6
Volume 3 • Issue 12 • 1000170 J Bioremed Biodeg ISSN: 2155-6199 JBRBD, an open access journal Open Access Rada et al., J Bioremed Biodeg 2012, 3:12 DOI: 10.4172/2155-6199.1000170 Open Access Research Article Keywords: Moisture; Respirometry; Respirometric index; Temperature Introduction Since 1999, with the Directive 1999/31/CE (Italian Council Directive 1999), the European Union (EU) has imposed limits for waste landfilling (lower heating value and biodegradable matter targets) and also technical requirements for waste treatment plants in order to decrease their negative effects on the environment [1]. To this concern, one of the most important parameters is the respiration activity. In Italy, with the transposition of the EU technical norm CEN/TR 15590 (Italian National Standard Body 2007) into the decree 205/2010, this parameter is also requested for the output of the mechanical- biological plants aimed to energy recovery through solid recovered fuel (SRF) generation [2,3]. e biomass respirability or the Respirometric Index (RI) is the amount of oxygen consumed from the microorganisms. Generally, oxygen consumption is referred to as an amount of volatile solids (VS, also called Organic Matter, OM) during a fixed time [4]. e respirometric tests based on oxygen consumption can be classified in static and dynamic methods depending on the oxygen aeration type [5] Absent aeration (static): O 2 consumption is usually measured via pressure differences or via direct measurements of the O 2 content in air or liquid medium [6-8]; in Italy static respirometry is established by the Italian Organization for Standardization with the norm UNI 10780 (Italian Standard Body 1998) [9]. Continuous aeration (dynamic): it requires the precise measurement of the air flow rate and the O 2 content at the inlet and outlet of the test devices [10,11]; in Italy, dynamic respirometry is established by the Italian Organization for Standardization with the norm UNI/TS 11184 (Italian Standard Body 2006) [12]. e static method limits oxygen circulation and dispersion in the substrate slowing the degradation processes of the organic matter. Moreover, it does not allow the removal of the exhausted air from the instrumentation. e pH decrease and the toxicity phenomena related to CO 2 or other fermentation gas accumulation cause the slowing of the biological activity. For those reasons, the static method underestimates the oxygen consumption. On the contrary, in the dynamic method, the measurement of the oxygen consumption is made taking into account the continuous aeration. Additionally, differences related to the respirometric index methods concern the time of reference for the assessment of its value. In Austria and in Germany, for instance, the reference parameter is the AT 4 which represents the respirometric activity calculated on four days and referred to dry matter (DM). To this concern, the limit for waste landfilling is respectively 7 mg O2 g DM -1 and 5 mg O2 g DM -1 [5]. In general, the dynamic method to which the Italian regulations refer is the Costech one, developed by DiProVe [13]. Parallel to this approach, in the University of Trento another system for measuring the dynamic respirability of a substrate, the AIR-nl Respirometer, has been developed [14,15]. Considering the significant differences of approach of the two instruments and more in general the fragmentation of the national regulations on respirometry, the present paper was developed with the aim of contributing to a correct adoption of both of the devices, in order to obtain only one respirability value from both of them. To this concern, in this paper, the relationship between the output of tests conducted in parallel with the two instruments on the same samples was analyzed. Differences between the results of the measurements and the obtainable parameters (RI 24 ) were underlined too. Some considerations on reference moisture for sample pre-treatment, test *Corresponding author: Rada EC, Department of Civil and Environmental Engineering, University of Trento, Via Mesiano 77, I-38050 Trento, Italy; E-mail: [email protected] Received August 01, 2012; Accepted October 10, 2012; Published October 27, 2012 Citation: Rada EC, Ragazzi M, Venturi M (2012) Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi- Continuous Aeration. J Bioremed Biodeg 3:170. doi:10.4172/2155-6199.1000170 Copyright: © 2012 Rada EC, et al. This is an open-a ccess article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi-Continuous Aeration Rada EC*, Ragazzi M and Venturi M Department of Civil and Environmental Engineering, University of Trento, Via Mesiano 77, I-38050 Trento, Italy Abstract In the sector of respirometry applied to solid waste many tools are available. Moreover, methods for assessing the oxygen consumption can be significantly different. In Italy, the sector is characterized by the adoption of dynamic approaches that are preferred to the static ones. In this paper, a comparison between the two respirometers more adopted in Italy is presented. The first one is the Costech respirometer, with continuous aeration and without temperature control. The second one is the AIR-nl respirometer, with temperature control and a semi- continuous aeration based on the principle of avoiding limiting conditions of oxygen. Results were obtained during an experimental research that involved on one hand the construction of an analytical model for evaluating the Respirometric Index (RI) for various substrates and on the other, the critical analysis of the obtained measurements. In this paper, the relationship between the output of tests conducted in parallel with two different instruments on the same samples is analyzed. Some differences between the results of the measurements and obtainable parameters (RI 24 ) are underlined. Considerations on reference moisture, temperature role, and respirometric index calculation method allow the understanding of the characteristics of the two methods in more detail. Journal of Bioremediation & Biodegradation J o u r n a l o f B i o r e m e d i a ti o n & B i o d e g r a d a t i o n ISSN: 2155-6199

