Column Care and Column Care and Maintenance Maintenance Nicola Jones Hichrom Limited

Column Care and Maintenance - Life · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Embed Size (px)

Citation preview

Page 1: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Care andColumn Care and

MaintenanceMaintenance

Nicola JonesHichrom Limited

Page 2: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Problems

Lifetime 63 %

Pressure 23 %

Reproducibility 18%

Performance 12%

LC/GC Survey

Page 3: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Problems

♦ Severe peak tailing

♦ High back pressure

♦ Lower efficiency

♦ Split peaks

Symptoms

0.0 2.5 5.0 7.5 10.0Time (min)

Page 4: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

The First Step

♦ The performance either with the

manufacturers conditions or your

own standard conditions

♦ Solvent compatibility

♦ Storage recommendations

A new column:

CHECK

Page 5: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

HPLC Columns

♦ Columns are the consumable part of

an HPLC system

♦ They do not last forever!

♦ Select the right column for the job

♦ Consider all operational variables

mobile phase, pH, temperature

Page 6: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Outline

♦ Simple preventative measures

♦ Column lifetime

♦ Pressure

♦ Performance

♦ Reproducibility

♦ Conclusion

Page 7: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Simple Preventative Measures

DETECTOR

Injection valve

Guardcolumn

Analytical column

DataSystem

Solventreservoir

PUMP

Inletfilter

Inlinefilter

Pre column

Sample Purification

Solvent Filtration

Page 8: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Simple Preventative Measures

♦ Use of high quality solvents – HPLC grade

♦ Removal of particulates- solvent filtration

- in-line filter

- sample purification

- RP to NP

- use of buffered mobile phase

♦ Implementation of washing protocols

Page 9: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Outline

♦ Simple preventative measures

♦ Column lifetime

♦ Pressure

♦ Performance

♦ Reproducibility

♦ Conclusion

Page 10: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Low pH– ligand cleavage– accelerated with T

OH

OH

HO

OH

OH

OHHO

HOHO

HOHO

Column Lifetime

High pH– silica dissolves– effected by T, mM buffer & type, mobile phase strength

Page 11: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

High pH

Mobile phase : 50%H2O, / 50%MeCN / 0.2% TEA, pH 11

Initial analysis

After 30 injections

column voided

Page 12: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

♦ Restricted pH range 2 < pH < 7

♦ Temperature < 50ºC

♦ Use of guard column

Is it possible to work outside these “ideal” conditions?

The “ideal” conditions

YES

Page 13: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

Low pH stability

Components; 1: Uracil, 2:Dimethyl phthalate, 3: Toluene, 4: Biphenyl, 5:Phenanthrene

Conditions:0.1% TFA in H2O – MeCN (50:50)pH 1.81.0 ml / min, 22oC150x4.6mm

Page 14: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

High pH stability

Components; 1: Uracil, 2:Dimethyl phthalate, 3: Toluene, 4: Biphenyl, 5:Phenanthrene

Conditions:25mM pyrrolidine in H2O - MeCN (50:50)pH 11.01.0 ml / min, 22oC

Page 15: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

High Temperature Application – 80oC

Camurus, Malmo Sweden

Sample: Cyclosporin A & UColumn: Ace C18 5um 250x3.0mm

Mobile phase: H2O / MeCN / MTBE / phosphoric acid (460:510:30:1)Flow rate: 0.8 ml / minTemperature: 80oCInjection Volumne: 20ulDetection: 210nm

Page 16: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

112 hours

3 hours

300 hours

381 hours

Page 17: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

High Temperature Application

Retention vs Analysis Time

0

10

20

30

0 200 400 600 800 1000

Analysis Time (hours)

Ret

enti

on

(m

inu

tes)

CsU

CsA

Page 18: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

High Temperature Application

Time (hrs) CsU CsA CsU CsA3 25.86 27.56 5790 75478 25.43 27.06 5308 767541 26.16 27.93 5615 769658 25.5 27.16 5539 7565112 26.12 27.86 5566 7624145 26.18 27.91 5499 7528212 25.37 27.02 5450 7501240 25.29 26.91 5600 7635300 24.7 26.32 5519 7491352 24.74 26.32 5453 7443381 24.86 26.53 5470 7327509 25.02 26.61 5195 6892592 24.39 25.97 5237 6976660 25.11 26.76 5194 6796750 25.68 27.29 4704 6081790 26.16 27.9 4832 6146

Retention (min) Plate count (N)

Page 19: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Lifetime

Flush with 100% H2O for 24 hours, 1.0ml/min

High Aqueous Conditions

Conditions: MeCN/50mM KH2PO4, pH3.5 (7:93)Components; 1: Uracil, 2:Dimethyl phthalate, 3: Toluene, 4: Biphenyl

