Upload
p-muranyi
View
219
Download
4
Embed Size (px)
Citation preview
ORIGINAL ARTICLE
Influence of relative gas humidity on the inactivationefficiency of a low temperature gas plasmaP. Muranyi1, J. Wunderlich1 and M. Heise2
1 Department of Food Technology, Fraunhofer Institute for Process Engineering and Packaging IVV, Freising, Germany
2 Department of Plasma Technology, Fraunhofer Institute for Laser Technology ILT, Aachen, Germany
Introduction
Plasma technology is an innovative and versatile method
for the treatment and modification of surfaces and is suit-
able for a variety of industrial applications, for example
etching, coating, cleaning and sterilization. The latter is
currently an active field of research because gas plasma
allows fast and safe sterilization of packaging materials
such as bottles, lids and films – without adversely affect-
ing the main properties of the packaging materials. This
approach is very interesting for heat-sensitive materials
such as plastics (e.g. polyethylene terephthalate and poly-
ethylene) which are becoming increasingly important in
the food, pharmaceutical and medical industries. Cur-
rently, ionizing radiation, ethylene oxide (EtO) and
hydrogen peroxide are widely used in medical fields for
the nonthermal sterilization of heat-sensitive items. The
disadvantages of high toxic EtO are residues absorbed on
the plastics and long storage times for venting. Energetic
radiation (e.g. gamma radiation) is costly and can modify
materials such as polypropylene (Moreira et al. 2004). In
the area of food packaging, hydrogen peroxide and per-
acetic acid in combination with moderate heat are the
main established methods for the sterilization of heat-sen-
sitive materials (Ansari and Datta 2003). Both agents are
harmful chemicals and it is therefore necessary to comply
with threshold limit values or maximum allowable con-
centrations making their use technologically elaborate.
Furthermore, it is challenging to remove residual chemi-
cals from the packaging, handling is difficult and from a
technical point of view it is very tricky to apply the cor-
rect amount of chemical in a homogeneous pattern.
These are some of the reasons why both hydrogen perox-
ide and peracetic acid technologies are inconvenient and
why new sterilization methods are required. Low temper-
ature plasma has the potential to replace or complement
current decontamination or sterilization technologies in
the medical and food industries. In previous studies we
have shown that a cascaded dielectric barrier discharge
(CDBD) inactivates a variety of micro-organisms within
Keywords
Aspergillus niger, Bacillus subtilis, CDBD, gas
humidity, low temperature plasma, spores.
Correspondence
Peter Muranyi, Department of Food
Technology, Fraunhofer Institute for Process
Engineering and Packaging (IVV),
Giggenhauserstr. 35, D-85354 Freising,
Germany. E-mail: [email protected]
2007 ⁄ 1015: received 29 June 2007, revised 7
November 2007 and accepted 10 November
2007
doi:10.1111/j.1365-2672.2007.03691.x
Abstract
Aims: To investigate the effect of relative gas humidity on the inactivation effi-
ciency of a cascaded dielectric barrier discharge (CDBD) in air against Aspergil-
lus niger and Bacillus subtilis spores on PET foils.
Methods and Results: The inactivation kinetics as a function of treatment time
were determined using synthetic air with different relative humidity as the pro-
cess gas. Spores of A. niger and B. subtilis respectively were evenly sprayed on
PET foils for use as bioindicators. In the case of A. niger, increased spore mor-
tality was found at a high relative gas humidity of 70% (approx. 2 log10). This
effect was more evident at prolonged treatment times. In contrast, B. subtilis
showed slightly poorer inactivation at high gas humidity.
Conclusions: Water molecules in the process gas significantly affect the inacti-
vation efficiency of CDBD in air.
Significance and Impact of the Study: Modifying simple process parameters
such as the relative gas humidity can be used to optimize plasma treatment to
improve inactivation of resistant micro-organisms such as conidiospores of
A. niger.
Journal of Applied Microbiology ISSN 1364-5072
ª 2008 The Authors
Journal compilation ª 2008 The Society for Applied Microbiology, Journal of Applied Microbiology 104 (2008) 1659–1666 1659
short-treatment times (Muranyi et al. 2007). Even bacte-
rial spores such as endospores of Bacillus atrophaeus can
be reduced by about five orders of magnitude in just a
few seconds.
To evaluate and validate the efficiency of sterilization
methods, selected test strains are used as biological indi-
cators. Characteristics of these micro-organisms are their
high resistance to the sterilizing agent and nonpathoge-
nicity. Examples of such test strains are Bacillus subtilis
for hydrogen peroxide, Geobacillus stearothermophilus for
superheated steam and Bacillus pumilus for gamma radia-
tion (Bernard et al. 1990). To find an appropriate test
germ for a newly developed sterilization method, it is
necessary to investigate a wide range of micro-organisms.
