P2031-KP Oxygen (O 2 ) Monitor SMART MONITOR OPERATION MANUAL (KEYPAD VERSION 20D)

Oxygen Sensor Manual

Embed Size (px)

Citation preview

Page 1: Oxygen Sensor Manual

P2031-KP Oxygen (O2) Monitor

SMART MONITOR

OPERATION MANUAL

(KEYPAD VERSION 20D)

Page 2: Oxygen Sensor Manual

2

MONITOR CONFIGURATION: CONFIG MODE While the Smart Monitor is in the Normal Gas Monitoring Mode, by pressing and holding the [CONFIGURE/4] Key for two seconds, the operator can access the configuration programming, CONFIG MODE, for the monitoring device. Provided a correct User Access Code is entered, a list of the configuration functions can be accessed that define all of the necessary programming parameters and settings for the monitor. A System Access Code of “111” may be used at any time that the program asks for the Access Code. Once the correct User Access Code is entered, the message “CONSPECV20D” will continuously scroll across the liquid crystal display (LCD). The operator should press the [CONFIGURE/4] Key once to confirm. The LCD will show “000” and wait for the operator to enter the desired configuration function code. The operator can use the Keys [1], [2] or [3] to increment/decrement the LCD display to the desired function code number. Key [4] can be used to toggle the increment/decrement mode. Below is a list of the Configuration Functions and associated Code Numbers:

CODE FUNCTION

001 Exit Back to Normal Gas Monitoring Mode

002 Change Address

(Only Necessary on Monitors Connected to Conspec Central Control System)

003 Change Monitoring Range

004 Change Calibration Input (Percent (%) of Full Scale Range)

005 Change Auto-Cal Duration (Time in Minutes)

006 Change “LO” Alarm Level (1st Alarm Set Point)

007 Change “LO-LO” Alarm Level (2nd Alarm Set Point)

008 Change Alarm Silence Duration (Time in Minutes – Factory Set at 20 Minutes)

009 Change Sensor-Fail Alarm Output Option

010 Change Alarm Latch Option

011 Manually Calibrate Gas Input

012 Not Used

013 Manually Calibrate Analog Output

014 Not Used

015 View Raw Analog to Digital Converter (ADC) Data

016 Reset/Restart O2 Monitor

017 Change User Access Code (Factory set at “001")

018 Test EEPROM (Conspec Personnel Use Only)

019 Test Internal Timer (Conspec Personnel Use Only)

All other codes (Including 012 and 014) will cause the Oxygen (O2) Monitor to return to the beginning of the CONFIG MODE where the message “CONSPECV20D” will continuously scroll across the LCD. If no keys are pressed for two minutes, the monitor will return to the Normal Gas Monitoring Mode automatically.

Page 3: Oxygen Sensor Manual

DETAILS OF CONFIGURATION FUNCTIONS

001 When this function is selected, the Oxygen (O2) Monitor will first verify the validation of the Alarm Settings. Specifically, checks that the LO-LO ALARM setting is less than the LO ALARM setting, otherwise the previous alarm settings will be restored. The Oxygen (O2) Monitor will then perform a save function to store all the parameters and settings into memory (EEPROM). When completed, the Oxygen Monitor will return to the Normal Gas Monitoring Mode.

002 When this function is selected, the message “ENTER ACCESSOR ADDRESS” will scroll

across the LCD. Press Key [4] to confirm. The LCD will then show the current address of the Monitor. Use the Keys [1], [2] or [3] to increment/decrement the address setting, the Key [4] can be used to toggle the increment/decrement mode. When finished, press the Key [4] twice to confirm.

003 When this function is selected, a message “ENTER INPUT RANGE” will continuously scroll

across the LCD. Press the Key [4] to confirm. The LCD will then show the current monitoring range. Use the Keys [1], [2] or [3] to increment/decrement the monitoring range, the Key [4] can be used to toggle the increment/decrement mode. When finished, press the Key [4] twice to confirm. The Smart Monitor will then return to the beginning of the CONFIG MODE where the message “CONSPECV20D” continuously scrolls across the liquid crystal display LCD.

