20
SPECTRO 2000 RS & SPECTRO 2000 RSP OPERATING MANUAL

Operating Manual 2000 RS & 2000 RSP

Embed Size (px)

Citation preview

Page 1: Operating Manual 2000 RS & 2000 RSP

SPECTRO 2000 RS

&

SPECTRO 2000 RSP

OPERATING

MANUAL

Page 2: Operating Manual 2000 RS & 2000 RSP

2

Thank You Very Much For Your Purchase Of This High Performance Instrument. This

Manual Will Introduce The Automatic SPECTRO 2000 RS & SPECTRO 2000 RSP

Microprocessor Spectrophotometer. Before Use, Please Read This Manual Carefully.

IMPORTANT

This Spectrophotometer Is Thoroughly Checked And Inspected Before Shipment. The

Voltage Power Switch At The Back Of The Spectro Has Been Set for 110V. 60Hz. (U.S.A.

Voltage Power). If The Voltage Power In Your Country Is 220V. 50Hz., You Can Convert

It, By Switching The Voltage Switch At The Back Of The Spectro To 220V.. Before Any

Action Or Use, Check The Spectro & Switch It to 220V., If Necessary. Use This

Procedure To Avoid Any Damage To The Instrument.

TABLE OF CONTENTS

Table Of Contents: Page: 2

Specifications: 3

Main Application & Installation 4

Abbreviations: 5

Spectro 2000 RS & Spectro 2000 RSP Front Panel: 6

Instructions To Use: 7

Operating Manual: 8

Example A: 12

Example B: 13

Example C: 14

Example D: 15

Example E: 16

Example F: 17

Page 3: Operating Manual 2000 RS & 2000 RSP

3

SPECIFICATIONS

Wavelength: 325 - 1100nm

BandPass: 6nm

Stray Light: 0.1%T At 340nm And 400nm

Wavelength Repeatability: + 0.5nm

Wavelength Accuracy: +1nm

Light Source: Tungsten Halogen Lamp

for 1000 hours

Grating Mirror: 1200 lines / 1mm

Data: Two Lines

Link To Computer & Printer RS232

Power: 110V. 60Hz. or 220V. 50Hz.

Transmittance T%: 1999

Absorption: 2000

Concentration: 2000

Dimension & Weight: 16 x 14 x 9 inches

20 lbs.

Page 4: Operating Manual 2000 RS & 2000 RSP

4

MAIN APPLICATIONS

The Spectro 2000 RS Spectrophotometer Is A High Technology Instrument, That Is Used

For Qualitative & Quantitatif Analysis To Measure Absorption (A), Percent

Transmittance (T), Or Concentration (C) with Wavelength Range of 325 -1100nm.

Spectro 2000 RS Is Best Suited For Various Fields Such As: Clinical Examination,

Biochemistry, Agricultural Research, Petrochemistry, Chemistry, Environmental

Protection, And Quality Control.

Spectro 2000 RS With Soft Key Has A Microprocessor With Memory, And Storage

Functions. It Provides Power Shut-Off Protection. ( It Will Be Shut Off Automatically

When It Is Not In Use)

The Spectro Is Also Equipped With Interface RS232 To Link To An IBM Compatible

Computer And Printer. The Spectro 2000 RSP Has A Built In Printer.

INSTALLATION AND REQUIREMENTS

1. This Instrument Should Be Installed In A Low Corrosive, Fume Free Environment,

With A Firm And Stable Table Or Surface.

2. Before Plugging In The Unit, The Power Should Be Checked At The Back Of The

Spectro 2000 RS, & Spectro 2000 RSP. The Spectro Voltage Power Switch Should Be

Checked To See If It Has To Be Converted According To The Lab. Power. (110V. or

220V. ) To Prevent Damage From The High Tech Instrument, Please Follow This

Procedure.

WARNING!!!

If The Voltage Power In Your Country Is 220V. 50Hz., And It Has Not Been Checked

Or Converted To 220V., Do Not Operate. If It Is Operated With The Wrong Voltage,

The Spectro Will Be Damaged.