We have found that conidiospores of Aspergillus niger
have high resistance to CDBD plasma treatment in air.
Therefore, A. niger conidiospores could be used for vali-
dation and optimization of the efficiency of plasma
decontamination in common. Generally, the aim of the
optimization process is the improvement of A. niger inac-
tivation. In the case of plasma there is the possibility of
varying parameters such as the process gas, pressure or
applied power density. Previous studies with various pro-
cess gases have shown that air seems to be very efficient
for destroying A. niger conidiospores (Muranyi et al.
2007) and for this reason we have tried to improve this
microbicidal effect by changing the gas humidity.
In this study, we investigated the mortality of two dif-
ferent micro-organisms, spores of B. subtilis DSM 4181
and A. niger DSM 1957, as a result of plasma treatment
with the CDBD set-up and as a function of the humidity
of the synthetic air process gas.
Materials and methods
Experimental set-up
The set-up for decontamination experiments has been
previously described in detail (Muranyi et al. 2007). The
respective lab-scale device is equipped with a so-called
CDBD which is a special kind of atmospheric pressure
plasma. It is suitable for the decontamination of flat
materials because of its geometric constraints (mm to
cm). A drawback regarding decontamination might be
that UV emitting plasmas are often not very chemically
reactive and vice-versa. By spatially combining two differ-
ent plasmas in one single set-up this problem is overcome
in the CDBD. In the first discharge gap (a closed, flat
quartz vessel) UV emitting excimer gases are utilized. In
our case a gas mixture of Xe and Br was used, which
emits monochromatic light of 282 nm. In the second gap
air or other oxidizing gases are applied directly to the
sample. As the first gap is only separated from the second
gap by a UV-transparent dielectric, both kinds of plasma
act on the sample at the same time. As Falkenstein
(1997a, 1997b) has shown, it is possible to increase the
total amount of hydroxyl radicals generated in a standard
barrier discharge in humid air or argon mixtures by illu-
mination with UV radiation at 254 nm. Moreover, the
discharge, consisting of many single micro-discharges,
more uniformly covers the treated surface area and this
was also confirmed for CDBD (Heise et al. 2004). Com-
pared with standard barrier discharges, we therefore
expect the CDBD to produce higher radical concentra-
tions (OH, O) in dry or humid air and this could
improve the microbicidal effect.
Conditioning of gas humidity
The water content of synthetic air was adjusted with a
special gas conditioning set-up (Fig.1). The device con-
tains closed loops of dry gas and wet gas, which are con-
trolled by a mass flow controller (Mass-Flo, 2000 sccm;
MKS Instruments, Berlin, Germany). For creating wet
gas, a dry synthetic air flow is fed through a water tank.
The final gas humidity is achieved by blending the two
gases and the ratio is determined by a humidity sensor
(Therm 8736; Ahlborn, Holzkirchen, Germany). Com-
mercial dry synthetic air [composition 80% N2 : 20% O2
(v ⁄ v)] from a gas bottle was fed with an initial pressure
of 1 bar into the gas conditioning device, resulting in a
gas flow of 2000 sccm. Experiments were carried out at
relative gas humidities of between 0% and 80%, increas-
ing in steps of 10%, and at a gas temperature of about
20�C. Dry synthetic air with 0% moisture was directly fed
in the CDBD set-up controlled by a gas flowmeter (ABB,
Gottingen, Germany). The CDBD set-up was directly
connected to the gas conditioning device by flexible pres-
sure tubing.
Spore preparation
Bacillus subtilis DSM 4181 and Aspergillus niger DSM
1957 were used as test strains and obtained from the Ger-
man National Resource Centre (DSMZ, Braunschweig,
Germany). Bacillus subtilis was selected because of the
highly resistant endospores and their function as a bioin-
dicator in hydrogen peroxide sterilization (Leaper 1984).
Bacillus subtilis DSM 4181 is recommended as test strain
against hydrogen peroxide by the German Association
VDMA (Industry Association for Food Processing
Machines and Packaging Machines), for validation of
aseptic filling machines in food industry (VDMA 2006).
Aspergillus niger was chosen because of its different mor-
phology and high UV resistance. Previous studies have
shown that both test strains have high resistance to
Influence of gas humidity on plasma inactivation efficiency P. Muranyi et al.