Bioremediation & Biodegradation - OMICS International · Rada et al., J Bioremed Biodeg 2012, 3:12f ... Some differences between the results of the measurements and obtainable

  • Upload
    vophuc

  • View
    216

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Bioremediation & Biodegradation - OMICS International · Rada et al., J Bioremed Biodeg 2012, 3:12f ... Some differences between the results of the measurements and obtainable

Volume 3 • Issue 12 • 1000170J Bioremed Biodeg ISSN: 2155-6199 JBRBD, an open access journal

Open Access

Rada et al., J Bioremed Biodeg 2012, 3:12 DOI: 10.4172/2155-6199.1000170

Open Access

Research Article

Keywords: Moisture; Respirometry; Respirometric index;Temperature

IntroductionSince 1999, with the Directive 1999/31/CE (Italian Council

Directive 1999), the European Union (EU) has imposed limits for waste landfilling (lower heating value and biodegradable matter targets) and also technical requirements for waste treatment plants in order to decrease their negative effects on the environment [1]. To this concern, one of the most important parameters is the respiration activity.

In Italy, with the transposition of the EU technical norm CEN/TR 15590 (Italian National Standard Body 2007) into the decree 205/2010, this parameter is also requested for the output of the mechanical-biological plants aimed to energy recovery through solid recovered fuel (SRF) generation [2,3].

The biomass respirability or the Respirometric Index (RI) is the amount of oxygen consumed from the microorganisms. Generally, oxygen consumption is referred to as an amount of volatile solids (VS, also called Organic Matter, OM) during a fixed time [4]. The respirometric tests based on oxygen consumption can be classified in static and dynamic methods depending on the oxygen aeration type [5]

• Absentaeration (static):O2 consumption is usually measured viapressure differences or via direct measurements of the O 2 content inair or liquid medium [6-8]; in Italy static respirometry is establishedby the Italian Organization for Standardization with the norm UNI10780 (Italian Standard Body 1998) [9].

• Continuousaeration(dynamic):itrequirestheprecisemeasurementof the air flow rate and the O 2 content at the inlet and outlet of thetest devices [10,11]; in Italy, dynamic respirometry is established bythe Italian Organization for Standardization with the norm UNI/TS11184 (Italian Standard Body 2006) [12].

The static method limits oxygen circulation and dispersion in thesubstrate slowing the degradation processes of the organic matter. Moreover, it does not allow the removal of the exhausted air from the instrumentation. The pH decrease and the toxicity phenomena related to CO 2 or other fermentation gas accumulation cause the slowing of the

biological activity. For those reasons, the static method underestimates the oxygen consumption. On the contrary, in the dynamic method, the measurement of the oxygen consumption is made taking into account the continuous aeration.

