MultiRAE Using the MultiRAE personal multigas monitor Firmware v 3.10 (rev A)

Multirae user training_310a

  • Upload
    lgray70

  • View
    165

  • Download
    2

Embed Size (px)

Citation preview

Page 1: Multirae user training_310a

MultiRAE

Using the MultiRAE personal multigas

monitorFirmware v 3.10 (rev A)

Page 2: Multirae user training_310a

MultiRAE features Turning on the MultiRAE Recommended Daily Start-up Procedure User displays Alarm modes Programming displays Calibration

Training Agenda:

Page 3: Multirae user training_310a

MultiRAE Features: Sensors

Oxygen: 0-30% Combustibles: 0-100% of LEL Toxic sensors: select two from CO, H2S,

SO2, NO, NO2, Cl2, HCN, NH3, PH3

VOC monitor: 0-2000 PPM

Page 4: Multirae user training_310a

MultiRAE Features: Reliability Extremely rugged for extensive field use Weather Proof Case Temperature range of -4oF to 113oF (-20oC to

45oC) 4-Way power: Li-Ion(NiCad), Alkaline or run

continuously on 110 VAC and 12 VDC Runs 10-12 hours with pump RFI protection against radio interference Intrinsically safe: Class I, Division I, Groups A,

B, C, D LEL Over-Range Protection

Page 5: Multirae user training_310a

MultiRAE Features: Rugged Case

Display

Charging Port

Wrist Strap

Buzzer & Exhaust

Serial Port

Page 6: Multirae user training_310a

Three buttons on a sealed membrane faceplate: • Y/+ (horn): clears & tests alarms• N/- (light): turns on manual backlight for 90

seconds• MODE (on)

MultiRAE Features: Simple Faceplate

Light Sensor

Alarm LEDs

Page 7: Multirae user training_310a

MultiRAE Features: Rugged Boot

Liquid Trap

Rugged Rubber Boot

Belt Clip

Page 8: Multirae user training_310a

MultiRAE Features: Internal Pump Makes remote sampling easy Sample draw over 100 feet! Liquid water trap for added protection Pump stall feature: when moisture is

detected or when pump is blocked the pump will shut off, protecting the MultiRAE from potential damage

~150cc/min at Low flow, ~250cc/min at High flow

Only use Teflon tubing when PID is installed

Page 9: Multirae user training_310a

MultiRAE Features: External Prefilter Always use the prefilter Replace when it looks very dirty Replace when MultiRAE is in pump alarm

with the filter on and you can clear the pump alarm with the filter off.

You wouldn’t leave your house without putting on your pants so

don’t use the MultiRAE without its external filter!

Page 10: Multirae user training_310a

Unplug MultiRAE from charger Hold “MODE” Key to turn on Alarm will beep once Watch display screen for messages such as:

• Sensors installed & their warranty expiration• Alarm limits• Last calibration date• User/Alarm/Datalog modes

Warm-up will take approximately 90 seconds

MultiRAE Start-up: Turning On

Page 11: Multirae user training_310a

If this screen appears during the start-up just acknowledge it with the “Y” key and follow the procedure for setting the clock.

The battery has gone dead and the MultiRAE has detected a mismatch between the date code on the sensors and its internal clock.

MultiRAE Start-up: Warnings TOX1 VOC TOX2

Incorrect YearLEL Check Clock!!! OXY

Page 12: Multirae user training_310a

This is a warning screen. As long as the sensor calibrates properly it is valid to continue using it. Acknowledge by pressing the “Y” key to continue warm-up.

For maximum MultiRAE availability, RAE Systems recommends replacing sensors at the end of their warranty period

TOX1 VOC TOX2

O2 InstalledLEL Warranty Expired OXY

MultiRAE Start-up: Warnings

Page 13: Multirae user training_310a

Multi Start-up: Check Sensor Zero TOX1 VOC TOX2

0 0 0LEL 0-3 20.6-21.2 OXY

After 90 sec. warm-up, the MultiRAE should display for all of the installed sensors

If there is no alarm, and the sensor readings are in the range shown above, then the MultiRAE is ready for use.

