77
Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device (EPID) Entwicklung einer dosimetrischen Kontrolle der täglichen Strahlenbehandlung mittels eines MV-Bildgebungssystems (EPID) Der Medizinischen Fakultät der Friedrich-Alexander-Universität Erlangen-Nürnberg zur Erlangung des Doktorgrades Dr. rer. biol. hum. vorgelegt von Mohammadsaeed Saboori aus Hamedan (Iran)

Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

  • Upload
    vuduong

  • View
    218

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

Development of a daily dosimetric control

for radiation therapy using an electronic

portal imaging device (EPID)

Entwicklung einer dosimetrischen Kontrolle der täglichen

Strahlenbehandlung mittels eines MV-Bildgebungssystems (EPID)

Der Medizinischen Fakultät

der Friedrich-Alexander-Universität

Erlangen-Nürnberg

zur

Erlangung des Doktorgrades Dr. rer. biol. hum.

vorgelegt von

Mohammadsaeed Saboori

aus Hamedan (Iran)

Page 2: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

Als Dissertation genehmigt von der

Medizinischen Fakultät der Friedrich-Alexander-Universität

Erlangen-Nürnberg

Vorsitzender des Promotionsorgans: Prof. Dr. Dr. h.c. J. Schüttler

Gutachter: Prof. Dr. R. Müller

Gutachter: Prof. Dr. B. Hensel

Gutachter: Prof. Dr. R. Fietkau

Tag der mündlichen Prüfung: 13.Mai 2015

Page 3: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

Table&of&content&&

!

Abstract:&.....................................................................................................................&1!

1! Introduction&..........................................................................................................&4!1.1! Facts!about!cancer!........................................................................................................................................!4!1.2! External!Beam!Radiotherapy!(EBRT)!...................................................................................................!6!1.2.1! Intensity+Modulated+Radiation+Therapy+(IMRT)+.........................................................................+8!1.2.2! Biological+point+of+view+of+radiotherapy+.........................................................................................+9!

1.3! Quality!concept!in!radiotherapy!...........................................................................................................!10!1.3.1! In+vivo+dosimetry+in+EBRT+...................................................................................................................+11!1.3.2! Methods+of+IVD+.........................................................................................................................................+12!1.3.3! Electronic+Portal+Imaging+Device+systems+...................................................................................+13!1.3.4! Applications+of+EPID+system+...............................................................................................................+15!

1.4! In!vivo!dosimetry!with!the!EPID!system!...........................................................................................!15!1.5! Purpose!of!study!..........................................................................................................................................!17!

2! Accuracy&of&a&commercial&treatment&planning&system&for&transit&dose&calculations&

at&large&distances&......................................................................................................&19!2.1! INTRODUCTION!...........................................................................................................................................!19!2.2! MATERIALS!AND!METHODS!..................................................................................................................!20!2.2.1! Equipment+..................................................................................................................................................+20!2.2.2! 2D+dose+map+calculation+using+TPS+................................................................................................+20!2.2.3! Reference+transit+dose+measurement+setup+................................................................................+21!2.2.4! Accuracy+of+transit+dose+calculation+by+TPS+...............................................................................+22!

2.3! RESULT!............................................................................................................................................................!24!2.3.1! Field+size+effect+.........................................................................................................................................+24!2.3.2! Phantom+thickness+effect+.....................................................................................................................+25!2.3.3! Air+gap+effect+.............................................................................................................................................+25!2.3.4! IMRT+plan+test+..........................................................................................................................................+26!

2.4! DISCUSSION!...................................................................................................................................................!28!2.5! Conclusion!......................................................................................................................................................!29!

3! 2D&Patient&transit&EPID&dosimetry&for&treatment&fraction&dose&verification&..........&31!3.1! Introduction!...................................................................................................................................................!31!3.2! Materials!and!methods!.............................................................................................................................!34!3.2.1! Equipment+..................................................................................................................................................+34!3.2.2! Transit+dose+distribution+predicted+at+the+EPID+level+.............................................................+35!3.2.3! EPID+dose+measurements+....................................................................................................................+36!3.2.4! Transit+dose+measurement+formula+...............................................................................................+41!3.2.5! ProofRofRprinciple+IMRT+plan+verification+...................................................................................+41!

3.3! Result!................................................................................................................................................................!42!3.3.1! Initial+system+parameters+...................................................................................................................+42!3.3.2! IMRT+plan+verification+..........................................................................................................................+45!

3.4! Discussion!.......................................................................................................................................................!50!3.4.1! Flood+field+correction+............................................................................................................................+51!3.4.2! Open+field+calibration+for+transmission+dosimetry+..................................................................+51!3.4.3! Field+size+correction+...............................................................................................................................+52!3.4.4! Phantom+thickness+correction+...........................................................................................................+52!3.4.5! OverRresponse+correction+....................................................................................................................+53!3.4.6! Transit+measured+data+analysis+.......................................................................................................+53!3.4.7! IMRT+plan+verification+..........................................................................................................................+54!

3.5! Conclusion!......................................................................................................................................................!55!

Page 4: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

4! Summary&and&conclusion&.....................................................................................&58!

5! Bibliography&........................................................................................................&61!

6! Acknowledgment&.................................................................................................&67!

7! Curriculum&Vitae&..................................................................................................&70!

Page 5: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

Abstract: Electronic Portal Imaging Devices (EPIDs) can be used to perform dose measurements

during radiation therapy treatments if dedicated calibration and correction procedures

are applied. The purpose of this study was to provide a new calibration and correction

model for an amorphous silicon (a-Si) EPID for use in transit dose verification of step-

and-shoot intensity modulated radiation therapy (IMRT). A model was created in a

commercial treatment planning system to calculate the nominal two-dimensional (2D)

dose map of each radiation field at the EPID level. The EPID system was calibrated and

correction factors were determined using a reference set-up, which consisted a patient

phantom and an EPID phantom. The advantage of this method is that for the calibration,

the actual beam spectrum is used to mimic a patient measurement. As proof-of-

principle, the method was tested for the verification of two 7-field IMRT treatment

plans with tumor sites in the head-and-neck and pelvic region. Predicted and measured

EPID responses were successfully compared to the nominal data from treatment

planning using dose difference maps and gamma analyses. Based on our result it can be

concluded that this new method of 2D EPID dosimetry is a potential tool for simple

patient treatment fraction dose verification.

Page 6: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

2

Zusammenfassung:

Die an den meisten Beschleunigern der Strahlentherapie angebrachten

Bildempfängersysteme (EPIDs, Electronic Portal Imaging Devices) können bei

entsprechender Kalibrierung auch als Dosimeter eingesetzt werden. Ziel dieser Arbeit

war es, ein System zur Kalibrierung und Korrektur des EPID-Signals (Detektor aus

amorphem Silizium) in Wasserenergiedosis zu entwickeln und bereitzustellen, womit

die transmittierte Dosis während der Strahlenbehandlung zur Verifikation relevanter

Parameter der Therapie herangezogen werden kann. Moderne avancierte

Behandlungstechniken bedienen sich der sogenannten intensitätsmodulierten

Strahlentherapie (IMRT), die sich im technisch einfacheren „step and shoot“-Verfahren

oder in dynamischer Technik durchführen lässt. Es ist bei entsprechender Erweiterung

des „field of view“ der CT-Bildgebung gelungen, die transmittierte Dosis in der Ebene

des EPID im vorhandenen Bestrahlungsplanungssystem vorauszuberechnen. Im

Vergleich der gemessenen 2D-Dosis (EPID dose map) mit der voraus berechneten kann

die Qualität sowohl der integralen Dosis als auch aller einzelnen Subfelder der IMRT

bewertet werden. Mit Hilfe einer praxisrelevanten Anordnung (Patientenphantom und

EPID-Phantom) konnten die wesentlichen Einflussgrößen (Geometrie, Feldgröße,

Streuverhalten und Einfluss auf die spektrale Energieverteilung) erfasst und in die

Kalibrierung zur Wasserenergiedosis übernommen werden. Als „proof-of-principle“

wurde die Methode an zwei typischen Lokalisationen (Kopf-Hals-Region und

Abdomen) getestet. Es wurden dazu typische IMRT-Einstellungen mit jeweils 7

Subfedern herangezogen. Die voraus berechneten und gemessenen 2D „dose maps“

wurden verglichen und die Differenzen (difference maps) dargestellt sowie die

sogenannte Gammaanalyse durchgeführt. Aufgrund der Ergebnisse erweist sich diese

neue Methode der 2D EPID-Dosimetrie als sehr empfindliche und praxistaugliche

Kontrolle und Verifikation der täglichen Patientenbehandlung in der Strahlentherapie.

Mit der Messung der transmittierten Dosis werden sowohl Parameter der Beschleuniger

(MU, Kollimatorwinkel, Lamellenpositionen, Gantrywinkel) als auch

Patientenparameter (Lagerung, Feldposition, Bewegung, Gewichtsverlust,

Organfüllzustand z.B. Blase Rektum) erfaßt.

Page 7: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

3

Page 8: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

4

1 Introduction

1.1 Facts about cancer

Cancer is a general name for considerable number of complex diseases that could

influence any part of the human body. Cancer is an illness when the abnormal cells

divide without control. Cancer cells have the potential to spread to other parts of body

through the blood and lymph systems (metastasis). According to the WHO report

(2014), there were 14.1 million new cancer cases, 8.2 million cancer deaths and 32.6

million people are living with cancer (within 5 years of diagnosis) in 2012 worldwide.

Figure 1-1 shows the most frequent types of cancer among men and women. It is

estimated that the annual cancer cases will increase from 14 million in 2012 to 22

million in next two decades (WHO, 2014). Cancer treatment needs careful decision of

Figure 1-1.: The most common cancer incidences and mortality in the world in 2012

based on the WHO report. For men, lung cancer and for women, breast cancer were the

most common forms of cancer in the world (WHO, 2014).

Page 9: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

5

one or more treatment modalities, such as surgery, chemotherapy and radiotherapy. The

goal of the treatment is either to relieve patients of tumor burden and pain (curative

treatment) or significant life prolongation while improving quality of life of the patients

(palliative treatment).

The choice of cancer treatment depends on many factors such as location, grade and the

stage of tumor. Radiotherapy is one of the cancer treatment modalities, either alone or

in combination with other modalities. Radiotherapy is utilizing ionizing radiation to

control cell growth. There are three main ways of using ionization radiation for patient

treatment (Figure 1-2). They are categorized to: systematic radiation therapy, internal

radiation therapy (Brachytherapy) and external beam radiation therapy (EBRT). In

systematic radiation therapy radioactive drugs, which are called radiopharmaceutical (in

liquid forms), are administrated to the patients to take part in the metabolism in order to

accumulate in the affected organ. Using radioactive iodine for treatment of

hyperthyroidism is a well-established example for that kind of treatment. In

brachytherapy, encapsulated radioactive sources are placed directly into or near the

volume to be treated. The main types of brachytherapy based on the source

implementation are: Intracavitary (sources placed into body cavities close to tumor),

interstitial (sources are implanted within the tumor volume), surface or mould (sources

are places over the tissue to be treated), intraluminal (sources are placed in a lumen)

and intravascular (a single source is placed into arteries) (IAEA, 2005).

Page 10: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

6

Figure 1-2.: Different modern radiotherapy modalities.

1.2 External Beam Radiotherapy (EBRT)

The most widely used modality of radiation therapy is EBRT (Figure 1-2). In this

modality the radiation is generated in a machine and guided through different elements

to the patient body. One of the most common types of machine that is widely used in

EBRT is a medical linear particle accelerator (shortened to linac, see Figure 1-3). Linacs

accelerate electrons using power RF microwave fields. For treatment, either the

electrons itself are directed to the patient, e.g., for treatment of superficial tumors or

they are used to generate MV X-rays in a target inside the head of accelerator.

The delivery of the administrated radiation dose to the patient’s body is achieved with

different methods and techniques. By utilizing three dimensional (3D) imaging systems,

a 3D treatment planning system and also 3D dose delivery, it it possible to conform the

dose volume to the prescribed target volume while keeping the dose to healthy organs

bellow their tolerance dose. This technique, which is called 3D conformal radiotherapy

(3D-CRT), is still used in many radiotherapy centers.

Radiotherapy

Brachytherapy

External beam

radiotherapy

Systematic

radiotherapy

EBRT techniques:

- 3D-CRT

- IMRT

- IGRT

- IMAT

- SRS/SRT

Type of implant:

- Intracavitary

- Interstitial

- Surface (mould)

- Intraluminal

- Intravascular

Targeted radiotherapy

Page 11: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

7

An important device to achieve such field conformation is the Multi Leaf Collimator

(MLC) which installed as the most down-stream device in the accelerator head. It is

made up of several individual leaves of about 5 mm at isocenter, which can move

individually to conform the field shape to the projection of the target volume. MLCs

were first introduced to market in 1980 to replace conventional field-shielding blocks

(Figure 1-4).

Recently, modern radiotherapy machines are equipped with different imaging devices

that are able to scan the patient anatomy before (or/and during) delivery of a treatment

fraction. This technique of dose delivery is called Image Guided Radiotherapy (IGRT).

It can help to reduce the radiation side effects while ensuring that the relative target

volume and some reference points are in the same geometrical position as in the

reference imaging data set used for treatment parameter optimization (treatment

planning).

By using a 3D localization system (stereotactic apparatus) and multiple narrow beams

there would be possible to have more precise dose delivery system. This type of EBRT

is called Stereotactic Radiation Therapy (SRT). Using this technique used for

Figure 1-2.: A medical linac system (ONCOR) equipped with portal imaging

system (manufactured by SIEMENS).

3

Page 12: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

8

intracranial lesions in a single fraction, it is called Stereotactic Radiosurgery (SRS)

(Mayles et al., 2007; Khan, 2010).

1.2.1 Intensity Modulated Radiation Therapy (IMRT)

Intensity Modulated Radiation Therapy (IMRT) is an advanced form of EBRT. IMRT

allows additional optimization of the dose distribution inside the planning target volume

(PTV) as well as in organs at risk (OARs), which are demanding individual restrictions

for dose exposition. In IMRT multiple inhomogeneous field overlaid gives a

homogeneous but much more conformal coverage of the target. A specialized planning

system is indispensable for this kind of optimization considering the potentials of the

individual Linac. IMRT techniques could be categorized into two groups: in the first

method; multiple static MLC shaped fields are utilized. This technique is called Step

and Shoot IMRT. In another IMRT method, dynamic MLC systems are utilized. In this

method leaves are in motion during radiation while on the step and shoot IMRT, leaves

move only when the beam is off. Figure 1-3 shows a conventional and an IMRT plan

created for a patient with oropharynx tumor. Combination of IMRT and arc

radiotherapy, called Intensity Modulated Arc Therapy (IMAT), came to reality with

actual hardware and software development. IMAT allows quicker and more precise

Conventional field

Blocks

Shaped Field at Isocentre Shaped Field at Isocentre

MLC Leaves in Linac Head

Figure 1-3.: Methods of field shaping devices; conventional field shaping

blocks and MLC system (www.medphysfiles.com).

4

5

Page 13: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

9

radiation treatment. While in IMRT techniques the degree of freedom in treatment

planning increased during the optimization (including couch, gantry, collimator,

energy,…), it is able to improve dose uniformity inside the target volume and decrease

the delivery time. Immobilization and position verification of patients play a key role in

the promising outcome of IMRT and IMAT treatments (Teoh et al., 2011).