Additionally, differences related to the respirometric indexmethods concern the time of reference for the assessment of its value. InAustriaandinGermany,forinstance,thereferenceparameteristheAT4 which represents the respirometric activity calculated on four days and referred to dry matter (DM). To this concern, the limit for waste landfilling is respectively 7 mgO2 gDM

-1 and 5 mgO2 gDM-1 [5].

In general, the dynamic method to which the Italian regulations refer is the Costech one, developed by DiProVe [13]. Parallel to this approach, in the University of Trento another system for measuring the dynamicrespirabilityofasubstrate,theAIR-nlRespirometer,hasbeendeveloped [14,15].

Considering the significant differences of approach of the two instruments and more in general the fragmentation of the national regulations on respirometry, the present paper was developed with the aim of contributing to a correct adoption of both of the devices, in order to obtain only one respirability value from both of them. To this concern, in this paper, the relationship between the output of tests conducted in parallel with the two instruments on the same samples was analyzed. Differences between the results of the measurements and the obtainable parameters (RI24) were underlined too. Some considerations on reference moisture for sample pre-treatment, test

*Corresponding author: Rada EC, Department of Civil and Environmental Engineering, University of Trento, Via Mesiano 77, I-38050 Trento, Italy; E-mail: [email protected]

Received August 01, 2012; Accepted October 10, 2012; Published October 27, 2012

Citation: Rada EC, Ragazzi M, Venturi M (2012) Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi-Continuous Aeration. J Bioremed Biodeg 3:170. doi:10.4172/2155-6199.1000170

Copyright: © 2012 Rada EC, et al. This is an open-a ccess article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi-Continuous Aeration Rada EC*, Ragazzi M and Venturi MDepartment of Civil and Environmental Engineering, University of Trento, Via Mesiano 77, I-38050 Trento, Italy

AbstractIn the sector of respirometry applied to solid waste many tools are available. Moreover, methods for assessing

the oxygen consumption can be significantly different. In Italy, the sector is characterized by the adoption of dynamic approaches that are preferred to the static ones. In this paper, a comparison between the two respirometers more adopted in Italy is presented. The first one is the Costech respirometer, with continuous aeration and without temperature control. The second one is the AIR-nl respirometer, with temperature control and a semi-continuous aeration based on the principle of avoiding limiting conditions of oxygen. Results were obtained during an experimental research that involved on one hand the construction of an analytical model for evaluating the Respirometric Index (RI) for various substrates and on the other, the critical analysis of the obtained measurements. In this paper, the relationship between the output of tests conducted in parallel with two different instruments on the same samples is analyzed. Some differences between the results of the measurements and obtainable parameters (RI24) are underlined. Considerations on reference moisture, temperature role, and respirometric index calculation method allow the understanding of the characteristics of the two methods in more detail.

Journal of Bioremediation & BiodegradationJo

urna

l of B

iorem

ediation & Biodegradation

ISSN: 2155-6199

Page 2: Bioremediation & Biodegradation - OMICS International · Rada et al., J Bioremed Biodeg 2012, 3:12f ... Some differences between the results of the measurements and obtainable

Citation: Rada EC, Ragazzi M, Venturi M (2012) Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi-Continuous Aeration. J Bioremed Biodeg 3:170. doi:10.4172/2155-6199.1000170

Volume 3 • Issue 12 • 1000170J Bioremed BiodegISSN: 2155-6199 JBRBD, an open access journal

Page 2 of 6

temperature, respirometric index calculation criteria were developed for understanding more in details the potential of the two methods.

Material and MethodsThe main characteristics of the two methods are described below.

