON/OFF irrigation or

Select a program: D1, D2, D3, D4 D1 is the most powerful setting for dense bone D4 is for detachment of soft tissue

Reduce or increase power within a program

Handpiece light ON/OFF

Progressive or full ON/OFF power

Save settings

b) Control unit back panelThe power receptacle with its grounding pin is used to connect theconsole to the power supply using a plug-in power cord.The foot pedal connector is used to connect the console to themultifunction control foot pedal.The fan , protected by a metal grille, keeps the Piezoelectric System at its optimum performance level.The irrigation pole holders .The power switch is used to switch the console ON or OFF. Potentiated equipment connector plug .

3

4

6

1

2

5

Piezoelectric System Synthes 3

4 42 3 6 1 5

28923-J10400A_28923-J10400A 3/9/11 10:42 AM Page 3

Page 5: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

c) Irrigation pumpsThe irrigation pumps are located on either side of the console andaccommodate Synthes irrigation tubing with cassettes. The irrigationcassettes are installed by lifting the pump covers and inserting thecassettes horizontally into the cassette housings.

d) Front panel The two cable connectors acceptthe connector of the PiezoelectricSystem handpiece cable.

e) Control foot pedalThe functions on the control foot pedal allow the user to work in asterile environment. Once the settings have been adjusted on theconsole, the user does not need to touch the LCD screen.

Foot pedal button definition

Handpiece control (ON/OFF or progressive)Flush/prime irrigationChange program (D1– D4)Select left or right handpiece Irrigation ON/OFF

6

7

1

2

3

4

5

6

7

5

4

3

2

1

Piezoelectric System Synthes 4

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 4

Page 6: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Caution: United States Federal Law restricts the use of this devicesolely to qualified, trained and competent health practitioners orpersonnel under their supervision.

Device usersAttention:– The Synthes Piezoelectric System must not be used if the patient

and/or the operator has a cardiac stimulator (pacemaker) or any otheractive implant (e.g. a cochlear implant).

– Use of the Synthes Piezoelectric System is restricted solely to qualified,trained and competent health care practitioners in the normal contextof their work.

– If you have received this device in error, please contact your SynthesConsultant.

Interactions– The device complies with applicable electromagnetic compatibility

standards. However, the user should ensure that any potentialelectromagnetic interference does not cause an additional risk(presence of radiofrequency emitters, electronic devices, etc.).

– The device is not designed to withstand shocks delivered by anelectric defibrillator.

– Do not attempt to connect to the Piezoelectric System connectors anyaccessories other than those supplied by Synthes.

Caution: The tips are specifically designed for the Synthes PiezoelectricSystem. The system is not compatible with other systems’ tips.

Piezoelectric System Synthes 5

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 5

Page 7: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

To reduce the risk of accidents, the precautions stipulated below mustbe taken.

Electrical connectionTo avoid risk of electric shock, this device must be connected to anelectrical power supply with a protective ground.

– The electric supply to which the device is connected must comply withthe standards in force in your country.

– If loss of electrical power during use is likely to generate an unacceptable risk, the device must be connected to a suitablepower source (e.g. UPS).

Using the device– Do not use the device if it appears to be damaged or faulty.

– Turn the device OFF before unplugging the power cord.

– To unplug the power cord, grip the cord plug and hold the wallsocket.

– Never use any irrigation solution containers other than those intendedfor suspension from the supplied brackets.

– The device must only be used with bottles or bags of physiologicalsaline or sterile water.

– The capacity of the irrigation solution containers used must not exceed one liter per bracket.

– When the device is not to be used for a long period of time, unplugthe console from the electric supply.

– Do not exert excessive force on the screen.

– Do not move the console during use.

Piezoelectric System Synthes 6

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 6

Page 8: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Overheating, burnsTo avoid any risk of burns or overheating, irrigation must always beswitched ON and active when the handpiece is in use.

The user shall regularly control the level of irrigation to ensure its coolantaction to the tip. A lack of irrigation presents a hazard to the patient.

Environment– Do not cover the console or obstruct the ventilation vents.

– Do not immerse the device in liquid, and do not use it outdoors.

– Do not tilt the console at an angle greater than 5°.

– Do not place the console near a heat source.

– Ensure that the cables and power cord are not in a traffic path.

– The device must be stored in an appropriate and safe place.

– The device is not designed for use in the presence of anaestheticgases or any other flammable gas.

– Do not expose the console to water vapor, or splashes.

– Condensation inside an electrical device may be dangerous.

– If the console must be moved from a cool place to a hot place, allow it to warm up to the new room temperature before use.

– The device is not designed to work near ionizing radiation.

– Do not insert metal objects into the device (risk of electric shock,short-circuit or emission of hazardous substances).