Figure 1-5.: a) Conventional plan for an Oropharynx patient with three open beams, the

yellow arrows represent beam with equal intensities. b) Simplified schematic IMRT

plan with multiple beams, the arrows represent different segments of each beam with

different intensities. The plan is optimized to minimize the dose to parotids and spinal

cord.

1.2.2 Biological point of view of radiotherapy

Radiation biology is not an issue of this thesis, but it is the most important reason for

development of EPID-dosimetry. Ionizing radiation leeds to various alterations in living

mamalian cells. In particular, it causes lesions and alterations at the genetic code

presented in the deoxyribonucleic acid (DNA). That may leed to cell death,

chromosomal alterations or only distinct mutations or activations in the genetic code.

Some of this reactions with the endpoint “cell death” are intended for the tumor cells

inside the CTV (clinical target volume) and are unintended in the surrounding healthy

tissue. The most importand factor for this kind of reactions is the absorbed dose inside

tissue respectively inside singel cells. So treatment planning and application aim at

a) b)

Page 14: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

10

realizing proper dose values inside CTV and vice versa dose values in healthy tissue as

small as possible.

Even so that the respond of living cells, tumor cells and healthy tissue, doesn’t diversify

too much, radiation therapy is possibly effective when the optimized dose distributions

are realized in the daily application over treatment periods. This principle demonstrated

in Figure 1-6 by two sigmoid curves, the first one is Tumor Control Probability (TCP)

(curve A) and the second one is the Normal Tissue Complication Probability (NTCP)

(curve B). Any variance of patient’s position and condition (e.g., body weight, pain,

filling conditions of intestine, colon, bladder etc.) may influence and downgrade the

quality of treatment and thus reduce the therapeutic window. Patient movement or only

movement of inner organs especially patient’s breathing have to be checked or

compensated during treatment. Daily checks using EPID dosimetry have the potential

for detecting most of these parameters mentioned above. It’s a valuable control for

getting the right radiation dose at the righ position (IAEA, 2005; Beasley et al., 2005).

Figure 1-6.: The therapeutic window in radiotherapy, the curve A represent the TCP,

curve B the probability of complications (IAEA, 2005).

1.3 Quality concept in radiotherapy

A high level of accuracy is an essential requirement in radiotherapy as a multi-step

complex procedure in order to reach the desired tumor control rates, in combination

with minimized complication rates. The Quality Assurance (QA) program in

radiotherapy diminishes uncertainties and inaccuracies in dosimetry, treatment

Therapeutic window

Page 15: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

11

planning, equipment performance, and treatment delivery meanwhile improving

dosimetric and geometric accuracy as well as precision of dose delivery. Uncertainty

should be managed prospectively and dose errors should be kept within acceptable

tolerances. QA programs ensure quality of radiotherapy treatment intrinsically for

patient safety and avoidance of accidental exposure. Consequently, patient safety is

automatically integrated with the QA program. Several QA guidelines have been issued

by a number of organizations such as International Atomic Energy Agency (IAEA),

American Association of Physicist in Medicine (AAPM) and European Society for

Therapeutic Radiology and Oncology (ESTRO) (AAPM, 1994; ESTRO, 1995; IAEA,

1998).

1.3.1 In vivo dosimetry in EBRT

One of the QA methods in EBRT to improve the quality of patient treatment is in vivo

dosimetry (IVD). In vivo dosimetry refers to measuring techniques for recording and

verifying the dose actually received by the patient during treatment. It helps to detect

the major errors and to assess clinically relevant differences between planned and

delivered dose for each individual patient. IVD could help to limit the escalation of a

problem to following treatment session of a particular patient. In addition, it could avoid

systematic errors which could influence other patients. Moreover, if no error is detected,

IVD provides a conformation record of correctly delivered dose within the expected

tolerance. Until now there is no general consensus in radiotherapy organizations for the

cost effectiveness of IVD and its routine implementation is not widespread. While few

errors are detected it has been argued that most treatments are done without any error

and only small number of patient could benefit from rectifying errors. However, recent

series of major accidents in radiotherapy centers would have been ruled out if

appropriate IVD methods had been utilized (IAEA, 2013). The IVD feed back role in

radiotherapy chain is presented in Figure 1-7.

Page 16: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

12

Figure 1-7.: The total radiotherapy procedures and role of in vivo dosimetry for the

assessment of out come of treatment.

1.3.2 Methods of IVD

There are different devices and dosimetric methods for patient IVD, which could be

used as real-time and passive dosimeters (methods that need processing after radiation).

Real time detector systems include diode, metal oxide semiconductor field effect

transistor (MOSFETs), and Electronic Portal Imaging Device system (EPID). These

types of noninterventional detectors are able to provide dosimetric information to

evaluate the dose during treatment session. In contrast, the passive detectors such as

Thermoluminescent Dosimeters TLDs, Optically Stimulated Luminescent Dosimeters

(OSLDs) and also Film (radiographic and radiochromic) are not able to provide

immediate measurement. The passive detectors require some finite time from minutes to

Localization -

CT simulation

Delineation of

volumes

Treatment

planning

Verification and

simulation

Patient set-up

Treatment

delivery

Follow-up

Record and

verify

In vivo

dosimetry

Machine QC

Beam data

collection

Page 17: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

13

hours for read-out. Table 1-1 summarizes the physical and dosimetric characteristics of

some of available detector system in EBRT (ESTRO, 2006; Mijnheer et al., 2013).

Table 1-1.: Summary of characteristics of various detectors for in vivo dosimetry in

EBRT.

Detector

Parameter

Diode MOSFET TLD OSLD Film EPID

Cables + + - - - -

Bias voltage + + - - - -

Temperature

dependence + +/- - - +/- -

Energy

dependence + + + + + +

Angular

dependence + + - -

Dose + + + + + +

Dose rate + + - - - -

Readout delay no no yes yes yes no

Main

advantage

Good

reproducibility,

immediate

reading

Immediate

reading, minor

fading

No cable, few

correction,

reusable

No cable,

reusable

2D dose

distribution,

high

resolution,

permanent

record

2D and 3D

dose

distribution,

permanent

record

Main

disadvantage

Cumbersome

calibrations,

cable

connection

Limited

lifetime, high

cost

Labor

insensitive,

specific

reading

equipment

Short life,

dependence on

accumulated

dose, specific

reading

equipment

Cost, specific

scanning

device

Cost, limited

availability of

commercial

software

Note: + means no dependence on parameter; - means dependence response to the

parameter; +/- means in some type of detector could be dependent

1.3.3 Electronic Portal Imaging Device systems

In radiotherapy, acquisition of images with therapy beam is called portal imaging

(Langmack, 2001). Modern digital imaging devices were developed as tools to acquire

portal images of the treatment area and to reduce patient setup errors. Utilizing port

films as control systems is time consuming and labor intensive and could decrease the

throughput in a busy radiotherapy clinic. Therefore the need for a fast and accurate

Page 18: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

14

portal imaging system to intensify verification of radiation therapy arouse the

development of on-line electronic portal imaging devices (EPIDs) (Herman et al.,

2001). An EPID system consists of a two-dimensional flat radiation detector, which is

mounted on a semi-robotic arm fixed to the Linac chassis (see figure 1-1). The collected

information from the detector system is transferred to the main computer system of the

linac for processing and control. The detected portal images are displayed to the

therapist typically in direct comparison to digitally reconstructed radiographs (DRR)

that are calculated based on the planning computed tomography (CT) scan representing

the nominal position of the patient. Currently there exist different types of detector

systems which are listed bellow:

- Fluoroscopic detectors;

- Ionisation chamber detectors;

- Amorphous silicon detectors (a-Si);

In the fluoroscopic detector system, a scintillator plate converts the high energy X-ray

to light. A camera captures a fraction of this emerging light and transforms it to a video

signal which is digitized and stored by the processing system. The EPID system with

ionisation chamber detectors is working based on a grid of several ionization chambers.

The 2D map of gathered signal will be converted to gray scale images. Currently the

most commonly applied type of EPID is the a-Si EPID. These types of EPIDs are able

to acquire high quality presenting high-resolution images while requiring very small

dose values. The main principle of the a-Si EPID is based on a two-step process. In the

first step, the X-rays are converted into visible light by means of a metal plate and

phosphor screen (typically copper, Cu, and terbium-doped gadolinium oxysulfide, Gd2

O2 S:Tb, respectively). In the second step, the generated photons are absorbed by the

underlying of a-Si photodiodes which create electrical signals (Blake et al., 2013a). A

simplified schematic drawing of the a-Si EPID system in which the MV X-ray converts

to electrical signal is shown in Figure 1-8.

3

Page 19: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

15

1.3.4 Applications of EPID system

Since EPID is a fast image acquisition tool, it has different applications in radiotherapy.

It could be utilize for some QC tests at linacs, and also for detecting geometric and

dosimetric errors. It should be mentioned that the geometric accuracy is limited to:

- Uncertainties in a particular patient set-up;

- Uncertainties in the beam set-up;

- Anatomical and physiological changes of patient.

Nowadays many clinics are utilizing EPID systems for verifying patient positioning

during radiotherapy treatment and also measuring patient organ motion during

treatment. In addition, EPID can be used as QA tool to check the functioning of

radiotherapy equipment (e.g. MLCs) and for beam and patient dosimetry (IAEA, 2005).

1.4 In vivo dosimetry with the EPID system

Several groups have developed different methods for checking the planned dose

administrated to the patients. They can be categorized in two groups; pre-treatment

(without patient) verification and in vivo dosimetry (with patient), each approach

having its own benefits and limitations. Pre-treatment verification can detect a large

number of errors in the radiotherapy chain before starting treatment, but if there is any

unexpected change at the time of treatment, it will be missed. It is not easy to translate

the influence of pre-treatment discrepancies on the real dose distribution delivered at the

day of occurrence (McDermott et al., 2007; van Elmpt et al., 2008). Possible sources of

Figure 1-4.: schematic drawing of the EPID system, interaction of X-ray with different layers

generate electric signals.

8

Page 20: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

16

errors could be changes in the linac output, incorrect accessory, MLC failure, patient

changes (tumor shrinkage, weight loss, organ motion, anatomical changes in patient

since planning CT) or even changes in patient setup, including obstruction from

immobilization devices (van Elmpt et al., 2008; Chytyk-Praznik et al., 2013). Many of

these errors could be detected if all parameters would be checked during the patient

treatment via patient transmitted dose.

In vivo dosimetry (IVD), as part of a QA program, could detect major errors and

indicate the relevance of the differences between planned and delivered treatments

(Mijnheer et al., 2013). Comprehensive overviews of techniques for IMRT verification

using point detectors, two-dimensional (2D) arrays, films, and also EPIDs, have been

provided by several authors (Low et al., 2011; Mijnheer et al., 2013). The EPID is an

excellent in vivo dosimetry system since it may be used to verify dose in all dimensions

(0, 1, 2, 3 and 4D).

EPIDs can be implemented as a tool for transmission (transit), or non-transmission

(non-transit) dosimetry, i.e. with or without an attenuation object (patient or phantom)

in the beam path. Both methods can be categorized into a forward and back-projection /

dose reconstruction approach. Figure 1-9 shows above mentioned approaches (van

Elmpt et al., 2008). In the forward approach, the verification of fluence or dose is done

at the position of the imager, while in the second approach measurements are made

outside the attenuating medium, and the verification is performed inside the patient or

phantom (Wendling et al., 2006; Wendling et al., 2009).

Page 21: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

17

Figure 1-9.: Three different arrangement of EPID dosimetry and verification of dose

distribution, a) non-transmission pre-treatment approach, b) non-transmission during

treatment approach, c) transmission approach for pre-treatment and also during

treatment verification.

1.5 Purpose of study

The purpose of this study was to develop a simple method to evaluate and use EPID

detectors as a dosimetric system distally of the patients treated with step-and-shoot

IMRT. Intense and subtle calibration was necessary for transferring the EPID imaging

information into dose values. By comparing the calibrated and transformed transit dose

images at the EPID level with corresponding dose distributions calculated with a

commercial TPS changes to the nominal conditions can be achieved. For this reason the

calculation matrix of the TPS must be extended to incorporate the complete geometry of

the patient plus the EPID. Based on measurements, the EPID images are calibrated and

transformed into dose ‘portal dose images’. No additional algorithms such as Monte

Carlo or convolution methods were necessary. As proof-of-principle, the feasibility of

this method for 2D transmission dosimetry has been tested on homogenous and

anthropomorphic phantoms. To our knowledge this is the first study using an a-Si EPID

in combination with a commercial TPS to verify transit dose distributions.

EPID

EPID

EPID

Patient Patient or

phantom

Page 22: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

18

Page 23: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

19

2 Accuracy of a commercial treatment planning system

for transit dose calculations at large distances

2.1 INTRODUCTION A treatment planning systems (TPS) is an essential tool in each radiotherapy

department. Besides their main task, calculating and optimizing a treatment plan, they

have potential for many other applications including the calculation of portal dose

distributions of patients undergoing radiotherapy. Several groups developed in-house

methods for dose prediction at large distances to compare them with measured EPID

data. One of the most common methods is using an independent dose calculation.

Several models have been developed to predict the dose from zero dimension (point

dose) to three dimensional dose distributions (Chytyk and McCurdy, 2009; Fidanzio et

al., 2010; Chytyk-Praznik et al., 2013). Using an independent dose calculation

algorithm instead of a commercial TPS might have several advantages, but on the other

hand needs extra resources to develop and process such a system. There are other

studies, which used existing TPS systems for pre-treatment verification of treatment-

plans (at gantry angle zero). Advanced TPSs are capable of performing are capable of

performing pre-treatment verification using phantoms, as well as of predicting the

transit-dose behind real patients (e.g., in EPID level). This method could be an easy and

fast solution because it does not need extra data collection for a new calculation system.

(Mohammadi et al., 2008; Tyler et al., 2013)

One of the intrinsic problems of a TPS for dose calculation in the extended distances

(e.g. EPID level) is that they are confined to do the calculation within the field of view

(FOV) defined by the CT in use, which usually can’t encompass the geometry of

EPIDs. Reich et al. used a horizontal setup; both phantoms (patient and EPID) were

located on the CT table. They used the TPS to calculate the transmitted dose in the

EPID phantom (Reich et al., 2006). This method is able only to calculate the dose with

gantry and table angle of 90 degree. As substitute to the last method, Mohammadi et al.

extended the acquired CT matrix slices from 512×512 pixels to 1024×1024 pixels.

They set the Hounsfield Units of the extended region in a way that they had density of

air (ρ = 0 g/cm3). In that study, EPID was modeled as virtual phantom (a uniform slab

with water density) and the EPID phantom was in the fixed position and for each beam

the CT images of patient were rotated in the opposite direction of gantry rotation

(Mohammadi et al., 2008).

Page 24: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

20

The goal of our study was to check the accuracy of the transit dose calculation by a

commercial TPS in comparison to the ion chamber measurements as gold standard. In

this chapter our method to overcome geometrical limit of the TPS for dose calculation

in the extended distance were explained. Several influential parameters, such as field

size and air gap size, in transmision dose measurement were studied. 14 IMRT fields

were used to test the accuracy of transit dose calculation by the TPS. A calibrated 2-

dimensional ion chamber array (as our reference instrument) was utilized to validate the

calculated transit dose at the EPID level.