The Costech respirometer is an adiabatic reactor with an available volume of 30 liters. This instrument has a flux regulator, a flow meter, an air humidification system, an oxygen meter at the inlet and at the outlet, temperature probes for internal and external measurements. An integrated system registers in continuous oxygen concentration,temperature and airflow.A system for leachate collection completesthe device in order to avoid troubles with condensation phenomena.

According to the method, 10-13 kg of the wet samples mustbe used for the tests [16]. Samples must be optimized in terms of moisture content (a standard of 75% of water content must be complied with). The obtained sample must be placed in the respirometer and continuously aerated using flow values able to guarantee an outlet O2 concentration higher than 14% (v/v). This value is compulsory for keepingaerobicconditions.Temperatureisnotkeptsteady.Anhourlydynamic respirometric index is assessed in mgO2 kg-1vs h-1 and comes from the air flow-rate, form the difference in the O 2 concentration in the inlet and outlet air flows and from the initial VS content.

TheAIR-nlrespirometerisanadiabaticreactorhavingoneprobefor temperature and one probe for oxygen measurements [14,15]. The reactor is kept at the chosen temperature by a thermostatic bath (30°C). The system of aeration and analysis is composed by a double pneumaticcircuit:inthefirstone,atlowflow,theoxygenconcentrationis measured in continuous, checking that its value is always higher than 18.5% (v/v); indeed the second one, at higher flow, guarantees the re-oxygenation by a continuous flow of air through the sample. This last circuit has two electro-valves that allow the control of O 2 concentration openingorclosingthecircuit.Anhourlydynamicrespirometricindexis assessed in mgO2 kg-1 vs h-1. Moisture in the sample is corrected to 55% before the test, and one kg of sample is used for the tests.

In order to make a comparison between the two instruments, some experimental data from the runs developed at the Trento University, regarding the oxygen consumption of compost samples were originally used in this paper. Two sets of data were used. The first one regarded the calculated values of RI24 and the reference average temperature of the tests. The second one consisted of detailed data concerning oxygen consumption for additional runs. Those data came from tests developed in parallel on the same substrate, with both the respirometric instruments working in parallel. The two equipments are presented in figure 1.

Data collected from the two dynamic respirometers regarding the oxygen consumption were disposed along a negative slope. The average gradient of each series of data linearly decreasing, referred to the average time, gave the value of the specific respirometric activity (mgO2 h

-1), that was referred to the volatile solids available in the system. The respirometric indexRIwasthencalculatedbythefollowingexpression(Equation1):

2 2

1 2

% % 1 132 1000O OP VRI

R T T VS t

⋅= ⋅ ⋅ ⋅ − ⋅ ⋅ ∆

(1)

Where,

P=pressure [atm] assumed equal to the atmospheric one;

V=free volume [liter];

T=temperature [K];

32=oxygen molecular weight [g];

VS=volatile solids [kg];

∆t =time interval [h];

R=0.0821 [l.atm.K-1.mol-1] constant of the perfect gas;

1 and 2=points between the ones the slope is calculated.

The respirometric index dynamics in the time was then represented by a continuous curve and was calculated through a mass balance using the slope.

Regarding data recorded with the Costech Respirometer, it was needed to report oxygen consumption to the reference temperature of30°C,inordertocomparethevalueswiththeonesfromtheAir-nlRespirometer. The literature proposes an expression (Equation 2) valid intherangebetween20and40°C[16]:

( )3030

C TC TRI RI ϑ ° −° = ⋅ (2)

Where,

T=instantaneous temperature at which the measure is done (with the temperature probe placed into the substrate)

ϑ=reference constant, equal to 1.08.

From the modified values, for each time step, the area under the respirometric curve was assessed and from that the sum until a stated instant. In practice, it was possible to assess the cumulative oxygen consumption referred to a fixed time.

After that, the RI24 calculation was made using the following expression(Equation3):

( 12 )

( 12 )24

b h

a h

RI dt

RIb a

+

=−

∫ (3)

where referring to a fixed time, a and b indicate respectively the initial time at -12 hours and the final time at +12 hours.

