Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
CHEM 3430Analytical Chemistry II
Winter Semester, 2004© University of Guelph
Lecturer: Dr. Perry Martos, CIH
2
Course Objectives
• The overall aim of this course isto develop an understanding ofthe fundamental aspects ofanalytical chemistry used intoxicological considerations:– Target analyte extraction and
concentration– Chromatography and detectors– Mass spectrometry– Hyphenated techniques– Spectroscopic techniques– Electrochemical approaches– Data analysis
3
Topics• Data Reduction and Analysis• GLP, Terms and Definitions• Target analyte extraction and concentration
– LLE, SPE, SPME
• Chromatography and Detectors– LC, GC, CE, size exclusion, affinity– FID, NPD, PFD, PID, ECD, MS, MS/MS, MSn
• Mass spectrometry– Quad, Tandem Quad, Ion Trap, Sector
• Hyphenated techniques– LC-MS, LC-MS/MS, GC-MS, GC-MS/MS,
CE-MS and CE-MS/MS
• Spectroscopic techniques– UV/VIS, Fluorescence, AAS, AES, AFS, ICP
• Electrochemical approaches– Potentiometry and Voltammetry
4
The Analytical ProcessSample
Preparation
Extraction andConcentration
Target AnalyteSeparation
Target AnalyteDetection
Quantification
The (accuate) Result
Qualtitative orQuantitative?
Qualification
5
Qualitative and Quantitative• Ideally, determine a target analyte both
qualitatively and quantitatively.• Another way to consider: provide absolute
confidence the target analyte is present,when detected, and at what concentration(within a confidence limit).
• The use of chromatography and MS/MSand MSn permits this.
• The use of several (different) detectorseither in series or in parallel, e.g., GCcoupled PID and FID.
• Different columns with the same detector,e.g., analysis of organochlorines by GC-ECD with both a 5% phenyl column and a35% cyano column. Analytes havedifferent retention times on the differentcolumns.
6
Target Analytes• In most analyses, we require what is
called the “trace” and “ultra-trace”detection and determination of targetanalytes.– Trace has usually been defined as ppm
and sometimes ppb (ppmv in air)– Ultra-trace detection is considered at
levels of ppt and ppq
• Sometimes those analytes are“insult” environmental pollutants,pesticides, drug residues in meats,pharmaceuticals in blood…
• Methods and instruments inanalytical toxicology must be bothsensitive and selective for targetanalytes.
7
Concentrations• In water, blood,
urine, foods, soil,etc., the units oftarget analyteconcentration aretypically reportedas (wet or dry):
• ppm
• ppb
• ppt
• ppq
gµgor
mLµg
gpgor
mLpg
gngor
mLng
gfgor
mLfg
• In gases, such asair, the units ofanalyteconcentration aresometimes inmass/volume andsometimes involume/volume -be careful aboutcomparing vol/voland mass/volume
• ppmv
• ppbv
LµL
LnL
8
Airborne Concentrations• The airborne concentration of an analyte
that is either a gas or a vapour is typicallyrepresented in one of three manners: ppmv,% by volume, or mg/m3. Here is whereyou must be careful. When the unit ofconcentration is reported as mg/m3, youmust be aware of both the temperature andpressure at which this assessment wasmade. RH is also important.
• Example: an air sample using a charcoal samplingbed and a mass flow controlled air sampling pumpis taken over eight hours in a factory where styreneis used. The mass of styrene on the charcoal bed isdetermined to be 10 µg using NIOSH method3500/3501. The volume of air sampled was 50mL/min for 480 min. What is the average eighthour styrene concentration in ppmv and mg/m3 ifthe pressure and temperature during sampling were720 mm Hg and 30°C. Compare to styrene’s TLV(Ministry of Labour and NIOSH).
9
Airborne Concentrations• Consider the following for the ppmv
calculation:– Convert the mass of styrene to the volume it
would occupy at the specified averagetemperature and pressure.
• Consider the following for the mg/m3
calculation:– The use of a mass flow controlled air sampling
pump does not require additional correctionssince these pumps increase or decrease involumetric flow rate to compensate fordecreases or increases in atmospheric pressure,respectively. With automatic temperaturecorrection, the opposite is true. A conventionalair sampling pump, i.e., not mass flowcontrolled, would require correction, relative tohow it was calibrated.
• For aerosols, mass/volume is typical.– Equivalent volume diameters
11
Terms and Definitions• Several “Current” References on terms and
definitions for analytical methods• Need to understand GLP (Good Laboratory
Practices)• All analytical methods used for defining analyte
concentrations of significance are movingtoward GLP standards. Example: test methodsfor pharmaceuticals and pesticides
• ISO– http://www.iso.org/iso/en/ISOOnline.openerpage
• SCC– http://www.scc.ca
• Harmonized requirements IUPAC/ ISO/ AOAC– http://www.scc.ca/publicat/canp/1598_e.doc
• Overall: The analytical toxicologist has accessto terms and definitions for analytical methods
13
Quantitative Description ofInstrument Performance:
Why is it required?