2.2 MATERIALS AND METHODS

2.2.1 Equipment

In this study oncor linacs (SIEMENS) with OPTIFOCUS TM 82- leaf MLC systems

using 6 and 15 MV photon beams were used. A calibrated 0.3cm3 air vented ionisation

chamber (IC)(Type 30016, PTW, Freiburg, Germany) and a 2D array detector system

(Matrixx IBA Dosimetry, Schwarzenbruck, Germany) were utilized. The Pinnacle TPS

(V9.4 by Philips Medical Systems, Fitchburg, WI, USA) was used to create treatment

plans and calculate the transit dose distributions. The phantom was scanned in a CT

SOMATOM Sensation open (Siemens Healthcare, Erlangen, Germany).

2.2.2 2D dose map calculation using TPS

To overcome the FOV limit in the Pinnacle planning system, a script has been written to

expand the CT images of patients and copy the treatment plan to the expanded CT

images (Figure 2-1). A second script, which is called planar dose script (PlDo), creates a

CT#expansion

air

EPID%model%

(water)

Figure 2-1.: CT slice expansion with extra surrounding air layer and EPID phantom

Page 25: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

21

virtual EPID phantom (as substitute for the real EPID) and calculates the dose maps for

each beam (Figure 2-2). The virtual EPID-phantom is created always at fixed distance

to isocenter and perpendicular to each beam. This virtual phantom with density of water

is positioned in the same position as the real EPID system. The Source to calculation

Plane Distance (SPD) and also the size of this virtual phantom could be defined by user

base on patient treatment condition. The PlDo script creates for each beam a virtual

EPID phantom. Afterwards the EPID dose-map is calculated, the phantom removed and

the procedure repeated for the next beam (Figure 2-3). Analysis of EPID responses to

different parameters is necessary for transit dose measurements but is beyond the scope

of this chapter and will be discussed in following.

Figure 2-3.: Workflow of the two scripts in the TPS to calculate the transit dose maps

for each beam at given angle.

2.2.3 Reference transit dose measurement setup

For studying the influence of different parameters on transit dose measurement, a

reference setup (patient/EPID phantom) was defined by varying distance and/or

dimension. The first phantom, which called patient phantom, is consisting of solid water

slabs (RW3, PTW, Freiburg, Germany) in dimension of 30×30×22cm! (Figure 2-4).

This phantom located at patient table while the isocenter of machine always is located at

Read beam parameters

(gantry angle, isocenter)

Create virtual

phantom for the

beam

CT images expansion (512*512 to

1024*1024)

Calculate 2D

Fluence map for

the beam

Next beam

Remove the virtual

phantom

Figure 2-2.: Representation of virtual phantom for one beam at 288 degree for a pelvic patient

Page 26: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

22

the center of this phantom. The second phantom, EPID phantom, is created by PlDo

script at the EPID position. The isocenter distances to measurement plane in the EPID

phantom were chosen variability between 20 and 60 cm. The patient/EPID phantom

configuration was used for all measurements in this study. The ion chamber located in

central axis of the beam in the RW3 EPID phantom in depth of 2 cm. To measure the

transit dose of IMRT fields, a 2D array ionisation chamber system was also positioned

perpendicular to the beam in the position of the EPID phantom.

2.2.4 Accuracy of transit dose calculation by TPS

2.2.4.1 Field size effect

In the current study, checking the accuracy of transit dose calculation in our TPS the

field size has been varied and the dose predicted by the TPS has been compared to ion

chamber measurements. The field size effect was studied in presence of attenuating

media as showed on the patient/EPID phantom setup in Fig. 2b. In this setup the EPID

model was fixed distance of 140 cm(source to measurement plane distance SPD). The

field sizes were ranging from 3×3 to!20×20!"!. The ion chamber measurements were

performed at the center of beam at 2 cm depth inside the RW3 phantom. To find the

best matching depth in the TPS, calculations have been done for fixed field sizes at

different depths 1 to 5 cm for both beam energies.

Beam iso-center = phantom center

Measurement plane

SAD=100cm

Air gap RT table

Patient phantom

EPID phantom

Figure 2-3.: Reference patient/EPID phantom set up 4

media as showed on the patient/EPID phantom setup in Figure 2-4. In this setup the EPID

Page 27: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

23

2.2.4.2 Phantom thickness effect

When the patient (phantom) thickness varies, the absorption and scattering pattern and

also the spectrum of the beam changes. Based on detector structure, there might be

increasing or decreasing response to these changes. In the current study, the accuracy of

Collaped Cone algorithm (CC) was tested by changes of phantom thickness. The

patient/EPID phantom (Figure 2-4) was used to study this effect with the field size of

10×10cm!!using 30 MU. The patient phantom thickness ranged from 10 cm to 34 cm

(in the steps of 4 cm). The transit dose calculations have been done by TPS under equal

conditions and were compared with the ion chamber measured values.

2.2.4.3 Air gap effect

The transmitted beam is a combination of primary photons and photons scattered from

the patient (phantom). Modifying the air gap size between patient (phantom) and

detector will change the proportional contribution of the scatter component. Hence the

change of this parameter should be taken into account for transit dose calculation and

measurements. In the current study the defined reference patient/EPID phantom (Figure

2-4) was used. The source to detector distance (SDD) ranged from 130 cm to 160 cm in

the increment of 10 cm.

2.2.4.4 Evaluation of transit dose calculation for IMRT fields

Finally for testing the accuracy of transit dose calculation by the TPS, two IMRT plans

for 6MV and 15MV beams have been created. Each plan consisted of 7 beams and the

field’s areas ranged from 66 to 133 cm2 and the numbers of monitor units ranged from

78.6 to 309.2 MU. The IMRT fields were delivered to the above-mentioned phantom

setup (Figure 2-4). By using PlDo script, the 2D transit dose for each beam was

calculated using the TPS. The 2D array ionization chamber detector system was used to

measure the transit dose of IMRT fields. The detector was placed in the identical

position as the EPID hardware and virtual EPID-phantom. The source to measurement

plane distance was 140 cm (SPD = 140 cm). The internal and external buildup layer for

the detector system was 2cm for both energies. The !-evaluation method was utilized to

compare the calculated and measured dose maps. A dose-difference criterion of 3% and

distance-to-agreement criterion of 3 mm was selected for the evaluation.

Page 28: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

24

2.3 RESULT

2.3.1 Field size effect

Figure 2-5 a) and b) show the normalized dose calculation of different field sizes in

depths 1 to 5cm at the virtual phantom in the TPS for 6 and 15MV beams. The

horizontal and vertical axes represent one side (cm) of quadratic field sizes. The results

of dose variation is normalized to a standard field of 10 cm (100 cm2). The calculated

values were compared to the measured dose by the IC at the depth of 2 cm in the patient

phantom. The measurement depth of 1cm is smaller than the !!"# for both beam

energies (1.6 cm for 6 MV and 3 cm for 15 MV), there seems to be more instability in

the calculation of dose at this small depth. On the other measured depths (2 to 5 cm), the

depth of 2cm has the closest accordance to the IC reading.

Figure 2-5.: a) and b) the field size effect for 6 and 15MV beams (calculated output

factor by TPS at different depths in comparison to measured output factor by ion

chamber).

0 5 10 15 200.7

0.8

0.9

1

1.1

1.2

1.3

1.4

field size (cm)

no

rma

lise

d r

esp

on

se

X6MV

IC

d=1cm

d=2cm

d=3cm

d=4cm

d=5cm

0 5 10 15 200.7

0.8

0.9

1

1.1

1.2

1.3

1.4

field size (cm)

no

rma

lise

d r

esp

on

se

X15MV

IC

d=1cm

d=2cm

d=3cm

d=4cm

d=5cm

a) b)

Page 29: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

25

2.3.2 Phantom thickness effect

Figure 2-6 shows the phantom thickness variation of calculated and measured EPID-

dose for 6 and 15 MV beams. The horizontal and vertical axes represent the phantom

thickness (in cm) and normalized dose values, respectively. All data were normalized to

the reference phantom thickness of 22cm. As expected, an exponential decrease in the

transmitted dose values was detected for IC measured and TPS calculated values

(Berry et al., 2010; Blake et al., 2014). As shown in Fig. 3.c, the lower energy (6MV)

has greater variation to the changes of phantom thickness than the 15 MV data.

2.3.3 Air gap effect

The IC response by changing the air gap is presented in Figure 2-7. The horizontal axis

represents the air gap and the vertical axis the normalized dose values. All dose

measurements and calculations are normalized to reference phantom setup and field size

(phantom thickness of 22 cm, air gap size of 27 cm and field size of 10×10!cm!).

10 15 20 25 30 350.6

0.8

1

1.2

1.4

1.6

1.8

phantom thickness (cm)

no

rma

lise

d r

esp

on

se

IC X6MV

TPS X6MV

IC X15MV

TPS X15MV

Figure 2-4.: The phantom thickness effect for 6MV and 15MV beams. Measurements

and calculation was done with the reference field size of 10×10 cm2.

2-6

(Berry et al., 2010; Blake et al., 2014). As shown in Figure 2-6, the lower energy (6MV)

Page 30: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

26

By Figure 2-7.: the air gap effect for 6 MV and 15 MV beams. The measurements and

calculations were done with the reference field size of 10×10 cm2.

By increasing the air gap less lower-energy photons from the phantom reach the

detector. As shown in fig 3.d. the air gap effect greatly depends on the gap size while it

is less sensitive to the energy of the beam.

2.3.4 IMRT plan test

Figure 2-8. a) and b) show the 2D transit measured and calculated dose map for one of

the tested IMRT fields (field FA7). The Fig.4. c. is the 2D !γ -evaluation map of this

field. In table 1 the detail of tested IMRT field which included the beam energy, number

of MU, field size area and also the number of segments presented. For matching the

calculations and measurements, all beam gantry position set to zero. The result of !γ -

evaluation of measured and calculated dose map of IMRT fields summarized in Table

2-1.

15 20 25 30 35 40 45 500.7

0.8

0.9

1

1.1

1.2

air gap size (cm)

no

rma

lise

d r

esp

on

se

IC X6MV

TPS X6MV

IC X15MV

TPS X15MV

Fig2-7

Fig 2-8

detector. As shown in figure 2-7 the air gap effect greatly depends on the gap size while it

is less sensitive to the energy of the beam.

the tested IMRT fields (field FA7). The Figure 2-8 is the 2D γ - evaluation map of this

Page 31: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

27

Figure 2-8.: Comparison of measured and calculated dose map of one IMRT field a)

measured 2D transit dose by Matrixx system, b) calculated transit dose map by the TPS,

c) 2D gamma evaluation result.

a) b)

c)

Page 32: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

28

Table 2-1.: Detail of tested IMRT fields and result of gamma evaluation for 14 IMRT

fields.

Field

name

Energy

(MV)

Number

of MU

Field area

(cm2)

Number of

segments

% area

γ>1 γ -mean Std. Dev.

FA1 6 249.9 114 12 2.54 0.22 0.33

FA2 6 173.0 109 8 4.10 0.24 0.36

FA3 6 103.7 69 6 1.33 0.11 0.26

FA4 6 266.7 128 13 3.82 0.26 0.36

FA5 6 114.1 66 7 1.61 0.12 0.27

FA6 6 135.4 109 6 2.55 0.22 0.34

FA7 6 187.3 118 8 2.08 0.23 0.33

FB1 15 309.2 127 17 6.87 0.26 0.41

FB2 15 87.5 68 7 3.11 0.15 0.03

FB3 15 168.1 126 9 6.96 0.28 0.42

FB4 15 153.1 121 9 6.85 0.27 0.42

FB5 15 126.8 133 7 6.69 0.29 0.40

FB6 15 217.0 123 11 7.19 0.26 0.41

FB7 15 78.6 72 5 2.87 0.15 0.32

2.4 DISCUSSION Ideally the dose calculation should match with the measured dose. McCurdy et al.

studied the accuracy of pencil beam algorithm for dose prediction portal imaging

system by comparing with Monte Carlo (MC) simulation (McCurdy and Pistorius,

2000). It was reported that the percentage error was roughly proportional to the field

size in the selected air gaps (0 to 40cm). For their clinical frequent field sizes (smaller

than 20×20!cm!), the maximum deviations between predicted and simulated scatter

pattern was not greater than 0.8%. Dahlgren et al. utilized the collapsed cone (CC)

algorithm to calculate the portal dose and used it as reference for comparison with

measured portal dose images (Dahlgren et al., 2006). Using the EPID with matrix ion

chamber for measuring portal images, on average; 99% of field area passed the gamma

criteria (3%/3mm). Lee et al. studied the dose change by variation of the field size for

the open beam in different depths of water (with out any attenuator in the beam) (Lee et

Page 33: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

29

al., 2009). They reported that the measured out put factor in the depth of 5 and 3cm of

water had the same response like the EPID system for 6 and 18MV beams. In the

current study using CC algorithm, the out put factor at different depths of virtual

phantom calculated. The calculated output factor at depth of 2cm has the most similarity

to the measured values by IC for both beam energies.

The impact of varying phantom thickness on transit dose measurement reported by

Blake et al. (Blake et al., 2014). They compared the EPID response to the MC simulated

response. They have shown that by increasing the phantom thickness; the deviation

between EPID measured values and expected values increased. In the current study, it

has been shown that there are slight differences in ionisation chamber measured dose

and calculated values by changes of phantom thickness. The other important issue in

transit dose study is the air gap size. By increasing the air gap, more lower energy

scattered photons don’t reach the detector region (Chytyk-Praznik et al., 2013). The

influence of air gap size on transit dose calculation by the CC algorithm in comparison

to MC simulation was reported by Dahlgren et al (2002). With their setup, for 6MV

beam the deviation was between -3% and 0.3% (at depth of 1.7cm) and for 15MV beam

the deviation was between -2.5% and 0.6% (at depth of 2.9cm). In the current study, the

influence of changing air gap size on the dose response in measured dose with

ionization chamber and collapsed cone algorithm studied. For both energies the air gap

effect is similar.

In our study for the tested IMRT field on the reference setup, the average of γ -value

(γ!"#$) and the mean percent of area that fail the gamma criteria (γ > 1) for the plan

with 6MV beam were 0.20± 0.06 and 2.58± 1.05, respectively. For the IMRT field

with 15MV beam the mean !γ value (γ!"#$) averaged to 0.24± 0.06, and the mean

percent of area that fail the gamma criteria were 5.79± 1.92 .

2.5 Conclusion It could be concluded that the commercial TPS system that was used in this study, is

able to correctly calculate the transit dose in the extended distance. On the other hand

this method could predict the 2D transit dose at EPID level and could be used as

reference to compare with 2D transit dose measured with the EPID system.

Page 34: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

30

Page 35: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

31

3 2D Patient transit EPID dosimetry for treatment

fraction dose verification

3.1 Introduction The search for better QA and EPID dosimetry methods is ongoing, there are no good,

sufficient commercially available options after more than 20 years of published ideas,

and a new, simple, robust technique is still needed. To improve accurate dose delivery

methods, many modern and complicated radiotherapy techniques were developed, such

as Intensity-Modulated Radiation Therapy (IMRT), Intensity-Modulated Arc Therapy

(IMAT) and stereotactic radiotherapy/radiosurgery (SRT/SRS). The more complicated

the technique is, the higher the demands for verification of the treatment. Independent

of treatment technique, there will always will occur random or systematic errors, which

could affect the outcome of radiation therapy. These errors might be due to anatomical

movement, physiological changes of organ position or dimension, patient mis-

positioning, or change in treatment machine parameters. Failures might accrue during

treatment preparation (inaccuracy in the delineation process) or during the treatment

delivery (van Herk, 2004; Margalit et al., 2011). In external beam radiotherapy, Image-

Guided Radiotherapy (IGRT) techniques have helped to manage changes in patient

anatomy and to check the reliability and reproducibility of radiation field geometry with

respect to the patient anatomy. However pre-treatment imaging techniques cannot

provide direct information about the accuracy of the delivered dose.

