John M. Boone, Ph.D., FAAPM, FSBI, FACR, FAIMBE Professor of Radiology & Biomedical Engineering University of California Davis Sacramento, California Changing Perceptions and Updated Methods for Mammography Dosimetry Andrew M. Hernandez, B.S., the brains behind all this Post-Graduate Researcher Biomedical Engineering Grad Group University of California Davis Sacramento, California

Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

John M. Boone, Ph.D., FAAPM, FSBI, FACR, FAIMBE

Professor of Radiology

& Biomedical Engineering

University of California Davis

Sacramento, California

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Andrew M. Hernandez, B.S., the brains behind all this

Post-Graduate Researcher

Biomedical Engineering Grad Group

University of California Davis

Sacramento, California

Page 2: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

• Different anode materials (W)

• Higher Tube Potentials (digital)

• Different filter materials (e.g. Al, Ag, Pd, etc.)

• New knowledge about breast geometry & composition

• Skin Thickness

• Breast Density (magnitude)

• Glandular distribution in the breast

• Breast Tomosynthesis (not addressed in this symposium)

Why are things changing?

Changing Perceptions and Updated

Methods for Mammography Dosimetry

Page 3: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

This morning’s theme:

A series of research vignettes

Examples of clinical utility

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

research clinical

application

Page 4: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Breast CT as the Backstory

Skin Thickness

Breast Density / the Myth

New Mammography Spectra

Density Heterogeneity

Summary

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

Page 5: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

5

Tony

Seibert

Karen

Lindfors

Shadi

Shakeri

Simon

Cherry

Katie

Metheany

Shonket

Ray

Nathan

Packard

Dandan

Zheng

Ramsey

Badawi

George

Burkett

Whit

Miller

Fareedah

Simon

John

Brock

Clare

Huang

Kai

Yang

Orlando

Velazquez

Hong

Zhou

Alex

Kwan

Jessie

Xia

Carey

Floyd

Tom

Nelson

Craig

Abbey

Norbert

Pelc

Bruce

Hasegawa

Elizabeth

Krupinski

Varian Imaging Systems

Larry Partain

Gary Vishup

John Pavkovich

Hussan Mostafavi

Gerhard Roos

Ed Seppi

Cesar Proano

Anita

Nosratieh

Nicolas

Prionas

Sarah

McKenny

Lin

Chen

Loren

Niklason

Jeff

Siewerdsen

Martin

Yaffe

Linda

Phelps

Laurie

Boling

John

McGahan

Andrew

Hernandez

John

Boone

Peymon

Gazi

Page 6: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Computer aided design / computer aided manufacture (CAD/CAM)

Cambria 2011

Bodega 2007

Albion 2003

Doheny 2014 6

Page 7: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Albion 2004

Doheny 2015 Cambria 2011

Bodega 2007

Page 8: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

8

Doheny: Mechanical Assembly

Page 9: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

2003 2008 2010

FDK Reconstruction Code

~42 minutes ~35 minutes ~20 seconds

>100x

9

Preprocessed Projection Images

Reconstructed breast CT images

graphics processor unit (GPU)

main motor

collimator wheel

filter wheel

angle encoder

generator control

X-ray tube elevation

PET 1

PET 2

thermal sensor

inter-locks

detector acquisition

devices

motors

System Integration

Raw Cone-Beam Projection Images

Page 10: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Pendant Geometry Imaging

(no compression)

Page 11: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Albion

Cambria

Bodega

Doheny

UC Davis Medical Center, Sacramento

11

Page 12: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

12

Doheny (2015)

Cambria (2011)

Bodega (2007)

Spatial Resolution Improvements

Page 13: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Clinical Imaging

• Patients: women with suspicion of breast cancer (BIRADS 4 & 5’s)

• First bCT scan: Nov 22, 2004

• >600 women on UC Davis scanners

• ~2000 bCT volume data sets

• ~260 have had contrast injection

• Radiation dose same as 2V mammography

• Image reconstruction 5123 or 10243

13

Page 14: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

14

bCT (no injected contrast)