Piezoelectric System Synthes 7

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 7

Page 9: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Maintenance– Before and after each procedure, ensure that all accessories are

cleaned and sterilized.

Accessories– The device and accessories have been designed and developed

together to ensure maximum safety and performance.

– Use of accessories from other manufacturers is a potential hazard tothe user, the patient and the device.

RepairsDo not attempt to repair or to modify the device without authorizationby Synthes. When the device is modified or repaired, specific checks andverifications must be performed to ensure that the device can be usedsafely.

In the case of a fault, contact your Synthes Consultant for repair. Yoursystem is warranted for defects for one year from the date of purchase.Use of unauthorized repairers could damage the system and void thewarranty. If in doubt, contact your Synthes Consultant.

Piezoelectric System Synthes 8

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 8

Page 10: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Installation

Important: Do not place the Piezoelectric System close to or on top of another device.Do not place the power cord and the foot pedal cable in a wire cover orin a cable guard.Do not operate the handpiece unless the irrigation is switched to the ONposition.Do not disconnect the handpiece cables when the device is switched ONand the foot pedal is pressed.Do not insert or remove cutting tips when the handpiece is ON.The handpiece, wrench and graphic case must be cleaned and sterilizedbefore each use. Do not use the system with parts other than those supplied by Synthes.Check the integrity of the device and its accessories before and aftereach use in order to detect any problems.

– Position the console on a fixed horizontal surface, or one with a slopeof no more than 5°.

– Check that the power switch is in the OFF position.

– Connect the power cord to the receptacle on the back of the console.If necessary, connect the potential equalization connector to theconsole.

– Connect the power cord to a grounded power outlet.

– Connect the foot pedal cable to the foot pedal connector on the backof the console.

– Position the foot pedal so that it is easily accessible for use.

– Insert the irrigation poles into the bracket holders.

– Connect the handpiece cable to the connector on the front of theconsole. The default setting is for the right connector.

– Ensure that the console is close enough to the work area to be usedwithout pulling on the cables or power cord.

– Place the handpiece on the sterile field.

Piezoelectric System Synthes 9

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 9

Page 11: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Installation (cont’d.)

The pump housings are designed toaccommodate Synthes irrigationtubing with cassettes.

– Lift the pump cover and insert theirrigation tubing cassettehorizontally into the cassettehousings.

– Insert the sterile spike into theconnection on the irrigation bag,taking care to not touch the spike.

– Connect the irrigation tubing tothe irrigation port on thehandpiece. Use the sterile clipsprovided to attach the tubing tothe handpiece cable.

– Suspend the irrigation bottles or bags from the irrigation poles.

– Adjust the position of the console to suit your angle of view.

Handpiece function– Thread the selected cutting tip

onto the handpiece. Use thewrench to gently tighten the cuttingtip to the handpiece. Turn thewrench clockwise to tighten. Becareful not to overtighten the tip,which could damage the threadson the tip or the handpiece. Toremove the tip, turn the wrench counterclockwise.

– Prime the irrigation circuit by pressing the flush/prime button on theconsole screen or the multifunction foot pedal.

– When the irrigation solution reaches the end of the tip, release thefoot pedal.

– Use the Piezoelectric System in accordance with good medicalpractices.

Piezoelectric System Synthes 10

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 10

Page 12: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Startup

Important: After turning the console off, wait four seconds beforerestarting it.

By default, when the device is switched on, the right-hand handpiece isselected if two handpieces are connected. If only one handpiece is connected, the device selects the correct side. If no handpiece is connected, the device selects the right side.

SettingsThe settings are saved when each program is configured and areretrieved whenever that program is selected.

Irrigation functionThe Piezoelectric System is not designed to administer drugs. It must beused only with bottles or bags of physiological saline or sterile waterwith an individual capacity not exceeding one liter.

The Piezoelectric System must only be used with Synthes suppliedirrigation tubing that is designed for the system.

a) Flush/primeThe flush/prime function is located on the LCD screen and on the footpedal.

Press button or the foot pedal button to switch on the flush/primefunction.

The flush/prime function stays on as long as the button is pressed.

The flush can be switched on while the device is being used.

Piezoelectric System Synthes 11

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 11

Page 13: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

21

b) Flow rate adjustmentThe irrigation flow rate is adjusted using buttons or .

Each time the corresponding button is pressed, the flow rate is adjusted.Press and hold buttons or to scroll the flow rate values.The adjusted value is displayed on the LCD screen and stored in thecurrent program.The flow rate can be adjusted while the device is being used.

c) Irrigation ON/OFF

Important: To avoid any risk of burns or overheating, irrigationmust always be switched ON and active when the handpiece is in use.

The irrigation is switched ON or OFF by pressing button or the footpedal button.