Different organizations such as IAEA, AAPM and ESTRO have provided several

comprehensive quality control (QC) and quality assurance (QA) recommendations, to

prevent the errors mentioned above in the radiation therapy chain of procedures (which

are rapidly developing). Most of these guidelines were written to support prevention of

failures prior to treatment. However, regarding the check of dose received by patients

during treatment, there are only a few international publications that make

recommendations. There are, however, various countries that have specific legal

requirements for patient QA, such as France and Sweden (AAPM, 2005; ESTRO, 2006;

IAEA, 2013).

Page 36: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

32

Besides the advantages of using an EPID as a dosimetry tool, there are several issues

and problems that should be considered. Some of these are related to the EPID structure

and its design, and should be considered in all forms of dosimetry measurement. First of

all, this device was initially designed as an on-line imaging device to improve patient

positioning. The EPID is constructed from high atomic number materials (far from

water equivalent), which results in a response very different from conventional

dosimetry media (e.g. water). As a consequence, EPIDs have an undesirable over-

response to low energy photons in x-ray beams (Teymurazyan and Pang, 2012).

Another problem related to the EPID structure is that it is attached to the linac by a

holding arm, which causes a non-uniform back-scatter pattern to some types of EPIDs

(Rowshanfarzad et al., 2010). The imaging software of the EPID corrects the individual

pixel sensitivity and the wash out of the beam profile shape to make a flat image (flood

field correction). For dosimetry, the beam shape should be reintroduced to the images;

while the sensitivity correction should be maintained (Greer, 2005). Moreover, the a-Si

detector collects signals in a time frame manner that potentially leads to ghosting and

lag effects (McDermott et al., 2006).

The problems mentioned above should be considered for all forms of a-Si EPID

dosimetry, alike transmission and non-transmission. However, when there is an object

(phantom or patient) in the beam path, additional effects occur, which should be

considered for transit dosimetry. The EPID is an energy dependent system and the

transit beam spectrum varies with size and composition of an attenuating object. The air

gap distance also influence the energy spectrum impinging the EPID because of

increasing loss of scattered low energy photons with increasing distance. These effects

might lead to an over-response in some regions, which should be corrected (Grein et al.,

2002; Nijsten et al., 2007a). Another possible object in the beam path is the treatment

couch, which might attenuate the beam differently, depending on the beam angle and

couch position. Several studies recommend modelling the treatment couch to improve

transmission dosimetry results (Ali et al., 2011; Berry et al., 2014).

Several methods have been developed, and some of them are commercially available, to

perform pre-treatment verification of IMRT and IMAT with an EPID using non-

transmission methods, i.e. without any attenuator (phantom) in the beam. In this way it

is possible to verify, for instance, the 2D absolute fluence of IMRT fields, which can be

done at any planned gantry angle (Nicolini et al., 2008; Bailey et al., 2012; Nicolini et

Page 37: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

33

al., 2013). Van Esch et al.(2004) developed an algorithm for predicting 2D portal dose

images at the EPID level in air (non-transmission dosimetry). They used a pencil beam

dose calculation algorithm, which was commissioned entirely based on EPID

measurements. Van Zijtveld (2009) also used the predicted dose images without a

patient in the beam, and then by using the patient’s planning CT scan and an “imaginary

equivalent homogeneous phantom”, included corrections of the primary transmission

and the scattered radiation from the patient.

Several groups have developed methods to check the patient dose by means of EPID-

based transmission dosimetry using point, 2D and 3D approaches. Piermattei et al.

(2012) reported the use of EPID in vivo dosimetry, in which they verified the isocenter

dose of patients treated with 3D conformal radiotherapy techniques with 6, 10 and 15

MV beams. Cilla et al. (2014) expanded this method by considering a new component

(fluence inhomogenity index) for IMRT fields. Nijsten et al. (2007b) performed in vivo

verification of the dose at 5 cm depth on the central beam axis. They correlated the dose

measured with an EPID with dose values in a patient by using a back-projection model

with a semi-empirical relationship. Berry et al. (2012) used the same algorithm as van

Esch et al., along with Monte Carlo simulations to simulate the effect of an attenuating

object in the beam (transmission dosimetry). This method is similar to the 'buildup

factor' calculations used in radiation protection. Nijsten et al.(2007a) developed a model

for dose calibration of a-Si EPIDs for 2D transit dosimetry, in which corrections for the

ghosting effect, image lag, and beam profile were implemented. In addition, a

convolution model was included to correct for the field size dependence and scattering

of the EPID response. Persoon et al. (2012) used this method to analyze the inter-

fractional trend of the patient dose throughout the course of treatment. In The

Netherlands Cancer Institute (NKI), an algorithm was developed that uses a

measurement-based “convolution/correction approach” to reconstruct the 3D dose

distribution in a patient from the EPID-based transmitted dose. They compared the

reconstructed EPID dose with the calculated treatment planning system (TPS) 3D dose

distribution (Wendling et al., 2009; Wendling et al., 2012). This technique is

implemented in NKI clinic as an in vivo treatment verification tool for IMRT/IMAT and

non-IMRT plans (Olaciregui-Ruiz et al., 2013).

McNutt et al.(McNutt et al., 1996a; McNutt et al., 1996b) created a model based on the

collapsed cone superposition (CC) algorithm to predict the portal dose images through

Page 38: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

34

patient CT scans in the extended phantom that included the EPID. The authors checked

their algorithm for a single 10cm×10cm field transmitted through different phantoms.

Dahlgren et al. (2002) also used an in-house developed version of the CC algorithm, to

predict the portal dose images of two different types of EPID (liquid-filled matrix ion

chamber and camera-based EPID). They tested the validity of this method for several

rectangular and MLC-shaped fields for several thicknesses of homogenous phantom

(Dahlgren et al., 2006). Reich et al.(2006) modelled the portal dose transmitted through

homogenous phantoms in a commercial TPS (Pinnacle, Philips Medical Systems). They

also used the CC algorithm in combination with the extended phantom and presented

the verification of dose distributions behind phantoms of different thicknesses irradiated

with rectangular fields of a 6 MV beam. The authors found that changes in thickness of

~1 mm corresponded to dose changes of 0.6%. These effects could be detected in the

transmitted dose distributions. Mohammadi et al.(2008) used this technique and

presented the result of pre-treatment verification of an IMRT plan for prostate case on

anthropomorphic phantom with 6 MV beam. In both studies a scanning liquid

ionization chamber (SLIC) EPID was used.

The purpose of this chapter is to develop a simple method to evaluate the dose at the

EPID detector behind patients treated with step-and-shoot IMRT using portal dose

images measured with an amorphous silicon (a-Si) EPID. In this method, transit dose

images at the EPID level are compared with corresponding dose distributions calculated

with a commercial TPS using extended CT scans. Based on measurements, the EPID

images are calibrated and corrected without using additional algorithms such as Monte

Carlo and convolution methods. As proof-of-principle, the feasibility of this method for

2D transmission dosimetry has been tested on homogenous and anthropomorphic

phantoms. To our knowledge this is the first study using an a-Si EPID in combination

with a commercial TPS to verify transit dose distributions.

3.2 Materials and methods

3.2.1 Equipment

Two identically built and commissioned Siemens Oncor AvantGard (Siemens Medical

Solutions, Concord, CA, USA) linacs were used in this study. Both were equipped with

the OPTIFOCUS TM

82-leaf MLC system and have 6 and 15 MV photon beam energies.

Page 39: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

35

Portal images were acquired using two OPTIVUE 1000ST Electronic Portal Imaging

Devices (EPIDs). The EPID system consists of an XRD Perkin Elmer a-Si detector

(XRD 1640 AL7), which has a frame rate of 3.5 Hz, an active imaging area of

1024×1024 pixels and a pixel resolution of 0.4 mm. To reduce the noise level, the EPID

images were resampled to a resolution of 2mm. Each stored image is an integral of

several frames in which the number of frames depends on the irradiation time (Greer

and Popescu, 2003).

The TPS used for this study was Pinnacle v9.4 (Philips Radiation Oncology Systems,

Fitchburg, WI, USA). The TPS was used to generate IMRT step-and-shoot plans.

Transit dose images at the EPID level were predicted in the TPS using the collapsed

cone (CC) algorithm.

3.2.2 Transit dose distribution predicted at the EPID level

It has been shown that the dose distribution of the CC algorithm results in acceptable

values of the transit dose at the EPID level. For this reason, the EPID was modelled in

the TPS as an “extended phantom”, whereby the virtual EPID is included in the CT data

set.

Calculation of the 2D transmitted dose in the planning system is limited to the field of

view (FOV) of the CT scanner. TPSs are optimized to perform the calculations within

the image matrix given by the CT scan, which usually makes calculation at EPID level

impossible. In our department we use a SIEMENS CT scanner (SOMATOM Sensation

open). To overcome the limited FOV problem in our TPS, in-house developed software

(Matlab v12, [Mathworks, Natick, Massachusetts, U.S.A.]) was used to expand each of

the CT matrix slices from 512×512 pixels to 1024×1024 pixels, and setting the

Hounsfield Units of the additional pixels outside the patient such that ρ = 0 g/cm3, to

simulate air. For patient treatment the source to EPID distance is usually 140 cm in our

clinic.

To estimate the transit dose at the EPID level in the TPS, the EPID was modelled as a

phantom, defined by a volume of water, 30cm×30cm×5cm, ρ=1.00g/cm3

. This EPID

model needed to be perpendicular to the beam direction and always at a fixed distance

from the radiation source. The source-to-surface distance (SSD) was set to 138cm. The

calculated transit dose map !Ctr!,! was calculated at a depth of 2cm in the phantom,

so the measurement plane was located at 140cm from the source. (Figure 3-1). The

Page 40: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

36

depth of 2 cm was chosen as a compromise value of the measured build-up depths

necessary for 6 MV (1.6cm) and 15 MV (3.0cm). The parameters and specify a certain

point in this 2D plane.

Figure 3-1.: Reference combination of patient/EPID setup up for (a) measurements and,

(b) TPS prediction model.

3.2.3 EPID dose measurements

3.2.3.1 Flood field correction

Since EPID systems were primarily invented for imaging, there are several corrections

implemented to improve image quality (Antonuk, 2002). Dark-field (DF), bad pixel

map (defective pixels) and flood field (FF) images are the main components for EPID

image correction. A DF correction is an image acquired without radiation to account for

the offset in response due to background influences. A bad pixel map compensates for

defected pixels by replacing their value with an average response of 'good neighbors'.

The gain of detector pixels is not uniform, they have different sensitivities and the FF

correction is necessary to normalise the pixel response over the detector area. This

correction is essential for imaging but it completely removes the beam profile

information (Greer, 2005). For dosimetry, this correction should be removed, while the

correction of the individual pixel sensitivity across the EPID should remain (Warkentin

et al., 2003). In the present study, FF images (gain files) that are normally used to

automatically correct EPID images were collected. For each energy, a 2D model flood

field correction, !!!"(!,!) , was fitted to the collected images. Since the pixel

Page 41: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

37

sensitivity map is incorporated in the flood field correction image in our system, a

model is necessary to remove the flood field without sensitivity correction. This model

was used in all subsequent measurements rather than the original gain image, which

also included noise and small local deviations.

3.2.3.2 Open field calibration for transmission dosimetry

Dose characteristics of a-Si EPIDs have been reported in detail by other groups

(McDermott et al., 2004; Winkler et al., 2005; Chen et al., 2006; McDermott et al.,

2006; Nijsten et al., 2007a). For transmission dose measurements, the recorded signal is

a combination of primary and scattered radiation. Scattering of the primary fluence

occurs in the patient and also in the EPID. The scattered fraction of transmitted dose

depends on many factors such as patient (or phantom) thickness, field size of the beam

and also size of the air gap between the patient and detector (Herman et al., 2001).

McDermott et al. (2004) showed that extra build-up material (2.5 mm copper) could

decrease the sensitivity variation of the EPID system, which depends on the air gap size.

But adding an extra build-up layer is not generally permitted by the manufacturers and

may raise warranty issues if the additions fall outside the original EPID specifications.

As previously stated, the goal of this study was to compare measured transmitted dose

to the calculated planar dose map at the EPID level at any planned gantry angle.

Therefore, in this study, we decided not to use extra build-up material.

A reference set-up was defined to calibrate the EPID for transit dose measurement,

whereby the EPID is calibrated using the transmitted beam spectrum (for measurement

and also in the TPS). The reference patient/EPID set-up in this study consisted of two

phantoms (Figure 3-1). The first phantom (patient phantom) was a 30×30×22cm3 RW3

(PTW Freiburg) homogeneous phantom in which the isocenter of the beam was always

located at the center of the phantom (at 11cm depth). The second phantom (EPID

phantom) was a 30×30×5cm3 RW3 phantom, which was located at the EPID position of

the linac, with 138cm SSD.

The linac was calibrated according to the TRS398 protocol (IAEA, 2000). The protocol

requires the calibrated linac to deliver 100 cGy at a depth of dmax when a beam of 100

MUs is delivered with a field size (FS) of 10×10cm2 (!!!"). To calibrate the EPID

system, the patient/EPID phantom setup (Figure 3-1) was used under the same linac

reference conditions (!!!") for each energy, also with 100 MUs. The patient-phantom

was positioned at an SSD of 89 cm. Under these reference conditions, the isocenter of

Page 42: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

38

the machine was located at the midplane of the phantom. A central axis calibration

factor (!") was defined at the center of the EPID phantom (!!,!!) as:

!" =

!ion !!!,!!!", !!, !!

!" !!!,!!!", !!, !!

(1)

With !ion(!!!,!"!", !!,!!) being the transmitted dose behind a 22cm thick phantom,

measured using an ionization chamber in the EPID phantom at a depth of 2cm.

!"(!!!,!"!", !!,!!) is the average central pixel value (in a 1×1cm2 region of interest).

According to some reports, the a-Si EPID response to the integral dose is linear up to

250 MU, which is sufficient for the scope of treatment plans, therefore a single

universal calibration factor was obtained to convert EPID pixel values (!") to dose

(Greer and Popescu, 2003). The CF was calculated for both 6 and 15 MV beam

energies.

3.2.3.3 Field size correction factor

In transit dose measurements, there are several scatter components that affect the EPID

reading. They include head scatter, phantom scatter and also scatter within the structure

of EPID itself. Furthermore, owing to their structure, EPID systems respond differently

to varying scatter conditions, compared to other detector devices such as ionization

chambers (Lee et al., 2009). In currecnt study, a TPS was used to calculate the transit

dose images using an EPID model. The response of the EPID as a function of field size

in the central region (12×12 pixels) was investigated and compared to the calculated

dose at depth of 2 cm of the EPID model in the TPS. The reason for this correction is

that the TPS is designed to calculate the dose-to-water based on ionization chamber

measurements. It is therefore necessary to convert the EPID response to dose-to-water

when comparing its response to TPS dose calculations. The field size correction factor

(!"cor) was defined as the ratio of calculated dose values !"(!!!,!"! , !,!) over the

measured values from the EPID system !"(!!!,!"! , !!,!!), normalized to that of the

reference field (FS=10×10cm2).

