54
Pencil Beam and Collapsed Cone Algorithm Calculations for a Lung-type Volume Using CT and the OMP Treatment Planning System Methods Measurements have been carried out in both phantom and a specifically designed phantom which simulated human lung volume. Samples were taken from the Lung Planning CT images for 15 patients using the Oncentra Masterplan OMP Treatment Planning System. The X-axis was, following convention, taken to be horizontal, and the Y-axis to be vertical; accordingly, abscissa and ordinate distances to the skin, heart and the lungs were measured (see figure 8). Figures 4 and 5 show typical CT images for a patient’s lungs, while Tables 1 and 2 give the beam information and dose information for typical patients. The X-ray images were taken using CT-SIM: Philips Brilliance Big Bore. A print out of the planning CT images was produced by the Oncentra Masterplan OMP treatment Planning system (see section 3.3). 1 |

Pencil Beam and Collapsing Cone

Embed Size (px)

Citation preview

Page 1: Pencil Beam and Collapsing Cone

Pencil Beam and Collapsed Cone Algorithm Calculations for a

Lung-type Volume Using CT and the OMP Treatment Planning System

Methods

Measurements have been carried out in both phantom and a specifically designed phantom

which simulated human lung volume. Samples were taken from the Lung Planning CT

images for 15 patients using the Oncentra Masterplan OMP Treatment Planning System.

The X-axis was, following convention, taken to be horizontal, and the Y-axis to be vertical;

accordingly, abscissa and ordinate distances to the skin, heart and the lungs were measured

(see figure 8). Figures 4 and 5 show typical CT images for a patient’s lungs, while Tables 1

and 2 give the beam information and dose information for typical patients. The X-ray

images were taken using CT-SIM: Philips Brilliance Big Bore. A print out of the planning

CT images was produced by the Oncentra Masterplan OMP treatment Planning system (see

section 3.3).

1 |

Page 2: Pencil Beam and Collapsing Cone

Figure 4: Lungs Image for Patient by CT-SIM: Philips Brilliance Big Bore

Beam InformationBeam 1ANT 2RPO 3RPO 4ARO

Nom. Acc. Pot.(MV or MeV) 6 6 6 6FX (cm) 8.2 8.2 8.6 8.6FY (cm) 9.4 10.6 8.2 8.6SSD (cm) 87.2 86.8 85.6 84.7

Gantry (degrees) 0 223 267 320Wedge Angle(degrees) 60/60 60/33 60/25

Dose Information: Absolute dose 5500 cGy (275 cGy / fraction)Number of Fraction 20 20 20 20

MU or min / FractionIN = 424.77

OUT = 0IN = 185.78OUT = 85.22

IN = 92.85OUT = 69.34

16.39

Table 1: Beam Information and Dose Information for Patient

2 |

Page 3: Pencil Beam and Collapsing Cone

Figure 5: Lungs Image for Patient by CT-SIM: Philips Brilliance Big Bore

Beam Information

Beam3LPO

THORAX4ANT 5MINI-ANT 6LAO

Nom. Acc. Pot.(MV or MeV)

6 6 6 6

FX (cm) 10.6 10.1 9.7 14.2FY (cm) 9.7 14.3 11 10.1SSD (cm) 86.3 87.2 87.2 82.4

Gantry (degrees) 120 0 0 60Wedge Angle(degrees) 60/25 60/28 60/60 60/9

Dose Information: Absolute dose 4000 cGy (267 cGy / fraction)Number of Fraction 15 15 15 15

MU or min / FractionIN = 2291.07

OUT = 1711.09IN = 1277.72OUT = 793.65

IN = 1974.08OUT = 0

IN = 476.08OUT = 1301.50

3 |

Page 4: Pencil Beam and Collapsing Cone

Table 2: Beam Information and Dose Information for Patient

Design of multi-block chest phantoms

The first phantom was introduced to the experiment as shown in figure 7, in order to reduce

the uncertainty within the results and to increase the accuracy all that because of the very

inhomogeneous lung region that may led to poor dose distribution.

Figure 7: for Design 1 of the Multiblock phantom (first phantom)

The specially designed phantom

Using measurements taken from 15 patients, who had previously been scheduled for lung

radiotherapy, a second phantom consisting of multi-block components was designed. A

multi-block phantom is essentially a phantom containing a number of blocks with different

shapes and materials used to form an approximate cross-section of the patient. This

facilitates taking measurements on the phantom volume to confirm the prescribed dose. A

plan for the phantom was designed using similar field parameters, for example collimator

settings, beam weightings, wedge fractions, and gantry angles as the clinical plan. The two

lungs are presented in a lateral position, as shown in Figure 8 the heart is represented in the

4 |

Page 5: Pencil Beam and Collapsing Cone

middle to reflect the correct anatomy and the lighter color in both Figures 7 and 8 represent

the lungs.

Figure 8: Design 2 for the Multiblock phantom (second phantom), where (S-L) is skin and lung, and

(L-H) is lung and heart.

Table 3 shows the average distance between the skin, lungs and heart of the patient from

the X-ray for the X and Y axes. The table also illustrates the maximum and minimum

values for the X and Y axes, as well as the range of maximum and minimum values. Figure

8 illustrates the distance for X and Y axes between the skin, lungs and heart in the Multi-

Block Phantom. The phantom blocks designed for the experiments were 30 cm in length,

having square or right angled triangular cross-sections and 4 cm sides. They were made of

an epoxy resin-based tissue-equivalent material to represent water (WT1, density=1.02

5 |

Page 6: Pencil Beam and Collapsing Cone

gcm-3), lung (LN10 density=0.27 gcm-3) and bone (IB7, density=1.13 gcm-3). Some of the

square blocks of WT1 were drilled to accommodate a 0.6 cm3 graphite ionisation chamber.

The phantoms were composed by putting the epoxy blocks within an adjustable wooden

frame in desired configurations. The frame was held together using a series of small

wooden pins with diameters of 5 mm.