1660 Journal compilation ª 2008 The Society for Applied Microbiology, Journal of Applied Microbiology 104 (2008) 1659–1666
ª 2008 The Authors
CDBD plasma treatment, especially conidiospores of
A. niger (Muranyi et al. 2007).
Bacillus subtilis spores were obtained after inoculation
of tissue culture flasks containing 400 ml PCA medium
(Plate count agar; Merck, Darmstadt, Germany), 10 mg
manganese sulphate and 1 ml B. subtilis cells from an
overnight culture followed by incubation at 30�C for
10 days. The spores were harvested by scraping in sterile
ringer solution (Oxoid, Wesel, Germany), washed several
times by centrifugation (10 000 g for 10 min at 20�C)
and re-suspended in sterile ringer solution. After this pro-
cess the spore suspension was heated at 80�C for 32 min
to kill vegetative cells and then stored at 4�C.
Conidiospores of A. niger were formed on YGC plates
(Yeast extract glucose chloramphenicol; Merck) which
were inoculated with a suspension of cell material of A.
niger. After an incubation period of 10 days at 30�C the
spores were harvested by tapping the plates with sterile
sea sand. For deposition of the sand particles, the agglom-
erates of sand and spores were suspended in sterile ringer
solution treated with ultrasound for 1 min. The superna-
tant containing the conidiospores was removed and
finally stored at 4�C until usage.
Bioindicator preparation
Bioindicators carrying B. subtilis endospores or A. niger
conidiospores were prepared according to a previously
described inoculation method (Muranyi et al. 2007).
Planar squares with an area of 10 · 10 cm2 were cut
from polyethylene terephthalate blank foils (Mitsubishi
Polyester, Wiesbaden, Germany) and used as carriers for
the test strains. For disinfection and cleaning, the foils
were immersed in a bath of ethanol (70% v ⁄ v) and then
dried. Because of the very low initial microbial count of
the material this procedure was adequate.
Spores of the test strains were deposited on the surface
of the foils using a special spray device. An appropriate
dilution of an aqueous spore suspension was placed in a
holding tank and fed to a two-substance nozzle (Dusen
Schlick, Coburg, Germany) via a gas flow of nitrogen.
The air cap of the nozzle was adjusted to give a fine mist
of spore suspension to ensure a homogenous distribution
on the surface of the PET foils, which was verified by
microscopy. To prevent backside contamination only a
sub-area of 4 · 4 cm2 was spray inoculated by covering
the PET foil with a sterile metal template. The atomizing
pressure was 2Æ5 bar and the spray time was varied to
deposit 0Æ01 g suspension on the surface of the PET foil.
In this way, the microbial concentration on the samples
was controlled. Aspergillus niger had an initial inoculation
density of approx. 105 CFU ⁄ sample and B. subtilis of
106 CFU ⁄ sample. For avoiding an influence of different
cell densities on the inactivation efficiency (e.g. because of
agglomeration), the initial microbial load on the PET foil
was kept nearly constant for each test strain. The differ-
ence between the inoculation density of A. niger and B.
subtilis depends on the lower initial microbial count of
the spore suspension for A. niger. Finally the PET foils
were removed from the spraying set-up and allowed to
dry for 1 h under a laminar air flow cabinet.
Plasma treatment and spore recovery from the samples
Bioindicators were placed in the CDBD plasma device on
the ground electrode. The plasma was ignited by applying
an electrical power of about 170 W. To prevent microbial
contamination, the CDBD set-up was arranged under a
Figure 1 Experimental set-up for
conditioning the gas humidity. In our study
commercial synthetic air was humidified in
the range between 0% and 80% relative
humidity (20�C). The initial pressure was
1 bar and the gas flow was adjusted to
2000 sccm (ml min)1).
P. Muranyi et al. Influence of gas humidity on plasma inactivation efficiency
ª 2008 The Authors
Journal compilation ª 2008 The Society for Applied Microbiology, Journal of Applied Microbiology 104 (2008) 1659–1666 1661
laminar air flow cabinet. The applied dose was varied by
changing the treatment time in the range of 1–7 s. Com-
mercial synthetic air adjusted to different relative humidi-
ties was used as the process gas. After plasma treatment,
the samples and the untreated references were stored in
sterile Stomacher� bags (Seward, Norfolk, UK) filled with
100 ml sterile ringer solution containing 0Æ1% (w ⁄ v)
Tween-80 (Merck). Spores were removed from the surface
of the foil by washing for one minute with a Stomacher�
blender. Determination of B. subtilis survivors was carried
out using Koch’s plate dilution technique and spiral plate
method (Eddy Jet; IUL Instruments, Konigswinter, Ger-
many) on PCA. In the case of very low bacterial counts
micro-filtration was used for detection. For recovery of A.
niger only the plate dilution technique with YGC agar
(Merck) was utilized. Finally, after incubation at 30�C for
3 days, colony forming units were counted and microbial
reduction was determined as follows: logarithmic count
reduction = log10 N0 ) log10 N.