!"cor !"! =!"(!22,!!!,!0,!0)!

!"(!22,!!!,!0,!0).×

!!ref!22,!!10,!0,!0

!!ref!22,!!10,!0,!0 .

−1

(2)

Where !"! is the side length of square field sizes from 3×3cm2 to 20×20cm

2.

Page 43: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

39

For square fields it is easy to obtain and use !"cor but in IMRT, fields consist of

multiple irregular shaped subfields. The purpose of this study was to verify IMRT

fields, and therefore an estimate of the field size was required. For this purpose the

Sterling formula (Sterling et al., 1964) was used to find the equivalent square field

(EQS) of each IMRT field segment. Unless stated otherwise, all field sizes referred to in

this manuscript are defined at the isocenter. A Matlab code based on an edge detection

algorithm was developed to calculate the area (!!EPID) and perimeter (!!

EPID) of each

segment based on EPID images at the EPID level. The area/perimeter ratio was then

scaled to determine the ESQ at the isocenter level.

ESQisocenter = 0.714×

!×!!

EPID

!!EPID

(3)

The factor 0.714 rescales the ESQ from EPID level (140cm) to isocenter level (100cm).

3.2.3.4 Phantom thickness correction

Changing patient (phantom) thickness leads to changes in the absorption and scatter

pattern of the beam inside the phantom, which results in various beam hardening effects

at the exit side of the attenuating medium. This effect could increase or decrease the

EPID response with respect to the TPS-calculated dose values (Pecharroman-Gallego et

al., 2011; Jung et al., 2012). The reference patient / EPID phantom set-up (Figure 3-1)

was used to investigate this effect. In all measurements, the imager was located at a

fixed distance from the beam source (140 cm). The center of the phantom was always

located on the central axis of beam, and at the isocenter. The phantom thickness

correction factor (!!"# ) was defined as the ratio of the calculated dose value

!"(!! ,!!!", !!,!!) for different phantom thicknesses !! and the corresponding values

determined from EPID images; !"(!! ,!!!", !!,!!). The ratio was normalized to the

corresponding reference values (!" = 10×10cm2, phantom thickness of 22cm).

!!"# !! =!"(!!,!!10,!0,!0)!

!"(!!,!!10,!0,!0).×

!!ref!22,!!10,!0,!0

!!ref!22,!!10,!0,!0 .

−1

(4)

where the phantom thickness (!!) range was 10 cm (for head and neck patients) to 34

cm (for pelvic patients), in steps of 4cm. For IMRT fields, a single thickness correction

is performed per segment, where the thickness is measured along the field axis passing

through the CT scan of the patient.

Page 44: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

40

3.2.3.5 Over-response correction

It is well reported that the a-Si EPID detectors have an over-response to lower energy

photons due to their high atomic number structures (McCurdy et al., 2001; Kirkby and

Sloboda, 2005; Blake et al., 2013b). This effect causes deviations in dose measurements

within non-water equivalent media (EPID) and in the calculated dose of the EPID plane

in the TPS. Differences exist inside and also outside the primary beam region, but in the

outer primary beam region there is a higher proportion of low energy photons, which

lead to higher over-response of the EPID system. Several authors, including Nijsten et

al. (2007a) and Greer (2005) investigated this effect. Each IMRT field consists of

several segments, and thus the summation of transmitted dose maps of all segments

could be largely affected by overestimated dose values in the out-of-field regions.

To overcome this problem, an empirical Over-Response Correction (ORC) filter was

defined. Firstly, the filter normalizes each pixel response !!!(!,!) of each segment to

the maximum reading of the entire detector area:

Figure 3-2.: a) Comparison of the flood field corrected EPID image and the TPS profile

for a 10×10 cm2 field for a 6MV beam at the EPID level using the patient/EPID

phantom set-up. Points M and N are used for the over-response correction. b) Linear

over-response correction. The horizontal axis is normalized EPID reading (nPVk !,! )

and the vertical axis is the ORC!.

nPVk !,! =

!!!(!,!)!

max !!! !,! |!∀!!,!

(5)

In the second step, a linear correction was developed based on experimental

observations. Two points (M and N, see Figure 3-2. a)) were determined in both

calculated and measured profiles. Point M, defined as the starting point of this

N

M

Page 45: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

41

correction, quantifies the position (in 2D) in which the normalized nPV of measured

and calculated profile is 35%. Point N is defined as the point in which the EPID

response is 50% higher than the TPS value. As shown in Figure 3-2.b), the line that

passes through these points when irradiating a symmetric 10x10 cm2 field is defined as

the linear over-response correction (ORC!) of the EPID:

ORC! !,! =

1 for$nPV! !,! ≥ 0.35

!0×nPV! !,! + !0 for$nPV! !,! < 0.35

(6)

In this formalism bo and co are the linear parameters of the line intersecting N and M,

determined for a symmetric 10×10cm2 field.

3.2.4 Transit dose measurement formula

Finally, by considering all factors mentioned above, the measured transit dose,

!k!"(!,!)for each segment ! is the product !"!(!,!) of the relevant segment with the

universal calibration factor (CF), flood field correction (FFC(x, y)), field size correction

of that segment (!!cor !"! ), ), thickness correction factor for the relevant beam angle

!!"# !! , and over-response correction of the segment ORC! !,! which calculated in

the matlab software.

!k!"!,! = !"!(!,!)×!"×FFC(x,y)×!!cor !"! ×!cor(!!)×ORC! !,! (7)

The total measured transit dose for each field, !tot, is the sum of the transmitted dose of

all relevant individually calculated segments, !tr!!,! , where ! is the total number of

segments in the IMRT field:

!tot = !k!"!,!

!

!!!

(8)

3.2.5 Proof-of-principle IMRT plan verification

To determine the accuracy of the method, verification measurements were carried out

for two treatment sites. In this study, as in clinical practice, 6MV beams were used for

head-and-neck and 15MV beams for prostate IMRT plans. All segments of each field

were individually corrected, summed by equation (8) and compared with calculated

transit dose. The plan verification was first performed on a cubic phantom at zero gantry

angle. To verify the method with respect to changes in geometry, homogeneity and

gantry angle, the verification procedure was also performed on an anthropomorphic

phantom under conditions similar to patient treatments, with the planned gantry angles.

Page 46: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

42

Calculated and measured dose were analyzed with global γ-evaluation method (dose-

difference criterion: 4%; distance-to-agreement criterion: 3 mm). For each field, gamma

1% (γ1%), the mean gamma (γmean) and the percentage of gamma values above 1 (γ >1)

were calculated (Stock et al., 2005).

3.3 Result

3.3.1 Initial system parameters

3.3.1.1 Flood field correction

Figure 3-3 shows the model of the fitted flood field correction, based on a Gaussian fit

of the machine flood field image to the specific height of measurements (140 cm).

Figure 3-4 shows the normalized cross-line EPID profiles of an open field (10×10 cm2)

on the patient/EPID phantom.

−200 −100 0 100 2000.95

1

1.05

1.1

1.15

1.2

1.25

off axis distance (mm)

Flo

od F

ield

Cor

rect

ion

fact

or (

FF

C)

FFC X6MV

FFC X15MV

Figure 3-1.: Flood Field correction for X6MV and X15MV at 140cm from the beam

source.

3

Page 47: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

43

Figure 3-4.: Normalized EPID cross-line profile of both energies for a field 10×10cm

2

on reference patient-EPID phantom set-up. a) 6MV photon, b) 15MV photon. The

dotted lines are the raw EPID profiles and the solid lines the EPID profiles corrected

with FFC.

3.3.1.2 Open field calibration for transmission dosimetry

The TPS calculations and IC measurements at depth of 2cm in the central EPID

phantom were compared and the average difference was ( 2.41± 0.03 )% and

(2.44± 0.11)% for 6 and 15 MV, respectively. The absolute dose calibration for both

energies was obtained under reference conditions (Figure 3-1) for the Patient/EPID

phantom set-up with a reference open field of 10×10 cm2, SSD 140 cm and 100 MU.

For 6 MV, the calibration factor CF was and for 15 MV.

3.3.1.3 Field size correction

For each energy, a second-degree polynomial curve was fitted to the normalized relative

EPID response:

!"cor !"! = !!×(!"!)

2+ !!× !"! + !! (9)

For the 6MV photon energy, the fit parameters were !! = 1.792×10!!!"

!! , and

!!! = −1.905×10!!!"

!! !!! = 1.172 and for 15MV, !!! = 7.705×10!!!"

!! ,

!! = −3.173×10!!!"

!! and !!! = 1.242. For IMRT fields, equation (3) was used to

calculate EQS for each segment. Every segment was then corrected individually based

on its measured field size.

Page 48: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

44

Figure 3-5.: a) The variation of the EPID response and the TPS calculated value with

field size, on the central axis beam (at a depth of 2 cm) for both energies. b) Field size

correction factor for both energies. All data are normalized to the corresponding

10×10 cm2 field.

3.3.1.4 Phantom thickness correction

Figure 3-5 illustrates the difference in EPID reading and TPS calculated values for both

photon energies with changing phantom thickness in the patient/EPID phantom on the

central axis. The phantom thickness variation was determined over the interval 10 to

34cm. Figure 5(c) presents the normalized phantom thickness corrections for both

energies.

10 15 20 25 30 350.8

0.85

0.9

0.95

1

1.05

1.1

1.15

Phantom thickness (cm)

Thic

kness c

orr

ection f

acto

r

TCor X6MV

TCor X15MV

10 15 20 25 30 35

0.6

0.8

1

1.2

1.4

1.6

Phantom thickness (cm)

Norm

aliz

ed r

esponse

Thickness variation response X15MV

TPS

EPID

10 15 20 25 30 35

0.6

0.8

1

1.2

1.4

1.6

Phantom thickness (cm)

Norm

aliz

ed r

esponse

Thickness variation response X6MV

TPS

EPID

Figure 3-2.: Variation of TPS values and EPID response with changing phantom thickness

for (a) 6MV and (b) 15MV photon beam. (c) The normalized phantom thickness correction

factor for both energies. All data are normalized to the value of the corresponding reference

22cm phantom thickness.

field size (cm)

field size correction factor

Thickness correction factors

e

x

6

6

3-6 c)

Page 49: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

45

For 6MV the relative difference in response was about ±2%, and for 15MV, there was

greater variation of about ±5%. To simplify use, a first degree polynomial was fitted to

each data set:

!cor !! = !!× !! + !! (10)

For 6MV photon energy, the fit parameters were !!! = −1.0269×10!!!"

!! and

!c! = !1.0242 , and for 15MV, !! = −3.7946×10!!!"

!! and !c! = !1.0851 . This

correction was necessary for measurements of beams transmitted through

anthropomorphic phantoms or patients.

3.3.1.5 Over-response correction

The normalized EPID and TPS profiles for a 10×10cm2 field (6MV beam) in the TPS

EPID model at a depth of 2 cm are shown in Figure 3-2. a). The EPID profile (after

flood field correction) in the irradiated area matches the calculated dose profile.

However, in regions outside the field there are significant differences. This effect

caused large deviations in IMRT field measurements, which consist of multiple

segments. Therefore the ORC! was used to correct each segment individually. The

parameters in equation (6) were obtained from the linear fit in Figure 3-2. b), resulting

in !! = 2 and !!= 0.3.

3.3.1.6 Transit dose measurement

Using above correction methods each segment is treated individually using equation (7).

For each IMRT field, all corrected transmitted measured dose distributions are summed

according to equation (8). Figure 3-7 is an example of the calibration and correction

steps for a segment of an IMRT field. This sample segment belongs to field A1 (Figure

3-8 and Table 3-1), a field from the head and neck plan (6MV).

3.3.2 IMRT plan verification

The first plan verified was a head-and-neck IMRT plan with 7 fields and 6MV beams.

The second plan was a pelvic IMRT plan with 7 fields and 15MV beams. The head-and-

neck plan was copied in the TPS and delivered to a slab phantom (thickness of 22cm)

and all beams were measured with the gantry angle set to zero. To test the effect of non-

flat geometry and a small degree of inhomogeneity, the pelvic plan was copied and

delivered to a pelvic phantom (with fields delivered at their planned gantry angle).

Page 50: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

46

Figure 3-7.: Calibration and correction steps of transmitted measured dose for each

single IMRT segment. a) EPID central raw cross-line profile. b) Calibrated EPID profile

and calculated dose profile (TPS). c) Correcting flood field (FFC). d) Field size

correction (!!cor ). e) Over-response correction (!"# ). f) Final 2D transmitted

measured dose.

3.3.2.1 Head-and-neck IMRT plan evaluation with a slab phantom

The number of segments, number of MUs and the field area of each field is presented in

Table 3-1for the head-and-neck case. The plan was delivered to the slab phantom

(thickness of 22cm) and the measured dose was compared with the planned dose for

each field. The results of the measurements and the gamma evaluation for one field

(A1) are presented in Figure 3-8.

DO

SE

(cGy

)

Page 51: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

47

Figure 3-8.: IMRT field evaluation for a 6MV head-and-neck IMRT field (A1). a) 2D

measured transit dose, b) 2D calculate transit dose. c) Inline and d) cross line profiles

are as indicated in a) and b); e) 2D gamma evaluation map (global gamma).

Distance (mm)

Dose (

cG

y)

TPS

−170 −100 0 100 170

−170

−100

0

100

170 0

2

4

6

8

10

12

14

16

18EPID

Distance (mm)

Dose (

cG

y)

−170 −100 0 100 170

−170

−100

0

100

170 0

2

4

6

8

10

12

14

16

18

−200 −150 −100 −50 0 50 100 150 2000

2

4

6

8

10

12

Distance (mm)

dose(c

Gy)

middle inline profile

TPS

EPID

−200 −150 −100 −50 0 50 100 150 2000

2

4

6

8

10

12

Distance (mm)

dose(c

Gy)

middle cross line profile

TPS

EPID

Gamma evaluation

Y(4

%/3

mm

)

20 40 60 80 100 120 140 160

20

40

60

80

100

120

140

160

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

2

a)

e)

d) c)

b)

Page 52: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

48

Table 3-1.: Results of the EPID dose verification of an IMRT plan (head-and-neck,

6MV beams, 7 fields) from slab phantom measurements.

field name number of

segments

number of

MUs

field area

(cm2)

γ1% γmean % area γ >1

A1 9 88.2 435.7 1.08 0.44 2.49

A2 7 75 336.1 1.00 0.37 1.02

A3 7 96 359.2 1.12 0.41 2.73

A4 13 148 426.7 1.10 0.44 2.32

A5 16 174 378.3 1.00 0.33 1.13

A6 8 96 335.4 1.08 0.44 2.09

A7 9 98 355.0 1.14 0.43 3.21

3.3.2.2 Pelvic IMRT plan evaluation with an anthropomorphic pelvic phantom

Figure 3-9 shows the pelvic IMRT plan with 15MV beams was verified with the

anthropomorphic pelvic phantom. This plan consisted of 7 beams at various gantry

angles. In this test the effects of varying geometry, couch effect and homogeneity were

included in the measured and calculated planar dose distributions at the EPID level. In

Table 3-2 the beam properties and the results of the gamma evaluations are presented. It

should be noted that all beams passed through the treatment couch (some of them

partially and some of them completely).