When irrigation is selected, the symbol is shown.

When irrigation is not selected, the symbol is shown. Irrigationmust always be switched ON when the handpiece is in use. Damage tothe tip or necrosis may result if irrigation is not used.

Light function

The light function is displayed on the LCD screen.

Pressing button switches the light function available for thePiezoelectric system functions ON or OFF.

The light function has a default light-off time-out of nine seconds.

The light function switches OFF when the time-out has elapsed.

If necessary, you can change thelight time-out by changing thesettings in the Toolbox section ofthe Console screen option.

1

Piezoelectric System Synthes 12

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 12

Page 14: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Piezoelectric System function

Important: The Piezoelectric System function settings can becustomized by the user. The Piezoelectric System function settings canbe adjusted whether or not a handpiece is connected to the console.

The program giving the highest power is D1.

ProgramPower Level Mode Main functions

Very powerful D1 Osteotomy, osteoplasty

Powerful D2 Osteotomy, osteoplasty

Medium D3 Osteotomy, osteoplasty

Low D4 Soft tissue detachment

Programs correspond to the following level of powers:

Program Mode D1

Level of Fine Tuning 1 2 3 4 5

Power value 82% 86% 90% 95% 100%

Frequency Modulation 60 Hz

Program Mode D2

Level of Fine Tuning 1 2 3 4 5

Power value 64% 68% 72% 74% 78%

Frequency Modulation 60 Hz

Program Mode D3

Level of Fine Tuning 1 2 3 4 5

Power value 44% 48% 52% 56% 60%

Frequency Modulation 60 Hz

Program Mode D4

Level of Fine Tuning 1 2 3 4 5

Power value 41% 46% 51% 56% 60%

Frequency Modulation 30 Hz

Piezoelectric System Synthes 13

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 13

Page 15: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

The Fine Tuning gives the possibility to adjust the level of powerenergy within each mode from 1 to 5.

Select the left or right Piezoelectric System function by pressing Synthesbutton or the foot pedal button.

Important: In progressive mode, for programs D1 to D3, thedelivered power ranges between power level 1 of program D3 and theconfiguration defined by the user (program and power level). Inprogressive mode, the power delivered by program D4 ranges onlybetween levels 1 and 5, according to the value defined by the user.

Select handpiece light ON/OFF, by pressing button .

Confirm the new settings by pressing button .

The saving of the various settings is confirmed by an audible signal anda visual signal (button flashing).

Repeat this sequence if necessary in order to adjust all the programs.

2

Piezoelectric System Synthes 14

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 14

Page 16: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Bone Density and Power Settings

Very high bone densityand thick cortical bone.

High bone density andthick cortical bone.

Intermediate bonedensity, thin corticalbone and densespongy bone.

Weak bone density, fineor even absent corticalbone and spongy bone.

Piezoelectric System Synthes 15

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 15

Page 17: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Toolbox

Important: Piezoelectric System has a Toolbox function used tomake various adjustments, including brightness, light off/time-out, audiovolume and factory configuration reset.

To access the Toolbox function, switch the device OFF, wait four seconds,switch device ON, then press button on the welcome screen within 2 seconds.

a) Audio volume adjustmentAdjust audio volume valueusing buttons or . Audiovolume is adjustable from 0% to100%. The relative value of theaudio volume is displayed.

b) Screen brightnessadjustmentAdjust the brightness valueusing buttons or .Brightness is adjustable from 30% to 100%. The brightnesslevel setting is displayed.

c) Light off/time-outadjustmentAdjust the time-out length value

by using buttons or .The length of the time-outfunction can be adjusted from nine (9) to eighteen (18) seconds.

d) Storing settingsTo store the modified settings, press button . The welcome screen willreappear.

e) Factory configuration resetPress button to restore the factory configuration settings. Thefactory configuration for the Piezoelectric System functions becomeseffective when button is pressed. The welcome screen will reappear.

Piezoelectric System Synthes 16

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 16

Page 18: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

f) Software versionThe version of the software is displayed at the bottom of the Toolboxscreen.

The Piezoelectric System is configured in the factory with the followingsettings:

Piezoelectric mode

SafetyThe console is equipped to detect malfunctions of the device. If aninternal operation error occurs, the transmission fault symbol isdisplayed in the error zone and the device beeps four times. To cancelthis default, simply switch OFF the device from the ON/OFF button. Waita few seconds and switch ON again to restart the device.

When the handpiece is disconnected, the handpiece missing icon isdisplayed in the error zone and the device beeps four times.

Programmode

Powerlevel Irrigation Main

functions

Very powerful D1 3 60 ml/min.

Osteotomy, osteoplastyPowerful D2 3 60 ml/min.

Medium D3 3 60 ml/min.