• To (quantitatively) describe instrumentperformance so that analytical measurements ofunknowns can result in a quantitative result.
• Example: The analysis of steroids in urine; theanalysis of caffeine.
• Control charting– Monitor method performance over time. Method
performance can be distinct from instrumentperformance. Ensures the capture of “drifts” in themethod.
14
Analytical MethodConsiderations
COST
SPEED NEEDS
• COST: Training, equipment...• SPEED: When is it required...• NEEDS: e.g., a required detection limit...• Overall: An equal balance among the three is
often required.
15
Basic Terms• Precision: Repeatability (within a lab with
one analyst using one method) andReproducibility (among labs with differentanalysts using the same method)
• Bias: Absolute or relative systematic error• Sensitivity: Calibration or analytical...• Response Factor: Response as a function
of analyte concentration• Limit of Quantitation (LOQ): A numerical
value usually directly related to LOD (>)• Detection Limit (LOD or MDL): Different
methods (statistical) to establish LOD• Not Detected: Signal not discernible from
noise using the given analytical system• Concentration Range: From LOQ to LOL• Selectivity: Coefficient of selectivity (%)
16
• Standard Deviation (s)for a small set of data.
• Rel. Standard Deviation (using arithmetic mean)
• %RSD or CV (Coefficient of variation)
• Standard Deviation of the Mean
• Variance
• Overall standard deviation - Uncertainty( ) ( ) ( ) ( )22
32
22
1 ... Nsssss +++=
2sNssm =
1
)(1
2
−
∑ −= =
N
xxs
N
ii
%100⋅= xsCV
xsRSD =
N
xN
ii∑ −
= =1
2)( µσ
Smallset ofdata
Population
Precision
17
• Variance (s1)2 of TEST METHOD #1 (sample)
• Variance (s2)2 of TEST METHOD #2 (sample)
• The ratio of variances (where S22 >S1
2)
21s
22s
21
22
SSF =
Gaussian(Normal)
Distributions
F-test: Comparison of Variances
18
• Mean of TEST METHOD #1 (sample)• Mean of TEST METHOD #2 (sample)• Are the means equal? The t-test we use is
dependent on whether the variances of the two testmethods are the same or different. Again, assumeGaussian Distributions. One-tailed t-test for equalvariances, and two-tailed for unequal variances.The two-tailed presents more relaxed constraints onestablishing equivalence.
1X
2X
t-test: Comparison of Means
19
Accuracy and Precision...
• Before we can appreciate the term Bias, we mustfirst understand the basic terms “accuracy” and“precision”.
• Accuracy is generally accepted as the closeness of a“result” to that of an expected value.
• Consider the following example:– 10 µL of each 1, 5, 20, and 100 ppm standards of
caffeine standards (USP grade) are injected into anLC-MS. An external standard calibration curve isgenerated.
– 10 µL of a “reference” (NIST traceable) samplecontaining caffeine of a concentration unknown tous, is then injected into the LC-MS. The response,once interpolated from the calibration curve, yields10.4 ± 0.2ppm (95%) in the sample.
– We find out later that the actual concentration of the“reference” caffeine standard is 10.1 ± 0.1ppm(95%).
– Is your result accurate? Why?
20
...Accuracy and Precision...
• We must appreciate that the so-called determinationof an analyte concentration in a sample is really, atbest, an “estimate” of its actual concentration.
• Associated with every determination is a level of“precision”.
• Consider the following continuing with our caffeinedetermination: (Note, the red bar is the “expected”concentration of the target analyte and is 10 ppm.)
5-
10-
15-ppm
• 10 measurements, withan averageconcentration of 10.1ppm, yet the range ofconcentrations is >20%RSD.
• Accurate Not Precise
21
...Accuracy and Precision...
5-
10-
15-ppm
• 10 measurements,with an averageconcentration of 12ppm and with>20% RSD.
• Not Accurate andNot Precise
5-
10-
15-ppm
• 10 measurements, withan averageconcentration of 12ppm, yet with <10%RSD.
• Precise Not Accurate
22
...Accuracy and Precision
5-
10-
15-ppm
• 10 measurements, with anaverage concentration of10.4 ppm, and with <2%RSD.
• PRECISE, but is itACCURATE?
• From this, we can see clues re. BIAS, t-testcomparison of means of results usingdifferent methods (ANOVA), LOD, LOQ,and even ruggedness.
Accuracy, Precision, ND, LOD, LOQ,Selectivity, Sensitivity, Linearity,Ruggedness
23
Bias and ErrorsPage 12-on and A2-A5 Skoog Holler Nieman
• BIAS: The systematic departure of the “measured”value from the “true” or “expected” value.