Page 53: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

49

Figure 3-9.: Evaluation of one field of the 15MV Pelvic IMRT plan (field B4 in Table

3-2). The plan was verified using the anthropomorphic pelvic phantom at the planned

gantry angles. a) 2D measured transit dose, b) 2D calculate transit dose. c) Inline and d)

cross line profiles are as indicated in a) and b); e) 2D gamma evaluation map (global

gamma).

Distance (mm)

Dose (

cG

y)

TPS

−170 −100 0 100 170

−170

−100

0

100

170 0

2

4

6

8

10

12

14

16

18EPID

Dose (

cG

y)

−170 −100 0 100 170

−170

−100

0

100

170 0

2

4

6

8

10

12

14

16

18

−200 −100 0 100 2000

2

4

6

8

10

12

14

16

off axis distance (mm)

dose(c

Gy)

middle inline profile

TPS

EPID

−200 −100 0 100 2000

2

4

6

8

10

12

14

16

off axis distance (mm)

dose(c

Gy)

middle cross line profile

TPS

EPID

Gamma evaluation

Y (

4%

/3m

m)

20 40 60 80 100 120 140 160

20

40

60

80

100

120

140

160

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

2

e)

a)

d) c)

b)

Page 54: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

50

Table 3-2.: Results of the EPID dose verification of an IMRT plan (pelvis, 15MV

beams, 7 fields) from anthropomorphic pelvic phantom measurements.

field

name

number

of

segments

Gantry

angle

phantom

thickness

number of

MU

field area

(cm2)

γ1% γmean % area γ >1

B1 8 30 24.5 116.9 627.0 1.40 0.47 5.49

B2 10 105 31.5 120.8 619.9 1.00 0.38 1.05

B3 8 130 29 87.0 582.4 1.38 0.50 6.44

B4 14 180 20 168.9 625.6 1.30 0.42 4.25

B5 9 230 29 105.1 560.5 1.30 0.38 3.48

B6 9 255 31.8 116.9 592.8 1.28 0.37 3.10

B7 7 330 24 91.9 588.0 1.18 0.42 2.37

3.4 Discussion

This dissertation deals with a method able to control and verify the administrated dose

distribution to patients during single treatment sessions (fractions). In this chapter the

feasibility of a method to verify patient transit dose during step-and-shoot IMRT

treatment was presented. The evaluation of measured dose is performed field-by-field at

the EPID level. The transit dose maps were calculated in the TPS using extended CT

images. This method has been used in previous studies for pre-treatment verification

(Reich et al., 2006; Mohammadi et al., 2008). In these studies, for non-zero gantry

angle measurements, the original CT images were rotated, and for each beam angle a

new rotated CT image set was imported and then transit dose maps of all beams were

calculated at zero gantry angle. This workflow manipulates the original CT data and

increases the risk of errors due to its complexity. In a non-transit study by Tyler et al.,

the XiO and Monaco treatment planning systems were used to calculate the fluence

maps in water at !!"#. In that study, the verification was performed with all gantry,

collimator and couch angles set to zero degrees; i.e it can only be used for pre-treatment

verification (Tyler et al., 2013). In our method, for each gantry angle, the predicted dose

Page 55: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

51

map was modeled in its nominal planned position, which makes it possible to use for

dosimetry during patient treatment.

3.4.1 Flood field correction

There are different methods to remove the flood field correction required for non-

dosimetry EPID image processing. A correction method was developed based on the

raw pixel response map and ion- chamber measurements at dmax (Greer, 2005). In

another study by Nijsten et al. (2007a) the dose profile in water was measured and the

data used for a deconvolution of the EPID images, while simultaneously the field size

dependence and beam profile correction were obtained. Since the raw EPID image is

divided by the flood field image in automatic processing, Wendling et al. removed the

flood field correction by multiplying it by each acquired dose image. (Wendling et al.,

2006) This is only possible when the sensitivity correction is performed separately to

the flood (and dark) field correction. In our study, as shown in Figure 3-2and Figure

3-3, the raw flood field images were used and a 2D FFC correction matrix was fitted to

these images. The advantage of this method is that while the original sensitivity

correction of the system is kept, the beam profile is put back to the images. In the

techniques of first two studies mentioned above, the corrections are based on a different

photon beam spectrum (in Perspex or water) in comparison to the clinical conditions

(only the EPID and no extra materials).

3.4.2 Open field calibration for transmission dosimetry

Calibration of the EPID for dosimetry has been done in many different ways. In similar

studies the EPID system has been calibrated without any object in the beam; e.g.

ionization chamber measurements at the EPID level were used as a reference (Wendling

et al., 2006; Nijsten et al., 2007a). In another study the dose at dref in a water phantom

(at a source-to-surface distance of 100cm) calculated by the TPS was used as reference

dose for calibrating the system (Lee et al., 2009). Berry et al. (2012) used the intrinsic

Varian calibration method, which is based on in-air calibration at the SSD of 100cm.

This method is not applicable for all therapy machines because the EPID cannot always

be positioned at 100cm from the beam focus. The problem with methods that calibrate

the EPID with only an open beam (without any attenuator), is that the photon beam

energy spectrum is different from the transmitted spectrum. The methods mentioned

above that calibrate the EPID with multiple correction steps (open and attenuated

3

4

Page 56: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

52

beams) have been proven to work, but are extremely complicated. The advantage of our

approach is that the calibration is performed with the transmitted beam from the

beginning (patient-EPID set-up), so while it is simple, the beam spectrum is still

representative of patient measurements. Without back-projecting the dose, there is no

need to isolate/remove the scatter within the EPID.

3.4.3 Field size correction

One of the most common methods for correcting for the field size effect is using a

convolution technique. In one study, ionization chamber measurements were used in a

mini-phantom at the position of the EPID, but without a full-scatter EPID phantom.

This method was designed for back-projection, so it was necessary to obtain a kernel

correction to remove the EPID scatter component, in combination with non-scatter and

full-scatter field size measurements. (Wendling et al., 2006). Nijsten et al. (2007a) also

used a ‘field size dependent kernels’ correction method, which was based on ionization

chamber measurements in water. In an other study, (Lee et al., 2009) determined the

changes in dose with respect to the field size at different water depths, and compared

them with corresponding EPID response. In the present study, an algorithm was

developed for plan verification for the forward transit approach. The effects of square

field size variations were studied for transit dose measurement in a patient / EPID

phantom combination (see figure 1b). One of the advantages of this method compared

to other studies is that the transit beam spectrum was used instead of that of the open

beam (without attenuator in beam path).

3.4.4 Phantom thickness correction

The energy spectrum of the photons that reach the EPID system changes dramatically

with the amount of attenuating material in the beam path to the detector. In one study,

the transmission through slab phantoms of varying thicknesses was studied for two field

sizes (10×10cm2 and 20×20cm

2), with three different beam energies (Pecharroman-

Gallego et al., 2011). According to the findings of this study, the beam hardening with

increasing phantom thickness could be described by adding an extra term in the

exponential attenuation, while there was only a very small deviation in the transmission

for different field sizes. (Jung et al., 2012) used Monte Carlo simulations and reported a

4 to 5% dose-response variation for thickness changes between 20 and 38cm. In our

study, a similar approach was used to correct for the thickness variation effect. This

Figure 3-1

Page 57: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

53

correction is relevant during transit dose measurements of patients, where there is a

wide range of patient thicknesses, especially at different gantry angles. For instance, the

radiological path length through the pelvic phantom varied from 21 to 32cm at different

gantry angles, mainly due to varying geometry, but with a small variation due to

inhomogeneity. The current factor corrects for mainly geometrical variation, and further

corrections for more extreme changes in homogeneity is the subject of future work.

3.4.5 Over-response correction

There are different methods to improve the over-response of EPID systems in regions

that receive a dose from lower energy photons. McDermott et al. (2004) used an

additional 2mm copper build-up layer for filtering out the low energy scattered photons.

This solution needs an extra modification of the EPID system and therefore it is not a

reasonable solution for all clinics. (Nijsten et al., 2007a) used an energy spectrum

correction based on a mean transmission correction within the treatment field. To obtain

a model for this energy spectrum correction, they acquired images during a dummy

session without a patient. This method needs extra measurement time and in addition

corrections for the out-of-field regions, based on the energy spectrum inside the field,

which could introduce dose differences in the out-of-field region.

In the present study, a new empirical correction method was introduced that corrects for

the over-response of the EPID in the out-of-field regions. The advantage of our method

is that it is a simple empirical approach, which corrects the pixel values in the out-of-

field region of each IMRT segment individually.

3.4.6 Transit measured data analysis

There are different methods to verify the delivered dose to the patient. Each method has

its advantages and disadvantages. An Italian group developed a method to verify the

isocenter point dose by measuring the transmitted dose with an EPID (Cilla et al.,

2014; Fidanzio et al., 2014). They compared the daily 2D transmitted images (without

calibration) and reported that they could detect changes in patient setup, machine output

and MLC behavior during each treatment. This method could be a fast method but the

limitation of this approach is that the 2D method relies on the first fraction of treatment.

If there would be an error in the first fraction, the reference would be wrong for the rest

of the treatment. A 2D approach was developed by Berry et al. (2012). They simulated

Page 58: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

54

the effect of attenuating material (phantom or patient), and modified a method

developed earlier (Van Esch et al., 2004), which is based on in-air verification. They

tested their model for pre-treatment verification using a fixed air gap and two ‘couch-to-

source distances’. In a follow-up study these authors reported the results of applying

this method for the verification of the transit dose of 9 patients treated with IMRT using

a 6MV photon beam (Berry et al., 2014). They reported that on average 95.7% of the

pixels passed the gamma evaluation with 5%/3-mm criteria for the beams that had no

interference with the couch. At the MAASTRO clinic in the Netherlands a 2D

calibration model for transit dose measurements was developed, which included an

energy spectrum correction (Nijsten et al., 2007a). They reported a satisfying gamma

evaluation across the entire EPID plane inside the field, but they found large dose

differences in out-of-field regions. They also determined their corrections for only one

SSD of 150cm. To use this method at other distances it needs an extra energy spectrum

correction while the patient scatter contribution to the EPID may also increase

significantly. In the presented new method, the patient scatter at different positions has

been taken into account by modeling and correcting the segments individually (see eq.

7) In addition, the over-response of the EPID, which could cause a problem in the out-

of-field region, has been taken into account. This method is simple and does not need

elaborate modeling time.

3.4.7 IMRT plan verification

Several studies have developed EPID transit dosimetry methods and similar results have been

reported. In the Netherlands Cancer Institute a back-projection method was developed for

verifying the patient treatment both pre-treatment and in vivo. The accuracy of the approach was

tested for a pre-treatment verification of a five-field IMRT prostate plan. In this test, the

midplane dose for each field was checked separately. The gamma criterion to compare EPID

versus film in this study was 2%/2mm. For all beams, the mean γ value was 0.4 and the mean

maximum γ index 1.2 and the mean % of pixels with γ>1 was 0.05% (Wendling et al., 2006).

This method is now applied routinely for the verification of all IMRT and IMAT treatments in

NKI (Olaciregui-Ruiz et al., 2013). Chung et al.(2011) used a 2D back-projection dose-

reconstruction algorithm to verify six conformal fields and also five IMRT fields in a slab

phantom using a 2D array detector (MapCHECK). With the gamma criterion of 3%/3mm, they

found for the IMRT fields (each with five segments) that three had a 100% gamma index

passing rate, and two had gamma index passing rates of 99.6% and 98.8%. In the present study

the global gamma evaluation of the 6MV head-and-neck IMRT plan resulted in γ1%, γmean and

Page 59: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

55

percent of area with γ>1 with mean values of [1.07 ± 0.06, 0.41 ± 0.04, 2.41 ± 0.81%] and

maximum values of 1.14, 0.44 and 3.2%, respectively. The 15 MV pelvic IMRT plan

verification was performed with an anthropomorphic phantom.

A similar approach was reported by (Mohammadi et al., 2008). According to their

results, for fields in the absence of the treatment couch, approximately 95% of the field

area passed the gamma criteria (2.54 mm and 3% of central axis dose). For the field

with a treatment couch in the beam path, the result reduced to 85%. In our study, for the

IMRT fields tested with the pelvic phantom at the planned gantry angles, the mean value

of γ!"#$, the mean value of γ!%, and mean percent of area with γ>1 were 3.74 ± 1.83,

0.42±0.05!!"# 1.26 ± 0.14 respectively. Therefore in this chapter the verification method

presented per field showed significantly better results compared to previously reported

methods that passed through the treatment couch.

Our method has been developed for treatment fraction dose verification and in this

paper the first results of a phantom study are reported. Issues that were investigated

concerned the dependence of different treatment plan parameters, such as field size,

phantom thickness, and gantry rotation on the measured transit doses, and as proof of

principle, the analysis of the results using gamma evaluation. It should be noted that

gamma evaluation alone is not sufficient and could miss clinically relevant dose errors

(Nelms et al., 2013). However, it is a very common tool that could help to detect major

errors. Compared to studies mentioned above, our method is simple and could be used

to detect errors outside pre-defined gamma alert criteria. Further work is in progress to

test the accuracy and sensitivity of our method with respect to other parameters such as

patient mis-positioning, changes in patient anatomy as well as errors in machine

parameters. In addition, the clinical implementation of this method for actual patient

treatments will be investigated.

3.5 Conclusion We developed a method for patient transit fraction dose verification of IMRT. In this

method an EPID was calibrated against an ionization chamber. The EPID system is

constructed from non-water equivalent materials, and is primarily designed as an

imaging device, therefore its response had to be corrected for 2D transit dose

measurement. In the protocol, corrected EPID data are compared to predicted dose

images calculated by the commercial TPS system used in our clinic. The presented

Page 60: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

56

method is simple and accurate, and is expected to form part of our QA program for

IMRT. In our study we have used global gamma evaluation criteria of 4% and 3 mm,

and it has shown that it could verify IMRT treatments delivered under clinical

conditions. Future studies including actual patient measurements at different treatment

sites are needed to determine its clinical potential.

Page 61: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

57

Page 62: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

58

4 Summary and conclusion In practice of radiation therapy, especially using advanced technologies like IMRT

(intensity modulated radiation therapy) and IMAT (intensity modulated arc therapy) a

daily check of series of treatment parameters is absolutely essential. Treatment

parameters include beam information, patient position and movement. This study

sought to show that EPIDs (electronic portal imaging devices), which often are installed

at treatment units, have got the potential for controlling patient position, quality of beam

shaping as well as the dose distribution administered to the PTV (planning target

volume).

Initially, the treatment planning systems (TPS) are designed for calculating and

optimizing the dose distribution in patients. In addition to this aim, they are able to

calculate the transmitted dose through the patient. The transmitted dose which is

impinging the EPID system could be measured separately for each beam direction

(projection) and any subfield. TPSs usually are optimized for using CT-data of patients

having a fixed frame for the field of view. Inside the treatment room the EPID is far out

of this matrix. For calculating the dose with the EPID, an expansion of this calculation

matrix of the TPS was necessary. Transit dose calculation should provide correct

information about both magnitude and position of dose distribution. Therefore

influential parameters in dose calculation and measurement in transit dose were reported

in this study.

Conclusion can be drawn that the commercial TPS system (Pinnacle), used in this

study, is able to calculate correctly the transit dose at extended distances. On the other

hand the pre-calculated transit dose distribution can be used for comparison with the

actually measured one at the EPID level under clinical conditions with a real patient.

