Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Embed Size (px)

Citation preview

Page 1: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Magnet Safety

Page 2: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Magnetic Fields

• main magnetic field is very strong• BUT static magnetic fields are less of a concern than changing

magnetic fields• moving quickly through a magnetic field, especially the head, is

a BAD idea -- like doing whole brain TMS on yourself

• some people experience dizziness, nausea, metallic tastes– BUT these were also reported in 45% of subjects when the magnet was

OFF!

• consent form lingo: “no known risks”– you can never prove anything is safe, only that something is unsafe

Page 3: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Magnet Safety: Big Things

Source: www.howstuffworks.com

Source: http://www.simplyphysics.com/flying_objects.html

“Large ferromagnetic objects that were reported as having been drawn into the MR equipment include a defibrillator, a wheelchair, a respirator, ankle weights, an IV pole, a tool box, sand bags containing metal filings, a vacuum cleaner, and mop buckets.”

-Chaljub et al., (2001) AJR

Page 4: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Very Serious Risk

Source: http://www.mrireview.com/docs/mrideath.pdf

Westchester NY, 2001

Page 5: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Magnet Safety: Little Things

Aneurysm clips can be pulled off vessels, leading

to death

Flying things can kill people.Even in less severe incidents, they can fly into the magnet and damage it or require an expensive shutdown.

Page 6: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Subject SafetyAnyone going near the magnet – subjects, staff and visitors – must be thoroughly screened:

Subjects must have no metal in their bodies:• pacemaker• aneurysm clips• metal implants (e.g., cochlear implants)• interuterine devices (IUDs)• some dental work (but fillings are okay)

Subjects must remove metal from their bodies• jewellery, watch, piercings• coins, etc.• wallet• any metal that may distort the field (e.g., underwire bra)

Females must not be pregnant or at risk of conceiving• Some institutions even require pregancy tests for any female, every session

Subjects must be given ear plugs (acoustic noise can reach 120 dB)

This subject was wearing a hair band with a ~2 mm copper clamp. Left: with hair band. Right: without.

Source: Jorge Jovicich

Page 7: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Fall-off of Magnetic Field

Page 8: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Very Serious Risk

Source: http://www.fmrib.ox.ac.uk/%7Epeterj/safety_docs/fda_primer.html

Page 9: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Magnet Safety1. Principal Investigators should be sure all lab members are aware of hazards.2. Make sure that anyone who is about to enter the magnet room has been filled

out consent and screening forms (subjects, lab members, visitors).3. Remove all metal, coins, credit cards etc. as soon as you enter the magnet area.4. Think! Train yourself to mini-screen yourself every time you approach the

magnet room.5. Do not enter the magnet room with any tools (e.g., scissors). Use only magnet-

friendly tools in the toolbox in the magnet room.

Do the metal macarena!

Page 10: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Specific Absorption Rate (SAR)

• excess energy heats body tissues• if body heats faster than natural cooling, temperature

rises• Specific Absorption Rate (SAR) = amount of heat

absorbed by body • magnets have SAR limits to prevent overheating

– limited to 1 degree rise in core body temperature – depends on body size, geometry, thermoregulation– depends on pulse sequences (e.g., larger flip angles =

greater SAR)

Page 11: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Other safety issues

• fire safety– always give subjects a panic button– make sure that subject can be evacuated quickly if needed– have an MR-compatible fire extinguisher available– operator must know safety protocols

• quenching– rapid decrease in magnetic field strength– helium boils off and can fill room (displacing oxygen)– only voluntarily initiated in extreme situations

• burns– do not loop coils– do not place electrodes on subjects’ skin

Page 12: Magnet Safety. Magnetic Fields main magnetic field is very strong BUT static magnetic fields are less of a concern than changing magnetic fields moving

Other safety issues

• claustrophobia– subject screening

• acoustic noise– soundproofing– earplugs– headphones