N0 is the mean of the initial microbial count of the
untreated reference samples, N is the mean of the number
of survivors after plasma treatment. For each tested
parameter, three samples were examined and data were
presented as averages.
Results
Inactivation of B. subtilis endospores as a function of gas
moisture content
Figure 2 shows the inactivation of B. subtilis endospores
using the CDBD plasma set-up as a function of the treat-
ment time and relative gas moisture. For comparison of
the different kinetics, results were graphically displayed as
N ⁄ N0. This is feasible because of the very similar initial
bacterial count of all the reference samples N0 (approx.
1 · 106 CFU ⁄ sample). It can be observed that the inacti-
vation of B. subtilis worsens with increasing gas humidity.
This effect is very evident after a short-treatment time of
1 s. To determine the extent of the humidity effect, we
have compared the results for the gas moisture limit
values. Table 1 displays the microbial log reduction
(log10 N0)log10 N) at 0% and 80% relative humidity as a
function of the treatment time and Fig. 3 shows the cor-
responding graph. After 1 s a mean microbial reduction
of 4Æ7 log10 was observed at 0% humidity and only
2Æ6 log10 at 80% humidity. This indicates that the spore
mortality of B. subtilis decreases by up to approx. 2 log10
by increasing the moisture level. Furthermore, spore mor-
tality was not proportional to the treatment time, which
indicates a bi-phasic progression of the kinetic as
reported by others (for example Purevdorj et al. (2003)).
A reason could be that plasma decontamination is a
method with limited penetration depth and therefore
stacked spores or surface layers can gradually slow down
the inactivation process.
Inactivation of A. niger conidiospores as a function of
the gas moisture content
The inactivation of A. niger conidiospores under the same
parameters ⁄ conditions as mentioned for B. subtilis is
shown in Fig. 4. Unlike for B. subtilis, an improvement of
A. niger inactivation with increasing relative gas humidity
70 80 1e–7
1e–6
1e–5
1e–4
1e–3
N/N
0
1e–2
1e–1
1e+0
60 50 40 30 20 10 0 8 Trea
tmen
t tim
e (s
)
7 6
5 4 3
2 1 0
Rel. gas humidity (%)
Figure 2 Inactivation of Bacillus subtilis endospores as a function of
the treatment time and the relative gas humidity of synthetic air. The
experiments were carried out with the cascaded dielectric barrier dis-
charge plasma set-up, equipped with a 282-nm excimer flat lamp.
The applied power was about 170 W. The initial bacterial count was
approx. 106 CFU foil)1.
Table 1 Mortality of Bacillus subtilis endospores after CDBD treat-
ment using synthetic air of different relative humidity
Treatment time (s)
Inactivation degree log10 (N0 ⁄ N)
Relative gas humidity (%)
0 80
1 4Æ7 ± 0Æ3 2Æ6 ± 0Æ3
3 5Æ1 ± 0Æ1 4Æ2 ± 0Æ7
5 5Æ0 ± 0Æ3 4Æ4 ± 0Æ1
7 4Æ9 ± 0Æ2 4Æ4 ± 0Æ2
CDBD, cascaded dielectric barrier discharge.
N0, initial microbial count of untreated reference samples.
N, final microbial count after CDBD treatment at 170 W and varying
times.
The spore mortality was calculated as the difference between the
untreated reference and the number of survivors after CDBD treat-
ment (log10 N0 ) log10 N).
Influence of gas humidity on plasma inactivation efficiency P. Muranyi et al.
1662 Journal compilation ª 2008 The Society for Applied Microbiology, Journal of Applied Microbiology 104 (2008) 1659–1666
ª 2008 The Authors
is observed. This effect becomes more evident at longer
treatment times. If we consider a treatment time of 7 s,
we observe that the microbicidal effect is most evident on
changing from 0% to 10% relative humidity and chang-
ing from 60% to 70% relative humidity. The highest inac-
tivation of 3Æ3 log10 was found for 70% relative gas
humidity and 7 s treatment time. This compared with
only 1Æ4 log10 at 0% relative humidity and the same treat-
ment time (Table 2, Fig. 5). This indicates an improve-
ment of approx. 2 log10 because of conditioning of the
relative gas humidity. Remarkable is that the inactivation
becomes slightly worse at the highest moisture level of
80% (see Table 2). To verify this result the experiment
was repeated with relative gas humidities of 65%, 70%
and 75% (Table 3, Fig. 6). Unfortunately the trial was
done with a new spore suspension of A. niger, meaning
that strict direct comparison of the two experiments is
not possible. The kinetics in this case also showed the
positive effect of high relative gas humidity on the inacti-
vation of A. niger spores, an effect which becomes more
evident at prolonged treatment times. The difference is
about 2 log10 at 70% relative gas humidity and 7 s treat-
ment time compared with dry air. This corresponds with
the results of the first experiment. A decrease in inactiva-
tion, as observed at 80% relative gas humidity, was not
observable at 75%.