0 1 2 3 4 5 6 7 8 9 10

Time - Minutes

0

100

200

300

400

500

600

700

Re

sp

on

se

- M

illiV

olts

Ace AQ

0 1 2 3 4 5 6 7 8 9 10

Time - Minutes

0

100

200

300

400

500

600

700

Re

sp

on

se

- M

illiV

olts

C18

0 1 2 3 4 5 6 7 8 9 10

Time - Minutes

0

100

200

300

400

500

600

700

Re

sp

on

se

- M

illiV

olts

Ace AQ

0 1 2 3 4 5 6 7 8 9 10

Time - Minutes

0

50

100

150

200

250

300

350

400

450

Re

sp

on

se

- M

illiV

olts

C18

Page 20: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Outline

♦ Simple preventative measures

♦ Column lifetime

♦ Pressure

♦ Performance

♦ Reproducibility

♦ Conclusion

Page 21: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Pressure

♦ Avoid pressure shocks

♦ Increase flow gradually upon equilibration

♦ Turn valve quickly

♦ Blocked frit, line

♦ Buffer precipitation

♦ Column contamination

Potential Problems

Good Practice – primarily associated with irregular silicas

Page 22: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Pressure Problems

♦ Check the pressure with & without the column

Determining the cause

♦ Wash the column

♦ Back flush the column

If the column pressure is high

If the system pressure is high

♦ Change in-line filter

♦ Flush the system

Page 23: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Cleaning

Flush with stronger solvents than your mobile phase

Reversed phase solvent choices in order of increasing strength

♦ Mobile phase without buffer salts

♦ 100% methanol

♦ 100% MeCN

♦ 75%MeCN / 25% IPA

♦ 100% IPA

♦ 100% methylene chloride

♦ 100% hexane

Page 24: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Cleaning

Column Contamination

Mobile phase : 20% H2O, / 80% MeOH1: Uracil, 2: Phenol, 3: 4-Chloronitrobenzene, 4: Toluene

0.0 2.5 5.0

2

41

3

Time (min)

QC analysis

2

41

3

0.0 2.5 5.0Time (min)

QC analysis after cleaning 100% IPA, 35 oC

USP Tailing Factor Component 4: 1.17

USP Tailing Factor Component 4: 1.32

Page 25: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Summary

To help prevent back pressure problems

♦ Use column protection

- guard cartridges

- in-line filters

♦ Filter samples

♦ Sample clean up

♦ Appropriate column flushing

♦ Filter buffered mobile phases

♦ Flush the system regularly

Page 26: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Outline

♦ Simple preventative measures

♦ Column lifetime

♦ Pressure

♦ Performance

♦ Reproducibility

♦ Conclusion

Page 27: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

♦ Peak tailing, broadening

♦ Loss of efficiency

Deterioration in performance

Can be caused by

♦ Column secondary interactions

♦ Column void

♦ Column contamination

♦ Column loading

♦ System dead volume

♦ Solvent effects

Page 28: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Si

Si

SiSiSi

Least Acidic Most Acidic

The Silica Surface

O

Si

OO Si

Si

O

O

OH

OHO

O

SiSi O

O Si

Si

HO

SiO

O

O Si

HO

O

OSi

O

SiO

O SiO

SiO

O

OH

O Si

Si

O

O

O

O

Si

O

OHO

Si

OOH

SISiSi

OHOH

OHOH

GeminalGeminal

HO

FreeFree

SiSiHOHO

MM++ChelatedChelated

SiSi

OHOH OHOH

SiSi

AssociatedAssociated

SiSiSiSi

OHOHOHOH

Least Acidic Most Acidic

Page 29: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Silica Bonding :Monomeric

+Si O H

Silica surface

C l Si

C H3

C H3

Monofunctional bonding reagent

C H

OSi Si

3

C H3 Bonded silica

Single point attachment

Reproducible bonding technique

Surface Chemistry

Page 30: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Silica Bonding :Endcapping

♦ remove residual silanols

♦ improve the chromatography of basic compounds

X S i

C H3

C H3

C H3

S i OH

S i O S i

C H3

C H3

S i O S i

C H3

C H3

S i O S i C H3

C H3

C H3

S i OH

S i O

S i OH

S i

C H3

C H3

S i O S i

C H3

C H3

+

Surface Chemistry

Page 31: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

Column Secondary Interactions

Mobile phase : 85% 25mM Na2HPO4 / 15% MeCN1: Phenylpropanolamine, 2: Ephedrine, 3: Amphetamine, 4: Methamphetamine, 5: Phenteramine

TIme (min) Time (min)0.0 2.5 5.0

54

32

1

54

3

2

1

0.0 2.5 5.0

No TEA 10mM TEA

USP Tailing FactorComponent 5: 1.14

USP Tailing Factor Component 5: 1.57

Page 32: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Time (min)

0.0 2.5 5.0

54

32

1

54

32

1

Time (min)

0.0 2.5 5.0

Performance

Column Secondary Interactions

Mobile phase : 85% 25mM Na2HPO4 / 15% MeCN1: Phenylpropanolamine, 2: Ephedrine, 3: Amphetamine, 4: Methamphetamine, 5: Phenteramine

pH 3.0 pH 7.0

USP Tailing Factor Component 4: 2.35

USP Tailing Factor Component 4: 1.33

Page 33: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

XTerra MS C18Ace 5 C18

Symmetry 5 C18 Waters Spherisorb 5 ODS2

N 0.1 = 63,400 pl/mAs = 1.13

Column Secondary Interactions pH 6.0

Mobile phase : 80% MeOH / 20% 25mM KH2PO41: Norephedrine, 2: Nortriptyline, 3: Toluene, 4: Imipramine, 5: Amitriptyline