Page 14: Multirae user training_310a

If any reading is outside these limits, wait a few minutes then check again.• If readings are in limits use the unit• If readings are still outside these limits

perform “Fresh air calibration”

Multi Start-up: Check Sensor Zero TOX1 VOC TOX2

0 0 0LEL 0-3 20.6-21.2 OXY

Page 15: Multirae user training_310a

MultiRAE Start-up: Lamp Alarm

“Lamp” display along with audible alarm indicates that PID lamp has failed to light

Press “Y/+” key to clear alarm, if it clears MultiRAE is ready for use

If alarm does not clear, wait a few minutes and check again

TOX1 VOC TOX2

0 0.0 0LEL 0 Lamp 20.9 OXY

Page 16: Multirae user training_310a

MultiRAE Start-up: Lamp Alarm

If after a few minutes the “Lamp” message remains, turn off MultiRAE and restart

If after restart “Lamp” message disappears, MultiRAE is ready for use

If after restart “Lamp” message remains, the PID Lamp needs service

TOX1 VOC TOX2

0 0.0 0LEL 0 Lamp 20.9 OXY

Page 17: Multirae user training_310a

3 Display Modes

Program Mode: 12 screens for most advanced users (default mode)

Display Mode: 7 screens and has password to get into programming

Text Mode: 3 screens and has password to get into programming

Page 18: Multirae user training_310a

Comparison of Display Modes

Program (12) Display (8) Text (3)Sensor ID Sensor ID Sensor ID w/OKPEAK PEAK BatteryMIN MIN PC CommSTEL STELTWA TWABattery Print Reading*Time/Date/Temp BatteryStart Datalog* PC CommLEL Gas * Can be turnedVOC Gas off w/ProRAEPrint Reading* SuitePC Comm

Page 19: Multirae user training_310a

MultiRAE: 12 Program Mode Review Instantaneous reading/Sensor NameInstantaneous reading/Sensor Name PeakPeak MinimumMinimum STELSTEL TWATWA Battery VoltageBattery Voltage Date/Time/Run Time/ TemperatureDate/Time/Run Time/ Temperature Datalog Mode/Start DatalogDatalog Mode/Start Datalog Display Gas C.F. LELDisplay Gas C.F. LEL Display Gas C.F. VOCDisplay Gas C.F. VOC Print Reading?Print Reading? Comm. PC?Comm. PC?

Page 20: Multirae user training_310a

Instantaneous Reading (Main Display)

MultiRAE will return to this screen in 60 seconds if left in any subsequent screen

This screen alternates between the instantaneous reading and the sensor name approximately every 3 seconds

TOX1 VOC TOX2

0 0.0 0LEL 0 20.9 OXY

Page 21: Multirae user training_310a

TOX1 VOC TOX2

CO VOC H2SLEL LEL OXY OXY

CO: Carbon Monoxide Sensor VOC: Photo Ionization Detector (PID) H2S: Hydrogen Sulfide Sensor LEL: Combustible gas sensor OXY: Oxygen sensor Tap “MODE” key to proceed

Instantaneous Reading (Main Display)

Page 22: Multirae user training_310a

Peak Reading Display

Holds Highest reading since MultiRAE was turned on

Tap “MODE” key to proceed to next screen

TOX1 VOC TOX2

34 584 11LEL 6 PEAK 20.9 OXY

Page 23: Multirae user training_310a

Minimum Reading Display

Holds Lowest reading since MultiRAE was turned on

Tap “MODE” key to proceed to next screen

TOX1 VOC TOX2

0 0.0 0LEL 0 MIN 13.5 OXY

Page 24: Multirae user training_310a

STEL Display

Short Term Exposure Limit (average for the past 15 minutes)

Displays “***” until MultiRAE has been on for 15 minutes

STEL only is calculated for VOC & toxics STEL alarm beeps once a second Tap “MODE” key to proceed

TOX1 VOC TOX2

0 0.0 0LEL STEL OXY

Page 25: Multirae user training_310a

TWA Display

Time Weighted Average is the accumulated reading of gas concentration divided by 8 hours since the MultiRAE was turned on

TWA only is calculated for VOC & toxics TWA alarm beeps once per second Tap “MODE” key to proceed

to next screen

TOX1 VOC TOX2

0 0.0 0LEL TWA OXY

Page 26: Multirae user training_310a

Battery Voltage Display

MultiRAE shuts down when battery voltage drops below 4.2 volts

Normal Full Charge is over 4.8 volts Tap “MODE” key to proceed to next

screen

TOX1 VOC TOX2

Battery = 5.1VLEL Shut off at 4.1V OXY

Page 27: Multirae user training_310a

Date/Time/Run Time/Temp. Display

Date Time of day Accumulated time in hours and minutes

since the MultiRAE was turned on Temperature in Centigrade (operating

range from -20oC to 45oC) Tap “MODE” key to proceed

TOX1 VOC TOX2

Feb 15, 01 12:00LEL ON=01:22 25oC OXY

Page 28: Multirae user training_310a

Datalog Mode Display

Manual: logs if “Y/+” key is pushed now Automatic: this screen is deleted and it logs