Treatment time (s)
Sur
vivo
rs (
log 10
mea
n C
FU
foil–1
)
876543210
7
6
5
4
3
2
1
0
Figure 3 Spore mortality of Bacillus subtilis as a function of treat-
ment time and relative humidity. Inactivation tends to decrease with
increasing gas moisture. This phenomenon is more pronounced
at short-treatment times e.g. 1 s and for the highest relative humidity
(80%). Relative humidity: d, 80%; ,, 0%.
70 80 1e–4
1e–3
N/N
0
1e–2
1e–1
1e+0
60 50 40 30 20 10 0 8 Trea
tmen
t tim
e (s
)
7 6
5 4 3
2 1 0
Rel. gas humidity (%)
Figure 4 Inactivation kinetics of Aspergillus niger as a function of the
treatment time and the humidity of synthetic air. The plasma treat-
ment was carried out with the cascaded dielectric barrier discharge
set-up. The initial microbial count was in the range of approx.
105 CFU foil)1.
Table 2 Mortality of Aspergillus niger conidiospores after CDBD
treatment using synthetic air of different humidity
Treatment time (s)
Inactivation degree log10 (N0 ⁄ N)
Relative gas humidity (%)
0 70 80
1 0Æ7 ± 0Æ2 1Æ2 ± 0Æ2 0Æ8 ± 0Æ0
3 1Æ0 ± 0Æ1 1Æ8 ± 0Æ2 1Æ5 ± 0Æ1
5 1Æ2 ± 0Æ3 2Æ4 ± 0Æ4 2Æ2 ± 0Æ4
7 1Æ4 ± 0Æ3 3Æ3 ± 0Æ4 2Æ6 ± 0Æ4
CDBD, cascaded dielectric barrier discharge.
N0, initial microbial count of untreated reference samples.
N, final microbial count after CDBD treatment at 170 W and varying
times.
Treatment time (s)
876543210
Sur
vivo
rs (
log 10
mea
n C
FU
foil–1
)
6
5
4
3
2
1
0
Figure 5 Inactivation of Aspergillus niger conidiospores as a function
of treatment time and relative humidity. It can be observed that spore
mortality increases with increasing humidity. This effect becomes
more pronounced at prolonged treatment times. Relative humidity: d,
0%; ,, 70%; n, 80%.
P. Muranyi et al. Influence of gas humidity on plasma inactivation efficiency
ª 2008 The Authors
Journal compilation ª 2008 The Society for Applied Microbiology, Journal of Applied Microbiology 104 (2008) 1659–1666 1663
Discussion
The aim of this study was to optimize the CDBD inacti-
vation efficiency in air against A. niger conidiospores by
modifying the relative gas humidity. The mould A. niger
was chosen because of its high resistance to CDBD
plasma treatment as described previously (Muranyi et al.
2007). To examine the influence of gas humidity on bac-
teria simultaneously, endospores of B. subtilis were
selected. Generally, the properties of gas plasmas can be
modified and optimized by varying the main parameters
such as pressure, power and process gas. Each alteration
of one of these parameters changes the whole plasma
chemistry and influences, for example, the electron den-
sity, concentrations of charged or reactive particles and
the amount of emitted UV radiation. In the case of the
CDBD set-up, only the power, process gas composition
and the type of excimer flat lamp can be changed. To
achieve the highest inactivation rates, trials were carried
out with a maximum power of approx. 170 W (for this
set-up). Previous studies with different process gases had
shown that laboratory air is very effective for inactivation
of micro-organisms (Muranyi et al. 2007). There is also
the opportunity to change the wavelength of the excimer
flat lamp via selection of the excimer gas mixture, but
this has been studied elsewhere. In this work, we have
tried to improve the decontamination efficiency by taking
synthetic air having a gas composition of 80% nitrogen
and 20% oxygen and humidifying it to simulate atmo-
spheric conditions. Compared with normal air, minor
components such as argon and carbon dioxide are not
present in synthetic air. Previous experiments with
A. niger DSM 1957 and B. atrophaeus DSM 2277 have
shown that their effect can however be neglected (Fig. 7).