• There are many sources of BIAS, and they may beadditive.
• Accuracy can be assessed quite simply usingcertified or standard reference materials that“match” your test standards and samples.
– Matrix effects can create artifacts resulting in BIAS,positive or negative.
•• SYSTEMATIC ERRORS can result in BIAS.SYSTEMATIC ERRORS can result in BIAS.• A: No BIAS; B: BIAS +ve
24
Errors...
• Three basic types:– Instrumental– Method– Personal
Overall:• Systematic or determinate• Random - indeterminate• Mistakes (human errors,
or prejudice) - determinate
Instrumental Errors• “Drift” in electronic circuits (e.g., improper zero)• Temperature control is unstable is subject to
ambient parameters.• Poor power supplies, e.g., other instruments on the
same power grid perturbing the instrument’s power.• Other systems in the area creating a field that
influences detector response, coincidentally whilethe samples are running, but not the standards.
• Regular calibration is required, the frequency ofwhich is for the most part empirically determined(from actual experiments).
In all cases, accuracy is most likely affected
25
...Errors...
Method Errors• Often introduced by non-ideal chemical behavior.• Loss of sol’n by evaporation.• Analyte losses upon unexpected adsorption or
absorption.• Contaminants.• Interferences (affects selectivity, e.g.,
electrochemical methods, AA)• Instability of reagents.• Difficult to detect the target analytes.• The solution is usually to prepare the standards in
the same matrix as the samples. This can be easilyaccomplished by Standard Addition, IsotopeDilution (deuterated analogues), and InternalStandard approach (preferably with deuteratedanalogues). This is where External StandardCalibration can result in BIAS.
• Use of CRM (certified reference materials),intralaboratory method validation, and verificationby other analysts really helps here.
26
...Errors...
Human (Personal) Errors• Made unknowingly and sometimes knowingly.• Prejudice w.r.t. reading meniscus, thermometers,
pH meters, peak integration (apex determinations),colour end points.
27
...Errors...
Sources of BIAS and VARIABILITY in the Lab• Sample storage (contamination, physical or
chemical degradation)• Sample handling (contamination during
preparation).• Sub-sampling• Weighing and volumetric devices• Solvent purity• Extraction yields - low and/or variable• Analyte concentration following evaporation• Clean-up steps (analyte losses)• Quality of reference standards• Instrument calibration• Instrumental - injection variability and/or
discrimination, matrix effects, changes in detectorresponse during the course of the sample analyses
• Different analysts• Environmental and electrical conditions in the lab
(T, RH).
28
...Errors
Assessing BIAS and Variability• Use of reference materials• Comparison to another method (preferably a
standard method)• Use of representative blank matrices and spikes• Use of true positive samples and spikes, which
have been confirmed by other analysts or lab• Within-run and between-run variability studies
33
A Look At Calibration Curves:
• Does r² adequately indicate linearity in alinear least squares regression?
• R² is often used to describe linearity, butbe very careful, since extreme bias canbe realized at low concentrations -example to follow.
• Comment: Use Response Factors (RF)(Response/unit concentration)
• What about weighting? (1/x)• What about non-linear functions? Look
at 2nd order:( )
arespyabb
Conc2
)int(42 −−±−=
( )slope
yrespConc int−=bmxY +=
cbxaxY ++= 2
34
• Linear Least Square Regression, Unweighted• Ideal Calibration Curve• Y-Int=0
Curve 1
y = 5.00000xR2 = 1.00000
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
0 500 1000 1500 2000 2500 3000 3500
[ ] ppm
Res
pons
e
ppm Arb Units1 55 2525 125
125 625625 31253125 15625
Curve 1
35
Curve 2
• Linear Least Square Regression, Unweighted• Here, Points 1, 2, and 3 are each increased by 100%• The calibration curve still looks good from the
perspective of r²
ppm Arb Units1 105 50
25 250125 625625 3125
3125 15625
Curve 2
y = 4.9872x + 34.165R2 = 0.9999
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
0 500 1000 1500 2000 2500 3000 3500
[ ] ppm
Res
pons
e
36
Curve 3
• Point 6 is increased by only 10%
y = 5.5067x - 69.415R2 = 0.9997
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0 500 1000 1500 2000 2500 3000 3500
[ ] ppm
Res
pons
e ppm Arb Units1 55 2525 125
125 625625 31253125 17188
Curve 3
37
Summary of Curve ComparisonsCurve 1 Curve 2 Curve 3
Arb Units Arb Units Arb Units ppm5 10 5 125 50 25 5
125 250 125 25625 625 625 1253125 3125 3125 625
15625 15625 17188 3125
Slope 5.00000 4.98720 5.50665Y-Int 0.00000 34.16524 -69.41482
R² 1.00000 0.99994 0.99967
Er% slope -0.26% 10.13%
Curve 1 Curve 2 Curve 3ppm ppm Er% Curve 1 ppm Er% Curve 2 ppm Er% Curve 3
1 1 0% -5 -585% 14 1251%5 5 0% 3 -36% 17 243%25 25 0% 43 73% 35 41%
125 125 0% 118 -5% 126 1%625 625 0% 620 -1% 580 -7%3125 3125 0% 3126 0% 3134 0%
• Interpolation of concentrations using the estimatedslopes and Y-intercepts
38
Use of Response Factors toUnderstand Curve Linearity
• Consider the following calibration curve. We havea value of 5 Units of “Bias” for all measurements.