N 0.1 = 21,800 pl/mAs = 1.83

N 0.1 = 100 pl/mAs = 5.93

N 0.1 = 24,400 pl/mAs = 1.33

Page 34: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

Column Void

Column: Partisil 10um SAX, 250x4.6mmEvaluation Conditions: 5mM KH2PO4, 1.0ml/min 1: CMP, 2: UMP

Initial performance

Av N = 24,000 pl/m

Performance after 100 injections

Av N = 3,200 pl/m

Page 35: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

Column Contamination

100% IPA100% MeOH

100% Ethyl acetate

New Column

Column Returned from Customer

Column After Flushing

Lichrospher 60A silica, 5um, 250x4.0mm, Heptane/ethyl acetate, (90:10) 1: butylbenzene, 2: methyl benzoate, 3: nitrobenzene

Page 36: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

System Dead Volume

Time (min)0.0 0.5 1.0 1.5 2.0

4

32

1

Time (min)0.0 0.5 1.0 1.5 2.0

4

32

1

10µl loop 50µl loop

Mobile phase : 85%H2O, 0.1% TFA / 15%MeCN, 1.0ml.min1: Phenylalanine, 2: 5-benzyl-3,6-dioxo-2-piperazine acetic acid, 3:Asp-Phe, 4:Aspartame

Page 37: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

Solvent Effects

Injection volume: 30ulMobile phase: 82% H2O / 18% MeCN1: Caffeine, 2: Salicylamide

Injection solvent 100% MeCN

12

Injection solvent mobile phase

1

2

Page 38: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

Solvent Effects

Mobile phase : MeOH / H2O, 0.1% TFA, (60:40)

Fresh TFA added to mobile phase

Current mobile phase

Page 39: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Performance

Temperature variation

Column: Inertsil ODS 3V 250x4.6mm, 5umMobile phase : 20% H2O / 80% MeOHComponents: 1: Uracil, 2: Dimethyl phthalate, 3: Toluene, 4: Biphenyl, 5: Phenanthrene

AmbientTemp

40 oC

♦ Performance increased

Peak 5: 57600 to 70000 pl/m

♦ Pressure reduced

1340 psi to 1050 psi

♦ Run time reduced20 min to 17.5 min

Page 40: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Summary

♦ Minimise system dead volume

♦ Match the guard column to the analyticalcolumn

♦ Keep the column clean

- pre-treat samples

- replace guard cartridge periodically

- implement washing protocol

To help gain optimum performance from your column

Page 41: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Outline

♦ Simple preventative measures

♦ Column lifetime

♦ Pressure

♦ Performance

♦ Reproducibility

♦ Conclusion

Page 42: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

EARLY CONVENTIONAL CONVENTIONAL

Reproducibility

Page 43: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

PARTIALLY VALIDATED FULLY VALIDATED

Reproducibility

Page 44: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

CONVENTIONAL BASE DEACTIVATED

Reproducibility

1.51.5

22

2.52.5

33

3.53.5

44

22 44 66 88 1010

Column NumberColumn Number

kk’’

+ 4 %

1.51.5

22

2.52.5

3.53.5

44

22 44 66 88 1010

Column NumberColumn Number

kk’’

± 50%

33 + 10%+ 10%

Page 45: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Reproducibility

Page 46: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Column Reproducibility

Page 47: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Batch Validation

Single Chromatographic Validation Test

Page 48: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Batch Validation

Multiple Chromatographic Validation - Test 1

Page 49: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Batch Validation

Multiple Chromatographic Validation - Test 2

c

Page 50: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Batch Validation

Multiple Chromatographic Validation - Test 3

Page 51: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Ace C18Ace C18 3.703.70Hichrom C18Hichrom C18 4.924.92Symmetry C18Symmetry C18 5.99 5.99 Luna C18(2)Luna C18(2) 7.047.04HypurityHypurity C18 C18 9.869.86

Validated Materials

YMC Pro C18, Waters XTerra & Inertsil ODS 3V are alsovalidated but the variation in hydrophobicity is not specified

k’(neutral molecule) ± (%)

NB: All data taken from manufacturers validation certificates

Page 52: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

Summary

♦ Good column to column reproducibility

♦ Good batch to batch reproducibility

♦ Increased testing by manufactures

♦ Use of validated columns provides long term

reproducibility

Page 53: Column Care and Maintenance - Life  · PDF fileColumn Care and Maintenance Nicola Jones ... ♦Blocked frit, line ... ♦ Select the column most suitable for your application

♦ Simple measures can be adopted to prolong column

lifetime

♦ Regular performance tests to monitor any deterioration

♦ Select the column most suitable for your application

♦ Consider all operational variables

- the column is not always the problem

Conclusion