Skin to Lung (cm)

Lung to heart (cm)

Heart (cm)

Heart to Lung (cm)

Lung to Skin (cm)

X,Y axis

Average for patient(Lateral)

2.8 4.2 4.2 3.3 3.1 17.4

Average for patient (Ant-post)

2.3 1.9 4.4 1.3 2.6 11.1

Average for Phantom(X)

4 8 4 8 4 28

Average for Phantom(Y)

4 4 6 2 4 20

MAX in X axis for patients (Lateral) 19MIN in X axis for patients (Lateral) 15MAX in Y axis for patients (Ant-post) 13.7MIN in Y axis for patients (Ant-post) 9.6RANGE in X axis for patients (Lateral) 4RANGE in Y axis for patients (Ant-post) 4.1Table 3: The area for Lateral and Ant-post in 15 Patients (average and range for 15 patients) and

average for Multiblock phantom

Figures 9 and 10 below depict various stages in the construction of the thorax phantom

within its frame. Expanded polystyrene spacer elements with triangular cross sections

stabilised the slanted surfaces.

6 |

Page 7: Pencil Beam and Collapsing Cone

Import and Plan

After scanning the multi-blocks phantom using CT - SIM: Philips Brilliance Big- Bore, the

planning CT images were sent to the Oncentra Masterplan (OMP) Treatment Planning

System. The Oncology Management System: Impac, MOSAIQ was used to transfer the

data from the OMP treatment planning system to the Linac before running the Linac to

determine the points’ ISO center, Beam Information and Dose Information, as shown in

figure 14, 15, 16 and 17 for the first and second phantom.

The phantoms were positioned on the Elekta Precise linac, isocentre and aligned with

lasers, and the ion chamber was placed at each dose point, for example Iso, DP1, DP2, DP3

and DP4 (see figure 12 and 13). Doses were measured for the dosimeters and chambers.

The field size and gantry angles chosen are typical of clinical plans for the same 15 patients

as used to design phantom 2. A field size of 10 x 10cm, was used for all fields. Gantry

angles of 00-3150-2700 and 00-600-1200 were used for phantom 1 and 2 respectively. Tables

4 and 5 show beam information for the first and second phantoms, respectively. The energy

7 |

Page 8: Pencil Beam and Collapsing Cone

used for the plans was 6MV because lung cancer is treated clinically with 6MV in HOF

Hospital 10 MV beam is not used because considered very high energy and risky to the

lungs. Wedges were used for beam one and three- the angle of the wedge is 60/60 for each

beam. Figure 12 and 13 show the plan for phantoms 1 and 2, with the isocentre and dose

points measured.

For the first phantom was generated using three 6 MV photon beams, all with a 10 x 10

cm2 field size, as shown in fig A.

Figure A. The plan used for first phantom.

The plan was isocentric and included an ANT beam with a 60º wedge and a right RAO

beam with no wedge. The third field was a right LAT oblique beam with a 60º wedge. The

first phantom was outlined and the total dose prescribed to the isocentre was 5492.8 cGy

(274.6 cGy / fraction).

8 |

Page 9: Pencil Beam and Collapsing Cone

For the second phantom was generated using three 6 MV photon beams, all with a 10 x 10

cm2 field size, as shown in fig B.

Figure B. The plan used for second phantom.

The plan was isocentric and included an ANT beam with a 60º wedge and a left LAO beam

with no wedge. The third field was a left LPO oblique beam with a 60º wedge. The second

phantom was outlined and the total dose prescribed to the isocentre was 3971.8 cGy (264.8

cGy / fraction).

9 |

Page 10: Pencil Beam and Collapsing Cone

Figure 12: Plan for the first phantom, showing isocentre and 3 dose points (DP1, DP2 and DP3)

(see appendix for large pictures).

Beam InformationBeam ANT RAO LATNom. Acc. Pot.(MV or MeV) 6 6 6Field size X (cm) 10 10 10Field size Y (cm) 10 10 10SSD (cm) 90 86 90Gantry (degrees) 0 315 270

Table 4: Beam information for the first phantom.

10 |

Page 11: Pencil Beam and Collapsing Cone

Figure 13: Plan for the second phantom, showing isocentre and 4 dose points (DP1, DP2, DP3 and

DP4) (see appendix for large pictures).

Beam InformationBeam ANT LAO LPONom. Acc. Pot.(MV or MeV) 6 6 6Field size X (cm) 10 10 10Field size Y (cm) 10 10 10SSD (cm) 90 88.5 91.3Gantry (degrees) 0 60 120

Table 5: Beam information for the second phantom.

11 |

Page 12: Pencil Beam and Collapsing Cone

Measurements on the Linac

For the experiment with the phantoms, a Farmer dosimeter and an Ionisation Chamber with

a volume of 0.6cc, both from N E Technology were used. A 6MV X-Ray beam with a SSD

of 100cm and a depth of 5cm was used, along with a field size of 10 x 10cm and a Set Dose

(SD) of 400MU. For the first phantom, on the first day, the experiments were conducted at

room temperature of 20.5 oC, and at a pressure of 765.5 mmHg. For the second phantom,

on the second day, the experiments were conducted with the temperature at 20.2 oC and

pressure 759.2 mmHg. A temperature correction factor of 0.9945 was calculated using

equation 2.3, and a Depth Dose Correction to dmax of 0.863 was used for the calculations,

which is a constant for a 6MV linac in experiment. Further, the value of Wρ (density

correction) was taken as 1.000, which is a correction for Perspex to water. The calibration

factor for the ion chambers were 0.794 and 1.034, as two different chambers were used for

the two sets of measurement. The ion recombination factor, Pion, was 1.0042 in both cases.

The following equations were used to calculate the dose delivered:

Dose = Reading x ND x Pion x Ø (P, T) x Wρ / %DD (2.1)

For Phantom: D (cGy) = Reading x ND x P ion x Ø (P, T) x daily calibration correction factor (2.2)

Wρ is the density correction factor;

%DD is the percentage Depth Dose.

ND Calibration factor ion chamber and electrometer.

Pion ion recombination.

12 |

Page 13: Pencil Beam and Collapsing Cone

First phantom- daily calibration correction factor = (400MU / 401.9cGy)

Second phantom- daily calibration correction factor = (400MU / 399.1cGy)

Readings were obtained from the Dosemeter and converted to dose (401.9-399.1) using Equation 2.1.

Ø (P, T) = (273 + T / 293) * (760 / P), (2.3)

Where Ø (P, T) is the temperature and pressure correction factor, given by equation 2.3

In the users’ beam, the correction factor for air temperature and air pressure Ø (P, T) is

given as:

Ø (P, T) = ;

and is applied to convert the measured signal to the reference conditions used for the

chamber calibration at the standards laboratory. Note that P and T (in oC) are chamber air

pressure and temperature, respectively, at the time of measurement, while Po and To (in oC)

are the normal conditions used in the standards laboratory.