• There is also detector saturation at the highest targetanalyte concentration, resulting in ~10% loss ofexpected response, common with LC-MS/MS andion trap MS
• Is this curve linear, and usable for quantitation atthe low concentration range?
y = 4.4932x + 74.4370R2 = 0.9995
0
2000
4000
6000
8000
10000
12000
14000
16000
0 500 1000 1500 2000 2500 3000 3500
[ ] ppm
Res
pons
e
ppm Arb Units1 105 3025 130
125 630625 31303125 14067
39
Analysis of Linear Calibration Curve DataUsing Response Factor
• U.S.EPA indicates that with an average RF <15%CV, the curve is linear, and with up to 25% CV onRF as acceptable for some methods.
• Note that the detection limit for this method willneed adjustment with the elimination of acalibration curve data point in the low end.
• It could be that this calibration curve is truly non-linear, which is quite acceptable. The onlycondition is that CV around each point must be“low”.
Following Q-testppm Arb Units RF
1 10 5 30 6.025 130 5.2125 630 5.0625 3130 5.0
3125 14067 4.5
slope 4.5 Ave. RF 5.1Y-int 94.1 Std. Dev. 0.5
R² 0.9995 CV 11%
ppm Arb Units RF1 10 105 30 6.025 130 5.2
125 630 5.0625 3130 5.03125 14067 4.5
slope 4.5 Ave. RF 6.0Y-int 74.4 Std. Dev. 2.0
R² 0.9995 CV 34%
Original Data
40
• What strategies can be employed to establish a clearunderstanding of the calibration curve?
• What about the Dixon Test to check the RF’s.Remember, a larger Confidence is actually a lessreliable result.
• The RF data provide strong clues with respect tothe detection limit (see later).
Calibration Curve Example
NXXXXQ
−−
=1
21explowest highest to fromRank
...,, 321 NXXXX
41
• Calibration curves must appropriately reflect theanalyte concentration range of interest in thesample.
• Develop the calibration curve with more evenlyspaced data points.
• Include a number of repeat calibration curve pointmeasurements for each data point.
• Develop the calibration curve over a period of time• Ensure full multi-point calibration curve data points
are developed for each set of unknown samples,with no more than about 6 samples betweenrunning the calibration curve points.
How to avoid some of the pitfallswith calibration curves
42
Approaches to Quantification
1. External Standard: A calibration curve is first developed.No consideration for matrix here. The samples are run, asclose to the same conditions as the standards. Sampleresponses are interpolated from the calibration curve.
2. Bracketing standards: Here, two standards around asample response are used to quantify the target analyteconcentration. Is just as accurate as external standardcalibration. Not usually accepted by regulatory bodies,but works well particularly for research applications,surveys.
3. Average Response Factors: Simple, but powerful. Here,provided the y-int is small, the Ave. RF can be quiteeffectively used, and is as accurate as external standardcalibration.
4. Internal Standard: A more elegant technique involvingintroducing a compound into the sample extract just priorto analysis. The ratio of responses of the internal standardto the known concentrations of target analyte is used todevelop a calibration curve. Then, in the sample, the ratioof IS response to target analyte response of unknownconcentration is interpolated from the curve. Goodcorrection for instrumental factors affectingquanitification.
43
Approaches to Quantification
5. Standard Addition: See Skoog Holler Nieman. Here, theactual background matrix is spiked with various (known)concentrations of the target analyte.
6. Isotope Dilution: The most elegant of all quanitificationprocedures. It involves adding an isotope of the targetanalyte to the sample prior to extraction. It therefore actsas a surrogate compound and internal standard, but mostimportantly, will elute at almost exactly the sameretention volume as the unlabelled analyte.There are various versions of isotope dilution, such asradio immunoassay and neutron activation.The ultimate isotope dilution would be to present theradioactive isotope of the target analyte to the test systemprior to incorporation into plant and animal tissue. Thiswould therefore quite thoroughly mimic the actualphysical-chemical environment of the target analyte.This strategy is actually used with new pesticideregistrations. Here, radioactive isotopes of the pesticidesare presented to plants. The fate of the pesticide can beeasily tracked.
44
Selectivity• This is a non-dimensional term, which can be
expressed in %, that can quantify the selectivity ofsystem for a specific target analyte. Let’s look atExample 1-2 in Skoog Holler Nieman (page 15).