004 When this function is selected, the message “ENTER CALIBRATE INPUT” will

continuously scroll across the LCD. Press the Key [4] once to confirm, and the LCD will show the current calibration input value (as a Percent of Full Scale Gas). Use the Keys [1], [2] or [3] to increment/decrement the calibration gas value, the Key [4] can be used to toggle the increment/decrement mode. The Calibration Input should be set as a Percentage of SCALE GAS (span gas) being used to FULL SCALE GAS (sensor range), rounded up to the nearest percent. For example, the input value of “084” [(20.9%/25%)=83.6%=84%] would be used if 20.9% O2 span gas were being used. When finished, press the Key [4] twice to confirm. If an invalid calibration input value is entered (i.e., the calibration input is outside the monitoring range), the message “BAD INPUT” will scroll across the LCD once and the previous calibration input value will be restored, the Smart Monitor will then return to the beginning of the CONFIG MODE where the message “CONSPECV20D” continuously scrolls across the LCD.

005 When this function is selected, the message “ENTER AUTO-CALIBRATION DURATION”

will scroll across the LCD. Press Key [4] once to confirm, and the LCD will show the current auto-calibration duration setting in minutes. Use Keys [1], [2] or [3] to enter the desired duration setting. Use Key [4] to toggle the increment/decrement mode. When finished, press Key [4] twice to accept. The Oxygen (O2) Monitor will then return to the beginning of CONFIG MODE where the message “CONSPECV20D” will scroll across the LCD.

006 When this function is selected, the message “ENTER LO ALARM” will scroll across the

LCD. Press Key [4] once to confirm. The LCD will then show the current LO Alarm setting. Use Keys [1], [2] or [3] to enter the LO ALARM setting. Use Key [4] to toggle the increment/decrement mode. When finished, press Key [4] twice to accept. If an invalid LO ALARM level is entered (i.e., the LO ALARM Level is outside the monitoring range), the message “BAD INPUT” will scroll across the LCD once and the previous LO ALARM setting will be restored. The monitor will then return to the beginning of CONFIG MODE where the message “CONSPECV20D” will scroll across the LCD.

Page 4: Oxygen Sensor Manual

4

007 When this function is selected, the message “ENTER LO-LO ALARM” will scroll across the LCD. Press Key [4] once to confirm. The LCD will then show the current LO-LO ALARM setting. Use Keys [1], [2] or [3] to enter the desired LO-LO ALARM setting. Use Key [4] to toggle the increment/decrement mode. When finished, press Key [4] twice to accept. If an invalid LO-LO ALARM is entered (i.e., the LO-LO ALARM setting is outside the monitoring range or greater than the LO ALARM setting), the message “BAD INPUT” will scroll across the LCD once and the old LO-LO ALARM setting will be restored. The Oxygen Monitor will then return to beginning of CONFIG MODE where the message “CONSPECV20D” will scroll across the liquid crystal display (LCD).

008 When the Oxygen (O2) Monitor goes into LO-LO ALARM, it will turn “ON” the Alarm Output.

Pressing the [SILENCE/3] Key once will temporarily disable the Alarm Output. The duration (in minutes) of the silencing time can be set by selecting this function. The message “ENTER ALARM QUIET DURATION” will scroll across the LCD. Press Key [4] once to confirm. The LCD will show the current Alarm Quiet Duration setting (Factory set at 20 Minutes). Use Keys [1], [2] or [3] to enter the new duration time (minutes). Use Key [4] to toggle the increment/decrement mode. If “000" is entered as the Alarm Quiet Duration, once the [SILENCE/3] Key is pressed the Alarm Output will be silenced as long as the monitor remains in an Alarm Condition. If a value between “001" and “255" is entered as the Alarm Quiet Duration, the Alarm Output will stay quiet for that number of minutes after the [SILENCE/3] Key is pressed. If, after the set duration, the monitor remains in an Alarm Condition, the alarm output will be turned “ON” again. If the Oxygen Monitor returns to the “NORMAL” condition before the duration expires, the timer will be reset. When finished, press Key [4] twice to accept. The Oxygen Monitor will return to the beginning of CONFIG MODE where the message “CONSPECV20D” will scroll across the LCD.