The RI24 values were obtained associating at each time the corresponding value of the integral average of RI in an interval of 24 hourscenteredontheconsideredtime.Accordingtothemethod,theRI value to be associated to each run is the highest of the RI24 values. The run can be stopped when the slope of the integral curve of RI24, determined on a period longer than 12 hours, is negative.

According to the AIR-nl method, the acceptability of the tests

Figure 1: Costech respirometric (left) and AIR-nl Respirometer (right).

Page 3: Bioremediation & Biodegradation - OMICS International · Rada et al., J Bioremed Biodeg 2012, 3:12f ... Some differences between the results of the measurements and obtainable

Citation: Rada EC, Ragazzi M, Venturi M (2012) Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi-Continuous Aeration. J Bioremed Biodeg 3:170. doi:10.4172/2155-6199.1000170

Volume 3 • Issue 12 • 1000170J Bioremed BiodegISSN: 2155-6199 JBRBD, an open access journal

Page 3 of 6

conducted with that respirometer depends on the behavior of the temperature value during the overall run; if temperature varies by morethan±2°C,themeasurementshouldberepeated.Indeed,AIR-nlrespirometer works at steady temperature.

For the comparison between the two devices, it was needed to modifydatatakenfromAIR-nltotheaveragetemperatureregisteredbyCostech, for the same analyzed sample. When the average temperature in a Costech test was higher than 40°C, the resulting respirometric value hadtobeconsiderednotacceptableforacomparisonaftertemperaturecorrection. The reason is related to Equation 2 that is suitable until 40°C.

The used samples were taken from different composting plants. For a better analysis of the stabilization, development sample ages were selected in the range of 7-45 days.

Additional data were generated from three samples analyzed inparallel recording the cumulative O2 consumption during the same period (66 hours).

ResultsThe available measurements and the calculation results are

presented in Table 1 (first series of data). It can be underlined that RI24 values measured with Costech are always lower than the ones from

AIR-nlaftertemperaturecorrection.ThisaspectclearlycontradictsthestatementthatdefinesthesystemAIR-nlasatoolthatisnotcompletelydynamic, involving “an underestimation of oxygen consumption” [17]. An important parameter is the moisture percentage. Some studiesshowed that the increased biological activity, underlined as maximum oxygen consumption, is found in correspondence with moisture values that can range from 50% to 70% [18]. From the moisture point of view, samples carriedoutwith theAIR-nlRespirometerhas theadvantageto be corrected to an optimal reference value of 55%, (as the method requires) providing an acceptability range of 2.5 percentage points [15]. The reference moisture value for the Costech method is equal to 75%, a value that cannot be considered optimal in this sense.

From data reported in table 1 that were considered acceptable, a graph about RI24,measuredwiththeAIR-nlandCostechrespirometers,wasobtainedandispresentedinfigure2.Asitcanbeseen,theobtainedcorrelation has a linear fitting, with a correlation coefficient equal to 0.9640. Taking into account that the same samples were used, the experimental values obtained with the Costech device were on average lower than those resultedwith theAIR-nl (indeed the slope resulted0.7).

For a deeper comparative analysis, results of additional tests performed in parallel with both respirometers are shown in table 2 and figure 3. It can be noticed that the linear relationship between the two

Sample Age

AIR-nl Costech

ΔT[°C]

Acceptability of the data:

TCostech>40°C

RI24

2O

VS

mg

kg h

Average temperature RID

[°C]

RI24 (reference T of Costech)

2O

VS

mg

kg h

RI24

2O

VS

mg

kg h

Average temperature RID

[°C]