Page 5: Operating Manual 2000 RS & 2000 RSP

5

ABBREVIATIONS

A = Absorption

T = Transmittance

C = Concentration

R = Reference

S1 = Sample 1

S2 = Sample 2

S3 = Sample 3

ESC = Escape

CHA = Change

FACTOR = Function & Factor

ACC = Accept

Lamp Energy = Need Lamp Adjustment

EXT Print = External Printer

INT Print = Internal Printer

INIT CONC = Initial Concentration

Page 6: Operating Manual 2000 RS & 2000 RSP

6

Front Panel of Spectro 2000RS and Spectro 2000RSP

Page 7: Operating Manual 2000 RS & 2000 RSP

7

INSTRUCTIONS TO USE

1. First Check The Voltage Switch At The Back Of The Spectro 2000 RS

& Spectro 2000 RSP To be Converted To The Lab. Power, Then Proceed To Plug In The

Power.

2. Turn On Power Switch

3. The Wording On The Display Will Show “Please Wait” (Instrument Is Warming)

The Spectro Will Make Noise During This Process While It Is Being Warmed Up.

4. After The Unit Has Warmed Up, (The Motor Sound has Ended) The Display Will

Show:

550nm Wavelength, A or T: ( Absorption or Transmittance), And R or S1 - S2 - S3

Sample Position.

5. Now You Can Command The Instrument By The 10 Soft Keys

Page 8: Operating Manual 2000 RS & 2000 RSP

8

10 KEYS

1- ATC To Choose Absorbance (A), Transmittance (T), Or Concentration (C)

2- ABC Adjusting 0A And 100%T

3- FACTOR = Function Or Factor Or Menu

4- Print

5- Move Cuvette Forward

6- Move Cuvette Backward

7- Enter = Enter Data

8- Decrease Wavelength

9- Increase Wavelength

10- ESC = Escape

Page 9: Operating Manual 2000 RS & 2000 RSP

9

OPERATING MANUAL

SPECTRO 2000 RS

SPECTRO 2000 RSP

The Spectro 2000 RS And Spectro 2000 RSP Are High Technology Instruments For

Different Analytical Tests Of Spectrophotometry That Can Be Used For Chemistry,

Biochemistry, Water Analysis, Environmental Research, & Medical Diagnostic Tests.

The Spectro 2000 And Spectro 2000 RSP Is A Microprocessor System With A Memory

And Storage Function. It Provides Power Shut Off Protection And Is Equipped With

Standard Interface RS232 For Connection To An IBM Compatible Computer & External

Printer.

Spectro 2000 RS & Spectro 2000 RSP Can Measure Absorption, Transmittance, And

Concentration For 4 Different Samples At One Time. There Are 10 Soft Touch Keys.

There Are 2 Lines On The LCD Display With Up To 31 Characters. The Spectro 2000 RS

& Spectro 2000 RSP Have 10 Keys Each.

There Are 2 Lines On The LCD Monitor. The Upper Line Shows The Current Wavelength

A or T or C Value, And Cuvette Position Reference or S1, S2, S3.

The Lower Line Shows The Menu And The Factor Of C Value, And Function Menu. The

“ESC” and “CHA” and “ACC” Correspond To “Escape”, “Change” & “Accept”, and

By Using “Enter”, Your Selection Is Approved. To Get The Menu, Use The Factor Key.

Before Plugging The Spectro To The Power, Please Check To Be Sure The Switch At The

Back Of The Spectro Is On 110V. 60Hz., Or 220V. 50Hz. (Whatever Your Lab Voltage

Power Is). Please Convert The Power Switch As The First Step.

Page 10: Operating Manual 2000 RS & 2000 RSP

10

1. ABS = Adjusting “0” A And 100%T

2. A/T/C = To Select One Of 3 Positions (A or T or C)

A = Absorption = (Log of T)

T = Transmittance

C = Concentration = (C=AxF)

“C” Has 2 Values MOL/L and MOL/ml

3. Factor or Function

Can get A Menu And Roll The Menu, But Cannot Get Into The Menu Of The Lower Level.