009 This option allows the operator to Enable or Disable the monitor to turn “ON” the LO-LO

ALARM output in case of a sensor-fail condition. When this function is selected, the message “SENSOR FAIL ENABLE 1=YES or 0=NO” will scroll across LCD. Press the [4] Key once to confirm. The LCD will show the current setting. Use Key [3] to enter the setting. Use Key [4] to toggle the increment/decrement mode. If “000” is entered, a sensor-fail condition will not trigger the Alarm Output. If “001” is entered, a sensor-fail condition will enable the monitor to turn “ON” the Alarm Output. When finished, press Key [4] twice to accept. The monitor will return to CONFIG MODE where the message “CONSPECV20D” will scroll across the LCD.

010 This option allows the operator to Enable or Disable the latching of the LO-LO ALARM

output. When selected, the message “LATCH ENABLE 1=YES or 0=NO” will scroll across the LCD. Press Key [4] once to confirm. The LCD will show the current setting. Use Key [3] to increment/decrement mode. If “000" is entered, the alarm output will not be latched, i.e., when the Oxygen (O2) Monitor returns to the “NORMAL” condition, the Alarm Output will not be latched and the alarm will be turned “OFF”. If “001" is entered, the alarm output will be latched, i.e., when the Oxygen (O2) Monitor returns to the “NORMAL” condition, the Alarm Output will be latched and the alarm will remain “ON”. In this mode, LATCH ENABLED (”001”), only by pressing the [SILENCE/3] Key will you be able to turn “OFF” the Alarm Output. This is not applicable for the sensor-fail alarm, no latching is available even if this setting is set to “001”. When finished setting this option, press Key [4] twice to accept. The Oxygen Monitor will return to the beginning of CONFIG MODE where the message “CONSPECV20D” will scroll across the LCD.

011 This function allows the operator to manually calibrate the P2031-KP, Oxygen (O2) Monitor.

Please refer to the section “MANUAL CALIBRATION” for details.

Page 5: Oxygen Sensor Manual

012 Not Used 013 This function allows the operator to define the analog output range. Please refer to the

section “CALIBRATE ANALOG OUTPUT” for calibration instructions. 014 Not Used 015 This function will display the Raw Analog to Digital Converter (ADC) data. Press Key [4]

once when finished to return to the beginning of CONFIG MODE. 016 This function will cause the P2031-KP, Oxygen (O2) Monitor to save all parameters to the

onboard EEPROM chip, then reset the monitor to its start-up condition. 017 When this function is selected, the message “CHANGE ACCESS CODE” will scroll across

the LCD. Press Key [4] once to confirm. The LCD will then show the current User Access Code (Factory set at “001"). NOTE: The System Access Code of “111" may be used if the User Access Code is forgotten. Use Keys [1], [2] or [3] to enter the new User Access Code. Use Key [4] to toggle the increment/decrement mode. When finished, press Key [4] twice to accept. If an invalid User Access Code is entered (i.e., a User Access Code greater than 255), then the message “BAD INPUT” will scroll across the LCD once and the old Access Code will be restored to memory. The Oxygen (O2) Monitor will then return to the beginning of CONFIG MODE where the message “CONSPECV20D” will scroll across the liquid crystal display (LCD).

018 This function is used to test the integral EEPROM chip. 019 This function is used to test the Internal Timer. NOTE: Whenever an Invalid entry is entered and the message “BAD INPUT” is displayed, the new

entry will be discarded and the original data will be restored to memory.

