Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
NPL/IPEM Reference Dosimetry Audits
Russell Thomas
Radiotherapy Standards Users’ Meeting5th June 2007
Background
NPL, at the invitation of IPEM, started conducting reference dosimetry audits in 1995 since then we have conducted:
• 40 MV photon audits• 6 electron audits (1996 code)• 8 electron audits (2003 code)• 6 kV photon audits (first one in 2003)
Audit Format
Basic structure for all reference dosimetry audits:
• Beam quality• Machine output• Field instrument calibration
+ specific areas such as ion recombination+ individual areas of concern
MV photon audit set up
MV photon beam quality indexMV photon beam quality index
Beam quality index (TPR)
0.9800
0.9850
0.9900
0.9950
1.0000
1.0050
1.0100
1.0150
0 10 20 30 40 50 60 70
Audit number
Rat
io N
PL/H
ost NPL/Host
0.998std dev 0.5%
MV photon field instrument calibration
Field Instrument calibration
0.9800
0.9850
0.9900
0.9950
1.0000
1.0050
1.0100
1.0150
1.0200
0 10 20 30 40 50 60 70
Audit number
Rat
io N
PL/
Hos
t
NPL/Host
1.001
std dev 0.7%
MV photon machine output
Machine output (cGy/MU)
0.9850
0.9900
0.9950
1.0000
1.0050
1.0100
1.0150
1.0200
1.0250
0 10 20 30 40 50 60 70Audit number
Ratio
NPL
/Hos
t
NPL/Host 1.003 std dev 0.7%
Nisbet/Host 1.003 std dev 1.0%
Thwaites/Host 1.003 std dev 1.5%
Depth dose measurements in water
Electron measurements in solid water
Electrons 2003 C of P
Machine output
0.994
0.996
0.998
1
1.002
1.004
1.006
1.008
1.01
0 1 2 3 4 5 6A ud it numb e r
Field instrument
0.998
1
1.002
1.004
1.006
1.008
1.01
1.012
1.014
0 1 2 3 4 5 6
Audit number
Ratio
NPL
/Hos
t
NPL/Host 1.003 std dev 0.3%
Nisbet/Host 0.994 std dev 1.8%
NPL/Host 1.002 std dev 0.3%
kV set up for HVL measurements
KV audit results
Beam Quality
0.9
0.95
1
1.05
1.1
0.5 1 1.5 2 2.5 3 3.5 4 4.5
Audit number
Rati
o N
PL
/Ho
st NPL/Host
1.019
std dev 4.8%
KV audit results
Machine output
0.965
0.97
0.975
0.98
0.985
0.99
0.995
1
1.005
1.01
1.015
0 1 2 3 4 5
Audit number
Rat
io N
PL
/Ho
st
Machine output
expected
Field instrument
0.99
0.995
1
1.005
1.01
1.015
0 1 2 3 4 5
Audit number
Ratio
NPL
/Hos
t
NPL/Host 0.999 std dev 0.7% NPL/Host 1.000 std dev 0.5%
What was our goal for 2004-07?So far…..
Region KV photon MV photon Electron 2003
A
B
C
D
E
F
G
H N/A
Helical Tomotherapy
• Clinicians like the dose distributions that tomotherapy can give …
Helical Tomotherapy is …
• … just an “optimised” integrated system to deliver IMRT
• alanine/EPR could be as useful in “ordinary” IMRT
Tomotherapy beam profiles
Open field at isocentre:
length: 40 cm
width: 5 cm, 2.5 cm or 1 cm
Jeraj et al., Med Phys 31 (2004) 396-404
Individual pellet doses (Cromwell)
Measured dose / TPS dose
0.95
0.96
0.97
0.98
0.99
1
1.01
1.02
1.031 7 13 19 25 31 37 43 49 55 61 67 73 79 85 91 97 103
109
115
121
127
133
139
5 cm beam
2.5 cm beam
1 cm beam
off-axis beam
static beam1.0 ± 2%
Current and Future work
• NPL offers alanine measurements in the NPL-tomo phantom as a routine service to any TomoTherapy user, with– on-site measurements by NPL for UK users– Postal service (phantom and dosimeters) for non-UK users
• NPL would like to apply alanine dosimetry to IMRT delivered by conventional linacs– Volunteers, please, to identify and help investigate issues– Will co-ordinate work with the RTSIG working group on IMRT
Where now for NPL Audits?
• DTI funding from October 2007?• Hospitals buying additional audits • Continue with kV, MV and Electron audits• HDR Brachytherapy• Smallfield and IMRT• Improve tie in with individual audit region• National electron audit of all radiotherapy departments• Measuring impact
Staff changes in NPL Radiation Dosimetry
Departures:Martin PalmerRebecca Nutbrown
Secondment:Russell Thomas (2 years)
Maternity Leave:Clare Lee (9 months)
Staff changes in NPL Radiation Dosimetry
Protection Level Calibration Service –MSP: Martin Kelly ext 8502
Deputy: Graham Bass
Therapy Level Photon Calibration Service –MSP: Graham Bass ext 6503
Deputy: Julia Pearce / James Manning
Therapy Level Electron Calibration Service –MSP: Graham Bass ext 6503
Deputy: Julia Pearce
NPL Reference Dosimetry Audits –Graham Bass ext 6503
Graham Bass is Captain Dosimetry!
Thank You!