Page 15: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

15

Mass Lesions Microcalcifications

Page 16: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

16

Contrasted Enhanced breast CT pre

pre post

post

Malignant

benign

AUC = 0.87

Page 17: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

17

Dedicated breast PET/CT

Whole-body PET/CT

Invasive Mammary Carcinoma

Page 18: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Clinical Implications

Page 19: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

19

CC

CC MLO

MLO

Two 2D mammograms Volumetric breast CT data

~500 contiguous images

MLO

Page 20: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for
Page 21: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Breast CT as the Backstory

Skin Thickness

Breast Density / the Myth

New Mammography Spectra

Density Heterogeneity

Summary

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

Page 22: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Past Monte Carlo Studies typically assumed a 4 mm

(or 5 mm) skin thickness for breast dosimetry

X Wu, GT Barnes, DM Tucker, Spectral dependence of glandular tissue dose in screen-film mammography, Radiology 179: 143-148: 1991 DR Dance, Monte Carlo calculation of conversion factors for the estimation of mean glandular dose, PMB 35: 1211-1219: 1990 X Wu, EL Gingold, GT Barnes, DM Tucker, Normalized average glandular dose in molybdenum target-Rhodium filter and rhodium-target-rhodium filter mammography, Radiology 193: 83-89: 1994 DR Dance, CL Skinner, KC young, et al., Additional factors for the estimation of mean glandular dose using the UK mammography dosimetry protocol, PMB 45: 3225-3240: 2000

Page 23: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Radiology

Medical Physics

Page 24: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

x-rays

W/Mo (50 mm)

skin

homogeneous

breast tissue

Page 25: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

25

Observation from breast CT images: Skin is not 4 mm thick on the breast

Page 26: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Medical Physics

Page 27: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Verified the spatial accuracy in three dimensions using a phantom…

phantom concept

z-dimension x-dimension y-dimension

results

bCT scanner

Page 28: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

average = 1.45 mm (s = 0.30 mm)

N = 100 breasts N = 51 women

Segmentation

Algorithm

Measurements Skin Thickness Results

Page 29: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Clinical Implications

Page 30: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing the skin thickness from 4.0 mm to 1.5 mm

increased the DgN values by about 17-18%

But this assumes a homogeneous breast composition

Page 31: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Breast CT as the Backstory

Skin Thickness

Breast Density / the Myth

New Mammography Spectra

Density Heterogeneity

Summary

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

Page 32: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

32

Page 33: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Areal glandular fraction (2D)

Volume glandular fraction (3D)

Page 34: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

X Wu, GT Barnes, DM Tucker, Spectral dependence of glandular tissue dose in screen-film mammography, Radiology 179: 143-148: 1991 DR Dance, Monte Carlo calculation of conversion factors for the estimation of mean glandular dose, PMB 35: 1211-1219: 1990 X Wu, EL Gingold, GT Barnes, DM Tucker, Normalized average glandular dose in molybdenum target-Rhodium filter and rhodium-target-rhodium filter mammography, Radiology 193: 83-89: 1994 DR Dance, CL Skinner, KC young, et al., Additional factors for the estimation of mean glandular dose using the UK mammography dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for monoenergetic and high-energy x-ray beams: Monte Carlo assessment, Radiology 213: 23-27: 1999 JM Boone, Normalized glandular dose (DgN) coeffiecients for arbitrary x-ray spectra in mammography: computer-fit values of Monte Carlo derived data. Med Phys 29: 869-875: 2001

All of these papers assumed that “aerial glandular density” was equal to “volume glandular density”

To be clear, there is no such thing as a 100% glandular breast

Only ~4% of women have a volume glandular fraction >50%

Page 35: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

SPIE

Page 36: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Original Breast CT image After 3D Median filtering After Adipose Flattening