The studies have shown that the inactivation of A. niger
conidiospores can be improved by about 2 log10 by
increasing the relative gas humidity. The highest spore
mortality of approx. 3Æ3 log10 was found at 70% relative
humidity and a treatment time of 7 s. As shown by the
inactivation kinetics, the effect was enhanced at prolonged
plasma exposure which indicates a time-dependent reac-
tion. Remarkable is the slightly worse mortality at 80%
relative humidity. This could indicate that there is an
optimum relative humidity in the range of 70% for inac-
tivation of A. niger conidiospores.
Table 3 Repeat experiment with a new suspension of Aspergillus
niger conidiospores and synthetic air of different humidity
Treatment
time (s)
Inactivation degree log10 (N0 ⁄ N)
Relative gas humidity (%)
0 65 70 75
1 1Æ0 ± 0Æ0 1Æ5 ± 0Æ4 2Æ1 ± 0Æ1 1Æ9 ± 0Æ2
3 1Æ4 ± 0Æ1 2Æ1 ± 0Æ2 2Æ3 ± 0Æ2 2Æ4 ± 0Æ0
5 2Æ0 ± 0Æ0 2Æ0 ± 0Æ1 3Æ1 ± 0Æ2 4Æ0 ± 0Æ4
7 2Æ3 ± 0Æ2 3Æ1 ± 0Æ2 4Æ0 ± 0Æ4 4Æ0 ± 0Æ5
CDBD, cascaded dielectric barrier discharge.
N0, initial microbial count of untreated reference samples.
N, final microbial count after CDBD treatment at 170 W and varying
times.
Sur
vivo
rs (
log 10
mea
n C
FU
foil–1
)
Treatment time (s)
876543210
1e+6
1e+5
1e+4
1e+3
1e+2
1e+1
1e+0
Figure 6 Inactivation kinetics of Aspergillus niger conidiospores on
being subjected to cascaded dielectric barrier discharge treatment as a
function of time and at relative humidities of 0%, 65%, 70% and
75%. The process gas was synthetic air and the applied power was
about 170 W. Relative humidity: d, 0% s, 65%; ., 70%; n, 75%.
Treatment time (s)
876543210
Sur
vivo
rs (
log 10
mea
n C
FU
foil–1
)
7
6
5
4
3
2
1
0
Figure 7 Inactivation kinetics of Aspergillus niger DSM 1957 coni-
diospores and Bacillus atrophaeus DSM 2277 endospores as a func-
tion of time for normal air and synthetic air. Relative humidity (20�C)
was kept constant for both gases. The treatment was done with the
cascaded dielectric barrier discharge at an applied power of about
180 W. The kinetics shows that there is no significant effect of minor
gases in normal air on the inactivation efficiency. (d) A. niger ⁄ room
air 47Æ5% RH; (s) A. niger ⁄ synthetic air 47Æ5% RH; (.) B. atro-
phaeus ⁄ room air 38% RH; (,) B. atrophaeus ⁄ synthetic air 38% RH.
Influence of gas humidity on plasma inactivation efficiency P. Muranyi et al.
1664 Journal compilation ª 2008 The Society for Applied Microbiology, Journal of Applied Microbiology 104 (2008) 1659–1666
ª 2008 The Authors
In contrast, the kinetics for B. subtilis endospores indi-
cated a slight worsening of inactivation at higher gas
humidity. Compared with dry air, the negative effect of
water vapour was most apparent (2 log10) for a treatment
time of 1 s and at 80% relative humidity.
We assume that the positive effect on inactivation
because of the increased gas humidity in the plasma
mainly arises from the decomposition of additional water
molecules into hydroxyl radicals as a result of impact by
other gas molecules or charged particles. The dissociation
of the O-H bond in a water molecule requires an energy
of approx. 5Æ1 eV. Such energy values are easily reached
by the accelerated electrons in the plasma. The photon
energy of the 282 excimer flat lamp (4Æ4 eV) is not suffi-
cient to dissociate the water molecules directly. However,
one main path of hydroxyl radical generation in a DBD
discharge is photodissociation of ozone into atomic sin-
glet oxygen and subsequent reaction of the atomic oxygen
with water molecules (Falkenstein 1997a). The maximum
absorption of ozone occurs at 255 nm, within the Hartley
absorption band (244–278 nm), so radiation emitted by
the 282 nm excimer surely contributes to the formation
of hydroxyl radicals.
The hydroxyl radicals are very electronegative and are
therefore strong oxidizing agents which can damage
microbial cells. The main targets of oxidative processes
include unsaturated fatty acids of the membrane lipid
bilayer and diverse enzymes and proteins of the cell ⁄ spore
(Laroussi 2005). The standard electrode potential of
hydroxyl radicals (+2Æ88 V) is much higher than other
species such as atomic oxygen (+2Æ42 V) or ozone
(+2Æ07 V). Furthermore, two hydroxyl radicals can react
to form hydrogen peroxide, which is also known to have
microbicidal effects. In this case, the effect is based on the
released atomic oxygen, which damages essential parts of
the cells: 2(ÆOH) fi H2O2 fi H2O + O. This mechanism
was also reported by Purevdorj et al. 2003.