The experiment in chapter three was carried out to calibrate the EPID system for transit

dose measurement. The EPID was calibrated against an ionization chamber. While the

EPID system is constructed from non-water equivalent materials, and is primarily

designed as an imaging device, its response had to be corrected for 2D transit dose

measurement. Influences of different treatment parameters such as beam energy, field

size and phantom thickness on the transit dose were investigated. In the protocol,

corrected EPID data are compared to predicted dose images calculated by the Pinnacle

TPS used in our clinic. As proof of principle measured transit dose maps using EPID

were compared with predicted maps from TPS.

Page 63: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

59

Since the invention of EPID systems, there were many studies to develop one universal

in vivo dosimetry method for patient treatment verification. Each work has advantages

and disadvantages, which were discussed in chapter two and three, but up to now the

EPID dosimetry methods are not yet gained wide clinical use because of limited

availability of commercial software solution (Mijnheer et al., 2013). The presented

method in this study is simple to calibrate and transform EPID response into transit dose

measurement. The calibrated EPID dose maps are comparable with the calculated dose

map by the planning system. By comparing the measured and calculated data it would

be possible to detect all deviations larger than the intrinsic error margin of the EPID-

TPS combination. For clinical quality management practice also error margins are

defined, which are usually bigger than the intrinsic ones.

Future research is necessary in multiple areas to have fast online EPID dosimetry

solution. They could be listed as bellow;

• The presented method has been tested using a slab phantom and also an

anthropomorphic pelvic phantom; it should be tested for dealing with bigger

inhomogeneity changes such as inside the thorax region.

• In order to evaluate the efficiency of the method it is recommended to simulate

clinically relevant errors (such as missing segment, wrong beam energy, patient

miss positioning,…).

• In the current report the physical principle of EPID transit dosimetry was

studied. It might be recommendable for acceleration and improvement of the

clinical procedure to create a graphical user interface capable of grabbing the

calculated data from the planning system and the measured data from EPID (this

part of work is on progress by a master student).

• The presented method has been tested on Siemens Oncor Linacs, which are

equipped with flat panel detectors manufactured by Perkin Elmer, Inc. The

calibration and correction methods of the EPID system could thus be used in a

similar way for the other machines that have identical detector systems (Elekta,

Vero4DRT). In future this method will also be tested with the Varian and

Vero4DRT machines in our clinic.

• There is a need to test EPID transit dosimetry for several patient treatment sites

to find the clinical applications and limits of this method.

Page 64: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

60

Page 65: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

5 Bibliography

AAPM!1994!Physical!aspects!of!quality!assurance!in!radiation!therapy.!!

AAPM!2005!Report!of!TG!62!of!the!Radiation!Therapy!Committee,!“Diode!in!vivo!

dosimetry! for! patients! receiving! external! beam! radiation! therapy,”! AAPM!

Report!No.!87!(Medical!Physics!Publishing,!Madison,!WI,!2005).!!

Ali!A!S,!Dirkx!M!L,!Breuers!M!G!and!Heijmen!B!J!2011!Inclusion!of! the!treatment!

couch! in! portal! dose! image! prediction! for! high! precision! EPID! dosimetry!

Med+Phys!38!377`81!

Antonuk! L! E! 2002! Electronic! portal! imaging! devices:! a! review! and! historical!

perspective! of! contemporary! technologies! and! research!Phys+Med+Biol!47!

R31`65!

Bailey!D!W,!Kumaraswamy!L,!Bakhtiari!M,!Malhotra!H!K!and!Podgorsak!M!B!2012!

EPID! dosimetry! for! pretreatment! quality! assurance!with! two! commercial!

systems!J+Appl+Clin+Med+Phys!13!3736!

Beasley!M,!Driver!D!and!Dobbs!H!J!2005!Complications!of!radiotherapy:!improving!

the!therapeutic!index!Cancer+Imaging!5!78`84!

Berry! S! L,! Polvorosa! C,! Cheng! S,! Deutsch! I,! Chao! K! S! and!Wuu! C! S! 2014! Initial!

clinical! experience! performing! patient! treatment! verification! with! an!

electronic! portal! imaging! device! transit! dosimeter! Int+ J+ Radiat+ Oncol+ Biol+

Phys!88!204`9!

Berry! S! L,! Polvorosa! C! S! and! Wuu! C`S! 2010! A! field! size! specific! backscatter!

correction!algorithm!for!accurate!EPID!dosimetry!Medical+Physics!37!2425`

34!

Berry! S! L,! Sheu! R! D,! Polvorosa! C! S! and!Wuu! C! S! 2012! Implementation! of! EPID!

transit!dosimetry!based!on!a!through`air!dosimetry!algorithm!Med+Phys!39!

87`98!

Blake!S!J,!McNamara!A!L,!Deshpande!S,!Holloway!L,!Greer!P!B,!Kuncic!Z!and!Vial!P!

2013a!Characterization!of!a!novel!EPID!designed!for!simultaneous!imaging!

and!dose!verification!in!radiotherapy!Medical+Physics!40!091902`11!

Blake!S!J,!McNamara!A!L,!Vial!P,!Holloway!L,!Greer!P!B!and!Kuncic!Z!2014!Monte!

Carlo! simulation! of! the! transit! dosimetric! response! of! an! a! `Si! electronic!

portal!imaging!device!Journal+of+Physics:+Conference+Series!489!012005!

Blake! S! J,! Vial! P,! Holloway! L,! Greer! P! B,! McNamara! A! L! and! Kuncic! Z! 2013b!

Characterization! of! optical! transport! effects! on! EPID! dosimetry! using!

Geant4!Med+Phys!40!041708!

Chen! J,! Chuang! C! F,! Morin! O,! Aubin! M! and! Pouliot! J! 2006! Calibration! of! an!

amorphous`silicon! flat! panel! portal! imager! for! exit`beam! dosimetry!Med+

Phys!33!584`94!

Chung! H,! Li! J! and! Samant! S! 2011! Feasibility! of! using! two`dimensional! array!

dosimeter! for! in!vivo!dose!reconstruction!via! transit!dosimetry! J+Appl+Clin+

Med+Phys!12!3370!

Chytyk! K! and! McCurdy! B! M! C! 2009! Comprehensive! fluence! model! for! absolute!

portal!dose!image!prediction!Medical+Physics!36!1389`98!

Page 66: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

62

Chytyk`Praznik!K,!VanUytven!E,! vanBeek!T!A,!Greer!P!B!and!McCurdy!B!M!2013!

Model`based!prediction!of!portal!dose!images!during!patient!treatment!Med+

Phys!40!031713!

Cilla!S,!Azario!L,!Greco!F,!Fidanzio!A,!Porcelli!A,!Grusio!M,!Macchia!G,!Morganti!A!G,!

Meluccio! D! and! Piermattei! A! 2014! An! in`vivo! dosimetry! procedure! for!

Elekta!step!and!shoot!IMRT!Phys+Med!30!419`26!

Dahlgren!C!V,!Eilertsen!K,! Jørgensen!T!D!and!Ahnesjö!A!2006!Portal!dose! image!

verification:! the! collapsed! cone! superposition! method! applied! with!

different!electronic!portal!imaging!devices!Phys+Med+Biol!51!335`49!

ESTRO! 1995! Recommendations! for! a! Quality! Assurance! Programme! in! External!

Radiotherapy!(Booklet!No.2).!!

ESTRO!2006!Methods!for!In!vivo!Dosimetry!in!external!radiotherapy,!2!ed,!physics!

for!clinical!radiotherapy,!Booklet!no.!1,!ESTRO,!Brussels.!!

Fidanzio!A,!Greco!F,!Mameli!A,!Azario!L,!Balducci!M,!Gambacorta!M!A,!Frascino!V,!

Cilla!S,!Sabatino!D!and!Piermattei!A!2010!Breast!in!vivo!dosimetry!by!EPID!J+

Appl+Clin+Med+Phys!11!3275!

Fidanzio!A,!Porcelli!A,!Azario!L,!Greco!F,!Cilla!S,!Grusio!M,!Balducci!M,!Valentini!V!

and! Piermattei! A! 2014! Quasi! real! time! in! vivo! dosimetry! for! VMAT!Med+

Phys!41!062103!

Greer!P!B!2005!Correction!of!pixel! sensitivity!variation!and!off`axis! response! for!

amorphous!silicon!EPID!dosimetry!Med+Phys!32!3558`68!

Greer!P!B! and!Popescu!C!C!2003!Dosimetric! properties! of! an! amorphous! silicon!

electronic! portal! imaging! device! for! verification! of! dynamic! intensity!

modulated!radiation!therapy!Med+Phys!30!1618`27!

Grein!E!E,!Lee!R!and!Luchka!K!2002!An!investigation!of!a!new!amorphous!silicon!

electronic!portal!imaging!device!for!transit!dosimetry!Med+Phys!29!2262`8!

Herman!M!G,!Balter!J!M,!Jaffray!D!A,!McGee!K!P,!Munro!P,!Shalev!S,!Van!Herk!M!and!

Wong! J!W!2001!Clinical! use! of! electronic! portal! imaging:! report! of!AAPM!

Radiation!Therapy!Committee!Task!Group!58!Med+Phys!28!712`37!

IAEA!1998!Quality!Assurance!in!Radiotherapy,!IAEA!TECDOC!989.!!

IAEA! 2000! Absorbed! dose! determination! in! external! beam! radiotherapy:! an!

international!code!of!practice!for!dosimetry!based!on!standards!of!absorbed!

dose! to! water! (Vienna:! IAEA)!IAEA!TRS`398! (www`

pub.iaea.org/mtcd/publications/pdf/trs398_scr.pdf).!!

IAEA! 2005! Radiation+ Oncology+ Physics:+ A+ Handbook+ for+ Teachers+ and+ Students,+

IAEA,+vienna+

IAEA!2013!Development!of!procedures!for!in+vivo!dosimetry!in!radiotherapy,!IAEA!

Human! Health! Report! No.! 8.! (International! Atomic! Energy! Agency,! IAEA,!

Vienna,!Austria,!2013).!!

Jung!J!W,!Kim!J!O,!Yeo!I!J,!Cho!Y!B,!Kim!S!M!and!DiBiase!S!2012!Fast!transit!portal!

dosimetry! using! density`scaled! layer! modeling! of! aSi`based! electronic!

portal!imaging!device!and!Monte!Carlo!method!Med+Phys!39!7593`602!

Khan!F!M!2010!The+Physics+of+Radiation+Therapy:!Lippincott!Williams!&!Wilkins)!

Kirkby! C! and! Sloboda! R! 2005! Consequences! of! the! spectral! response! of! an! a`Si!

EPID!and!implications!for!dosimetric!calibration!Med+Phys!32!2649`58!

Langmack!K!A!2001!Portal!imaging!Br+J+Radiol!74!789`804!

Lee! C,! Menk! F,! Cadman! P! and! Greer! P! B! 2009! A! simple! approach! to! using! an!

amorphous!silicon!EPID!to!verify!IMRT!planar!dose!maps!Med+Phys!36!984`

92!

Page 67: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

63

Low!D!A,!Moran!J!M,!Dempsey!J!F,!Dong!L!and!Oldham!M!2011!Dosimetry!tools!and!

techniques!for!IMRT!Med+Phys!38!1313`38!

Margalit!D!N,!Chen!Y!H,!Catalano!P!J,!Heckman!K,!Vivenzio!T,!Nissen!K,!Wolfsberger!

L!D,!Cormack!R!A,!Mauch!P!and!Ng!A!K!2011!Technological!advancements!

and!error!rates!in!radiation!therapy!delivery!Int+J+Radiat+Oncol+Biol+Phys!81!

12!

Mayles!P,!Mayles!P,!Nahum!A,!Nahum!A!E,!Rosenwald!J!C!and!Rosenwald!J!C!2007!

Handbook+of+Radiotherapy+Physics:+Theory+and+Practice:!Taylor!&!Francis)!

McCurdy!B!M,!Luchka!K!and!Pistorius!S!2001!Dosimetric!investigation!and!portal!

dose!image!prediction!using!an!amorphous!silicon!electronic!portal!imaging!

device!Med+Phys!28!911`24!

McCurdy!B!M!and!Pistorius!S!2000!Photon!scatter!in!portal!images:!accuracy!of!a!

fluence!based!pencil!beam!superposition!algorithm!Med+Phys!27!913`22!

McDermott!L!N,!Louwe!R!J,!Sonke!J! J,!van!Herk!M!B!and!Mijnheer!B!J!2004!Dose`

response! and! ghosting! effects! of! an! amorphous! silicon! electronic! portal!

imaging!device!Med+Phys!31!285`95!

McDermott!L!N,!Nijsten!S!M,!Sonke!J! J,!Partridge!M,!van!Herk!M!and!Mijnheer!B!J!

2006!Comparison!of!ghosting!effects! for! three!commercial!a`Si!EPIDs!Med+

Phys!33!2448`51!

McDermott! L! N,! Wendling! M,! Sonke! J! J,! van! Herk! M! and! Mijnheer! B! J! 2007!

Replacing! pretreatment! verification! with! in! vivo! EPID! dosimetry! for!

prostate!IMRT!Int+J+Radiat+Oncol+Biol+Phys!67!1568`77!

McNutt! T! R,! Mackie! T! R,! Reckwerdt! P! and! Paliwal! B! R! 1996a! Modeling! dose!

distributions!from!portal!dose!images!using!the!convolution/superposition!

method!Med+Phys!23!1381`92!

McNutt! T! R,! Mackie! T! R,! Reckwerdt! P,! Papanikolaou! N! and! Paliwal! B! R! 1996b!

Calculation!of!portal!dose!using!the!convolution/superposition!method!Med+

Phys!23!527`35!

Mijnheer! B,! Beddar! S,! Izewska! J! and! Reft! C! 2013! In! vivo! dosimetry! in! external!

beam!radiotherapy!Med+Phys!40!070903!

Mohammadi! M,! Bezak! E! and! Reich! P! 2008! Verification! of! dose! delivery! for! a!

prostate!sIMRT!treatment!using!a!SLIC`EPID!Appl+Radiat+Isot!66!1930`8!

Nelms!B!E,!Chan!M!F,! Jarry!G,!Lemire!M,!Lowden! J,!Hampton!C!and!Feygelman!V!

2013! Evaluating! IMRT! and! VMAT! dose! accuracy:! practical! examples! of!

failure!to!detect!systematic!errors!when!applying!a!commonly!used!metric!

and!action!levels!Med+Phys!40!111722!

Nicolini!G,!Clivio!A,!Vanetti!E,!Krauss!H,!Fenoglietto!P,!Cozzi!L!and!Fogliata!A!2013!

Evaluation!of!an!aSi`EPID!with!flattening!filter! free!beams:!applicability!to!

the! GLAaS! algorithm! for! portal! dosimetry! and! first! experience! for!

pretreatment!QA!of!RapidArc!Med+Phys!40!111719!

Nicolini!G,!Fogliata!A,!Vanetti!E,!Clivio!A,!Vetterli!D!and!Cozzi!L!2008!Testing!the!

GlaaS! algorithm! for! dose!measurements! on! low`! and! high`energy! photon!

beams!using!an!amorphous!silicon!portal!imager!Med+Phys!35!464`72!

Nijsten! S! M,! van! Elmpt!W! J,! Jacobs!M,! Mijnheer! B! J,! Dekker! A! L,! Lambin! P! and!

Minken!A!W!2007a!A!global!calibration!model!for!a`Si!EPIDs!used!for!transit!

dosimetry!Med+Phys!34!3872`84!

Nijsten!S!M!J!J!G,!Mijnheer!B!J,!Dekker!A!L!A!J,!Lambin!P!and!Minken!A!W!H!2007b!

