IceCube DOM Calibration. Jim Braun. Motivation. DOM-Cal – Big Picture. Application runs on DOM Can calibrate all DOMs in parallel Stores calibration data on DOM flash memory Java client Coordinates calibration Reads calibration data from DOM Produces DOM-Cal XML calibration files - PowerPoint PPT Presentation

Citation preview

Page 1: IceCube DOM Calibration

April 20, 2005 1

IceCube DOM Calibration

Jim Braun

Page 2: IceCube DOM Calibration

April 20. 2005 2

Motivation

Page 3: IceCube DOM Calibration

April 20. 2005 3

DOM-Cal – Big Picture

• Application runs on DOM– Can calibrate all DOMs in parallel– Stores calibration data on DOM

flash memory• Java client

– Coordinates calibration– Reads calibration data from DOM– Produces DOM-Cal XML

calibration files• Calibrator class reads XML

file, applies calibration constants

• All code in dom-cal project on glacier

– Authors: John Kelley, Jim Braun, Kael Hanson

DOM hub

DOM-CalClient

XMLFiles

Database

Calibrator.java

Analysis

Page 4: IceCube DOM Calibration

April 20. 2005 4

Calibration Routine

• ATWD Calibration– Reconstruct PMT voltage waveform from ATWD data– Requires:

• Voltage calibration of ATWD

• Measurement of amplifier gain for ch0, ch1, and ch2

• Frequency calibration of ATWD

• Baseline measurement

• PMT Gain Calibration– Find relationship between PMT gain and applied voltage

• PMT Transit Time Calibration (NEW!)– Find PMT transit time as a function of applied voltage

Page 5: IceCube DOM Calibration

April 20. 2005 5

ATWD Calibration

• Bootstrap process!

V = 5 x bias_dac / 4096V = 0.0001220 x (0.4 x disc_dac – 0.1 x bias_dac)f = 20MHz

ATWDFrequencyCalibration

ATWDVoltage

Calibration

PulserCalibration

BaselineCalibration

Amplifier ch0Calibration

Amplifier ch1Calibration

Amplifier ch2Calibration

Muons

Mainboard LED

Page 6: IceCube DOM Calibration

April 20. 2005 6

Pulser Calibration

• Determine relationship between pulser DAC setting and peak voltage– Use known relationship between discriminator DAC and voltage– Set discriminator DAC, adjust pulser DAC until 50% of pulses

cross discriminator threshold

– At this point, the pulser peak voltage corresponds to known discriminator voltage.

– Peak voltage distribution is very narrow

Page 7: IceCube DOM Calibration

April 20. 2005 7

Pulser Calibration

• Repeat for ~10 discriminator voltages• Relationship is very linear

• Now know pulser peak voltage given DAC setting

• Will use this relationship in amplifier calibration

Page 8: IceCube DOM Calibration

April 20. 2005 8

ATWD Voltage Calibration

• Determine relationship between ATWD value and signal voltage

• Use known bias DAC voltage relationship– Bias is independent of amplifier gain– Set bias, record average ATWD value

• For each bin (0-127) of each signal channel (0-2) of each ATWD (0-1)

• O(100) samples

– Apply linear fit to ATWD value vs. voltage data

• Pedestal patterns are eliminated

Page 9: IceCube DOM Calibration

April 20. 2005 9

ATWD Voltage Calibration

• Now know voltage of any ATWD bin given a channel number, bin number, and value

• Requires 768 linear fits!

• ATWD response not entirely linear• Need to calculate and subtract small baseline offset for

each channel during precision measurements

Page 10: IceCube DOM Calibration

April 20. 2005 10

Baseline Calibration

• New in DOM-Cal 5.0• Measures average baseline offset for each ATWD

channel• Need to measure baseline whenever the internal state of

the DOM changes• Known to affect baseline:

– Analog multiplexer– Mainboard LED power supply– PMT high voltage

• Affects low gain channel the most– Main source of Hagar’s ch1:ch2 charge discrepancy

Page 11: IceCube DOM Calibration

April 20. 2005 11

Baseline Calibration

• We care most about affect of high voltage and residual baseline from imperfect ATWD calibration

• Take baseline data both with HV off and HV at values spanning DOM operating points

• Store all data points• Use baseline value closest to

operational HV when calibrating TestDAQ data

• For gain calibration, domcal chooses only the voltages where baseline calibration points exist

Page 12: IceCube DOM Calibration

April 20. 2005 12

Amplifier Calibration• Calibrate high gain channel (0) with pulser

– Pulser too weak to accurately calibrate lower gain channels• Set pulser peak to a known voltage, record peak voltage after amplification

in ch0– Use ATWD ch0 calibration data to find peak voltage

• Maximize ATWD sampling speed to better localize peak

• Record mean and error of O(250) ch0 peaks

• Ratio of mean voltage and known pulser voltage yields amplification factor

Page 13: IceCube DOM Calibration

April 20. 2005 13

Amplifier Calibration

• Need source of high amplitude pulses to calibrate ch1 and ch2 amplifiers

• Use PMT signals! – new in DOM-Cal 5.0– Muons work well at surface– Mainboard LED needed in deep ice

• LED power supply shifts baseline, need to recalibrate

• For ch1:– Select pulses which have an ATWD peak value of

600-800 counts• Too few ch0 counts -- too much error in ch1 peak voltage• Too many ch0 counts -- ch0 nonlinearity becomes significant

Page 14: IceCube DOM Calibration

April 20. 2005 14

Amplifier Calibration

• For ch1:– Record ratio of ch1 peak

voltage to ch0 peak voltage for O(250) iterations

– We know ch0 gain, so ch1 gain is given by the product of ch0 gain and voltage ratio.