Low D4 3 60 ml/min. Soft tissue detachment

Piezoelectric System Synthes 17

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 17

Page 19: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Shutting down the deviceUpon completion of the medical procedure:– Remove the bottles or bags of irrigation from the brackets.

– Remove the irrigation line spike from the bottle or bag.

– Immerse the irrigation line spike in a vessel containing distilled orsterile water.

Important: – Rinse to remove saline solution from the handpiece by running the

prime/flush function until the vessel and the irrigation line arecompletely empty.

– Remove the irrigation tubing clips.

– Disconnect the irrigation tubing from the handpiece and dispose in a safety container for soiled medical items.

– Disconnect the handpiece from the console.

– Disconnect the ultrasonic single-use tips and dispose in a safetycontainer for soiled, sharp, medical instruments.

– Switch the device OFF (O).

Piezoelectric System Synthes 18

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 18

Page 20: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Cutting TipsThe cutting tips are divided into four categories:– Cutting saw tips

– Cutting diamond tips

– Cutting scalpel tips

– Decorticating tips

These cutting tips can be used for osteotomy, osteoplasty, decorticating,drilling, shaping, and smoothing of bones and teeth in a variety ofsurgical procedures, including general orthopaedic, otolaryngological,maxillofacial, oral, hand, foot, neurosurgical, spine, and plastic/reconstructive surgery.

Caution: The tips are specifically designed for the Synthes PiezoelectricSystem. The system is not compatible with other manufacturers’ tips.

Principles of utilization

Important: An efficient use of these ultrasonic tips requires veryslight pressure to be effective.

The power of the Piezoelectric System applied to each cutting tip shouldbe set on the console screen according to the recommended mode, finetuning and irrigation level. Refer to the User’s Manual for recommendedpower settings.

To avoid any risk of burns or overheating, irrigation must always beswitched ON and active when the handpiece is in use. The user shallregularly control the level of irrigation to ensure its coolant action to thetip. A lack of irrigation represents a hazard to the patient. The irrigationlevel is given as a minimum and must be increased when necessary.

Piezoelectric System Synthes 19

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 19

Page 21: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Duration of use Cutting tips are provided sterile, for single use only, and can be usedcontinuously, as long as the maximum power and minimum irrigationsettings are strictly followed.

Cutting saw and diamond tips The intended uses of the tips are osteotomy and osteoplasty.

Use a “brush-stroke” movement:– Place the tip perpendicular to the surface

to be treated

– Move the tip with a slow back and forthmovement, in line with the cut

– Apply only very slight pressure — never use force

NEVER USE THE TIP AS A LEVER: use only the appropriate elevator to lift the bone graft.

Cutting scalpel tipsThe intended uses of this tip category are osteotomy, osteoplasty andbone harvesting.Use a repeated, short pulling movement – Apply slight pressure

– Never use force

– Never use the tip as a lever

Decorticating tipsThe intended use of this tip category is soft tissue decortication.

During its use, it is essential to keep contact with the soft tissue and the bone.– Apply slight pressure

– Never use force

– Never use the tip as a lever

1

2

Piezoelectric System Synthes 20

90°

1

2

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 20

Page 22: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

03.000.401S – Saw

Description: Saw, 20.9 x 14.1 x 4.0 x 0.6 mm

Ultra-sharp saw, with four sharpened teeth for in-depth cutting ofdense cortical bone.

Treatment examples: Le Fort osteotomies, SARPE, craniotomies, bonegraft harvest, sagittal split osteotomies, reconstruction, arthroplasty, andcraniosynostosis.

The laser marks at 3, 6, and 9 mm from its extremity,facilitate the evaluation of bone defect dimensionand control cutting depth.

Piezoelectric System Synthes 21

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

60

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 21

Page 23: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

03.000.402S – Saw

Description: Saw, 20.1 x 21.4 x 4.0 x 0.6 mm

Ultra-sharp saw, with four sharpened teeth for in-depth cutting ofdense cortical bone.

Treatment examples: Le Fort osteotomies, SARPE, craniotomies,posterior wall of the sinus and sagittal split osteotomies, or nervedecompression.

The laser marks, at 3, 6, 9, 12, and 15 mm from itsextremity, facilitate the evaluation of bone defectdimension and control cutting depth. This saw isparticularly effective for thick cortical bone.

03.000.403S – Saw

Description: Saw, left, 17.8 x 15.2 x 4.0 x 0.6 mm

Left angled bone saw, with four sharpened teeth for in-depth cuttingof dense cortical bone.

Treatment examples: Le Fort osteotomies, SARPE, craniotomies, posteriorwall of the sinus and sagittal split osteotomies, or nerve decompression.