Selectivity, Sensitivity, Signal toNoise, Detection Limit
45
Sensitivity...
• Ability of instrument to detect small changes intarget analyte concentration.
• Two factors can be used to represent sensitivity.– Slope of the calibration curve - recall this!
– Reproducibility (precision) of the measurementcan affect the sensitivity.
• Sensitivity, precision and calibrationcurves are related. The calibration curvesin the previous examples are optimal inthat s or σare negligible relative to thesignal. Also, when two calibration curveshave equal slopes, the one exhibiting betterprecision is more sensitive.
High
Low
Res
pons
e
Conc.
46
...Sensitivity
• When the confidence interval around particularmeasurements is low, we find that sensitivity iscompromised. Recall, sensitivity is the ability ofthe system to discriminate between analyteconcentrations, so “noisy” measurements results inlower confidence about sensitivity.
High
Low
Res
pons
e
Conc.
47
• IUPAC defines sensitivity as calibration sensitivity– This is the slope of the curve at the
concentration of interest. Recall ResponseFactors, where RF=Response/Conc. Note, noconsideration for precision with this approach.
Quantitative Definition of Sensitivity
blSmcS +=
Signal for“Blank”
Slope
Concentration
Signal
Res
pons
e
Conc.
48
• This definition accounts for noise at a specificconcentration. The slope can also be replaced witha response factor.
Analytical Sensitivity
sSm
=γ
StandardDeviation of
themeasurement
Slope
AnalyticalSensitivity
• Consider the use of an analytical balance. ASTMprovides guidelines on the acceptable threshold ofnoise on the sensitivity of the balance (USPreference).
• Amplification of the signal won’t be the answersince this will also increase the Ss.
• A disadvantage is that γ is concentration depend.
49
• This definition accounts for noise at a specificconcentration. The slope can also be replaced witha response factor.
Signal to Noise Ratio (S/N)(Chpt. 5, Skoog Holler Nieman)
sx
NS=
StandardDeviation of
themeasurement
MeanResponse
Signal toNoise Ratio
• This definition should look familiar...
RSDNS 1=
• What is NOISE… recall previous discussions.Simply, it is unwanted signal that can affect theproper detection and quantification of a desiredsignal. The S/N ratio has a direct impact on LOD.
50
• Noise has a direct effect on detection limit. Thedetection limit represents that concentration ofsignal which is no longer reproducibly or reliablydiscernable from noise, at which point no accuratestatement about analyte concentration can be made.Summary: we can see a signal due to the analytebut can’t exactly quantify the analyte concentration.
Noise and Detection Limit...
• Let’s review:
Not Detectable
LOD or MDL
LOQ (blank + 10 σ blank)
Estimated Analyte Concentration Using Test Method
Increasing Confidence
51
• ACS - basic definition of detection limit (seeequation below). The ACS approach is generallyaccepted; however, it has numerous limitationswhich we will explore. Note that k is usually takenas 3.
...Noise and Detection Limit...
blSblm kSS +=Standard
Deviation for“Blank”
Mean blanksignal
MinimumDistinguishable
AnalyticalSignal
mk
mSSC blSblm
m =−
=
Slope
Detection Limit
52
• Reducing noise or improving the signal will givebetter detection limits.
• For example, signal averaging (see below) appearsto pull the signal out of the noise. What ishappening here is due to the fact that “Noise isRandom”. Therefore, with enough acquired data,the noise will eventually cancel itself out. Considerthe concept of constructive and destructive waves.
• Below is an example of the effect of S/N ratio onthe NMR spectrum of progesterone.
...Noise and Detection Limit...
53
• Consider a 10 W light bulb, blue light.• In absolute darkness, that bulb, when turned on, would be
quite simple to detect. In fact, depending on the detector,say your eye, you should even be able to tell thedifference between 5 W, 10W and 15 W light bulbs.
• What if the background was contaminated, say with lightfrom one 1 W yellow light bulb.
• Now, what if the background was contaminated withlight from fifty 1 W yellow light bulbs. Would you stillbe able to see the 10 W blue light bulb with thatbackground?
The Blue Bulb...
YellowFilter
O RV
BGY
54
• It must be appreciated that detection limits are not“fixed” values, and will change with time, analyst,equipment, etc.
• Detection limits require that we can distinguishbetween a background or blank signal and a signalfrom a target analyte.
• A quantitative determination is typically taken as asystem with less than 10% RSD, less for moreserious applications such as pharmaceutical.
• The ACS has defined three important levels formeasurement of data
– LOD (blank signal+3σblank)– RDL (blank signal+6σblank)– LOQ (blank signal+10σblank)
More on S/N& Detection Limits...
55
• It must be appreciated that detection limits are not“fixed” values, and will change with time, analyst,equipment, etc.