Page 6: Oxygen Sensor Manual

6

AUTO CALIBRATION A soft touch keypad and liquid crystal display (LCD) is provided on the front cover of the enclosure for calibrating the P2031-KP, Oxygen (O2) Monitor. Regular calibration and testing is mandated to ensure the sensor is operating properly. To calibrate the Oxygen (O2) sensor, follow the procedures below:

1. Press the [CALIBRATE/5] Key and hold until the display scrolls the message “ENTER ACCESS CODE”. Press the [4] Key one time, to stop the scrolling of “ENTER ACCESS CODE”. Three Zeros will appear on the display “000” The program is waiting for the input of the User Access Code from the keypad. Press Key [3] for the ones Press Key [2] for the tens Press Key [1] for the hundreds Press Key [4] one time to reverse the number Press Keys [3], [2] or [1] to change the number The default User Access Code from the factory is “001”. After the Access Code is entered, press the [4] Key two times to accept.

2. The RED “SPAN” LED turns ON and will stay on for the duration of SPAN calibration.

3. Inject the span gas (typically 20.9% oxygen, O2) into the sensor via the calibration fitting at a

flow rate of 0.5 LPM.

4. Maintain the flow of the span gas to the sensor for as long as the RED "SPAN" LED is turned ON.

5. After the pre-defined span calibration period, the RED "SPAN" LED will turn OFF, and the

GREEN “ZERO” LED will turn ON.

6. Remove the span gas, and inject the zero gas (100% nitrogen, N2) into the sensor via the calibration fitting at a flow rate of 0.5 LPM.

7. Maintain the flow of the zero gas to the sensor for as long as the GREEN “ZERO” LED is

turned ON.

8. If the calibration procedure is performed properly, the GREEN “ZERO” LED will turn OFF after the pre-defined calibration period. The new calibration data will automatically be stored to the EEPROM for future reference and the monitor will return to the normal gas-monitoring mode.

9. If the new calibration data differs from the previous data by more than 20%, both the "SPAN"

and "ZERO" LEDs will be turned “ON” for approximately five seconds. This indicates that the calibration was not performed properly or the sensor cell has deteriorated beyond calibration standards. The message “BAD INPUT” will scroll across the LCD and the previously stored data will be restored to memory.

10. An unaccepted calibration is normally due to either using the incorrect scale gas or the sensor

cell has become defective and is providing less than 75% of the previous calibration data. In the second case, the sensor cell should be replaced and the unit should be re-calibrated to ensure proper operation.

Page 7: Oxygen Sensor Manual

MANUAL CALIBRATION: FUNCTION “011” CAUTION: Do not enter Function “011” unless an actual gas calibration is to be performed. Stepping through Function “011” without applying the proper calibration gasses will render the device inoperable until a full gas calibration is completed.

Press the [4] Key and hold until the display starts scrolling, “ENTER ACCESS CODE”. Press the [4] Key one time, to stop the scrolling of “ENTER ACCESS CODE”. Three zeros will appear on the display “000” The program is waiting for an input from the keypad. Press Key [3] for the ones Press Key [2] for the tens Press Key [1] for the hundreds Press Key [4] one time to reverse the number Press Keys [3], [2] or [1] to change the number The default Access Code from the factory is “001” After Access Code is entered, press the [4] Key two times to accept. Press the [4] Key one time to stop the scrolling of “CONSPECV20D” Three zeros will appear on the display “000” Press Key [3] one time to make “001” on the display Press Key [2] one time to make “011” on the display Press Key [4] two times to accept. The message “ZERO INPUT” continuously scrolls across the LCD. Press the [4] Key one time to confirm “ZERO INPUT”.

The display will show the current ADC data, apply the zero gas (100% nitrogen (N2) gas) to the sensor via the calibration fitting at a flow rate of 0.5 LPM. Allow the displayed value to stabilize. This will take approximately two to three (2-3) minutes.

Press the [4] Key one time to accept the Zero Scale Gas reading.

Make sure the zero gas (nitrogen) is off.

The message “SCALE INPUT” and “FULL SCALE INPUT” scrolls across the LCD. Press the [4] Key once and select the correct gas input type as it scrolls across the

LCD. Use “SCALE INPUT” for any span gas less than full-scale gas.

Example: For a typical 2031-KP, Oxygen (O2) Monitor set at a monitoring range of 0-25%: 25% O2 is FULL SCALE GAS so select the “FULL SCALE INPUT” function. 20.9% O2 is SCALE GAS so select the “SCALE INPUT” function.