Structure Of 5 Menus Is Explained As Follows: (See Example For How To Set-Up)

(1) PRINT SET-UP 1.1 EXT PRINT

1.2 INT PRINT 1.21 PAPER FEED

1.22 ALLOW PRINT

EXT PRINT Refers To EXTERNAL PRINT; INT PRINT Refers To INTERNAL PRINT.

(2) WAVELEG SET-UP (2.1 INITIAL WAVELG=550nm)

(3) INT CONC 3.1 MOL/L

3.2 MOL/ML

3.3 PPB

3.4 PPM

4. LAMP ENERGY Presented By 8 Black Blocks

5. RS232 SET-UP 5.1 BAUD=9600

5.2 BAUD=4800

5.3 BAUD=2400

5.4 BAUD=1200

Page 11: Operating Manual 2000 RS & 2000 RSP

11

4. PRINT

When The Instrument Initializes Its Set-Up, It Defaults To External Print. If You Want To Change It, You Must Get Into The Menu To Change The Set-Up. Once Properly Set-Up, The Current State Can Be Printed

Out. If It Is At A/T/C Or In Menu. THE NEXT TIME WHEN THE INSTRUMENT TURNS ON, IT WILL

STILL HAVE THE PREVIOUS TIME’S SET-UP. This Means That If The Last Time The Instrument Was

Set At PRINT, This Will Become The New Default Until It Is Changed.

Print Includes Internal Print And External Print. When It Is Set-Up To INT PRINT, You Can Set-Up The

Paper Feed And Allow To Print Items (Refer To “ Example” For How To Set-Up). If It Is Set-Up To Feed

Paper, Touch “PRINT” And The Printer Will Go Two Lines Without Printing (Refer To “Example”). If It

Is Set-Up To Allow Print, Touch “PRINT”, The Printer Will Print Out The Display And Current State

(Refer To “Example”). If It Is In A Menu Of BAUD SET-UP, To Print Current Baud Set-Up Is ON, And

Others Are OFF.

If It Is At A/T/C, Touch PRINT, It Prints Current Wavelength, Cuvette Position And A/T/C Value. If It Is At

C. It Prints Factor, Too. If You Want To Cancel The Current Set-Up Or Select Other Set-Up, You Must

Return To The Menu To Make The Set-Up Again.

5 And 6. ESC And Enter

ESC And ENTER Keys Correspond To Esc And (CHA, ACC) In The Menu. “CHA” Appeared At The

Lower Right Corner Means Whether To Want To Change, “ACC” Means Whether To Want To Accept,

“ESC” Means Whether To Want To Return.

When The Instrument Is In Menu, It Only Responds To Multi-Functional Keys FACTOR, ESC, ENTER And

PRINT Keys. If Any Other Key Is Touched, The Buzzer Will Send Out A Long Beeping Sound. When 2

Lines Are On The Display, The Cuvette Holder Will Not Move. A/T/C & Esc Should Be Touched To Clear

The Second Line. Then The Cuvette Can Be Moved. When It Is At A/T/C, ESC And ENTER Keys Are Close-

Up. If Touched, The Buzzer Will Send Out A Long Beeping Sound. When It Is At Master Menu, To Touch

ESC Key, It Skips Out Of The Menu Set-Up And Return To A/T/C. If It Is At Sub-Menu, To ESC Key, It

Returns To The Menu Of The Lower Level.

When It Is At Master Menu, The Lower Right Corner Of The Display Screen Will Show “CHA”. If The

Enter Key Is Touched, The Sub-Menu Set-Up Will Be Received.

Only At The Lowest Level Of The Master Menu. FACTOR”, will “ACC” Appear, Which Shows Whether

To Want To Select This Item. If ENTER Is Touched, That Means This Item Is Accepted, And At The Same

Time, It Returns To The Menu Of The Upper Level.

Page 12: Operating Manual 2000 RS & 2000 RSP

12

When It Is At FACTOR Of C Value And Initial Wavelength Set-Up Of The Menu, The Lower Right Corner

Of The Screen Will Appear “CHA”.