The temperature of the air in a chamber cavity should be taken as that of the phantom and

this is not necessarily the same as the temperature of the surrounding air. For measurements

in a water phantom the chamber waterproof sleeve should be vented to the atmosphere in

order to obtain a rapid equilibrium between the ambient air and the air in the chamber

cavity.

The ionisation chamber measurements were taken on linear accelerator A (Lin A). The

Linac was used to deliver the 6MV X-ray beam to each phantom separately. During this

process, the ionisation chamber was inserted within the phantom at each dose point, (Iso,

DP1, DP2, DP3 and DP4). Moreover, the radiation beam were delivered as per the plans in

13 |

273 .2+T273 .2+T0

⋅P0

P

Page 14: Pencil Beam and Collapsing Cone

figure 12 and 13. The readings were taken with the ionisation chamber are shown in table 8

and 9. These measurements were used with equation 2.2 to calculate the dose in cGy. The

percentage difference between the measured dose and the dose calculated using the PB and

CC algorithms was calculated.

Table 6 shows the two models for Dosemeter and Chambers used to obtain the dose, while

Table 7 summaries the parameters used in the experimental measurements.

Manufacturer Description Part No. Serial No. Local Description

Dosemeter:NE Technology Farmer 2570/1 B 944 & 1297 Field 2 & Field

3Chambers:

NE technology 0.6 cc thimble & graphite

2571 & 2571 A 1884 & 2921 Mk3 & Mk4

Table 6: Dosemeter and ionisation chambers used for the experimental measurements.

Sample depth dose chart for a 6 MV X-ray beam for a treatment distance of 100 cm SSDFirst day Second day

Field Size 10 x 10 cm 10 x 10Source to surface Distance (SSD) 100 cm 100 cm

Depth 5 cm 5 cmSet Dose (SD) 400 mu 400 mu

ND (calibration factor ion chamber & electrometer) 0.794 1.034Pion (ion recombination) 1.0042 1.0042

Temperature 20.5 Centigrade 20.2 Centigrade

Pressure 765.50 mmHg 759.20 mmHgØ (P,T) (Temperature & Pressure correction) 0.9945 1.0017

Wρ (density Correction) 1.000 1.000Depth Dose Correction to dmax 0.863 0.863

Table 7: Parameters used in experimental measurements.

14 |

Page 15: Pencil Beam and Collapsing Cone

4. Results and Discussion

4.1 Results

As describe in the Materials and Methods a Set Dose (SD) of 400MU was used. Readings

were obtained from the Dosimeter and converted to dose after accounting for daily

calibration correction factor for first and the second phantoms. Using Equation 2.1 and

Equation 2.2.the calculated values of 400MU=401.9cGy and 400MU=399.1cGy were

obtained respectively.

Table No 8 and Table No 9 summarize the Beam Information for Beams 1, 2 and 3 for

Phantom 1 and Phantom 2 respectively.

Table No 8:

Table No 1 demonstrates percentage difference between Pencil Beam (PB) with Experiment

Measured and Collapsed Cone (CC) with Experiment Measured for first phantom. The results are

as follows:

a. For Isocenter:

i. Experiment Measure for Isocenter ANT Beam one is 106.93 while for CC it is

107.9 and for PB it is 106.9. The % difference between the experiment measure

and that of CC is -0.02 % and between the experiment measure and that of PB is

0.9%.

ii. Experiment Measure for Isocenter RAO Beam two is 60.9 while for CC it is 61

and for PB it is 61.1. The % difference between the experiment measure and that

of CC is 0.1 % and between the experiment measure and that of PB is 0.3%.

15 |

Page 16: Pencil Beam and Collapsing Cone

iii. Experiment Measure for Isocenter LAT Beam three is 104.7 while for CC it is

105.7 and for PB it is 106.9. The % difference between the experiment measure

and that of CC is 0.9% and between the experiment measure and that of PB is

2.1%.

b. For DP 1:

i. Experiment Measure for DP1 ANT Beam one is 112.05 while for CC it is

112.7 and for PB it is 112.4. The % difference between the experiment

measure and that of CC is 0.5 % and between the experiment measure and

that of PB is 0.3%.

ii. Experiment Measure for DP1 RAO Beam two is 64.07 while for CC it is

64.7 and for PB it is 65.3. The % difference between the experiment

measure and that of CC is 0.9 % and between the experiment measure and

that of PB is 1.9%.

iii. Experiment Measure for DP1 LAT Beam three is 95.88 while for CC it is

97.3 and for PB it is 98.6. The % difference between the experiment

measure and that of CC is 1.4 % and between the experiment measure and

that of PB is 2.8%.

c. For DP2:

i. Experiment Measure for DP2 ANT Beam one is 98.7 while for CC it is 78.3

and for PB it is 80. The % difference between the experiment measure and

16 |

Page 17: Pencil Beam and Collapsing Cone

that of CC is 0.5 % and between the experiment measure and that of PB is

2.7%.

ii. Experiment Measure for DP2 RAO Beam two is 90.7 while for CC it is 71.1

and for PB it is 73.3. The % difference between the experiment measure and

that of CC is -0.6 % and between the experiment measure and that of PB is

2.4%.

iii. Experiment Measure for DP2 LAT Beam three is 117.74 while for CC it is

119.7 and for PB it is 120.6. The % difference between the experiment

measure and that of CC is 1.6 % and between the experiment measure and

that of PB is 2.4%.

d. For DP3:

i. Experiment Measure for DP3 ANT Beam one is 139.04 while for CC it is

139.6 and for PB it is 140.8. The % difference between the experiment

measure and that of CC is 0.4 % and between the experiment measure and

that of PB is 1.2%.

ii. Experiment Measure for DP3 RAO Beam two is 20.12 while for CC it is 27

and for PB it is 28.4. The % difference between the experiment measure and

that of CC is 34.1 % and between the experiment measure and that of PB is

41.1%.

iii. Experiment Measure for DP3 LAT Beam three is 6.7 while for CC it is 6.1

and for PB it is 6.7. The % difference between the experiment measure and

17 |

Page 18: Pencil Beam and Collapsing Cone

that of CC is -8.9 % and between the experiment measure and that of PB is

0%.

Table 8: Dosimeter Readings and percentage difference between PB with measured and CC with measured for first phantom.