The laser marks, at 3 and 6 mm from its extremity, facilitate theevaluation of bone defect dimension and control cutting depth. Thissaw is particularly effective for thick cortical bone.

Piezoelectric System Synthes 22

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

60

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

60

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 22

Page 24: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

03.000.404S – Saw

Description: Saw, right, 17.8 x 15.2 x 4.0 x 0.6 mm

Right angled bone saw, with four sharpened teeth for in-depth cuttingof dense cortical bone.

Treatment examples: Le Fort osteotomies, SARPE, craniotomies,posterior wall of the sinus and sagittal split osteotomies or nervedecompression.

The laser marks, placed at 3 and 6 mm from its extremity, facilitate theevaluation of bone defect dimension and control cutting depth. Thissaw is particularly effective for thick cortical bone.

03.000.405S – Scalpel round

Description: Scalpel, round, 22.45 x 12.6 x 3.9 mm dia. X 0.7 mm

Circular scalpel for osteoplasty and harvesting of bone particles or chips.

Treatment examples: Osteoplasty, periodontal surgery, sinus lift andcysts, debridement, callus removal, bunionectomy, exeresis.

Tip angulation: 120°

During bone graft, this tip shapes the recipient site(osteoplasty) to ensure the graft stability andfacilitate its integration. Used during sinus lift, itcollects bone shavings of the vestibular bonewindow. The aspirated autogenous bone will beharvested in a bone filter to be integrated later intothe bone filling material. In the presence of a cyst,the tip planes the site until the cystic follicle isperfectly exposed.

Piezoelectric System Synthes 23

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

60

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

60

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 23

Page 25: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

03.000.406S – Scalpel flat

Description: Scalpel, flat, 21.6 x 11.1 x 3.0 x 0.5 mm

Curved scalpel for osteoplasty and harvesting of bone particles or chips.

Treatment examples: Craniotomy, alveolar ridge augmentation,debridement, callus removal, bunionectomy.

Tip angulation: 90°

The osteoplasty tip cleans therecipient site, smooths surfacedefects and eliminates adheringperiosteum (fibrocellular layer) tobest adapt the recipient site tothe graft. It can be used forremodeling the graft, for

eliminating any secant or aggressive zone and for separating the corticalbone graft from subjacent medullary tissue during grasp of the graft.

Piezoelectric System Synthes 24

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

60

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 24

Page 26: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

03.000.407S – Scalpel

Description: Scalpel,17.7 x 15.5 x 1.2 x 0.5 mm

Scalpel for performing osteoplasty and dissecting.

Treatment examples: extraction of impacted teeth, removal ofremaining root fragment.

Tip height: 26.3 mm

During dissection, this tip allows deep penetrationwith precaution. The practitioner has to makesweeping movements. The ultrasonic micro-oscillations entail the resection of bone wall withoutaffecting peripheral tissues.

03.000.408S – Spade Round

Description: Spade, round, 24 x 11.05 x 4.0 x 0.65 mm

Non-cutting spatula for membrane elevation and for disengagementof anatomical structures.

Treatment example: sinus lift, membrane, dura or cyst removal.

Tip angulation: 120°

This tip has to stay in contact between the bone andsoft tissue. The elevation is undertaken at the apical,mesial, then distal part. It can be used for separatinga cyst from the osseous wall of the cystic cavity.

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

80

Piezoelectric System Synthes 25

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

30

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 25

Page 27: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

03.000.409S – Diamond

Description: Diamond, 24.6 x 12.85 x 2.6 x 0.6 mm

Diamond-coated tip for vestibular bone window cutting and forattenuating of sharp angles.

Treatment examples: BSSO, Le Fort, sinus lift, craniosynostosis,decompression (orbital, spinal, hand or foot).

Tip height: 30 mm

This tip is used to perform less aggressive boneincisions than saws. It is used for cutting andattenuating sharp angles while protecting thenearby anatomical structures. The practitioner hasto perform a constant (longitudinal) sweeping ofthe tip on the surface to be incised. The tipremodels all the secant bone zones likely to damagea membrane or surrounding soft tissue.

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

60

Piezoelectric System Synthes 26

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 26

Page 28: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

03.000.410S – Diamond, round

Description: Diamond, 21.9 x 12.4 x 1.8 mm

Diamond-coated tip for smoothing bone, precise osteoplasty anddecompression.

Treatment examples: BSSO, Le Fort, sinus lift, craniosynostosis,decompression (orbital, spinal, hand or foot).

Ball diameter: 1.5 mm

Tip height: 34.5 mm

This diamond tip performs very fine bone incisions.It is intended for cutting very thin bone, precisionosteoplasty and nerve decompression. The tipremodels all the secant bone zones likely to damagea membrane or surrounding soft tissue.