data when as soon as the MultiRAE on Scheduled: logs data at a preset time Will datalog 20,000 points (64 hrs of data) Entering programming pauses

datalogging

TOX1 VOC TOX2

Start Datalog?LEL OXY

Page 29: Multirae user training_310a

Datalog Mode Display

a small “L” is displayed in the left center of the MultiRAE screen indicates datalogging

Will datalog 20,000 points (64 hrs, 5 channels, 1 min interval)

User can choose one of 4 datalog modes (Manual, Automatic, Periodic & Scheduled in ProRAE Suite)

TOX1 VOC TOX2

0 0.0 0LEL 0 L 20.9 OXY

Page 30: Multirae user training_310a

LEL Units Display

Target Gas for LEL displayed here (MultiRAE is currently measuring LEL in units of methane)

Correction Factor can only be changed in programming mode

Tap “MODE” key to proceed to next screen

TOX1 VOC TOX2

LEL gas =LEL Methane OXY

Page 31: Multirae user training_310a

VOC Units DisplayTOX1 VOC TOX2

VOC gas =LEL Isobutylene OXY

Target Gas VOC is displayed here (MultiRAE is currently measuring VOC in units of isobutylene)

Correction Factor can only be changed in programming mode

Tap “MODE” key to proceed to next screen

Page 32: Multirae user training_310a

Print Reading Display

If the optional external printer is attached pressing the “Y/+” key will print out the current reading.

If the optional printer is not installed it will display “wait” for 20 seconds and then return to the “Print Reading?” screen.

This screen can be disabled in ProRAE Suite

TOX1 VOC TOX2

Print Reading?LEL OXY

Page 33: Multirae user training_310a

Communicate with PC Display

If “Y/+” key is pushed then MultiRAE will display “Monitor will Pause. OK?”

If “Y/+” key is pushed again the MultiRAE will display “Ready...” until it receives a signal from the computer or the “MODE” key is pressed.

Tap “MODE” key to proceed

TOX1 VOC TOX2

CommunicateLEL with PC? OXY

Page 34: Multirae user training_310a

Communicate with PC Display

When the MultiRAE is in this communication standby mode it stops monitoring gas concentrations and stops datalogging.

Datalogging must be manually restarted when exiting unless automatic datalogging is in use.

TOX1 VOC TOX2

Ready...LEL OXY

Page 35: Multirae user training_310a

MultiRAE Alarm Signals

The MultiRAE will provide audible and visual alarms to alert users to unsafe states

It is extremely important to note that during the following conditions the alarm signals are disabled: When entering the “Communicate with PC?”

standby mode. When entering the “Calibrate Monitor”” menu When viewing data in the “View or Change

Datalog?” menu Momentarily when printing data to

the optional serial printer.

Page 36: Multirae user training_310a

MultiRAE Alarms: High

“High” display along with 3 beat/second audible/visible alarm and flashing display backlight indicates that VOC sensor (PID) has exceeded high alarm setpoint (100 ppm VOC default).

Press “Y/+” key to clear if latching alarm

TOX1 VOC TOX2

0 High 0LEL 0 20.9 OXY

Page 37: Multirae user training_310a

MultiRAE Alarms: Low

“Low” display along with 2 beat/second audible/visual alarm and flashing display backlight indicates that Oxygen sensor has gone into low alarm (less than 19.5% Oxygen default)

Press “Y/+” key to clear if latching alarm

TOX1 VOC TOX2

0 0 0LEL 0 Low OXY

Page 38: Multirae user training_310a

MultiRAE Alarms: STEL

“STEL” display along with a 1 beat/second audible/visual alarm indicates indicates that VOC sensor has exceeded STEL alarm setpoint (25 ppm VOC default)

This alarm will only clear after 15 minutes in clean air or the MultiRAE is turned off!