ANT Beam one

Beam 1% Different

RAO Beam two

Beam 2% Different

LAT Beam three

Beam 3% Different

Total Dose at Isocentre

MU or min / Fraction IN = 451.9 78.4 IN = 487.8

Iso

PB IN = 106.9 - 0.02% 61.1 0.3% IN = 106.9 2.1% 274.9CC IN = 107.9 0.9% 61 0.1% IN = 105.7 0.9% 274.6

Reading from Dosemeter 135.5 77.2 132.7

experiment Measure 106.93 60.9 104.7 272.53

DP1

PB IN = 112.4 0.3% 65.3 1.9% IN = 98.6 2.8%CC IN = 112.7 0.5% 64.7 0.9% IN = 97.3 1.4%

Reading from Dosemeter 142 81.2 121.5

experiment Measure 112.05 64.07 95.88

DP2

PB IN = 80 2.7% 73.3 2.4% IN = 120.6 2.4%CC IN = 78.3 0.5% 71.1 - 0.6% IN = 119.7 1.6%

Reading from Dosemeter 98.7 90.7 149.2

experiment Measure 77.88 71.57 117.74

DP3

PB IN = 140.8 1.2% 28.4 41.1% IN = 6.7 0%CC IN = 139.6 0.4% 27 34.1% IN = 6.1 - 8.9%

Reading from Dosemeter 176.2 25.5 8.5

experiment Measure 139.04 20.12 6.7Percentage difference = [(PB/Measure) * 100] – 100%Percentage difference = [(CC/Measure) * 100] – 100%

For the First Phantom the isocentre plans includes an ANT beam with a 60º wedge, a right

RAO beam with no wedge and a LAT oblique beam with a 60º wedge. The Isocentre dose

for PB and CC algorithms were provided by the OMP treatment planning system. The first

phantom was outlined and the total dose prescribed to the isocentre for CC was 5492.8 cGy

(274.6 cGy / fraction) and for PB the total dose prescribed was 274.9 cGy / fraction. (Table

No 8). Dosimeter and ionization chamber were used to arrive at the Iso- Reading value of

135.5 cGy. Equation 2.2 was used to obtain the Iso Measure value of 106.93 cGy (Fig No

14 & 15, Table No 8).

18 |

Page 19: Pencil Beam and Collapsing Cone

Figure 14: First phantom calculated with Pencil Beam (PB)

Figure 15: First phantom calculated with Collapsed Cone (CC)

19 |

Page 20: Pencil Beam and Collapsing Cone

Table No 9:

Table No 9 demonstrates percentage difference between Pencil Beam (PB) with Experiment

Measured and Collapsed Cone (CC) with Experiment Measured for Second phantom. The results

are as follows:

a. For Isocenter:

i. Experiment Measure for Isocenter ANT Beam one is 89.3 while for CC it is 90

and for PB it is 90.5. The % difference between the experiment measure and that

of CC is 0.7 % and between the experiment measure and that of PB is 1.3%.

ii. Experiment Measure for Isocenter LAO Beam two is 117.7 while for CC it is

119.3 and for PB it is 120.6. The % difference between the experiment measure

and that of CC is 1.3 % and between the experiment measure and that of PB is

2.4%.

iii. Experiment Measure for Isocenter LPO Beam three is 50.5 while for CC it is

55.5 and for PB it is 55.6. The % difference between the experiment measure

and that of CC is 9.9% and between the experiment measure and that of PB is

10.09%.

b. For DP1:

i. Experiment Measure for DP1 ANT Beam one is 94.7 while for CC it is 94.4

and for PB it is 95.1. The % difference between the experiment measure and

that of CC is -0.1 % and between the experiment measure and that of PB is

0.6%.

20 |

Page 21: Pencil Beam and Collapsing Cone

ii. Experiment Measure for DP1 LAO Beam two is 121.9 while for CC it is

121.6 and for PB it is 121.4. The % difference between the experiment

measure and that of CC is -0.2 % and between the experiment measure and

that of PB is -0.4%.

iii. Experiment Measure for DP1 LPO Beam three is 50.03 while for CC it is

50.9 and for PB it is 50.8. The % difference between the experiment

measure and that of CC is 1.7 % and between the experiment measure and

that of PB is 1.5%.

c. For DP 2:

i. Experiment Measure for DP2 ANT Beam one is 82.3 while for CC it is 82.5

and for PB it is 83.5. The % difference between the experiment measure and

that of CC is 0.2 % and between the experiment measure and that of PB is

1.4%.

ii. Experiment Measure for DP2 LAO Beam two is 125.6 while for CC it is

125.6 and for PB it is 127.5. The % difference between the experiment

measure and that of CC is -0.1 % and between the experiment measure and

that of PB is 1.5%.

iii. Experiment Measure for DP2 LPO Beam three is 52.8 while for CC it is

55.5 and for PB it is 55.6. The % difference between the experiment

measure and that of CC is 5.1 % and between the experiment measure and

that of PB is 5.3%.

21 |

Page 22: Pencil Beam and Collapsing Cone

d. For DP3:

i. Experiment Measure for DP3 ANT Beam one is 106.3 while for CC it is

107.4 and for PB it is 110.4. The % difference between the experiment

measure and that of CC is 1.03 % and between the experiment measure and

that of PB is 3.8%.

ii. Experiment Measure for DP3 LAO Beam two is 100.7 while for CC it is

101.9 and for PB it is 102. The % difference between the experiment

measure and that of CC is 1.1 % and between the experiment measure and

that of PB is 1.2%.

iii. Experiment Measure for DP3 LPO Beam three is 34.9 while for CC it is 54

and for PB it is 55. The % difference between the experiment measure and

that of CC is 54.7 % and between the experiment measure and that of PB is

57.5%.

e. For DP4:

i. Experiment Measure for DP4 ANT Beam one is 132.5 while for CC it is

136.7 and for PB it is 139.3. The % difference between the experiment

measure and that of CC is 3.1 % and between the experiment measure and

that of PB is 5.1%.

ii. Experiment Measure for DP4 LAO Beam two is 20.8 while for CC it is 24.4

and for PB it is 26.4. The % difference between the experiment measure and

22 |

Page 23: Pencil Beam and Collapsing Cone

that of CC is 17.3 % and between the experiment measure and that of PB is

26.9%.

iii. Experiment Measure for DP4 LPO Beam three is 38.7 while for CC it is 39

and for PB it is 39. The % difference between the experiment measure and

that of CC is 0.7 % and between the experiment measure and that of PB is

0.7%.