Final Segmented Image

After Adipose Smoothing

glandular tissue

adipose tissue skin

Page 37: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

CT Number

Nu

mb

er o

f V

oxe

ls

CT Number

Nu

mb

er

of

Vo

xels

Original Breast CT image After Segmentation Algorithm

initial threshold final threshold

air

glandular

tissue

adipose

tissue

air

glandular

tissue

adipose

tissue

Page 38: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

skin glandular

adipose

air

Page 39: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Medical Physics

Page 40: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

excluding skin:

The Volume Glandular Fraction (VGF)

including skin:

glandular

glandular + adipose VGF =

glandular

skin + glandular + adipose VGF =

N = 191 (bCT only)

Page 41: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

N = 2831

Average = 19.3% Validation of Toronto versus UC Davis density assessment techniques

Page 42: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

3.5%

Median (~16% VGF)

Page 43: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

3.5%

Median (~16% VGF)

10%

Page 44: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for
Page 45: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Clinical Implications

Page 46: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

28 kV W / Rh Thickness = 49 mm 157 mAs

Page 47: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

SSD

SCD

SID

KISL = 0.99

Page 48: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

W/Rh @ 28 kV HVL = 0.498 mm Al Output (at 65.5 cm) = 4.8 mR/mAs

Half Value Layer Radiation Output

Tungsten / Rhodium Combination

Page 50: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

DgN12% = 0.290 mGy/mGy

DgN50% = 0.254 mGy/mGy

12.5% Breast Density

50% Breast Density

28 kV HVL = 0.498 mm Al

Page 51: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Incident Air Kerma to Breast:

4.8 mR/mAs × 157 mAs × KISL = 746 mR = 6.51 mGy [EAK]

Dose (50%) = 6.51 mGy × 0.254 mGy/mGy = 1.65 mGy

Dose (12%) = 6.51 mGy × 0.290 mGy/mGy = 1.89 mGy

14.4% increase due to more accurate glandular fraction assessment

1.17 × 1.144 = 1.34 or 34 % greater dose

skin thickness

glandular fraction

Page 52: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Breast CT as the Backstory

Skin Thickness

Breast Density / the Myth

New Mammography Spectra

Density Heterogeneity

Summary

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

Page 53: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

• Limited kV range: 18 to 42 kV

• Derived from measurements on x-ray tubes of the past

PHILIPS

MICRODOSE

GE

SENOGRAPHE CARE

HOLOGIC

SELENIA DIMENSIONS

SIEMENS

MAMMOMAT INSPIRATION

Why do we need another spectral model

Page 54: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

GE SENOGRAPHE

Rh Mo e-

Be window

• large & small focal spot on each target

HOLOGIC DIMENSIONS

W e-

Be window

• large & small focal spot

SIEMENS MAMMOMAT

Mo W

e-

Be window

• large & small focal spot on each target

W

Be window

PHILIPS MICRODOSE

e-

• large focal spot

Monte Carlo Simulation Geometry

central ray

detection

plane

Page 55: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

24 cm

27 cm

12 cm

18 cm

LARGE

14 cm

25 cm

MEDIUM SMALL

30

cm

19 cm

Detection Plane Size

Page 56: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

HOLOGIC: W–Be (630 μm) HOLOGIC: W–Rh (50 μm)

• 0.4 % difference in HVL, averaged across all vendors, after conventional filtration is applied.

=> Detection plane size does not affect spectral shape

Page 57: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

GE: Mo – Be (690 μm) GE: Mo – Mo (30 μm)

• 1.8 % difference in HVL, averaged across all vendors, after conventional filtration is applied.

=> Focal spot size does not affect spectral shape

Page 58: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

W–Be Wo–Rh (50 μm)

• 0.3 % difference in HVL, for W anode systems , after conventional filtration is applied.

Mo–Be Mo–Rh (25 μm)

• 0.4 % difference in HVL, for Mo anode systems , after conventional filtration is applied.