1 21 1748 30.4 5254 2952 44.7 14.3 No2 21 2112 30.8 5923 2171 44.2 13.4 No3 21 1749 30.8 4830 2683 44.0 13.2 No4 7 2366 31.0 3697 2728 36.8 5.8 Yes5 21 1795 30.6 1851 1591 31.0 0.4 Yes6 14 1239 30.5 1490 1047 32.9 2.4 Yes7 7 2560 30.8 9693 3851 48.1 17.3 No8 21 1073 30.4 1185 909 31.7 1.3 Yes9 14 1253 30.4 2663 1480 40.2 9.8 No10 21 600 30.4 399 177 25.1 5.3 Yes11 14 1533 30.5 2184 1525 35.1 4.6 Yes12 21 682 30.4 482 281 25.9 4.5 Yes13 21 880 30.7 1125 944 33.9 3.2 Yes14 45 1294 30.3 2340 1236 38.0 7.7 Yes15 21 1110 29.7 2102 1341 38.0 8.3 Yes16 0 2754 31.1 5256 3741 39.5 8.4 yes17 0 2824 30.4 8423 3589 44.6 14.2 No18 7 1899 30.6 6406 3151 46.4 15.8 No19 21+3 1203 30.4 3296 1564 43.5 13.1 No20 21 1296 30.5 2734 1280 40.2 9.7 No21 14 1403 30.4 7062 3501 51.4 21.0 No22 14 1738 30.8 4800 2656 44.0 13.2 No23 21 934 30.4 1000 967 31.3 0.9 Yes24 21 898 32.2 1879 1083 41.8 9.6 No25 15 957 30.7 3433 1631 47.3 16.6 No26 16 2701 31.1 21742 5953 58.2 27.1 No27 - 2735 31.4 16433 3853 54.7 23.3 No28 - 1769 31.3 1194 790 26.2 5.1 Yes

Table 1: Data used in comparing the two respirometers and the results of described calculations (original data in bold; RID=respiration index (dynamic)).

Page 4: Bioremediation & Biodegradation - OMICS International · Rada et al., J Bioremed Biodeg 2012, 3:12f ... Some differences between the results of the measurements and obtainable

Citation: Rada EC, Ragazzi M, Venturi M (2012) Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi-Continuous Aeration. J Bioremed Biodeg 3:170. doi:10.4172/2155-6199.1000170

Volume 3 • Issue 12 • 1000170J Bioremed BiodegISSN: 2155-6199 JBRBD, an open access journal

Page 4 of 6

measurements is very high (there is a correlation coefficient of about 0.9940), and this analysis provides further support to the analytical method developed in a previous study [19]. Form figure 2 it can be noticed that there is no difference between the two respirometers concerning the ratio between cumulative oxygen and RI24.

The numerical interpolation relation showed a slope very similar to the one theoretically assessed from the organic matter composition in Venturi et al. [19]. The slope of the previous experimental work was 0.01432; the slope of the straight line for the present research shown in figure 4 is equal to 0.01425. Considerations about the intercept of the straight interpolation were made: this value resulted verysmall (especially when compared to typical respirometric values of

1,000÷3,000 2O

VS

mg

kg h⋅ and because of that it was chosen to bring it to

zero in order to obtain a new relation.

It can be seen that the coefficient of determination does not suffer a significant decrease reaching the value 0.9639. Using the derived relationship (Equation 4) and recalling that the proposed limit of acceptability for admission to landfills by the Costech method showed a value of 1,000 2O

VS

mg

kg h⋅, it is possible to obtain the reference value for the

AIR-nlrespirometerbyreversingtherelationship.Thecorrespondingvalue is equal to 1411.19 2O

VS

mg

kg h⋅.