For the Second Phantom

The isocentric plan includes an ANT beam with a 60º wedge and a left LAO beam with no

wedge. The third field is a left LPO oblique beam with a 60º wedge. The second phantom

was outlined and the total dose prescribed to the isocentre was 3971.8 cGy (264.8 cGy /

fraction) and for PB the total dose prescribed was 266.7 cGy. Dosimeter and ionization

chamber were used to arrive at the Iso- Reading value of 135.5 cGy. Equation 2.2 was used

to obtain the Iso Measure value of 106.93 cGy.

The Isocentre dose for PB and CC algorithms were provided by the OMP treatment

planning system. The first phantom was outlined and the total dose prescribed to the

isocentre for CC was 5492.8 cGy (274.6 cGy / fraction) and for PB the total dose

prescribed was 274.9 cGy / fraction. (Table No 8). Dosimeter and ionization chamber were

used to arrive at the Iso- Reading value of 85.7 cGy and the ISO Measure value of 89.3 cGy

was obtained using Equation 2.2 (Fig No 16 & 17, Table No 9)

23 |

Page 24: Pencil Beam and Collapsing Cone

Figure 16: Second phantom calculated with Pencil Beam (PB)

Figure 17: Second phantom calculated with Collapsed Cone (CC)

24 |

Page 25: Pencil Beam and Collapsing Cone

Table 9: Dosimeter Readings and percentage difference between PB with measured and CC with measured for second phantom.

ANT Beam one

Beam 1% Different

LAO Beam two

Beam 2% Different

LPO Beam three

Beam 3% Different

Total Dose at Isocentre

Isocentre% Different

MU or min / Fraction IN = 382.4 144.1 IN = 229.2Iso PB IN = 90.5 1.3% 120.6 2.4% IN = 55.6 10.09% 266.7 3.5%

CC IN = 90 0.7% 119.3 1.3% IN = 55.5 9.9% 264.8 2.8%Reading from Dosemeter

85.7 113.0 48.5

experiment Measure 89.3 117.7 50.5 257.5DP1 PB IN = 95.1 0.6% 121.4 - 0.4% IN = 50.8 1.5%

CC IN = 94.4 - 0.1% 121.6 - 0.2% IN = 50.9 1.7%Reading from Dosemeter

90.7 117.0 48.0

experiment Measure 94.5 121.9 50.03DP2 PB IN = 83.5 1.4% 127.5 1.5% IN = 55.6 5.3%

CC IN = 82.5 0.2% 125.4 - 0.1% IN = 55.5 5.1%Reading from Dosemeter

79.0 120.5 50.7

experiment Measure 82.3 125.6 52.8DP3 PB IN = 110.4 3.8% 102 1.2% IN = 55 57.5%

CC IN = 107.4 1.03% 101.9 1.1% IN = 54 54.7%Reading from Dosemeter

102.0 96.7 33.5

experiment Measure 106.3 100.7 34.9DP4 PB IN = 139.3 5.1% 26.4 26.9% IN = 39 0.7%

CC IN = 136.7 3.1% 24.4 17.3% IN = 39 0.7%Reading from Dosemeter

127.2 20.0 37.2

experiment Measure 132.5 20.8 38.7Percentage different = [(PB/Measure) * 100] – 100%Percentage different = [(CC/Measure) * 100] – 100%

25 |

Page 26: Pencil Beam and Collapsing Cone

4.2 Discussion

4.2.1 Comparison between Pencil beam (PB) VS Collapsed Cone (CC):

Table 10 shows a comparison between PB and CC data for the First Phantom. As can be

seen in the above table, most of the beam values calculated by the two algorithms show a

variation of between 0 to 3%, except for the DP3 Doses in the case of the RAO and the

lateral beam 3, which show a variation in the range of 5% and 10%, respectively.

Table 10: Comparison of PB and CC algorithms for the First Phantom

Beam ANT Beam 1 RAO Beam 2 LAT Beam 3  PB CC % PB CC % PB CC %

Iso Dose (cGy/Fraction)

106.9 107.9 -0.9 61.1 61 0.1 106.9 105.7 1.1

DP1 Dose (cGy/Fraction)

112.4 112.7 -0.2 65.3 64.7 0.9 98.6 97.3 1.3

DP2 Dose (cGy/Fraction)

80 78.3 2.1 73.3 71.1 3.1 120.6 119.7 0.7

DP3 Dose (cGy/Fraction)

140.8 139.6 0.8 28.4 27 5.1 6.7 6.1 9.8

Figures 14 and 15 demonstrate position of DP 1, DP 2 and DP 3 related to the beams. It is

evident that DP 3 is closer to ANT Beam 1 but away from the RAO Beam 2 and LAT beam

3. The beams 2 and beam 3 reach DP 3 at a tangent. In case of Iso dose there is little

difference in the algorithm for Beam 2 and Beam 3. For DP 1 there is no significant

difference in the algorithm for any of the beams. In case of DP 2 there is a slight variation

in the algorithm for Beam 1 and Beam 2 with a difference of 2.1 % and 3.1 % respectively.

This variation in algorithm is expected and can be explained from the fact that Beam 1 and

26 |

Page 27: Pencil Beam and Collapsing Cone

Beam 2 have to pass through air. DP3 Beam 1 passes through only 3cm of water and no air

giving an accurate algorithm. Whereas in case of DP 3 Beam 2, there is a difference of 5.1

% suggesting that DP3 is situated in the low dose Penumbra. The algorithms are less

accurate in low dose areas with an absolute dose difference of less than 1.5 cGy, (Figure 14

and 15). Beam 3 does not pass through DP3 giving PB and CC algorithms values of 6.7

cGy and 6.1 cGy respectively demonstrating a difference of 9.8 % (Table 10). These

observations are similar to a retrospective treatment planning study conducted by

(ASPRADAKIS et al 2006)3, to evaluate the differences in the dose distributions and

monitor units predicted by CC and PB algorithms. They observed that the calculated dose

in unit density medium was within1% for the CC model and up to 2% for PB. In contrast in

low density medium and under full scatter conditions, CC overestimated the dose by 1%

whereas PB overestimated the dose by 9%. A negative value obtained while calculating the

percent difference is suggestive of a CC dose.