Page 59: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

W–Be Wo–Rh (50 μm)

• 0.3 % difference in HVL, for W anode systems , after conventional filtration is applied.

Mo–Be Mo–Rh (25 μm)

• 0.4 % difference in HVL, for Mo anode systems , after conventional filtration is applied.

For a given anode composition, vendor-specific geometrical

differences do not affect spectral shape

Page 60: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

A single mammography system geometry can be used

for all commercial systems

methodology for new spectral model adopted from:

8.5 keV (Lα)

10.0 keV (Lβ)

bremsstrahlung

W anode

17.5 keV (Kα)

20.0 keV (Kβ)

bremsstrahlung

Mo anode

20.5 keV (Kα)

23.0 keV (Kβ)

23.5 keV (Kβ)

bremsstrahlung

Rh anode

Page 61: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Molybdenum, Rhodium, and Tungsten Anode Spectral

Model using Interpolating Cubic Splines

Mo–Mo (30 μm) Rh–Rh (25 μm) W–Ag (50 μm)

• 20 to 60 kV (1 kV intervals)

• 0.5 keV energy resolution

• Minimal filtration (0.77 mm Be)

MASMICS, RASMICS, TASMICS

Page 62: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Mean glandular dose coefficients (DgN) for

contemporary mammography systems

x-ray spectra DgN(E)

& =>

DgN

Page 63: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Breast CT as the Backstory

Skin Thickness

Breast Density / the Myth

New Mammography Spectra

Density Heterogeneity

Summary

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

Page 64: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Glandular dose is the metric of interest !

“Detailed information will have to be obtained on the amount and distribution of

gland tissue in many individual cases before individual risk estimates can be made”

Page 65: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

x-rays

skin skin

heterogeneous

breast tissue

homoogeneous

breast tissue

Heterogeneous

(VGF = 20%)

Homogeneous

(VGF = 20%)

x-rays

Page 66: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

10 – 43% overestimation using structure phantoms

9 – 59% overestimation using unstructured phantoms

27% overestimation using simulated mechanical

compression of bCT images in 20 patients.

Page 67: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

0

20

40

60

80

100

0.0 0.2 0.4 0.6 0.8 1.0

0 1

0

10

20

30

40

50

60

0.0 0.2 0.4 0.6 0.8 1.0

GF

sag

itta

l (%

)

Relative y position Relative radial distance

0

20

40

60

80

100

0.0 0.2 0.4 0.6 0.8 1.0

GF

co

ron

al

(%)

Relative z location

RG

F (

%)

MEDICAL PHYSICS

Page 68: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Huang S.-Y., Boone J.M. et al. “The characterization of breast anatomical metrics using dedicated breast CT” Med Phys. 38 (4), April 2011

r=0 r=1

RG

F1 (%

)

Courtesy of S.Y. Huang

0

20

40

60

80

100

0.0 0.2 0.4 0.6 0.8 1.0

0

20

40

60

80

100

0.0 0.2 0.4 0.6 0.8 1.0

0

20

40

60

80

100

0.0 0.2 0.4 0.6 0.8 1.0

Relative radial distance (r)

RG

F2 (%

) R

GF

3 (%

)

Relative radial distance (r)

COM

Relative radial distance (r)

Radial Glandular Fraction (RGF)

Page 69: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

0

20

40

60

80

100

0.0 0.2 0.4 0.6 0.8 1.0

0 1

0

10

20

30

40

50

60

0.0 0.2 0.4 0.6 0.8 1.0

GF

sag

itta

l (%

)

Relative y position Relative radial distance

0

20

40

60

80

100

0.0 0.2 0.4 0.6 0.8 1.0

GF

co

ron

al

(%)

Relative z location

RG

F (

%)