• Normally distributed data will allow you to betterestimate the contribution of noise to your desiredsignal.
... More on S/N& Detection Limits
56
• The null hypothesis: No difference between theblank and sample (non-detect) (µA=µB).
• The alternative hypothesis: The sample response isgreater than the blank (detection) (µA>µB).
• Type I Decision Error:– Rejecting the null hypothesis when it is true (false
positive) (α)• Type II Decision Error:
– Accepting the hypothesis when it is false (falsenegative) (β)
• So, two errors are possible: False Positives andFalse Negatives depending on the hypothesis.
Type I and Type II Decision Errors
61
Summary
• Significant effort is made to understand thecontribution of noise on the signal from a targetanalyte.
• Chapter 5 (Skoog) brings us into the next level ofunderstanding noise, how to manage it, and how topull out signals from what would otherwise beuseless data.
62
Analytical Instrumentation• Some critical features of an analytical
instrument:– detect the target analytes in discrete manner
(say following chromatography)– little to no background signal without sample
and/ or standard– responsive to analyte - increase in instrument
response with increasing mass loading of targetanalyte
– deliver data to the human in a discernablemanner
• Almost all instrumentation is integratedwith op amps (operational amplifiers)
• The instrument takes physicalmeasurements and converts to digital -significant advantages to working in digitaldomain
66
More on Domains
• Non Electrical• Length• Mass• Pressure• (Digital)• Light Intensity
• Electrical• Analog• Digital• Time
68
• Magnitude of current, voltage, charge,power
• Continuous in both amplitude and time• Analog signals are very susceptible to
electrical noise.• Noise: Signals that can interfere with the
desired signal in an analytical process.
Analog Domains
69
• Here the fluctuating signals are plotted as afunction of time….f(t), as in FTNMR,FTIR, FTMS.
• These data can be transformed tofrequency via Fourier Transformation
Time Domains
70
• Can convert Time-Domain Signals into Digital-DomainSignals
• Data are stored as a series of numbers-encoded as either 0or 1, also ON or OFF.
• Information can be stated as in Fig. 1-5 (Logic LevelSignal) Number 14 - explain.
• A better method is to encode data into Binary Numbers -stored as a bit.
Digital Domains
72
• Resistance Ohms
• Current Amps
• Voltage Volts
• Capacitance Farads
• Energy Joules
• Power Watts
• Coil (Inductance) Henry’s
Useful Electrical Terms
3
2
sAmkg⋅⋅
c)(Charge/seA
32
2
sAmkg⋅⋅
2
42
mkgsA
⋅⋅
3
2
smkg ⋅
22
22
sAmkg⋅⋅
2
2
smkg ⋅
73
Overall Noise• We can expect that individual aspects of an
analytical measurement, when combined, result inan “overall noise”, which can be generallydescribed with the equation “sum of variances”.
• e.g. G=glassware, C=Chemical, I=Instrument.• S (overall) can be minimized by ensuring the the
individual noise components are controlled:– Class A glassware– Controlled environmental conditions– Well-designed instrumentation
2222
22222
...
...
nICG
nICG
SSSSS
SSSSS
+++=
+++=
Signals & Noise,Signals & Noise,Noise FiltersNoise FiltersChapter 5, Skoog
74
Instrumental Noise
• Four primary groups of InstrumentalNOISE:– Thermal (Johnson)
– Shot
– Flicker
– Environmental
fkTRvrms ∆⋅= 4
fIeirms ∆⋅= 2
fFN
1=
75
Thermal Noise
• Generally speaking, caused by electron flow (or othercharged carriers) through components in an instrument.Basically, thermal agitation of charge carriers(electrons/holes).
– E.g. Through resistors, capacitors, electrochemicalcells.
• Thermal noise: fkTRvrms ∆⋅= 4• Boltzmann’s constant: 1.38x10-23 J K-1
• T is temperature, K• R is resistance, Ω• ∆f is Bandwidth, Hz (range of frequencies)
• Note that the equation shows that thermal noise isactually independent of the actual frequency. This noiseis sometimes referred to as “White” Noise.
• Cooling key components reduces both T and R, therebyreducing thermal noise. So, cooling from RT to 77 K(with liq N2) will halve the white noise.
• Note that narrowing bandwidths makes the instrumentsluggish.
76
Shot Noise
• Generally speaking, caused by electron flow (or othercharged carriers) through junctions in an instrument.
– Across pn junctions, across the vacuum from anodeto cathode, in photocells…later.
• Shot noise: fIeirms ∆⋅= 2• I is the average direct current, A• e is the charge on the electron, 1.60x10-19 C• ∆f is the bandwidth of frequencies being considered.
• Note that the equation shows that shot noise is actuallyindependent of the actual frequency. This noise issometimes referred to as “White” Noise.