These Are Possible Options. If The ENTER Key Is Touched, It Can Change The Factor And The Initial

Wavelength. When Changing the Wavelength. Push The Wavelength Button A Few Times.

When the Option Of making a change Is Reached, ENTER Must Be touched to ensure receipt of the data.

If the ESC Key is touched, It will cancel the current set-up and return to the value before making the set-

up.

7 and 8 Multi-Functional

Function 1:When it is at A/T/C, Wavelength can be changed. Range of wavelength is 325nm to 1100nm

Function 2: When it is at Menu, It can roll the Menu.

Function 3: When it is at FACTOR set-up of C and in the initial wavelength of the menu, the value wanted

to set-up will automatically add 1, or subtract 1, yet it must be in “ACC” state, which means only when

“ACC” is appeared at the lower right corner of the screen, can the value add 1 automatically. Otherwise,

the function corresponds to Position 1.

9 & 10 MOVE SAMPLE

When the Instrument is turned on, and after initialization, the cuvette is at the R position. R represents

reference position S1, S2, S3 represents cuvette 1,2 and respectively.

If touch , cuvette position will move R-S1-S2-S3-R at the same time, cuvette position will be displayed on

the screen.

If touch , cuvette position will move S3-S2-S1-R-S3 at the same time, cuvette position will be displayed

on the screen.

When Putting The Reference Cuvette In The First Cuvette Place. 0A And 100%T By A/T/C should be

adjusted.

Page 13: Operating Manual 2000 RS & 2000 RSP

13

STORAGE MEMORIES:

The Microprocessor System Will Store The Memory Of The Spectro 2000 RS & Spectro 2000 RSP, With

The Same Set Up That Was In The Instrument Memory When The Spectro Was Last Turned Off.

When The Instrument Is Turned On Again, All Of The Previous Set Up That Was Last Used Will Still Show

On The Display.

It Means That When The Instrument Was Last Turned Off, If It Was Set Up At INT PRINT, It Will Still Be

Set Up At INT PRINT When The Instrument Is Put On The Next Time, Or If When The Instrument Was

Turned Off, And The Wavelength Was Set Up At 600nm, The Spectro Will Show 600nm The Next Time The

Instrument Is Turned On.

Page 14: Operating Manual 2000 RS & 2000 RSP

14

SUMMARY OF OPERATION

1. Convert The Voltage Power According To The Country Voltage Power (110V. or 220V.)

2. Turn The Power Switch On At The Back Of The Spectro

3. Display Shows: Spectro 2000 “Please Wait”

4. Sounds Of The Spectro Warming Up & Adjusting Will Now Be Heard.

Wait Until The Spectro Is Completely Warmed Up,( THE SOUND WILL STOP) To Adjust The Cuvette

Holder.

5. Display Shows:

6. If Different Wavelength Is Needed: Change The Wavelength By ( )

7. If A or T Is Needed:

Push The Soft Key A/T/C To See On The Display A or T or C.

8. Put The Cuvette With The Reference Solution In The First Cell Holder. The Sample For Testing In The

Cell Holder Should Be Placed In Cell Holder 2, 3, or 4. The Display Will Show R (Reference) Or Either

S2 - S3 - S4, By Pushing The Key S. Move (Sample Move).

9. The Spectro Will Show Setting Blank . When Putting The Reference In The First Cell Holder, Adjust

A.000 And T100.00%. By Touching The Soft Key A/T/C A Few Times.

10. If The Second Line (Menu) Comes To The Screen Unwillingly, Push The A/T/C or ESC Key To Remove

The Second Line.

(Menu) Is The Second Line Of The Display That Exists In The Memory Of The Spectro.