Wallhaußer (1995) pointed out generally that in the
case of dry air (<32% relative humidity) a twofold dose
of UV radiation is necessary for sterilization compared
with air of 72% relative humidity. This agrees with our
results with A. niger conidiospores which showed
improved inactivation at 70% relative humidity. Further-
more, Wallhaußer (1995) reported that increased
amounts of water vapour (>80% relative humidity)
reduce the sterilization effect because of the poorer trans-
missibility of UV radiation in such air and protective
water films around the micro-organism cells. We also
observed reduced inactivation of A. niger conidiospores at
80% relative humidity. However, regarding the reduced
inactivation of A. niger spores, and especially B. subtilis
spores, at elevated gas humidities, we do not believe that
limited transmissibility or protective water films play a
major role, because Ansari and Datta (2003) and Hale
and Querry (1973) reported a UV light penetration depth
in water of up to 30 cm. The adjusted gap of our plasma
set-up has only a width of 1 mm. Therefore these aspects
must be further investigated.
Different results for Bacillus species were obtained in
the study of Purevdorj et al. (2003). In the case of B.
pumilus spores and a low pressure plasma set-up, they
observed increased spore mortality by more than two
orders of magnitude by moistening the air. The reasons
for this variance could be the high resistance of B. subtilis
endospores to oxidative agents, especially the tested strain
DSM 4181 (hydrogen peroxide bioindicator), wherefore
the positive effect of additional reactive species is not
obvious in our case. However, this is just an assumption
because plasma chemistry in low pressure systems is
rather different.
Black pigmented A. niger conidiospores have especially
high UV resistance and require dosage levels 20–100 times
higher (Ansari and Datta 2003), but they are also sensitive
to oxidation. The results of our study indicate that an
accumulation of reactive species from decomposed water
molecules could be responsible for the improved inactiva-
tion of A. niger at prolonged treatment times.
Also interesting are the results of Maeda et al. (2003)
regarding the inactivation of E. coli using atmospheric gas
plasma with a humidified working gas in the 0–70% rela-
tive humidity range. They reported survival rates as a
function of humidity which corresponds to a standardized
normal distribution. This means no inactivation at 0%
and 70% relative humidity, and maximum inactivation at
43% which corresponds to normal room conditions. It
therefore seems that the effect of moisture in air on the
microbiological targets depends not only on the amount
of water vapour but also on the microbiological species.
To achieve the highest positive inactivation effect of a gas
plasma, it hence makes sense to validate the influence of
gas humidity for every microbiological application.
Falkenstein (1997a) has investigated the micro-dis-
charge behaviour of dielectric barrier discharges in water–
air mixtures under UV illumination, which is similar to
our CDBD set-up. It was shown on the one hand that
increasing water concentration in air results in hydroxyl
radical production which increases oxidation rates. On
the other hand, the water vapour reduces the number of
micro-discharges. The reason for this effect is the reduc-
tion of the surface resistance of the dielectric caused by
the adsorption of water molecules (Falkenstein 1997a;
Falkenstein and Coogan 1997). The additional UV radia-
tion of the excimer flat lamp leads again to more and
smaller micro-discharges with lower residual OH-radical
concentration; but as these are nonlinear effects, an
increased total concentration of OH is ultimately
P. Muranyi et al. Influence of gas humidity on plasma inactivation efficiency
ª 2008 The Authors
Journal compilation ª 2008 The Society for Applied Microbiology, Journal of Applied Microbiology 104 (2008) 1659–1666 1665
observed. Nevertheless, it could be that an increased
amount of water vapour finally reduces the plasma
homogeneity of the CDBD (Falkenstein 1997b). These
assumptions do however have to be confirmed. Measure-
ments carried out with a normal dielectric barrier dis-
charge have shown that increasing gas humidity (80%,
24�C) reduces the discharge homogeneity (Franken et al.
2003). This could be a reason for the observed decrease
in B. subtilis endospore inactivation at high relative gas
humidity. Which of these effects dominate in the process
depends, among other things, on the properties of the
chosen micro-organism.