• Reducing the bandwidth is the key to reducing this noise.• But remember that narrowing bandwidths makes the
instrument sluggish!
77
Flicker Noise
• Generally speaking, directly related to the inverse of thefrequency being studied.
• Flicker noise:f
FN1
=
• 1/f is the inverse of frequency being studied.
• The origins of flicker noise are not understood.• This noise can be reduced by using AC or modulated
signals…later.• Only really significant at f<100Hz
F lic k e r N o ise D e p e n d e n c e
0
1
2
3
4
5
6
7
8
9
1 0
0 0 .1 0 .2 0 .3 0 .4 0 .5
1 /f
Vol
tage
(V, D
.C.)
78
Environmental Noise
• The low frequency noise is “Flicker” and the rest isradiant energy, motors, mechanical vibrations, AM, FM,TV.
• The reduction of all forms of noise is important inachieving the ultimate detection limit.
• Two “quiet” regions. From ~3 Hz to 60 Hz, and fromabout 1KHz to about 500 kHz (AM frequency).
79
Ensemble Averaging
• Successive (replicate)data stored as anarray, and added.
• NMR and FTIR• Skoog Holler Nieman
p107
nNS∝
Software Techniquesto Remove Noise(To Increase S/N)
80
S/N
• Nyquist (sampling) Theorem– Data acquisition frequency must be at least 2X (usually
10X) the highest frequency of the signal– delta t is the time interval between signal samples
tf
∆=
21
nNS∝
81
Boxcar Averaging
• Assumes that noise is random, and that the signal variesslowly with time.
• Average of a small number of adjacent data points, yieldsa response this is better than 1 point. (Nyquist SamplingTheorem).
82
Shielding and Grounding
• Noise arising fromenvironmentallygeneratedelectromagnetic radiationcan be reduced byshielding and grounding.
• Electromagnetic radiationis absorbed by the shieldrather than by theenclosed conductor,particularly true when thecable is carrying smallcurrents.
• Ground everythingthrough a single point toavoid capacitivecoupling.
Hardware Techniquesto Remove Noise(To Increase S/N)
83
Difference andInstrumentation Amplifiers
• Noise rejection circuits.• Later when we learn about operational amplifiers.• Before we learn about amplifiers, we have to cover basic
RC circuits (resistance-capacitance).
84
Modulation
• Hollow cathode lamp in Atomic Absorption• Amplification of DC signals is often troublesome due to
Flicker Noise• Low frequency signals are often converted to higher
frequency signals - MODULATION
• Modulation is also used in IR spectroscopy (wavelength)modulation (FTIR).
• We will learn more about this later.
85
Signal Chopping:Chopper Amplifiers & Lock-In Amplifiers
• Somewhat similar concepts - chopper• Signal is converted to square waves (Chopper)• Can detect modulated signals by following reference ν (LIA)
– Can recover signals even when S/N<1.• Atomic Absorption is a good example (Ch. 9, Skoog)
HollowCathode
Lamp
Chopper
MirroredReference
(split)
Impingeson Sample
Sample
ModulatedSignal
HardwareandSoftware
86
Analog FiltersLow Pass Filter
Skoog Chapter 2
• Common to use Low Pass Filters.• Attenuates high frequency noise and passes low
frequency signals.• This concept is used with speakers - woofer requires low
pass filter (5 to 200 Hz) (could use a loop as well). Thetweeters require high pass filter (12 to 20 kHz).
88
Characteristics & Propertiesof an Op Amp
• An OP AMP is usually made up of:– resistors, capacitors, diodes, transistors
• It is an active device, with a power supply ofusually ± 15 V
• High gain• Fast response• Large input impedance• Small output impedance• Zero output for zero input• Cheap
• Op Amp is a term describing original use of thedevice - mathematical operations.
– Voltage amplification– Current to voltage conversion– Inverting signals– Comparing signals– Pulse shaping and formation– …many more uses….
The power supply isimplied here
89
Characteristics & Propertiesof an Op Amp
• Difference mode: difference between the two inputsis amplified and appears as the output.
• E.g., an Op Amp with a GAIN of 1x104, and aninput voltage of 1 mV, produces an output voltageof 10V. ( )−+ −⋅= vvGAINvo )(β
• Typical open loop GAIN: 104 to 106
• Typical input impendances: 106 to 1013 Ω• Typical output impendances: 1 to 10 Ω• Less than 0.1 mV output voltage for 0 V input• If -ve battery terminal is connected to the inverting
input, then the output is +ve.• If ac is connected to -ve terminal, then output is
180° out of phase.
90
Limitations of OPEN LOOPOperational Amplifiers
• Even though the Gain is high (vo/vi=Gain), theprimary limitation of the “open loop” Op Amp isthe roll-off of GAIN as the signal input frequencyincreases. Limits the usefulness of the OpAmp.