NOTE: Function “004” Calibration Input, should be set to the percentage of SCALE GAS (concentration of gas being used) to FULL SCALE GAS (sensor full scale range) rounded up to the nearest percent. For example, the input value of “084” [(20.9%/25%)=84%] would be used if 20.9% O2 span gas were being used as in the above example.

The display now shows the current ADC data value, apply the span gas to the sensor via a calibration fitting at a flow rate of 0.5 LPM. Wait until the value has stabilized, this will take approximately two to three (2-3) minutes. Press Key [4] one time to accept the reading.

Remember to shut off the span gas. Press Key [4] one time to stop the scroll of “CONSPECV20D” Three zeros will appear on the display “000”

If you want to change or view another function, enter the new number now, if you are finished, Press Key [3] to display for “001”, this is the code to exit to normal gas monitoring

mode. Press the [4] Key two times to exit.

Page 8: Oxygen Sensor Manual

8

CHANGE ACCESSOR ADDRESS: FUNCTION CODE “002”

Press [4] and hold until the display starts to scroll “ENTER ACCESS CODE”. Press [4] one time to stop the scroll of “ENTER ACCESS CODE”. Three zeros will appear on the display “000”. The program is waiting for an input from the keypad. Press Key [3] for the ones Press Key [2] for the tens Press Key [1] for the hundreds Press Key [4] one time to reverse the number Press Keys [3], [2] or [1] to change the number The default Access Code from the factory is “001” After the User Access Code is entered, press the [4] Key two times to accept. Press Key [4] one time to stop the scroll of “CONSPECV20D” Three zeros will appear on the display “000” Press Key [3] two times to display “002” on the LCD. Press Key [4] two times to accept this input Press Key [4] one time to stop the scrolling of “ENTER ACCESSOR ADDRESS”. Press Keys [3], [2] or [1] to change the Accessor Address. If you do not exit here, the monitor will not exit automatically, and this unit will not send the correct reading to the central station. After the new Accessor Address is entered, Press Key [4] two times to accept the new address. Press Key [4] one time to stop the scroll of “CONSPECV20D” Three zeros will appear on the display “000” If you want to change or view another function, punch in the number now, if you are finished, Press Key [3] for one “001`” this is the code to exit. Press Key [4] two times to exit

Page 9: Oxygen Sensor Manual

9

ANALOG OUTPUT CALIBRATION: FUNCTION CODE “013”

Note: the Analog Output Option is not available on late model, smart monitor units ordered for Conspec Trunk Communication Systems. When the 0xygen (O2) Monitor is first installed, if the Analog Output feature is to be used, it is required to set up the Analog Output ranges which represent the Zero Scale and Full Scale gas readings. Follow the instructions below for set-up of these two ranges:

1. In normal gas monitoring mode, press Key [CONFIGURE/4] and hold for two (2) seconds, the message “ENTER ACCESS CODE” scrolls across the LCD. Press Key [4] once to confirm. The LCD then shows “000”, waiting for operator to enter the User Access Code (factory set at “001“). Use Keys [1], [2] or [3] to enter the user access code. Key [4] can be used to toggle the increment/decrement mode. When finished, press Key [4] twice to accept. The message “CONSPECV20D” continuously scrolls across the LCD. Press Key [4] once to confirm.

2. LCD will then display “000”, waiting for the operator to select the Configuration Function. Use

Key [2] and [3] to select function “013”. Press Key [4] twice to Accept.

3. The message “CALIBRATE ZERO SCALE ANALOG OUTPUT” will scroll across the LCD. Press Key [4] once to confirm

4. The LCD will then show the current Zero Scale Analog Output Setting. Connect an ammeter

with the RED (+) probe to the Terminal Block marked “SIG”, and the BLACK (-) probe to the Terminal Block marked “COM”. Notice that the ammeter will show the Analog Output that corresponds to the current setting.

5. Use the Keys [1], [2] or [3] to enter the Zero Scale Output setting. Key [4] can be used to toggle

the increment/decrement mode. Note the reading changes on the Ammeter as the setting is changed. When the ammeter shows the desired Analog Output, which represents the Zero Scale reading, i.e. 4mA, press Key [4] twice to accept.