11. Set Up The Factor For Reference Solution.

12. Put The Cell Holder 2, 3, or 4 And Test The Absorption, Transmission, Or Concentration. .

Page 15: Operating Manual 2000 RS & 2000 RSP

15

PRINTING WITH THE SPECTRO 2000 RSP

Example A:

If want to set-up internal print, do the following:

1. Touch F to get into the menu.

2. The first item of the menu is PRINT SET-UP,

3. Touch ENTER to get into print set-up. This item includes internal print and external print, that is

EXT PRINT and INT PRINT,

4. Touch “F” or multi-function key ( up/down ) to get internal print. The internal print includes paper feed

and allow print, that is PAPER FEED and ALLOW PRINT,

5. Touch “F” or multi-function key ( up/down ) to get allow print.

6. Touch ENTER key again, to accept this item.

7. Touch “ESC” a few times to clear the Display.

8. Touch “PRINT” to print information on the Display.

Page 16: Operating Manual 2000 RS & 2000 RSP

16

Example B

If want to set wavelength to 550nm-600nm when the instrument is turned on, do the following:

1. Touch FACTOR to get into the menu.

2. Touch FACTOR or multi-functional key ( up/down ) to get WAVELG SET-UP,

3. Touch ENTER again to set-up the initial wavelength

4. Touch ENTER key again, “CHA” at the lower right corner of the screen, changes to “ACC” which

means to start accept.

5. Memory Storage Touch multi-function key ( up/down ) to get the initial wavelength to 600nm, then

touch ENTER to accept 600nm. Next time when the instrument turns on, it’s wavelength will be at 600nm.

Setting is completed and as it is shown in the figure. At that time, if ESC is touched, it will return to the

inital set-up wavelength 550nm.

Page 17: Operating Manual 2000 RS & 2000 RSP

17

Example C

If want to set-up concentration to MOL/ML, do the following:

1.Touch FACTOR to get into the menu.

2. Touch FACTOR or multi-functional key ( up/down ) to get “INIT CONC”.

3. Touch again ENTER to get the first item of the menu of lower level, that is “MOL/L”

4. Touch FACTOR or multi-function key ( up/down ) to get the second item “MOL/ML”

Touch again ENTER to accept this item. At that time, at the upper line of the display, C is followed by a

unit ml which represents MOL/ML

Page 18: Operating Manual 2000 RS & 2000 RSP

18

Example D

If want to test the lamp energy, do the following:

1. Touch FACTOR to get into the menu.

2. Touch FACTOR or multi-functional key to get “LAMP ENERGY”.

3. Touch again ENTER to get the lamp energy test. At that time, the lower line of the display will appear

several black blocks to represent the lamp energy. There will be A Maximum Of 10 black blocks. In case

“ERR1” appears at the lower line, it means that the lamp energy is too strong. There is no need to change

the lamp position. In case, “ERROR” appears at the lower line, it means that the lamp energy is too weak.

It is necessary to adjust the lamp position, If The Lamp Energy Is Too Weak.

4. To adjust the lamp position use a screwdriver, and push the Lamp back and forth until most of the black

blocks at the lower line reach the color Black completely (Until the Black Square Blocks are completely

colored in). Which means that the position of the lamp has been adjusted to its optimal position.

Page 19: Operating Manual 2000 RS & 2000 RSP

19

Example E

If want to set RS 232 Baud, do the following:

1. Touch FACTOR to get into the menu.

2. Touch FACTOR or multi-functional key ( up/down ) to get “RS232 SET-UP”

3. Touch ENTER to get the Baud selection. The first item is BAUD 9600.

4. Touch FACTOR or multi-function key ( up/down ) to reach the Baud wanted to set.

5. When turn to the BAUD that want to set, touch the ENTER key to select that BAUD Rate..

Page 20: Operating Manual 2000 RS & 2000 RSP

20

Example F

If want to set-up the factor, do the following:

1. Touch A/T/C Key to get C. Factor (0.1-2000)

2. “FACTOR = 1.0” will appear in the lower line of the screen

3. If ENTER key is touched “ACC” appears at the lower right corner of the screen, In order to change the

factor.

4. Touch the multi-function key factor value will add 1 or subtract 1 automatically

5. If reach the value that want to set, touch the ENTER key to accept. The factor set-up is the value that

was accepted. If do not want to accept, touch ESC to return to the previous value.