• The roll-off is substantial, and is corrected byproviding “feed-back” across vo and vi
• The closed loop Gain, although lower at somefrequencies relative to the open loop, is constantover an extremely large range of frequencies.
91
CLOSED LOOPOperational Amplifiers
• This operational amplifier uses negative feedback.• It has the effect of making the Gain of the circuit
independent of the OP AMP (proof later).• Output is 180° out of phase with input.
Point S is the Summing Point.Rf is the feedback resistanceRi is the input resistanceii is the input currentif is the feedback currentiS is the current flowing into the amplifier
vi
+
-
Rf
ifRi
voiiis
S
92
• The current through Ri is given by:
• The current through Rf is given by:
• Because the input impendance of the Op Amp is so high,is is negligible (current flowing into Op Amp).
• Therefore,
• Recall that:
• So, solving for v-
• v- is negligible relative to vout and vin , and because β isso large
• Therefore,
The Inverting Op AmpSkoog Holler Nieman page 56 (sect 3B-1)
vi
+
-
Rf
ifRi
voiiis
S
( )i
ini R
vvi −−=
( )f
outf R
vvi −= −
fi ii ≅( )
f
out
i
in
Rvv
Rvv −
=− −−
)( +− −−= vvvout β
βoutvvv −= +−
0vout ≈β
f
out
i
in
Rv
Rv −
=in
out
i
f
vv
RR
−=
93
vo=
vi from v+
-ii
Summary of Closed Loop Op Amp Applications
vi+- vo
Zi
Zf
Zi ZfFunction
VoltageFollower
CurrentFollower
Voltage Amp. or ÷
Differentiation
Integration
Summation
dtvCR
1 t
0i
i∫−
dtdvCR i
f−
R1R2R3R4
+++−
4
4
3
3
2
2
1
1f R
vRv
Rv
RvR
i
fi R
Rv ⋅−
94
Summary of Closed Loop Op Amp Applications
• Op Amps can be used for the measurement andamplification of transducer signals
• Op Amps are also used for:– Inverting– Comparators– Filters– Switching– Waveform generation
• The amplification/division Op Amp is mostcommonly used
vi
+
-
Rf
ifRi
voiiis
S
• If vi=1V and Rf=100Ω and Ri=10 Ω, then vo= – 10V• If vi=1V and Rf=10Ω and Ri=100 Ω, then vo= – 0.1V
i
fio R
Rvv ⋅−=
95
Voltage Follower Circuit
+
-
vivo
• This circuit is used for the accurate measurement ofpotential (voltage). Recall that we don’t want to introduce aloading error, so the Op Amp is ideal for potentialmeasurements because of its extremely large inputimpendance (~1012Ω). Application to pH measurement.
• This is a non-inverting amplifier.
−+ ≅ vv that know We oi vvvv == −+ and that and
oi vv = therefore,• This circuit has unit voltage gain, but provides extremely
large power gain related to input and output impendances.
oi vv =
RvviP
2==
o
i2
i
i
o
2o
input
output
ZZ
vZ
Zv
PP
Gain =⋅==
96
+-
Rf
if
vo
ix
Current Follower CircuitSHN p58
• Examples of current measurements include: phototubes,photomultipliers, flame ionization and electron capturedetectors (pA current outputs), voltammetry.
• These circuits provide current to voltage conversion.
fx ii ≅
fxffo RiRiv −=−=
f
ox R
vi −=∴
mL
mRR
REr+
−=%
• So, if Rf is 100 kΩ and ix is 1µA, then vo is 100 mV. Thisvoltage can be easily measured or amplified, as required.
• Loading error is important to consider with these devices...
97
Integration of Voltage Using an Op Amp
• This circuit can integrate a signal that varies as afunction of time. For example, it is used in aphotodiode array detector to integrate the magnitudeof voltage loss following direct exposure of the n-psilicon to light energy.
• There is a “reset” switch that will discharge thecapacitor, and have the circuit ready for the nextmeasurement.
vi
+
- vo
Cf
Ri
HoldSwitch
ResetSwitch
1. Open the hold switch and close the reset to discharge thecapacitor
2. Open the reset switch and close the hold switch3. Open the hold switch• This is a ramp generator
V
Time
98
• With the hold switch closed (the capacitor begins tocharge), we know that:
• The current in the capacitor is given by:
• We can write:
• Integrating from initial voltage and times to finalvoltage and times, recognizing that both initialvoltage and initial times are zero (0).
• Remember, both Ri and Cf are constant, so they arenot part of the integral; however, vi is dependent ontime, so it remains in the integral - which, after all,makes sense.
Integration of Voltage Using an Op Amp
fi ii =
dtdvCi o
f −=
dtdvC
Rv
Rvi o
i
i
i
ii −=∴= dt
CRvdvi
io −=
dtvCR
1dv22O t
0i
i
V
0o ∫−=∫ dtv
CR1v
t
0i
io ∫−=