6. The message “CALIBRATE FULL SCALE ANALOG OUTPUT” will scroll across the LCD.

Press Key [4] once to confirm. The LCD will then show the current setting for the Full Scale Analog Output. Use Keys [1], [2], or [3] to enter the Full Scale Analog Output Setting. Key [4] can be used to toggle the increment/decrement mode. Note the reading changes on the ammeter as the setting is changed. When the ammeter shows the desired Full Scale Analog Output reading, i.e. 20mA, press Key [4] twice to accept.

7. The monitor will return to the beginning of the CONFIG MODE where the message

“CONSPECV20D” will scroll across the LCD.

8. To Exit, press Key [4] once to confirm. Then, use Key [3] to enter the Exit Code, “001” and press Key [4] twice to accept. The unit will return to the normal gas-monitoring mode.

Page 10: Oxygen Sensor Manual

10

STARTUP OPERATION Once the SMART MONITOR is powered up, there is a one-minute idle startup time during which there will be no display on the liquid crystal display (LCD) and no communications to the Primary System. After the idle time, the SMART MONITOR will be in the NORMAL GAS MONITORING MODE. It is recommended that the operator review each CONFIGURATION FUNCTION to ensure all the parameters and settings are correct. In addition, it is recommended to perform a manual calibration the first time the unit is installed.

KEY FUNCTIONS IN NORMAL GAS MONITORING MODE [REVIEW / 1] KEY (HOLD FOR TWO (2) SECONDS) VIEW SETUP CONFIGURATION AND CALIBRATION DATA [SILENCE / 3] KEY SILENCE LO-LO ALARM [CONFIGURATION / 4] KEY (HOLD FOR TWO (2) SECONDS) ENTER INTO CONFIG MODE [CALIBRATION / 5] KEY (HOLD FOR TWO (2) SECONDS) INITIALIZE AUTO-CALIBRATION (AFTER PASSWORD IS ENTERED)

KEY FUNCTIONS DURING CONFIG MODE

[REVIEW / 1] KEY INCREMENT/DECREMENT THE 100TH DIGIT [2] KEY INCREMENT/DECREMENT THE 10TH DIGIT [SILENCE / 3] KEY INCREMENT/DECREMENT THE 1ST DIGIT [CONFIGURATION / 4] KEY CONFIRM OR TOGGLE INCREMENT/DECREMENT MODE OR PRESS TWICE TO ACCEPT DATA

CONNECTORS J1 SPEED BOARD ADAPTER CONNECTION J2 KEYPAD CONNECTION

J3 AUTO CALIBRATION AND ALARM OUTPUTS J5 ANALOG OUTPUT

COM COMMON (NOT USED) PIN 1 SIGNAL

OUT1 ZERO GAS-INPUT PIN 2 V+

OUT2 SPAN GAS-INPUT PIN 3 COMMON

OUT3 (LO-LO ALARM) J6 ANALOG INPUT

OUT4 (LO ALARM) PIN 1 SIGNAL

W2 JUMPER INSTALLED (STATUS LED POWER) PIN 2 V+

PIN 3 COMMON

Page 11: Oxygen Sensor Manual

11

IMPORTANT

IF THIS UNIT IS INSTALLED IN AN INTRINSICALLY SAFE (I.S.) AREA, THE JUMPER “W4" MUST BE ON. FAILURE TO COMPLY WILL RESULT IN ERRATIC READINGS FROM THE MONITOR. THIS JUMPER “W4" IS LOCATED ON THE P2065 BOARD IN THE UPPER-RIGHT-HAND CORNER OF THE BOARD (CIRCLED).

WHEN THIS UNIT IS NOT IN SERVICE, REMOVE THE CONNECTION FROM THE J2 SOCKET. FAILURE TO DO SO WILL RESULT IN DISHARGING OF THE BATTERY. IF THE BATTERY IS DEAD, THE (NO) CELL WILL TAKE LONGER TO STABLIZE.