TOX1 VOC TOX2

0 STEL 0LEL 0 20.9 OXY

Page 39: Multirae user training_310a

MultiRAE Alarms: TWA

“TWA” display along with a 1 beat/second audible/visual alarm indicates indicates that VOC sensor has exceeded the TWA alarm setpoint (10 ppm VOC default)

This alarm will only clear after moving to clean air and then turning off the MultiRAE

TOX1 VOC TOX2

0 TWA 0LEL 0 20.9 OXY

Page 40: Multirae user training_310a

Alarms: LEL OFF

“OFF” display along with a 3 beat/second audible/visual alarm indicates that LEL sensor has been exposed to too much gas and has shut off to protect the sensor.

Move the meter to clean air and press the “Y” key to reset.

TOX1 VOC TOX2

0 0 0LEL OFF 20.9 OXY

Page 41: Multirae user training_310a

Alarms: NEG

“NEG” display along with a 1 beat/second audible/visual alarm indicates indicates that VOC sensor has drifted negative (below zero) because the sensor was zeroed in an area containing VOCs.

Perform “Fresh Air Calibration” in an area clear of all VOCs.

TOX1 VOC TOX2

0 NEG 0LEL 0 20.9 OXY

Page 42: Multirae user training_310a

MultiRAE Alarms: Pump

“Pump” display along with 3 beep audible alarm indicates that pump has stopped due to line clog

Pump alarm is a latching alarm Press “Y/+” key to clear alarm and restart

pump

TOX1 VOC TOX2

0 0.0 0LEL 0 Pump 20.9 OXY

Page 43: Multirae user training_310a

MultiRAE Alarms: Low Battery

A flashing “Bat” display along with a 1 beep alarm every 10 seconds indicates that the battery voltage has dropped below 4.35 V and it will shut down in 20-30 minutes

Quickly complete confined space entry and charge MultiRAE or install alkaline battery

TOX1 VOC TOX2

0 0.0 0LEL 0 Bat 20.9 OXY

Page 44: Multirae user training_310a

MultiRAE Alarms: Memory Full

A flashing “MEM” display along with a 1 beat/second audible/tactile/visual alarm indicates that the datalog memory is full.

Clear datalog in “View or Change Datalog?” If you get this alarm you might want to

consider changing “Select memory full type?” to “Wrap around.”

TOX1 VOC TOX2

0 0.0 0LEL 0 MEM 20.9 OXY

Page 45: Multirae user training_310a

Hold “MODE” and “N/-” keys for 5 sec. Hold “MODE” and “N/-” keys for 5 sec. to get in Programming Modeto get in Programming Mode

If MultiRAE asks a question “?”If MultiRAE asks a question “?” Answer “Y” or “N”Answer “Y” or “N”

To Accept or EscapeTo Accept or Escape Use “MODE” KeyUse “MODE” Key repeatedly pushing the “MODE” key repeatedly pushing the “MODE” key

will always eventually return user to will always eventually return user to main displaymain display

Getting Into Programming

Page 46: Multirae user training_310a

Calibrate Monitor?Calibrate Monitor? Change Alarm Limits?Change Alarm Limits? View or Change Datalog?View or Change Datalog? Change Monitor Setup?Change Monitor Setup? Change Sensor Configuration?Change Sensor Configuration?

Choose (Y) to accept or (N) to move onChoose (Y) to accept or (N) to move onref. Appendix A-2 in MultiRAE Manualref. Appendix A-2 in MultiRAE Manual

Programming Menus

Page 47: Multirae user training_310a

Fresh Air Calibration?Fresh Air Calibration?

Make sure air is clean!Make sure air is clean! Multiple Sensor Calibration? Multiple Sensor Calibration? Single Sensor Calibration?Single Sensor Calibration? Modify Span Gas Value? Modify Span Gas Value? Change LEL/VOC Span Gas? Change LEL/VOC Span Gas?

Choose (Y) to accept or (N) to move on, Choose (Y) to accept or (N) to move on, MODE to escapeMODE to escape

Calibrate Monitor?

Page 48: Multirae user training_310a

At “Multiple Sensor Calibration” press At “Multiple Sensor Calibration” press “Y/+” key“Y/+” key

For CO/HFor CO/H22S/LEL/Oxygen “quad” gas S/LEL/Oxygen “quad” gas

calibration the next screen should calibration the next screen should appear as aboveappear as above

Multiple Sensor CalibrationTOX1 VOC TOX2

CO - - - H2SLEL LEL OK? OXY OXY

Page 49: Multirae user training_310a

press “Y/+” key to display this screen press “Y/+” key to display this screen Attach calibration gas regulator to Attach calibration gas regulator to

mixed gas cylinder and attach mixed gas cylinder and attach calibration hose to MultiRAEcalibration hose to MultiRAE