Table 11: A comparison between PB and CC data for the second Phantom

Beam ANT Beam 1 LAO Beam 2 LPO Beam 3  PB CC % PB CC % PB CC %

Iso Dose (cGy/Fraction)

90.5 90 0.5 120.6 119.3 1.08 55.6 55.5 0.1

DP1 Dose (cGy/Fraction)

95.1 94.4 0.7 121.4 121.6 -0.1 50.8 50.9 -0.1

DP2 Dose (cGy/Fraction)

83.5 82.5 1.2 127.5 125.4 1.6 55.6 55.5 0.1

DP3 Dose (cGy/Fraction)

110.4 107.4 2.7 102.0 101.9 0.09 55.0 54.0 1.8

DP4 Dose (cGy/Fraction)

139.3 136.7 1.9 26.4 24.4 8.1 39 39 0.0

27 |

Page 28: Pencil Beam and Collapsing Cone

Table no 11 shows the beam values calculated by the two algorithms. For Iso dose the

variation is not significant for all the beams. In case of DP 1 LAO Beam 2 and LPO Beam

3 show a variation of -0.1%. For DPI 2 all the beams have an accurate algorithm with a

percent difference of 1.2 and 1.6 for Beam 1 and Beam 2. In case of DP3, Beam 1 has a

difference of 2.7 and Beam 3 has a difference of 1. 8 %. In case of DP4 LAO Beam 2,

shows a maximum variation of 8%. This could be explained by the fact that point DP4 is

located at the edge of beam 2 in the penumbra region.

4.2.2 Comparison between algorithms and experimental data:

Phantom 1

In order to investigate the comparative accuracy of the Pencil Beam and Collapsed Cone

algorithms, the percentage differences were calculated by dividing the dose for each

algorithm by the dose calculated from equation 2.2. Figure 18(a – d) illustrates the accuracy

of each algorithm for each beam for First Phantom.

Fig: 18 a: Difference between PB and CC algorithm for Beam 1 (ANT) Phantom 1

28 |

Beam 1 (ANT)

Dose Points

Measured PB Series CC Series PB - CC

ISO -0.028 0.907 -0.935DP1 0.312 0.58 -0.268DP2 2.722 0.539 2.183DP3 1.264 0.402 0.862

ISO DP1 DP2 DP3-0.50

0.51

1.52

2.53

-0.0280.312

2.722

1.2640.907

0.580.5390.402

Beam 1 (ANT)

PB SeriesCC Series

Dose Point measured

perc

enta

ge d

iffer

ence

s

a)

Page 29: Pencil Beam and Collapsing Cone

The difference between PB and CC for Beam 1 (ANT) is maximum at DP2 where it is

found to be 2.183. For rest of the dosage points it is less than 1.

Fig: 18 b: Difference between PB and CC algorithm for Beam 2 (RAO) Phantom 1

The difference between PB and CC for Beam 2 (RAO) ranges from 0.164 to 6.96. It is

maximum at DP3 where it is found to be 6.96 showing a wide variation in the algorithm by

Pencil Beam (PB).

Fig 18 c: Difference between PB and CC algorithm for Beam 3 (LAT) Phantom 1

29 |

ISO DP1 DP2 DP3-100

1020304050

0.3281.9192.417

41.15

0.1640.983 -0.656

34.19

Beam 2 (RAO)

PB SeriesCC Series

Dose Point measuredperc

enta

ge d

iffer

ence

sb) Beam 2 (RAO)

Dose Points

MeasuredPB

SeriesCC

Series PB - CCISO 0.328 0.164 0.164DP1 1.919 0.983 0.936DP2 2.417 -0.656 3.073DP3 41.15 34.19 6.96

ISO DP1 DP2 DP3

-10-8-6-4-2024 2.1012.8362.429

00.9551.4810.66

-8.955

Beam 3 (LAT)

PB SeriesCC Series

Dose Point measured

perc

enta

ge d

iffer

ence

s

c) Beam 3 (LAT)

Dose Points

Measured

PB Series

CC Series PB - CC

ISO 2.101 0.955 1.146DP1 2.836 1.481 1.355DP2 2.429 0.66 1.769DP3 0 -8.955 8.955

Page 30: Pencil Beam and Collapsing Cone

For Beam 3 (LAT), the difference ranges from 1.146 to 8.955. The difference is maximum

for DP 3 (8.955) suggesting a very wide variation in the PB algorithm.

Phantom 2

The same exercise is repeated for Phantom 2, and the graphs are again plotted for Beam 1,

(ANT) Beam 2 (LAO) and Beam 3 (LPO) as shown in Figures 19 a – c.

Fig 19 a: Difference between PB and CC algorithm for Beam 1 (ANT) Phantom 2

The difference between PB and CC for Beam 1 (ANT) ranges from 0.56 to 2.823. It is

maximum at DP3 where it is found to be 2.823 showing a slight variation in the algorithm

by Pencil Beam (PB).

Fig 19 b: Difference between PB and CC algorithm for Beam 2 (LAO) Phantom 2

30 |

Beam 1 (ANT)

Dose Points

Measured

PB Series

CC Series PB-CC

Iso 1.343 0.783 0.56

DP1 0.634 -0.105 0.739

DP2 1.458 0.243 1.215

DP3 3.857 1.034 2.823

Dp4 5.132 3.169 1.963

Iso DP2Dp4-1

1

3

5

1.3430.634

1.458

3.8575.132

0.783-0.1050.243

1.034

3.169

Beam 1 (ANT)

PB SeriesCC Series

Dose Point measuredperc

enta

ge d

iffer

ence

s

a)

Iso DP1 DP2 DP3 Dp4-5

0

5

10

15

20

25

30

2.463-0.41 1.5121.29

26.92

1.359 -0.246 -0.159 1.191

17.3

Beam 2 (LAO)

PB SeriesCC Series

Dose Point measured

perc

enta

ge d

iffer

ence

s

b)

Beam 2 (LAO)

Dose Points Measured

PB Series

CC Series PB-CC

Iso 2.463 1.359 1.104

DP1 -0.41 -0.246 -0.164DP2 1.512 -0.159 1.671DP3 1.29 1.191 0.099Dp4 26.92 17.3 9.62

Page 31: Pencil Beam and Collapsing Cone

Difference for Beam 2 (LAO) is maximum in DP4 with a variation for PB to the tune of

9.62.

Fig 19 c: Difference between PB and CC algorithm for Beam 3 (LPO) Phantom 2

The difference between PB and CC for Beam 3 (LPO) ranges from -0.199 to 2.87, it is

maximum at DP3 where it is found to be 2.87 showing a slight variation in the algorithm by

Pencil Beam (PB).

