MEASURING INSTRUMENT MULTI - HANDY 2020. Connect the sensor Switch on the device Start to measure. Only as complete kit available. TECHNICAL DATA. 2 Measuring inputs for analogue sensors; Signals: 0 - 20 mA, 4 - 20 mA 1 Calculating channel for difference - PowerPoint PPT Presentation

Citation preview

Page 1: MEASURING INSTRUMENT MULTI - HANDY 2020

MEASURING INSTRUMENT

MULTI - HANDY 2020

Page 2: MEASURING INSTRUMENT MULTI - HANDY 2020

Only as complete kit available

Connect the sensor

Switch on the device

Start to measure

Page 3: MEASURING INSTRUMENT MULTI - HANDY 2020

TECHNICAL DATA

2 Measuring inputs for analogue sensors; Signals: 0 - 20 mA, 4 - 20 mA

1 Calculating channel for difference

Measuring rate: 1 ms (1 kHz)

Resolution: 12 Bit

Page 4: MEASURING INSTRUMENT MULTI - HANDY 2020

TECHNICAL DATA

LCD display (75 * 50 mm), adjustable background illumination

Measuring value memory: 128 KB, 1 series ofmeasurement, 60.000 measuring values, 2 Bytes /

measuring value

Easy handling- 4 soft keys

Page 5: MEASURING INSTRUMENT MULTI - HANDY 2020

HANDLING

Initialization menu appears for approx. 2 seconds

2020

Page 6: MEASURING INSTRUMENT MULTI - HANDY 2020

MAIN MENU

F1 (Peak) Min. / Max. mode

F2 No function

F3 (Start) Starting of storage

F4 Illumination on / off

Measuring value centered if only one ISDS sensor is connected

Battery display divided into 10 steps

Page 7: MEASURING INSTRUMENT MULTI - HANDY 2020

MAIN MENU

Measuring values of all connected sensors are shown

If the measuring values have the same variable, a third value will be displayed

This is used for the calculation of difference

Page 8: MEASURING INSTRUMENT MULTI - HANDY 2020

MIN / MAX MODE

F1 (Meas) Min. / Max. and normal mode

F2 (Del) Delete Min. / Max. memory

F3 (Start) Starting of storage

F4 Illumination on / off

Display if only one sensor is connected

Page 9: MEASURING INSTRUMENT MULTI - HANDY 2020

MEMORY MENU

F1 (Peak) Min. / Max. and normal mode

F2 No function

F3 (Stop) Stop storage

F4 Illumination on / off

Only real measuring channels can be stored. Representation of calculated channels of a storage only possible with HydrocomsysWin.

Display of storage status in 50 steps

Storage can be transferred via USB to a PC.

New storage overwrites the old storage.

Page 10: MEASURING INSTRUMENT MULTI - HANDY 2020

SETUP MENU

F1 (Down) One line down

F2 (Up) One line up

F3 (Slct) Select actual parameter

F4 (OK) Storage of parameter and return to the main menu

Activation via soft keys F1, F2 and F2

Languages: Italian, German, English, Spanish, French, Dutch, Swedish, Polish. Possibility to program your own language, e.g. Cyrillic

ISDS: On / Off

Unit: Selection between SI and US

Page 11: MEASURING INSTRUMENT MULTI - HANDY 2020

ISDS

If no ISDS sensor is connected to the measuring instrument, only a pressure measurement (0 - 20 mA) can be carried out.

The ISDS-function has to be switched off.

Adjustable measuring ranges: 6 - 60 - 100 - 200 - 400 - 600 bar

Page 12: MEASURING INSTRUMENT MULTI - HANDY 2020

ISDS

Without ISDS, 3 channels (2x measuring inputs, 1x calculated channel) are displayed and both measuring inputs are stored.

All ISDS sensors with output signal 0/4-20 mA are recognized automatically:

Temperature sensor Volume flow rate sensor with f/I-converter (NEW)

Page 13: MEASURING INSTRUMENT MULTI - HANDY 2020

RESET MENUF1 (No) Stop the reset, return to the

main menu

F2 No function

F3 No function

F4 Carry out a reset and return to the main menuActivation via soft keys F1, F2 and F3

After system breakdown, a software reset can be carried out

All parameters are adjusted according to manufacturing adjustments and the memory will be completely deleted

Page 14: MEASURING INSTRUMENT MULTI - HANDY 2020

COMMUNICATION WITH PC

HydrocomsysWin Basic: Blocks a certain part of the functions if MH 2020 is connected

Communication with PC via USB 2.0 interface

Extendable to full version by ordering release code which has to be entered in: Extras -> Configuration

HydrocomsysWin: Extras – Configuration

Page 15: MEASURING INSTRUMENT MULTI - HANDY 2020

COMMUNICATION WITH PC

Interface: RS232 has to be selected

Baudrate: 57600 has to be selected

COM: In this case COM 3 has to be selected

HydrocomsysWin: Extras – Configuration Windows: System configuration - Hardware

Page 16: MEASURING INSTRUMENT MULTI - HANDY 2020

MEASURING KIT

Details will follow…

HydrocomsysWin Basic:

Software as download on your homepage available

Included pressure transducer HT-PD:

Suitable only with instrument MH 2020

Page 17: MEASURING INSTRUMENT MULTI - HANDY 2020

MEASURING KITFollowing kits are available:

Description Measuring range Part no.

2020-pp-600-600 2 x 0 to 600 bar (0 to 60 MPa) 3160-18-18-69.00

2020-pp-400-400 2 x 0 to 400 bar (0 to 40 MPa) 3160-15-15-69.00

2020-pp-200-200 2 x 0 to 200 bar (0 to 20 MPa) 3160-10-10-69.00

2020-pp-60-600 1 x 0 to 60 bar (0 to 6 MPa)

1 x 0 to 600 bar (0 to 60 MPa)

3160-21-18-69.00

2020-pp-60-60 2 x 0 to 60 bar (0 to 60 MPa) 3160-21-21-69.00

2020-pp-6-60 1 x -1 to 6 bar (-1 to 0,6 MPa)

1 x 0 to 60 bar (0 to 6 MPa)

3160-32-21-69.00