24 240.7086Costech AIR nlRI RI −= ⋅ (4)

By combining the two relations (Equation 2 and Equation 4), the followingexpression(Equation5)wasobtained:

( ) ln ln0.7086_

_

T TCostech measured Costech AIR nl

Costech calculated

RIe

RIϑ − ⋅ −− = (5)

Where,

RICostech_measured is the respirometric measurement presented in table 1;

RICostech_calculated is obtained after the measurement with AIR-nlrespirometer, referring to Costech;

TCostech is the average temperature for the Costech respirometric runs;

TAIR-nlistheaveragetemperaturefortheAIR-nlrespirometricruns;

Taking into account the equation 5, in figure 5 values obtained from table 1 were reported. It can be seen that the indicated deviations are not correlated to the difference of average operating temperature of the two respirometers. Data reported in table 1 refer to the output of the software for the overall respirometric curve.Thus, the calculation ofRI24 with the same methodology for both instruments was not possible. Indeed, for the Costech respirometer the RI24 was calculated using the followingexpression(Equation6):

24

124

hh

RIDRID ==

∑ (6)

and for AIR-nl respirometer, the RI24 was calculated using the

followingexpression(Equation7):( 12 )

( 12 )24

b h

a h

RI dt

RIb a

+

=−

∫ (7)

where a and b indicate the initial time at -12 hours, respectively, the final time at +12 hours, with peak time reference.

In order to evaluate the magnitude error that was made, it was decided to use oxygen consumption for calculating the RI24 from

AIR-NL COSTECH

Sample

Reference time

Cumulated O2

RI24Reference

timeCumulated

O2RI24

[h] [mgO2]2O

VS

mg

kg h

⋅ [h] [mgO2] 2O

VS

mg

kg h

1 66 118919 2186 66 233259 3617.562 66 83644 1597 66 52304 1052.893 66 48388 1016 66 53185 1057.34

Table 2: Results of data on the tests performed in parallel with both respirometers.

Figure 2: Graphic correlation, without zero intercept, of the respirometric mea-surements performed with the Costech and the AIR-nl.

Figure 3: Correlation between cumulated oxygen and RI24 value.

Figure 4: Graphic correlation, with zero intercept, of respirometric measure-ments performed with Costech and AIR-nl.

Page 5: Bioremediation & Biodegradation - OMICS International · Rada et al., J Bioremed Biodeg 2012, 3:12f ... Some differences between the results of the measurements and obtainable

Citation: Rada EC, Ragazzi M, Venturi M (2012) Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi-Continuous Aeration. J Bioremed Biodeg 3:170. doi:10.4172/2155-6199.1000170

Volume 3 • Issue 12 • 1000170J Bioremed BiodegISSN: 2155-6199 JBRBD, an open access journal

Page 5 of 6

Costech measurements, with the methodology suggested by the expression used to calculate the RI24forAIR-nl(Equation7).

The expression (Equation 8) used to determine the error in percentagewas:

24 24[ ] [ ]

24 [ ]

% 100AIR nl Costech

AIR nl

RI RIE

RI−

−= ⋅ (8)

where 24 [ ]CostechRI and 24 [ ]AIR nlRI−

indicate the values obtained for

RI24 with the first and second expression considered by the two methodologies.

For the three data available for this elaboration (Table 2), the obtained values were equal to -0.0529%, -0.0038% and -0.5128%.They can be considered negligible.

In any case, the following analytical method was considered useful for the comparison of the two calculation methods. For this method, three data from the respirometric activity in consequential moments were considered (Figure 6a). Having to calculate, according to the reference methodology of AIR-nl, the integral of the respirometriccurve, b1, b2 and b3 will indicate the measured values (which correspond to the “bases” of trapezoids for the area calculation), h1 and h2 will indicate the time interval between the measurements (corresponding totheheightsof thetwotrapezoids)andA1andA2 the values of the two areas.

Figure 5: Lack of correlation between the ratio _

_

Costech measured

Costech calculated

RIRI

and the dif-

ference of average operation temperature for the two respirometers.

b1 b2 b3

A1 A2

h1 h2

b1 b2 b3

A1 A2

h1 h2

b4

h3

A3

a b

Figure 6: Reference scheme for the analytical calculation.