Using both the phantoms the difference in the algorithm for PB and CC is fairly

large in DP3 (Phantom 1) and DP 4 (Phantom 4) for Pencil Beam. These findings are

consistent with the conclusions of ASPRADAKIS et al (2006)3, who reported that PB

overestimated the dose by 9%. In this experiment it is apparent that PB tends to

overestimate the algorithm and therefore Collapsing Cone is a much preferred algorithm.

Nisbet et al (2004)27, had similar conclusion while comparing the accuracy of Pencil beam

with that of Collapsing Cone. They recommend usage of Collapsing Cone algorithm while

clinical treatment planning situations where lung is present.

31 |

Beam 3 (LPO)

Dose Points Measured

PB Series

CC Series PB-CC

Iso 10.09 9.9 0.19

DP1 1.539 1.738 -0.199

DP2 5.303 5.113 0.19

DP3 57.59 54.72 2.87

Dp4 0.775 0.775 0

Iso DP2Dp4

020406080

10.091.5395.303

57.59

0.7759.9 1.7385.113

54.72

0.775

Beam 3 (LPO)

PB SeriesCC Series

Dose Point measured

Perc

enta

ge d

if-fe

renc

esc)

Page 32: Pencil Beam and Collapsing Cone

Conclusion

1. This study was conducted to compare and contrast two algorithms Pencil Beam (PB)

and Collapsing Cone (CC).

2. Collapsing Cone is found to be more accurate when measured on two phantoms

suggesting that Collapsing cone is a much accurate algorithm for clinical treatment

planning scenario.

3. The experiment clearly shows that Pencil beam tends to overestimates the dose by 9.62

%.

4. Based on this study it is recommended that Collapsing Cone is used as the treatment

algorithm. Conclusions drawn in this study are consisted with findings of other studies.

32 |

Page 33: Pencil Beam and Collapsing Cone

References

[1] Ang, K. & Garden, A. S. 2006. Radiotherapy for Head and Neck Cancers: Indications

and Techniques. United States: Lippincott Williams & Wilkins

[2] Arnfield, MR, CH Siantar, J. Sirbers, et al. 2000. The Impact of Electron Transport on

the Accuracy of Computed Dose. Med Phys 27(6): 1266-74

[3] Aspradakis, M., McCallum, H. M. and Wilson, N. 2006. Dosimetric and Treatment

Planning Considerations for Radiotheraphy of the Chest Wall. The British Journal of

Radiology, 79 (2006), 828-836

[4] Chetty, IJ, B. Curran, JE Cygler, et al. 2007. Report of the AAPM Task Group No. 105:

Issues Associated with Clinical Implementation of Monte Carlo-based Progams (105)

[5] Chin, L. & Regine, W. 2008. Principles and Practice of Stereotactic Radiosurgery.

United States: Springer Science+Business Media, LLC

[6] Cox, J. 2007. Image-Guided Radiotherapy of Lung Cancer. United States: Taylor and

Francis

[7] Fotina, Winkler, Kunzler, Reiterer, Simmat and Georg (2009). Advanced Kernel

Methods VS. Monte Carlo-based Dose Calculation for High Energy Photon Beams.

Radiotherapy and Oncology. 93 (2009) 645-653

[8] Hasenbalg, F., Neuenschwander, H., Mini, R. et al., 2007. Collapsed cone convolution

and analytical anisotropic algorithm dose calculations compared to VMC++ Monte Carlo

simulations in clinical cases. Phys Med Biol, 52. pp. 3679–91.

33 |

Page 34: Pencil Beam and Collapsing Cone

[9] Khan, F. 2010. The Physics of Radiation Therapy. Philadelphia: Lippincott Williams &

Wilkins, a Wolters Kluwer Business

[10] Krieger, T. and Sauer, O. A. (2005). Monte Carlo- versus pencil-beam-/collapsed-

cone-dose calculation in a heterogeneous multi-layer phantom. Phys. Med. Biol. 50 (2005)

859-868

[11] Li, A. (2011). Adaptive Radiation Therapy. United States: Taylor and Francis Group,

LLC

[12] McGarry, C., O’Toole, M. and Cosgrove, V. 2010. Characterising intensity-

modulated radiation therapy (IMRT) software following upgrades in a commercial

treatment planning system. Journal of Radiotherapy in Practice, 9 , pp 209-221

doi:10.1017/S1460396910000130

[13] Papanikolaou, N. and Stathakis, S. 2009. Dose-calculation Algorithms in the Context

of Inhomogeneity Corrections for High energy Photon Beams. Texas: University of Texas

Health Sciences Center

[14] Rong, Y. 2008. The Validation of a Dynamic Adaptive Radiotherapy System. United

States: ProQuest LLC

[15] Schmitter, M. (2007). The Study and Analysis of Heterogeneity Correction

Algorithms Used in the Delivery of Step and Shoot and Sliding Window IMRT. United

States: ProQuest

[16] Schulz, R. & Agazaryan, N. 2011. Shaped-Beam Radiosurgery: State of the Art.

London; New York: Springer Heidelberg Dordrecht

34 |

Page 35: Pencil Beam and Collapsing Cone

[17] Small, W. 2008. Combining Targeted Biological Agents with Radiotherapy: Current

Status and Future Directions. United States: Demos Medical Publishing, LLC.

[18] Spirydovich, S. 2006. Evaluation of Dose Distribution for Targets Near

Inhomogeneities in Photon Beam Radiation Therapy. United States: ProQuest Information

and Learning Company

[19] Tyagi, N., A. Bose, and I. J. Chetty. 2004. Implementation of the DPM Monte Carlo

on a Parallel Architecture for Treatment Planning Applications. Med Bio 50(12): 2779-98

[20] Vanderstraeten, B., W. Duthoy, W. De Gersem, W. De Neve, and H. Thierens. 2006.

[18f] Fluro-deoxy-glucose Positron Emission Tomography ([18F]FDG-PET)

[21] Yue, X. 2007. Regulation of Biochemical Activities and Growth Cone Collapse by

Ephrin-A5 in the Hippocampal Neurons. United States: ProQuest Information and Learning

Company

[22] Ahnesjo, A. 1989. Collapsed cone convolution of radiant energy for photon dose

calculation in heterogeneous media. Medical Physics. 16(4): 577-592.