MEDICAL PHYSICS

Page 70: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

coronal view

a

b

T

10 %tile 50 %tile 90 %tile

Size Dependence

25 %tile 75 %tile 50 %tile

Density Dependence 1.5 mm skin thickness

cranial-caudal view

c

a

Phantom Design

Page 71: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

rVGF º VGF for a given contoured region

max rVGF min rVGF

Modeled RGFs in compressed breast phantoms

Page 72: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

HOMOGENEOUS

(VGF = 20%)

SPATIALLY-INDEPENDENT HETEROGENEOUS

(VGF = 20%)

y = 0.999x – 0.0001

R2 = 0.999

Validating Methodology

Page 73: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Photon Energy (keV)

DgN

(E)

(m

Gy/m

Gy)

Photon Energy (keV)

DgN

(E)

(m

Gy/m

Gy)

Photon Energy (keV)

DgN

(E)

(m

Gy/m

Gy)

Photo

n F

luence

31 kV (Mo-Mo)

2.0E+6

1.5E+6

2.0E+6

1.5E+6

0

Photo

n F

luence

28 kV (Mo-Mo)

2.0E+6

1.5E+6

2.0E+6

1.5E+6

0

Photo

n F

luence

32 kV (Mo-Rh)

2.0E+6

1.5E+6

2.0E+6

1.5E+6

0

SMALL

MEDIUM

LARGE

VGF = 17.0%

VGF = 12.6%

VGF = 7.0%

homogeneous

heterogeneous

DgN(E): heterogeneous vs. homogeneous

Page 74: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Density Dependence

-36% -29% -38%

-35% -36% -39%

-34% Mo

-37% Mo

-23% W

-26% W

Size Dependence

pDgN: heterogeneous vs. homogeneous

Page 75: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

10% inferior

centered

10% superior

x-ray tube

-6%

-31%

-49%

Asymmetric shifts in glandular distributions

Page 76: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Breast CT as the Backstory

Skin Thickness

Breast Density / the Myth

New Mammography Spectra

Density Heterogeneity

Summary

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

Page 77: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

• The only breast density metric that matters in breast

dosimetry is volume glandular fraction

• A new understanding of breast geometry along with

updating anode/filter/kV parameters will make clinical

breast dosimetry more accurate

• Skin thickness would be important with homogeneous

breast tissue, but less so with heterogeneous models

Page 78: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Changing Perceptions and Updated Methods

for Mammography Dosimetry

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

• The mammography spectra presented here will be

made available in spreadsheet format (after it’s

published) by request

• The DgN values described in a PMB publication (200

pages) will be provided by request

• Heterogeneous breast models represent the next

generation in breast dosimetry

Page 79: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Future Directions

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

• Generate more realistic breast shapes under compression

for Monte Carlo studies

• Model MLO projection different from CC

• With some tomosynthesis systems going to much wider

angles, a full angled-beam MC analysis will be necessary

• The cycle of CAP Research CAP Research CAP

continues

Page 80: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

Future Directions

AAPM Spring Clinical Meeting

Salt Lake City, Utah March 2016

New AAPM Task Group on Breast Dosimetry

• Ioannis Sechopoulis

• David Dance

• Ken Young

• R. Vanegen

• John Boone

Page 81: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

New AAPM Task Group on Breast Dosimetry

• Ioannis Sechopoulis

• David Dance

• Ken Young

• R. Vanegen

• John Boone

Page 82: Changing Perceptions and Updated Methods for ...amos3.aapm.org/abstracts/pdf/113-31143-379492-117897...dosimetry protocol, PMB 45: 3225-3240: 2000 JM Boone, Glandular breast dose for

John M. Boone, Ph.D., FAAPM, FSBI, FACR, FAIMBE

Professor of Radiology

& Biomedical Engineering

University of California Davis

Sacramento, California

Changing Perceptions and Updated Methods

for Mammography Dosimetry

Andrew M. Hernandez, B.S., the brains behind all this

Post-Graduate Researcher

Biomedical Engineering Grad Group

University of California Davis

Sacramento, California