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

60

Piezoelectric System Synthes 27

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 27

Page 29: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

03.000.411S – Sinus lift

Description: Sinus lift, 22.9 x 10.1 x 5.0 x 0.4 mm

Plateau tip, non-cutting, used for membrane elevation on thewindow’s edges.

Treatment examples: Sinus lift, decorticating.

Tip height: 31 mm

Plateau diameter: 5 mm

This non-cutting tip is used forelevating membranes. It isessential to keep permanentcontact between the membranelining and the bone.

MaximumRecommendedMode

Maximum Fine Tuning

MinimumIrrigationin ml/min

50

Piezoelectric System Synthes 28

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 28

Page 30: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Cleaning and Sterilizing

These recommendations are for reprocessing, specifically cleaning,Synthes Piezoelectric System reusable medical devices, including thePiezoelectric System handpiece-cable assembly and the flat wrench.

Cautions– Do not use steel wool or abrasive cleaners.

– Avoid solutions containing iodine or high chlorine content.

– Contaminated or used Synthes accessories should not be loaded in agraphic case and cleaned in a mechanical washer.

– Metallic handpiece nosepiece, LED ring, light diffuser and handpiece-cable assembly cannulation require particular attention duringcleaning.

– All devices must be thoroughly cleaned.

– Synthes doesn’t recommend or support the flash sterilization ofSynthes instruments.

– Synthes instruments are critical devices and must be terminallysterilized prior to use.

– The sterilization parameters are only valid for devices that areadequately cleaned.

– The following parameters are only valid for properly installed,maintained, calibrated and compliant reprocessing equipment.

– The recommended cleaning method for Piezoelectric handpiece andaccessories is manual and mechanical. Do not place SynthesPiezoelectric handpiece-cable assembly in an ultrasonic cleaner.

– Use a soft, lint-free cloth moistened with a neutral pH cleaner to wipethe console and foot pedal. Do not immerse.

Piezoelectric System Synthes 29

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 29

Page 31: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Processing instructions

Point of use care– Wipe blood and/or debris from the devices throughout the surgical

procedure, to prevent it from drying onto the surface.

– After the operation, remove the single-use tip at the front of thePiezoelectric handpiece assembly and dispose of it in the appropriatecontainer.

Important: In the operating room, rinse the irrigation line of thePiezoelectric System handpiece-cable assembly using distilled or sterilewater. Operate the irrigation pump for 20 seconds after each use toprevent the drying of debris inside the handpiece. Disconnect anddispose of irrigation tubing and clips in proper container.

Soiled devices should be separated from noncontaminated devices toavoid contamination of personnel or surroundings.

Devices should be covered with a towel dampened with purified waterto prevent blood and/or debris from drying.

Containment and transportationSoiled devices should be transported separately from noncontaminateddevices to avoid contamination.

Piezoelectric System Synthes 30

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 30

Page 32: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Preparation for decontamination

It is recommended that devices should be reprocessed as soon aspossible following use. Synthes devices should be reprocessed no morethan two hours after use.

Disassemble the handpiece prior to cleaning – The front part of the Piezoelectric handpiece assembly, the nose cone,

must be unscrewed.

– Remove the light diffuser.

– Disconnect the LED ring from the Piezoelectric handpiece, pulling itgently.

The handpiece, cable and handpiece components (nose cone, LED ringand light diffuser) should be manually precleaned prior to processing inan automated washer.

Soak and/or rinse heavily soiled devices prior to cleaning, to loosen anydried soil or debris. Use a neutral pH enzymatic cleaner solution to soakdevices. Follow the enzymatic cleaner manufacturer's instructions for use,for correct exposure time, temperature, water quality (i.e. pH, hardness)and concentration. Use cool tap water to rinse devices.

Synthes Piezoelectric devices must be first cleaned separately from theirgraphic cases. Lids should be removed from graphic cases for thecleaning process, if applicable. It is the responsibility of the end user toensure that all equipment used to reprocess Synthes devices is properlyinstalled, validated, maintained and calibrated.

Light diffuserNose cone Piezoelectric

handpiece-cable assemblyLED ring

Piezoelectric System Synthes 31

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 31

Page 33: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Cleaning — manual methodEquipment: Soft-bristled brushes, lint-free swab, soft, lint-free cloth,syringes, pipettes or water jet, neutral enzymatic cleaner and/or neutraldetergent solution with a pH≤ 8.5.

1) Rinse soiled device under running cold tap water for a minimum oftwo minutes. Use a soft-bristled brush and/or lint-free swab, or soft, lint-free cloth to assist in the removal of gross soil and debris.

2) Use a syringe, pipette or water jet to flush the cannulation with aneutral pH enzymatic cleaner solution.