In summary, we assume in the case of our CDBD set-
up and using synthetic air as the process gas those high
gas humidities of, for example 70% could be optimal for
the inactivation of oxidation-sensitive micro-organisms
such as A. niger. By the use of germs such as B. subtilis,
where the inactivation pathway is based mainly on UV
radiation, the process will be slowed down by high water
vapour content. We believe that a possible effect of air
transmissibility or water films on UV penetration can be
neglected in this set-up. The influence of gas humidity on
plasma homogeneity has to be evaluated. The solution for
practical applications could be an arrangement of two
CDBD units in series, each with different process gases
(dry air and humidified air) to extend the target decon-
tamination spectrum. In all cases, it is important to adjust
the relative humidity of the plasma gas to a value which
results in optimum inactivation conditions for as many
strains as possible.
Acknowledgement
This work was partially funded by the European Commis-
sion under contract number G1RD-2002-00747 (COLA-
PE). The authors would also like to express their
gratitude to Mrs Berger and Mrs Steinbuchl for technical
support and to Dr Rieblinger for useful discussions.
References
Ansari, M.D.I.A. and Datta, A.K. (2003) An overview of sterili-
zation methods for packaging materials used in aseptic
packaging systems. Trans IChemE 81, 57–65.
Bernard, D.T., Gavin, A., Scott, V.N., Shafer, B.D., Stevenson,
K.E., Unverferth, J.A. and Chandarana, D.I. (1990) Valida-
tion of aseptic processing and packaging. Food Technol-
Chicago 44, 119–122.
Falkenstein, Z. (1997a) The influence of ultraviolet illumina-
tion on OH formation in dielectric barrier discharges of
Ar ⁄ O2 ⁄ H2O: the Joshi effect. J Appl Phys 81, 7158–7162.
Falkenstein, Z. (1997b) Influence of ultraviolet illumination on
microdischarge behaviour in dry and humid N2, O2, air,
and Ar ⁄ O2: the Joshi effect. J Appl Phys 81, 5975–5979.
Falkenstein, Z. and Coogan, J.J. (1997) Micro-discharge behav-
iour in the silent discharge of nitrogen–oxygen and water–
air mixtures. J Phys D: Appl Phys 30, 817–825.
Franken, O., Pietsch, G., Saveliev, A., Hulka, L., Neiger, M.,
Roth, M., Liu, S., Soloshenko, I. et al. (2003) Grundlegende
Erforschung plasmagestutzter Verfahren zur Entkeimung von
Packstoffen fur Lebensmittel. Aachen: Rheinisch-Westfali-
sche Technische Hochschule, Forschungsbericht BMBF-
Projekt FKZ 13N7609 ⁄ 9, 2003.
Hale, G.M. and Querry, M.R. (1973) Optical constants of
water in the 200-nm to 200-lm wavelength region. Appl
Opt 12, 555–563.
Heise, M., Lierfeld, T., Franken, O. and Neff, W. (2004) Single
filament charge transfer and UV-emission properties of a
cascaded dielectric barrier discharge (CDBD) set-up.
Plasma Sources Sci Technol 13, 351–358.
Laroussi, M. (2005) Low temperature plasma-based steriliza-
tion: overview and state-of-the-art. Plasma Process Polym
2, 391–400.
Leaper, S. (1984) Comparison of the resistance to hydrogen
peroxide of wet and dry spores of Bacillus subtilis SA22.
J Food Technol 19, 695–702.
Maeda, Y., Igura, N., Shimoda, M. and Hayakawa, I. (2003)
Inactivation of Escherichia coli K12 using atmosphere gas
plasma produced from humified working gas. Acta Bio-
technol 23, 389–395.
Moreira, A.J., Mansano, R.D., Pinto, T., de, J.A., Ruas, R.,
Zambon, L., da, S., Valero da Silva, M. et al. (2004) Sterili-
zation by oxygen plasma. Appl Surf Sci 235, 151–155.
Muranyi, P., Wunderlich, J. and Heise, M. (2007) Sterilization
efficiency of a cascaded dielectric barrier discharge. J Appl
Microbiol 103, 1535–1544.
Purevdorj, D., Igura, N., Ariyada, O. and Hayakawa, I. (2003)
Effect of feed gas composition of gas discharge plasmas on
Bacillus pumilus spore mortality. Lett Appl Microbiol 37,
31–34.
VDMA (2006) Aseptic Packaging Machines for the Food Indus-
try – Minimum Requirements and Basic Conditions for the
intended Operation (in German). VDMA-Guideline 8742.
Wallhaußer, K.H. (1995) Praxis der Sterilisation, Desinfektion –
Konservierung. Thieme: Stuttgart S. 334.
Influence of gas humidity on plasma inactivation efficiency P. Muranyi et al.
1666 Journal compilation ª 2008 The Society for Applied Microbiology, Journal of Applied Microbiology 104 (2008) 1659–1666
ª 2008 The Authors