[23] Thomas, S. 2009. Treatment planning algorithms: Photon and Electron calculations.

Scope, 18(1). pp. 21-27.

[24] Buzdar, S. A., Afzal, M. and Todd-Pokropek, A. 2010. Comparison of pencil beam

and collapsed cone algorithms, in radiotherapy treatment planning for 6 and 10 mv photon.

J Ayub Med Coll Abbottabad. 22(3), 152:154.

[25] Dobler, B. et al. 2006. Optimization of extracranial stereotactic radiation therapy of

small lung lesions using accurate dose calculation algorithms. Radiation Oncology. 1(45).

35 |

Page 36: Pencil Beam and Collapsing Cone

[26] Pearson, M., Atherton, P., McMenemin, R., McDonald, F., Mazdai, G., Mulvenna, P.

and Lambert, G. 2009. The Implementation of an Advanced Treatment Planning Algorithm

in the Treatment of Lung Cancer with Conventional Radiotherapy. Clinical Oncology. 21:

168-174. doi:10.1016/j.clon.2008.11.018

[27] Nisbet, A. et al., 2004. Dosimetric verification of a commercial collapsed cone

algorithm in simulated clinical situations. Radiotherapy and Oncology. 73: 79-88.

[28] United States Cancer Statistics: 1999–2007 Incidence and Mortality Web-based

Report. Centers for Disease Control and Prevention, National Program of Cancer

Registries (NPCR).

[29] Jemal, A., Siegel, R., Ward, E., Murray, T., Xu, J. and Thun, M. J., 2007. Cancer

Statistics, 2007. CA: A Cancer Journal for clinicians. 57:43-66

[30] Shiu, A, S., Hogstrom, K, R. 1991. Pencil‐beam redefinition algorithm for electron

dose distributions. Med. Phys. 18, 7

[31] Ahnesjö A 1989 Collapsed cone convolution of radiant energy for

photon dose calculation in heterogeneous media Med. Phys. 16 577-592.

[32] Ahnesjö A, Saxner M and Trepp A 1992a A pencil beam model for

photon dose calculations. Med. Phys. 19 263-273.

[33] Knöös T, Nilsson M and Ahlgren L 1986 A method for conversion of

Hounsfield number to electron density and prediction of macroscopic

pair production cross-sections Radiother. Oncol. 5 337-345.

36 |

Page 37: Pencil Beam and Collapsing Cone

[34] Nucletron. Oncentra Master Plan v 1.5 SP1. Physics Reference

Manual. CE (Conformité Européenne), UK. REF 192.724 ENG-02

[35] Fotina, I. et al., 2011. Clinical comparison of dose calculation using the enhanced

collapsed cone algorithm vs. a new Monte Carlo algorithm Strahlenther Onkol, 187. pp.

433–41.

[36] Wilcox, E. E. et al., 2010. Comparison of planned dose distributions calculated by

Monte Carlo and Ray-Trace algorithms for the treatment of lung tumors with CyberKnife:

A preliminary study in 33 patients. International Journal of Radiation Oncology, Biology

& Physics, 77(1). pp. 277-284.

[37] A Aspradakis, M. M., Morrison, R. H., Richmond, N. D. & Steele, A., 2003.

Experimental verification of convolution/superposition photon dose calculations for

radiotherapy treatment planning. Physics in Medicine and Biology, (48)17. pp. 2873-2893.

[38] Almond, P. et al., 1999. AAPM’s TG-51 protocol for clinical reference dosimetry of

high-energy photon and electron beams. Journal of Medical Physics, 26(9). pp. 1847-1870.

[39] Fippel, M., 1999. Fast Monte Carlo dose calculations for photon beams based on the

vmc electron algorithm Medical Physics, 26. pp. 1466–75.

[40] Haryanto, F., Fippel, M., Bakai, A. et al., 2004. Study on the tongue and groove effect

of the Elekta multileaf collimator using Monte Carlo simulation and film dosimetry.

Strahlenther Onkol, 180. pp. 57–61.

37 |

Page 38: Pencil Beam and Collapsing Cone

[41] Heintz, B., Hammond, D., Cavanaugh, D. & Rosencranz, D., 2006. SU‐FF‐T‐09: A

Comparison of MatriXX, MapCHECK and Film for IMRT QA: Limitations of 2D

Electronic Systems. Medical Physics, 33(6). pp. 2052-2054.

[42] Korreman, S. S., Pedersen, A. N., Nottrup, T. J., Specht, L. & Nystrom, H., 2005.

Breathing adapted radiotherapy for breast cancer: comparison of free breathing gating with

the breath-hold technique. Radiother Oncol, 76. pp. 311–8.

[43] Muir, B. R. & Rogers, D. W. O., 2010. Monte Carlo calculations of kQ, the beam

quality conversion factor. Medical Physics, 37. pp. 5939-5950.

[44] Pallotta, S., Marrazzo, L. & Bucciolini, M., 2007. Design and implementation of a

water phantom for IMRT, arc therapy, and tomotherapy dose distribution measurements.

Medical Physics, 34(10). pp. 3724-3731.

[45] Pedersen, A. N., Korreman, S., Nystrom, H. & Specht, L., 2004. Breathing adapted

radiotherapy of breast cancer: reduction of cardiac and pulmonary doses using voluntary

inspiration breathhold. Radiother Oncol, 72. pp. 53–60.

[46] Richmond, N.D., Turner, R., Dawes, P.D.K., Lambert, G.D. & Lawrence, G.L., 2003.

Evaluation of the dosimetric consequences of adding a single asymmetric or MLC shaped

field to a tangential breast radiotherapy technique. Radiother Oncol, 67. pp. 165–70.

[47] White, D.R., Martin, R.J. & Darlison, R., 1977. Epoxy resin based tissue substitutes.

The British Journal of Radiology, 50(599). pp. 814–821.

38 |

Page 39: Pencil Beam and Collapsing Cone

[48] Wilkinson, M. & Aspradakis, M.M., 2004. A study of electronic compensation

methods in breast radiotherapy planning. In: IPEM Annual Scientific Meeting. York, UK:

IPEM.

[49] Seaby, A.W., Thomas, D. W. , RYDE, S. J. S. , Ley, G. R. and Holmes D, Design of amultiblock phantom for radiotherapy dosimetry applications. The British Journal ofRadiology, 2002. 75: p. 56–58.

39 |