3) Soak device in a neutral pH enzymatic cleaner solution for a minimumof ten minutes. Follow the enzymatic cleaner manufacturer'sinstructions for use for correct exposure time, temperature, waterquality (i.e. pH, hardness) and concentration.

4) Rinse device with cool water for a minimum of two minutes.

5) Manually clean device for a minimum of five minutes in a freshlyprepared neutral pH enzymatic cleaner solution. Use a soft-bristledbrush to remove soil and debris on the nose cone, light diffuser, LED ring, handpiece-cable assembly and the threaded fitting. Gentlyuse a lint-free swab to remove soil and debris on the inside of thenose cone, and LED ring. Clean device under water to preventaerosolization of contaminants. Note: Fresh solution is a newly-made, clean solution.

6) Rinse device thoroughly with deionized or purified water for twominutes.

7) Manually clean device for a minimum of two minutes in a neutral pHdetergent solution. Use a soft-bristled brush for the nose cone, lightdiffuser, LED ring, handpiece-cable assembly and the threaded fitting.Gently use a lint-free swab for the inside of the nose cone, opticalguide and LED ring. Clean device under water to preventaerosolization of contaminants.

8) Use a syringe, pipette, or water jet to flush the cannulation withdeionized or purified water and rinse device thoroughly withdeionized or purified water for two minutes.

9) Visually inspect device. Repeat the manual cleaning procedure untilno visible soil remains on devices.

10) Perform a final rinse on device, using deionized or purified water.

11) Dry device using a soft, lint-free cloth or clean compressed air.

Piezoelectric System Synthes 32

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 32

Page 34: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Cleaning – mechanical methodNote: Ultrasonic cleaning may cause further damage to devices thathave prior surface damage.

Equipment: Soft-bristled brushes, lint-free swab and/or soft, lint-freecloth, syringe, pipettes, and/or water jet, ultrasonic cleaner, washer/disinfector, neutral enzymatic cleaner and/or neutral detergent with a pH ≤ 8.5. Note: Ultrasonic process – flat wrench and nose cone only.1) Disassemble device if applicable. (Refer to page 31 for disassembly

information.)

2) Rinse soiled device under running cold tap water for a minimum ofone minute. Remove gross soil using a soft- bristled brush, a lint-freeswab or soft, lint-free cloth.

3) Manually clean device for a minimum of two minutes in a freshlyprepared neutral pH enzymatic cleaner solution. Follow the enzymaticcleaner manufacturer's instructions for the correct dilution, temperature,water quality (i.e. pH, hardness) and exposure time. Use a soft-bristledbrush and/or lint-free swab to remove soil and debris. Clean deviceunder water to prevent aerosolization of contaminants. Note: Fresh solution is a newly-made, clean solution.

4) Rinse device using cool to lukewarm running tap water for aminimum of one minute. Use a syringe, pipette or water jet to flushcannulation.

5) Prepare a fresh detergent solution using a neutral pH enzymaticcleaner. Follow the enzymatic cleaner manufacturer’s instructions forthe correct dilution, temperature, water quality (i.e. pH, hardness) andexposure time. Ultrasonic process — flat wrench and nose cone only. Note: Fresh solution is a newly-made, clean solution.

6) Clean only flat wrench and nose cone ultrasonically for 15 minutes.

7) Rinse device in cool to lukewarm running tap water for a minimum oftwo minutes. Use a syringe, pipette or water jet to flush thecannulation.

8) Visually inspect device. Repeat Steps 2 – 7 until no visible soil remainson device.

Piezoelectric System Synthes 33

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 33

Page 35: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Mechanical method – (after preclean)Use the following cycle parameters:

Thermal Disinfection

For automated cleaning, thermal disinfect at 93°C for a minimum of 5 minutes.

Drying

If a dry cycle is not included in a mechanical washer or if the device isnot processed in a mechanical washer:

– Dry each device thoroughly inside and out to prevent rusting andmalfunction.

– Use a soft, lint free cloth to avoid damage to the surface.

– Pay special attention to threads or areas where fluid can accumulate.

– Open and close devices so that all areas are reached.

– Dry hollow parts using the air jet with clean compressed air.

Piezoelectric System Synthes 34

Cycle Time (minutes)

Temperature/Water

Type of Detergent

Pre-wash 2 Cold tap water N/A

Wash I 2 Cool to warm tap water

Neutral enzymaticpH ≤ 8.5

Wash II 5 Warm tap water(>40°C)

Detergent pH ≤ 8.5

Rinse 2 Warm DI or PURW (>40°C)

N/A

Dry 40 90°C N/A

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 34

Page 36: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Sterilization

Synthes nonsterile Piezoelectric system devices must be terminallysterilized prior to use.