Routine! individualised! patient! dosimetry! using! electronic! portal! imaging!

devices!Radiotherapy+and+Oncology!83!65`75!

Page 68: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

64

Olaciregui`Ruiz! I,! Rozendaal! R,! Mijnheer! B,! van! Herk! M! and! Mans! A! 2013!

Automatic!in!vivo!portal!dosimetry!of!all!treatments!Phys+Med+Biol!58!8253`

64!

Pecharroman`Gallego!R,!Mans!A,!Sonke!J!J,!Stroom!J!C,!Olaciregui`Ruiz!I,!van!Herk!

M!and!Mijnheer!B! J!2011!Simplifying!EPID!dosimetry! for! IMRT! treatment!

verification!Med+Phys!38!983`92!

Persoon!L!C,!Nijsten!S!M,!Wilbrink!F!J,!Podesta!M,!Snaith!J!A,!Lustberg!T,!van!Elmpt!

W!J,!van!Gils!F!and!Verhaegen!F!2012!Interfractional!trend!analysis!of!dose!

differences!based!on!2D!transit!portal!dosimetry!Phys+Med+Biol!57!6445`58!

Piermattei! A,! Greco! F,! Azario! L,! Porcelli! A,! Cilla! S,! Zucca! S,! Russo! A,! Castro! E!D,!

Russo!M,!Caivano!R,!Fusco!V,!Morganti!A!and!Fidanzio!A!2012!A!National!

project! for! in! vivo! dosimetry! procedures! in! radiotherapy:! First! results!

Nuclear+ Instruments+ and+ Methods+ in+ Physics+ Research+ Section+ B:+ Beam+

Interactions+with+Materials+and+Atoms!274!42`50!

Reich! P,! Bezak! E,!Mohammadi!M! and! Fog! L! 2006! The! prediction! of! transmitted!

dose!distributions!using!a!3D!treatment!planning!system!Australas+Phys+Eng+

Sci+Med!29!18`29!

Rowshanfarzad!P,!McCurdy!B!M,!Sabet!M,!Lee!C,!O'Connor!D!J!and!Greer!P!B!2010!

Measurement! and! modeling! of! the! effect! of! support! arm! backscatter! on!

dosimetry!with!a!varian!EPID!Med+Phys!37!2269`78!

Sterling! T! D,! Perry! H! and! Katz! L! 1964! Automation! of! Radiation! Treatment!

Planning.! Iv.! Derivation! of! a! Mathematical! Expression! for! the! Per! Cent!

Depth!Dose!Surface!of!Cobalt!60!Beams!and!Visualisation!of!Multiple!Field!

Dose!Distributions!Br+J+Radiol!37!544`50!

Stock!M,!Kroupa!B!and!Georg!D!2005!Interpretation!and!evaluation!of!the!gamma!

index!and!the!gamma!index!angle!for!the!verification!of!IMRT!hybrid!plans!

Phys+Med+Biol!50!399`411!

Teoh!M,!Clark!C!H,!Wood!K,!Whitaker!S!and!Nisbet!A!2011!Volumetric!modulated!

arc! therapy:!a! review!of!current! literature!and!clinical!use! in!practice!Br+J+

Radiol!84!967`96!

Teymurazyan! A! and! Pang! G! 2012! Monte! Carlo! simulation! of! a! novel! water`

equivalent!electronic!portal!imaging!device!using!plastic!scintillating!fibers!

Med+Phys!39!1518`29!

Tyler!M,!Vial!P,!Metcalfe!P!and!Downes!S!2013!Clinical!validation!of!an! in`house!

EPID!dosimetry!system!for!IMRT!QA!at!the!Prince!of!Wales!Hospital!Journal+

of+Physics:+Conference+Series!444!012043!

van!Elmpt!W,!McDermott!L,!Nijsten!S,!Wendling!M,!Lambin!P!and!Mijnheer!B!2008!

A!literature!review!of!electronic!portal!imaging!for!radiotherapy!dosimetry!

Radiother+Oncol!88!289`309!

Van!Esch!A,!Depuydt!T!and!Huyskens!D!P!2004!The!use!of!an!aSi`based!EPID!for!

routine! absolute! dosimetric! pre`treatment! verification! of! dynamic! IMRT!

fields!Radiother+Oncol!71!223`34!

van!Herk!M!2004!Errors!and!margins!in!radiotherapy!Semin+Radiat+Oncol!14!52`64!

van! Zijtveld!M,!Dirkx!M,! Breuers!M,! de!Boer!H! and!Heijmen!B! 2009!Portal! dose!

image! prediction! for! in! vivo! treatment! verification! completely! based! on!

EPID!measurements!Med+Phys!36!946`52!

Warkentin!B,!Steciw!S,!Rathee!S!and!Fallone!B!G!2003!Dosimetric!IMRT!verification!

with!a!flat`panel!EPID!Med+Phys!30!3143`55!

Page 69: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

65

Wendling!M,!Louwe!R! J,!McDermott!L!N,!Sonke! J! J,! van!Herk!M!and!Mijnheer!B! J!2006!Accurate! two`dimensional! IMRT!verification!using!a!back`projection!EPID!dosimetry!method!Med+Phys!33!259`73!

Wendling!M,!McDermott! L!N,!Mans!A,! Olaciregui`Ruiz! Í,! Pecharromán`Gallego!R,!Sonke! J! J,! Stroom! J,! van!Herk!M!and!Mijnheer!B! J!2012! In!aqua!vivo!EPID!dosimetry!Med+Phys!39!367`77!

Wendling!M,!McDermott!L!N,!Mans!A,!Sonke!J!J,!van!Herk!M!and!Mijnheer!B!J!2009!A!simple!backprojection!algorithm!for!3D! in!vivo!EPID!dosimetry!of! IMRT!treatments!Med+Phys!36!3310`21!

WHO!2014!Cancer!Facts!sheet!No297.!World!Health!Organization.!February!2014.!Retrieved!10!June!2014.!!

Winkler! P,! Hefner! A! and! Georg! D! 2005! Dose`response! characteristics! of! an!amorphous!silicon!EPID!Med+Phys!32!3095`105!

Page 70: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

66

Page 71: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

67

6 Acknowledgment

It is only through the support and encouragement of numerous people that the

completion of a Ph.D program and a doctoral thesis can be accomplished. Foremost, I

would like to express my sincere gratitude to Prof. Dr. Reinhold G. Müller for the

continuous support of my Ph.D study and research, for his patience, motivation,

enthusiasm, and immense knowledge. His guidance helped me in all the time of

research and writing of this thesis.

I will take the privilege to express my sincere gratitude to Prof. Dr. Rainer Fietkau; the

director of radiotherapy department for the pleasant, cooperative and well-equipped

working environment. And many thanks for his permissions to use all equipment and

also for his support for joining professional international meetings.

I would like to thank Prof. Dr. Christoph Bert for his helps and guidance. His advises

assisted me in further development of my writing and communication skill, as well as

the ability to carry out research on a more effective and professional level.

I would like to express my deepest gratitude to Dr. Manfred Schmidt, for his excellent

guidance, caring, patience, and providing me with an excellent atmosphere for doing my

experiments.

I really appreciate Prof. Ben Mijnheer for his great help for my careers and also

research work. He was always available to me with supportive idea since 2004 that I

start my career in radiotherapy field. Although he was in the Netherlands (The

Netherlands Cancer Institute), I felt him beside me.

Thank is due to Dr. Mohammad Mohammadi for initial idea and his support for Matlab

programing.

A warm thank you to Markus Kellermeier, whom I had the pleasure of sharing an office

with him for five years. Thanks for your patience, honesty, approachability.

Furthermore I would also like to acknowledge with much appreciation the crucial role

of the medical physics team of radiotherapy department of Erlangen university clinic,

who supported me to use all required equipment and the necessary materials to

complete the task; Dr. Michael Lotter (the Godfather, who knows everything and kindly

help everyone), Dr. Ulrike Lambrecht, Dr. Nils Achterberg, Dr. Stefan Speer, Stephan

Kreppner and Wilhelm Stillkrieg, Lukas Kober, Dr. Andreas Richter, Alexander Weiss,

Antje Brückner and also my former colleague Dr. Andreas Klein.

Page 72: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

68

I would also like to thank for the helps and supports of the research member of medical

physic team specially Dr. Indra Yohannes for his technical support for the planning

system issues, and also Jens Wölfelschneider, Tobias Brandt, Josefin Hartmann, Heru

Prasetio, Felix Böckelmann, Marco Serpa, Barbara Witulla and Dr. Stefan Vasiliniuc.

I would also like to thank my parents, and my brother. They were always supporting me

and encouraging me with their best wishes.

Last but not least I would like to thank my best friend, my wife Rahil. She was always

there cheering me up and stood by me through the good times and bad.

Page 73: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

69

Page 74: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

70

7 Curriculum Vitae PERSONAL

INFORMATION

Mohammadsaeed Saboori

Sex: Male | Date of birth 1977 |

Nationality: Iranian | Marital status: Married

EDUCATION AND

TRIANING

12/2010–Present Ph.D. (Candidate) Medical Physics

Universitätsklinikum Erlangen, Strahlenklinik, Erlangen

(Germany)

09/2002–11/2005 M.Sc. Medical Physics

Tarbiat Modarres University, Faculty of Medical Science, Tehran

(Iran)

Topic of thesis: Performing diode technical specification and

correction factors for patient dose verification using diode

dosimeter during External Radiotherapy of Prostate cancer.

09/1997–03/2002 B.Sc. Applied Physics (majoring in Nuclear Physics)

Shahid Beheshti University, Faculty of Science, Tehran (Iran)

PROFESSIONAL

CERTIFICATE

24/02/2012 Certified Medical Physic Expert (MPE)

Anerkennung Fachkunde zur Medizinphysik-Experte

WORK

EXPERIENCE

05/2009–Present Medical Physicist

Strahlenklinik; Universität Erlangen, Erlangen (Germany)

11/2003–04/2009 Medical Physicist

Radiotherapy Oncology Department, Shohada Hospital, Shahid

Beheshti University of Medical Science, Tehran (Iran)

07/2006–03/2009 Medical Physicist and Responsible Health Physicist (part time)

Radiotherapy Oncology Department, Payambaran hospital, Tehran

(Iran)

PUBLICATION

AND

PRESENTATION

Publication 2D patient transit EPID dosimetry for treatment fraction dose

verification, M. Saboori, M. Schmidt, C. Bert and R. G. Müller,

(submitted to Phys. Med. Biol. on Sep 2014)

Page 75: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

71

Publication Radiation dose distribution under the area protected using a

Cerrobend block during external beam radiotherapy: a film

study,M. Mohammadi, A. Taherkhani, and M. Saboori, Journal of

Radiotherapy in Practice / Volume 13 / Issue 04 / December 2014,

pp 456-464

Publication Evaluation of the physical characteristic of Cerrobend blocks used

for radiation therapy, A. Taherkhani, M. Mohammadi, M. Saboori,

V. Changizi, International Journal of Radiation Research. 2010; 8

(2) :93-101.

Poster presentation QC tests and initial evaluation of transit dose measurement in

Vero4DRT using EPID system, M. Saboori, M. Schmidt, R.G.

Müller, C. Bert, EPI2k14,13th International Conference on

Electronic Patient Imaging, Aarhus, Denmark, 31 Aug - 3 Sep

2014

Poster presentation An experimental method for IMRT transit dose verification using

an a-Si EPID system, M. Saboori, M. Schmidt, M. Mohammadi,

C. Bert, R. Muller, ESTRO 33, Vienna, Austria from 04-08 April

2014

Poster presentation Specific patient exit Dose verification base on treatment planning

system and experimental data, M. Saboori, M Schmidt, R. Mueller,

M. Mohammadi, AAPM 54th annual meeting, Charlotte, North

Carolina, USA, 29 Jul- 2 Aug 2012

Poster presentation Validation of a model for EPID transit dose prediction for two

dimensional IMRT verification,M. Saboori,M. Schmidt, R. Müller,

M. Mohammadi,The 18th congress of radiation oncology german

society, (DEGRO 2012),Wiesbaden, Germany, 7-10 June 2012,

Poster presentation Experimental validation of a simple model for EPID transit dose

prediction by M. Saboori, M. Schmidt, M. Tzivaki, R. Müller, M.

Mohammadi, ESTRO 31 conference, Barcelona, Spain, 09-13 May

2012

Poster presentation Methods for determination of shielding requirements in HDR

Brachytherapy bunkers, M. S. Saboori, M. Abbaci, A. S.

Meigooni, B. H. Malayeri, A. J. Arfaee, ESTRO 25 meeting,

Leipzig, Germany, October 9 - 12, 2006.

Poster presentation Correction factors for in-vivo diode dosimetry in measurement of

Entrance Dose, M.S. Saboori, A. Jabari Arfaei. Sixth Iranian

Congress of Medical Physics, Mashhad, Iran, May 2004

PROFESSIONAL

TRAININGS AND

MEETINGS

EPI2k14, 13th

International Conference on Electronic Patient

Page 76: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

72

Imaging, Aarhus, Denmark, 31 Aug - 3 Sep 2014.

ESTRO 33,Annual ESTRO meeting, Vienna, Austria, 4 -8 Apr

2014

IMRT/VMAT Comprehensive course, IBA Dosimetry

International Competence Center (ICC), Schwarzenbruck,

Germany, 9-11 Oct 2013.

VMAT - RapidArc course, The Netherlands Cancer Institute,

Amsterdam, The Netherlands, 12-14 Sep 2013

Dose Modelling and Verification for External Beam

Radiotherapy, (ESTRO teaching course) Florence, Italy, 10 - 14

Mar 2013

World congress of brachytherapy, (ESTRO meeting),

Barcelona, Spain, 10-12 May 201

Image-guided Radiotherapy in Clinical Practice (ESTRO

teaching course), Amsterdam, The Netherlands, 6 – 10 Nov 2011

EPI2kX, 11th International Conference on Electronic Patient

Imaging, Leuven, Belgium, 7-8 Jun 2010

IMRT and other conformal techniques in practice, (ESTRO

teaching course) Ghent, Belgium, 2 - 6 May 2010

Modern brachytherapy techniques, (ESTRO teaching course)

Istanbul, Turkey, 9 - 13 Mar 2008

Joint ICTP-IAEA Activity on Imaging in Advanced

Radiotherapy Techniques, Trieste-Italy, 20-24 Oct 2008.

AAPM TRAINING COURSE "Clinical Dosimetry and

Quality Assurance for Advanced Radiation Therapy

Techniques", Prague, Czech Republic, 16-20 Jun 2008.

Calculation and Verification for External Beam Therapy,

(ESTRO teaching course) Budapest, Hungary, 29 Apr - 3 May

2007

8th

European School of Medical Physics, Archamps, France, 27

Oct - 29 Nov 2005, The school's programs that I participated

were:(3rd week: Medical Computing,4th week: Physics of Modern

Radiotherapy, 5th week: Brachytherapy)

Workshop on Medical Applications of Synchrotron Radiation

MASR2004, Trieste, Italy, 23-25 Sep 2004, (Organized by IAEA

Page 77: Development of a daily dosimetric control for radiation ...PhD+thesis.pdf · Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device

73

and UNESCO in International Center for Theoretical Physics)

College on Medical Physics, Trieste, Italy, 30 Aug – 22 Sep 2004,

(Organized by IAEA and UNESCO in International Center for

Theoretical Physics)