Turn on calibration gasTurn on calibration gas Follow instructions on screenFollow instructions on screen

Multiple Sensor CalibrationTOX1 VOC TOX2

Apply Mixed GasLEL OXY

Page 50: Multirae user training_310a

Multiple Sensor CalibrationTOX1 VOC TOX2

No Gas Flow...LEL OXY

If you get this screen check for gas flow, the MultiRAE will not perform span calibrations unless it senses calibration gas

Page 51: Multirae user training_310a

Multiple Sensor CalibrationTOX1 VOC TOX2

Apply gas or hitLEL any key to start OXY

Try again after you have verified that the gas is correct and it is flowing properly

Follow instructions on screenFollow instructions on screen Disconnect regulator from gas cylinder Disconnect regulator from gas cylinder

when complete calibrationwhen complete calibration

Page 52: Multirae user training_310a

VOC can only be calibrated VOC can only be calibrated individually, other sensors can be individually, other sensors can be calibrated individually if necessarycalibrated individually if necessary

At “Single Sensor Calibration” press At “Single Sensor Calibration” press “Y/+” key“Y/+” key

Use “MODE” key to Use “MODE” key to select VOCselect VOC

Single Sensor CalibrationTOX1 VOC TOX2

CO VOC H2SLEL LEL pick OXY OXY

Page 53: Multirae user training_310a

Attach calibration gas regulator to VOC (Isobutylene) cylinder

Attach calibration hose to MultiRAE and make sure it is tight

With cursor on VOC press “Y/+” key Turn on calibration gas Follow instructions on screen Repeat if necessary for other gases Disconnect regulator

Single Sensor Calibration

Page 54: Multirae user training_310a

Modify Span Value Allows you to change the calibration gas

values to use other cal gases Do not change if you are using the RAE

Systems cal gas supplied with your meter Change LEL/VOC Span Gas?

Allows you to change the calibration gas from methane (LEL) & isobutylene (VOC)

Do not change if you are using the RAE Systems cal gas supplied

Calibration

Page 55: Multirae user training_310a

Change High alarm limit? 3 Beeps per second (“High”)

Change Low alarm limit? 2 Beeps per second (“Low”)

Change STEL alarm limit? 1 Beep per second (“STEL”)

Change TWA alarm limit? 1 Beep per second (“TWA”)

Change Alarm Limits?

Choose (Y) to accept or (N) to move on, MODE to escape

Page 56: Multirae user training_310a

View or Change Datalog? Reset Peak and Minimum? Clear All Data? Change Datalog Period? Select Data Type? View Datalog? Enable/Disable Datalog? Select memory full type?

Choose (Y) to accept or (N) to move on, MODE to escape

Page 57: Multirae user training_310a

Select Memory Full Type? Wrap around (default): after the

memory becomes full the newest data is written over the oldest data (First in First Out or FIFO). This is an excellent feature to provide “background” datalogging that is only used in case of an incident.

Stop: datalogging will stop when the memory is full and the “MEM” alarm will sound.

Page 58: Multirae user training_310a

Change Monitor Setup? Change Site ID? Change ID Mode?

If enabled the MultiRAE will pause on start-up until correct ID is entered

Change User ID? Change Alarm Mode? Change User Mode?

“Program” is most advanced “Display” less screens & password “Text” 3 screens with “OK” & password

Page 59: Multirae user training_310a

Change Monitor Setup? Change Real Time Clock? Change Backlight Mode? Change Password? Change Pump Speed? Change Averaging Method? Change Display Language? (English,

Spanish) Set Temperature Unit? (Co or Fo)

Choose (Y) to accept or (N) to move on MODE to escape

Page 60: Multirae user training_310a

Change Sensor Configuration?

Change LEL/VOC Gas Selection (C.F.)?

Enable/Disable Sensor? Change Dilution Ratio? Change PID Lamp?