• For ch2– We now know ch1 gain, use

previous procedure to find ch2 gain

– Slow with muons (<1 Hz)– Slow when discriminator rate is

high– Use LED if necessary

Page 15: IceCube DOM Calibration

April 20. 2005 15

ATWD Frequency Calibration

• Select mainboard oscillator in ATWD analog multiplexer channel (channel 3)

• At various sampling speed DAC values, count number of bins between positive zero crossings in ATWD waveform

• Average O(100) clock waveforms• Assuming oscillator operates in

spec @ 20MHz, ATWD frequency is given by 20MHz * #bins

• Not quite linear – newer version will sample closer to 850

DAC value

Page 16: IceCube DOM Calibration

April 20. 2005 16

Gain Calibration

• Capture PMT single photoelectron pulses in ATWD– Glass radioactivity emits enough light

• Apply ATWD calibration to get PMT V(t) waveform

• I(t) given by V(t)/50• I(t) pulse integrated from –4 bins to +8 bins of pulse

maximum (~-14ns - +28ns), yielding SPE charge

• Repeat O(5000) times, histogram charge data, and apply nonlinear fit

Page 17: IceCube DOM Calibration

April 20. 2005 17

Gain Calibration

Charge (pC)

DiscriminatorEdge

Noise

Single-PhotoelectronPeak

Muons, etc.

Fit Exponential +Gaussian

Page 18: IceCube DOM Calibration

April 20. 2005 18

Gain Calibration

Repeat from 1200V to 1900V in 100V intervals

1400V

1600V 1700V

1500V

Page 19: IceCube DOM Calibration

April 20. 2005 19

Gain Calibration

Mean SPE charge vs. voltage is a power law

Gai

n (c

harg

e /

e)

Voltage

Linear log - log fit yields operating point

The number of photoelectrons for any pulse at a given HV is now determined

Page 20: IceCube DOM Calibration

April 20. 2005 20

Java Client

• In dom-cal project on glacier– Main class: icecube.daq.domcal.DOMCal

• Run with no args for usage instructions

• Reads calibration data from DOM flash– Stores calibration data in local XML files– Stores data in domprodtest database

• Can initiate calibration• Can run calibrations on many DOMs and DOM

hubs in parallel• Most will never need to use the java client

Page 21: IceCube DOM Calibration

April 20. 2005 21

DOM-Cal XML Files

• Hopefully, users won’t need to know much about the XML files or database structure either!– Access provided through calibration applications

• XML files are reasonably easy to read if needed– Contain:

• DOM hardware ID (No name…..sorry Mark)

• Temperature

• Date

• DAC settings and ADC readings

• Calibration information– Linear fit data

– Baseline data

– Gain histogram data

Page 22: IceCube DOM Calibration

April 20. 2005 22

DOM-Cal Calibrator

• Icecube.daq.domcal.Calibrator java class provides access to calibration data.– I3DOMCalibration equivalent object in IceTray

• Create a Calibrator for each DOM:– new Calibrator(XML_File);

• Most important routine:– atwdCalibrateToPmtSig()

• Takes an array of raw ATWD data and applies calibration to yield true voltage signal

• Other methods to access raw data, described in javadoc– http://www.amanda.wisc.edu/jbraun/domcaldoc/

Page 23: IceCube DOM Calibration

April 20. 2005 23

DOM-Cal Results

• Currently running DOM-Cal v4.3 at pole

• Baseline shift apparent in data from String 21 and from last year’s FAT runs

• Analog multiplexer enabled in all string-21 runs – just recently disabled

Page 24: IceCube DOM Calibration

April 20. 2005 24

DOM-Cal Results

• DOM-Cal 5.0 results are encouraging

• Baseline calibrated to zero for all channels

• No analog multiplexer effects

• Amplifier gains now calculated much more accurately

• Pulse heights now agree between all three ATWD channels

Page 25: IceCube DOM Calibration

April 20. 2005 25

DOM-Cal Results

• The bad news– Probably won’t see much gain when analyzing String 21 data,

even with DOM-Cal 5.0

• Local coincidence readout rate is a few Hz

• Interval is long enough for baseline to drift

• Hopefully there is a firmware fix!

Page 26: IceCube DOM Calibration

April 20. 2005 26

DOM-Cal 6.0++

• Other features DOM-Cal may deal with:– PMT transit time – almost finished!

– Ch0 time offset, bandwidth limitation

– Signal droop• Evidence time

constant can be easily measured (Chris W.)