The following are the recommendations for the sterilization of thecomponents stored in the graphic case (handpiece and cable, handpiececomponents (nose cone, light diffuser, and LED ring) and flat wrench).

– Dry times may be highly variable due to differences in packaging materials(e.g. nonwoven wraps), environmental conditions, steam quality, devicematerials, total mass, sterilizer performance and varying cool downtimes. The user should employ verifiable methods (e.g. visualinspections) to confirm adequate drying.

Inspection

– Synthes instruments should be inspected after processing (any methodpreviously listed).

– The recommendations provided have been validated by the medicaldevice manufacturer as being capable of preparing a non-sterile Synthesimplant or reusable medical device. It remains the responsibility of theprocessor to ensure that the processing is actually performed, usingequipment, materials and personnel in the reprocessing facility, andachieves the desired result. This requires validation and routinemonitoring of the process. Likewise, any deviation by the processorfrom the recommendations provided should be properly evaluated for effectiveness and potential adverse consequences.

– All users should be qualified personnel with documented expertise,competency and training. Users should be trained on hospital policiesand procedures along with current applicable guidelines and standards.

– Users should wear appropriate personal protective equipment (PPE) whenprocessing devices.

Cycle Type Min SterilizationExposure Time (minutes)

Min SterilizationExposure Temperature

Min Dry Time

Prevacuum 10 minutes 132°C (270°F) 20 minutes

Piezoelectric System Synthes 35

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 35

Page 37: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Additional Information

This information is provided in accordance with ANSI/ AAMI ST81, ISO17664, AAMI TIR 12 and AAMI ST 77.

Manufacturer contact

For further information, contact Synthes Customer Service Department at1-800-523-0322 or the Sterility Assurance Department at 610-719-6836.

This quick reference chart is not intended to be a replacement of thePiezoelectric System User's Manual.

Please refer to the User's Manual for complete details on the system.

Piezoelectric System Synthes 36

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 36

Page 38: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Handpiece not connected

Transmission fault

Toolbox

Reduce a value

Increase a value

Reduce flow rate

Increase flow rate

Flush/prime

Irrigation

Light

SYNTHESPiezoelectric System

Piezoelectric function

Foot pedal ON/OFF mode

Foot pedal progressivemode

Storing of settings

Piezoelectric program

Confirmation

Audio volume

Brightness

Time-out length

Factory configuration

Refer to the accompanying documentation

Class BF

I Class 1

CE marking

Do not dispose of in household waste

Year of manufacture

Alternating current

Control foot pedal

0 Device power OFF

I Device power ON

Handpiece ON

Degree of protectionagainst liquid of theControl foot pedal

!3ZG5

MEDICAL EQUIPMENTUL60601-1

CAN / CSA C22.2 No. 601.1

WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL60601-1 AND CAN / CSA C22.2 No. 601.1

Piezoelectric System Synthes 37

Symbols

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 37

Page 39: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Piezoelectric System Set (01.001.598)

Graphic Case60.001.400 Single Level Modular Graphic Case,

for Piezoelectric System (includes lid)

60.001.401 Attachment Tray, for Piezoelectric System

Instruments05.001.400 Console, for Piezoelectric System

(includes power cord 05.001.140)

05.001.401 Hand Piece, for Piezoelectric System

05.001.402 Foot Pedal, for Piezoelectric System

05.001.404 Flat Wrench, for Piezoelectric System

Also available05.001.140* Medical Grade Power Cord, three prongs, clear plug

60.001.402 Lid, for Piezoelectric System05.001.417 LED Ring, for Handpiece,

for Piezoelectric System

05.001.418 Light Diffuser Cone for Handpiece, for Piezoelectric System

05.001.419 Irrigation Pole, for Piezoelectric System

05.001.420 Nose Cone, for Piezoelectric System

Note: Trauma customers add .97 to the part number, CMF customers add .98 (ex: 05.001.400.98)

* Part number 05.001.140 does not require an extension.

Piezoelectric System Synthes 38

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 38

Page 40: Piezoelectric System. Ultrasonic handpiece with cutting ...synthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes North America... · Piezoelectric System. Ultrasonic handpiece with cutting

Distributed bySynthes (USA)1302 Wrights Lane EastWest Chester, PA 19380Telephone: (610) 719-5000To order: (800) 523-0322Fax: (610) 251-9056www.synthes.com

Synthes (Canada) Ltd.2566 Meadowpine BoulevardMississauga, Ontario L5N 6P9Telephone: (905) 567-0440To order: (800) 668-1119Fax: (905) 567-3185

© 2011 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 3/11 J10400-A

Manufactured by

SATELEC A Company of ACTEON Group17 av. Gustave Eiffel BP 3021633708 MERIGNAC cedex France

28923-J10400A_28923-J10400A 3/9/11 10:43 AM Page 39