Micro MEsh GASeous Detectors (MicroMegas). RD51 Electronics school CERN 3 – 5 February 2014. What are Micromegas ?. Micromegas are parallel-plate chambers where the amplification takes place in a thin gap, separated from the conversion region by a fine metallic mesh - PowerPoint PPT Presentation

Citation preview

Page 1: Micro  MEsh GASeous  Detectors (MicroMegas)

Micro MEsh GASeous Detectors

(MicroMegas)RD51 Electronics school

CERN 3 – 5 February 2014

Page 2: Micro  MEsh GASeous  Detectors (MicroMegas)

-800 V

-550 V

What are Micromegas ? Micromegas are parallel-plate

chambers where the amplification takes place in a thin gap, separated from the conversion region by a fine metallic mesh

The thin amplification gap (short drift times and fast absorption of the positive ions) makes it particularly suited for high-rate applications

03 - 05 February 2014RD51 Electronics School 2

The principle of operationof a MicroMegas chamber

Conversion & drift space

MeshAmplificationGap 128 µm

(few mm)

Page 3: Micro  MEsh GASeous  Detectors (MicroMegas)

UV

Mask

PCB250 µm 150 µm

Drift

128 µm

5 mm

Pillars (Ø ~300μm)

03 - 05 February 2014RD51 Electronics School 3

The bulk MicroMegas technique

Page 4: Micro  MEsh GASeous  Detectors (MicroMegas)

Bulk MicroMegas structure

03 - 05 February 2014RD51 Electronics School 4

Standard configuration Pillars every 5 (or 10) mm Pillar diameter ≈350 µm Dead area ≈1.5 (0.4)% Amplification gap 128 µm Mesh: 325 lines/inch

Pillar distance on photo: 2.5 mm

Page 5: Micro  MEsh GASeous  Detectors (MicroMegas)

µ

HV = -550 V

HV = -850 V

MicroMegas as μTPC

03 - 05 February 2014RD51 Electronics School 5

Page 6: Micro  MEsh GASeous  Detectors (MicroMegas)

Operating parameters Chambers are operating with an Ar:CO2 (93:7) gas mixture

(same gas as MDTs, safe and cheap, no flammable components)

High Voltage (moderate HV requirements) Mesh: -500 V (amplification field 40-50 kV/cm) Drift-electrode: -800 V (~600 V/cm) Currents in nA range

03 - 05 February 2014RD51 Electronics School 6

Page 7: Micro  MEsh GASeous  Detectors (MicroMegas)

Performance requirements Spatial resolution: ~60 m Angular resolution: ~0.3 mrad Good double track resolution Trigger capability Efficiency: > 98%

03 - 05 February 2014RD51 Electronics School 7

Page 8: Micro  MEsh GASeous  Detectors (MicroMegas)

Read-out strip

Mesh support pillars

PCB

Sparks in the chamber

Mesh

Sparks between mesh and readout strips may damage the detector and readout electronics and/or lead to large dead times as a result of HV breakdown

03 - 05 February 2014RD51 Electronics School 8

Page 9: Micro  MEsh GASeous  Detectors (MicroMegas)

Read-out strip Insulator

Resistive strip0.5-5 MΩ/cm

Read-out strip

Embedded resistor15-45 MΩ 5mm long+500V

PCB

PCB

To avoid spark effect the readout strips were covered with the 64 µm thick insulator layer with resistive strips on top of it connected to the +HV via discharge resistor and mesh is connected to GND

Resistive MicroMegas chambers

CHAMBER R11 R12 R13

HV resistor (MΩ) 15 45 20

Resistance along strip (MΩ/cm)

2 5 0.5

Resistive characteristics of the chambers

03 - 05 February 2014RD51 Electronics School 9

Page 10: Micro  MEsh GASeous  Detectors (MicroMegas)

-300V

128 µm

5 mm

First 2D chambers with the “spark protection” resistive strips

R16, R17, R18

Pitch: all strips – 250 µm;Width: resistive – 60 µm

Y-strips – 100 µmX-strips – 200 µm

X-stripsY-strips

+500VGND

Resistive MicroMegas chambers

03 - 05 February 2014RD51 Electronics School 10

Page 11: Micro  MEsh GASeous  Detectors (MicroMegas)

MicroMegas mesh currents and HV dropin neutron beam

Standard MM:Large currentsLarge HV drops, recovery time O(1s)Chamber could not be operated stably

R11:Low currentsDespite discharges, but no HV dropChamber operated stably up to max HV

Gas: Ar:CO2 (85:15) Neutron flux: ≈ 106 n/cm2/sec

Standard chamber vs resistive

HV valuesMesh currents

03 - 05 February 2014RD51 Electronics School 11

Page 12: Micro  MEsh GASeous  Detectors (MicroMegas)

Spark signals resistive vs standard

03 - 05 February 2014RD51 Electronics School 12

Sparks measured directly on readout strips through 50 OhmSeveral spark signals plotted on top of each other to enhance the overall characteristics

R12 shows 2-3 order of magnitude less signal and shorter recovery time than standard MM

10 µs -600

-

400

-2

00

0

Page 13: Micro  MEsh GASeous  Detectors (MicroMegas)

Induced charge

CA

AmpCopper strip

Resistive strip

C1C3

C2

C4

R1

-HV Mesh

C1 – capacitance Mesh to groundC2 – capacitance R-strip to groundC3 – capacitance R-strip to readout stripC4 – capacitance readout strip to groundCA – input capacitance of preamplifier

Equivalent scheme of resistive MicroMegas chambers

03 - 05 February 2014RD51 Electronics School 13