Choose (Y) to accept or (N) to move on, MODE to escape

Page 61: Multirae user training_310a

MultiRAE: Power Off

Hold Mode Key for full 5 seconds Audible alarm will beep and display

will read “Power-down in ...5 seconds”

Leave MultiRAE on charger when not in use

Page 62: Multirae user training_310a

MultiRAE: Smart Charging

Plug 12 VDC charger into charge port Displayed for Nicads only not Li-Ion

• Pushing “N/-” MultiRAE will charge• Don’t push anything and MultiRAE will

switch to charge in 10 sec.• Full charge can take up to 8 hours

TOX1 VOC TOX2

Start BatteryLEL deep discharge? OXY

Page 63: Multirae user training_310a

MultiRAE: Charging

Pushing “Y/+” will get this screen Pushing “Y/+” again will get

“Discharging...” screen for NiCad batteries only

TOX1 VOC TOX2

LEL Are you sure? OXY

Page 64: Multirae user training_310a

MultiRAE: Charging

Discharging time depends on amount of battery charge left, up to 8 hours

MultiRAE turns on to drain battery & will automatically start charging at end of discharge

Discharge Nicads monthly

TOX1 VOC TOX2

Discharging.....LEL Battery=4.5V OXY

Page 65: Multirae user training_310a

MultiRAE: Trickle Charging

High charge once plugged in When fully charged it will switch to

trickle charge to maintain the battery Turning off and on power to charger

will reset charge to high and may burn-out battery if done repeatedly

TOX1 VOC TOX2

Charging.....LEL Battery=5.8V OXY

Page 66: Multirae user training_310a

MultiRAE: Alkaline Adapter

Accepts 4 AA alkaline batteries Provides 12-14 hours of duration Make sure that cable runs in groove

between batteries or else the pack will not fit

Page 67: Multirae user training_310a

Fig 7.1, Page 7-1, MultiRAE Manual

MultiRAE: Battery Replacement

Page 68: Multirae user training_310a

MultiRAE: Sensors

Fig. 7.2, page 7-5 of MultiRAE Manual

Page 69: Multirae user training_310a

MultiRAE: Sensor Replacement Make sure that sensors/dummy sensors fill

all places to prevent gas leakage Replace gas plate and tighten screws Replace battery Turn on MultiRAE After warm up period, block pump inlet

• If pump does not stall, inspect pump hose connections and reconnect if necessary

• If pump stalls, replace MultiRAE cover and calibrate monitor

Page 70: Multirae user training_310a

MultiRAE: When to Clean PID Sensor PID works but creeps up to 10+ ppm even

in clean air after “Fresh Air Calibration” Rapid movements of MultiRAE sends PID

into alarm due to dirt movement in sensor PID responds to moisture Always check for clean inlet filters and

tubing prior to cleaning PID

Page 71: Multirae user training_310a

MultiRAE: PID Sensor Clean/Replace Open MultiRAE Unplug battery Remove 4 screws holding gas plate Carefully lay gas plate back into battery

compartment without disconnecting hoses to pump

Remove silver shielding cap from PID sensor

Gently pull out PID sensor (white Teflon)

Page 72: Multirae user training_310a

MultiRAE: PID Sensor Clean/Replace Dip entire PID sensor into lamp cleaning

solution for at least 3 minutes (an ultrasonic cleaner is recommended, but a shot glass is the perfect sized container)

Dry sensor thoroughly (warm air will speed up this process) overnight is best

Page 73: Multirae user training_310a

MultiRAE: PID Lamp Clean/Replace To remove lamp from MultiRAE, grasp

lamp by sides and pull out. Try not to touch the flat lamp surface.

If lamp works, use a cotton swab to clean the flat lamp surface with cleaning solution

Replace lamp, DO NOT TOUCH FLAT! Install battery and turn on MultiRAE

• If after warm-up, lamp lights proceed with reassembly

• If lamp does not light, replace lamp

Page 74: Multirae user training_310a

MultiRAE: PID Lamp Clean/Replace Replace white Teflon PID sensor Replace silver PID shielding cap Make sure that all sensors (in TOX1 and

TOX2 spots) are in place to prevent gas leakage

Replace gas plate and tighten screws Replace battery Turn on MultiRAE

Page 75: Multirae user training_310a

MultiRAE: PID Lamp Clean/Replace After warm up period, block pump inlet

• If pump does not stall, inspect pump hose connections and reconnect if necessary

• If pump stalls, replace MultiRAE cover and calibrate monitor, do not overtighten screws

Page 76: Multirae user training_310a

MultiRAE: Gas Plate Cleaning If PID still creeps after cleaning PID sensor

it is recommended to clean the gas plate Follow instructions to remove pump from

gas plate Immerse gas plate in lamp cleaning

solution without pump attached (a ultrasonic cleaner will speed cleaning)

Dry gas plate with warm air Follow instructions for pump replacement

Page 77: Multirae user training_310a

Using ProRAE SuiteConfiguration from Computer

Datalogging

Page 78: Multirae user training_310a