ThevalueofA1andA2areexpressedaspresentedinequation9:( )1 2 1

1 2b b h

A+ ⋅

= and ( )2 3 22

b b hA

2+ ⋅

= (9)

Thus, the RI24 value in the case of only three data may be calculated usingthefollowingexpression(Equation10):

( ) ( )( ) ( )

1 2 1 2 3 2 1 1 2 1 2 2 3 21 224 3

1 2 1 2 1 22 2b b h b b h b h b h b h b hA A

RIh h h h h h

+ ⋅ + + ⋅ ⋅ + ⋅ + ⋅ + ⋅+= = ==

+ ⋅ + ⋅ +

( )1 1 3 22

1 22 2b h b hb

h h⋅ + ⋅

= +⋅ + (10)

If there is an additional point of measurement (Figure 6b) it is possible to express the areaA3 in a similarway and in this case thecalculation of RI24 for four datacanbeexpressedas(Equation11):

1 2 324 4

1 2 3

A A ARI

h h h+ +

= =+ +

( ) ( ) ( )( )

1 2 1 2 3 2 4 3 3

1 2 32b b h b b h b b h

h h h+ ⋅ + + ⋅ + + ⋅

=⋅ + +

( ) ( )( )

1 2 2 2 3 3 1 1 4 3

1 2 32h h b h h b b h b h

h h h+ ⋅ + + ⋅ + ⋅ + ⋅

=⋅ + +

(11)

Atthispoint,itispossibletoevaluatetheRI24 expression (Equation 12)reportedtothen-thmeasure:

( )1

1 1 1 12

24 1

12

n

n n i i ii

nn

ii

b h b h b h hRI

h

− −=

=

⋅ + ⋅ + ⋅ +=

∑ (12)

In the case of measurements made with the Costech respirometer, there is a constant interval of time between a measurement and another, equal to 1 hour. If, for simplicity, the same hypothesis is taken also for theAIR-nl,wewillhave ih h 1 hour∆= = . The total time is 24 hours, resulting, therefore, in the presence of 25 measures.

Because of these simplified observations and assumptions, the expression changes, becoming simpler (a change of indices is done for a comparison with the method Costech):

( )

( )24

1 1 1 1 1 252 2 2

24

2 2 2

2 24 48 48

n n

n i n i ii i i

n AIR nl

b b h b h b b b b b bRI

h

+ += = =

+ ⋅∆ + ⋅ ∆ + + ⋅ + + ⋅= = =

⋅ ⋅∆

∑ ∑ ∑

(13)The characteristic relation for Costech (Equation 14) can be

expressedas:24

1 124 24 24

n

i ii i

Costech

b bRI = == =

∑ ∑ (14)

WhileforthereferencemethodofAIR-nl,24areaswith25measureswere used for the calculation (the extremities of analysis interval at -12 hours and +12 hours from the initial peak, are identified), for the method implemented by Costech a simple arithmetic average for 24 hours is done, considering 24 measures. For a comparison between the two methods of calculation, it is possible to calculate the difference betweenthederivedexpressions(Equation.15):

( )24 24 24n AIR nl CostechRI RI RI−

∆ = − =

1 1 1 1 12 1 2 2 1 1

2 2 2 2

48 24 48 48

n n n n

n i i n i ii i i i n

b b b b b b b b bb b+ +

= = = = ++ + ⋅ + + ⋅ − ⋅ − ⋅

−= − = =

∑ ∑ ∑ ∑

(15)The result can be expressed with a more usual notation (Equation

16):

Page 6: Bioremediation & Biodegradation - OMICS International · Rada et al., J Bioremed Biodeg 2012, 3:12f ... Some differences between the results of the measurements and obtainable

Citation: Rada EC, Ragazzi M, Venturi M (2012) Critical Analysis of Two Respirometric Methods for Solid Substrates Based On Continuous and Semi-Continuous Aeration. J Bioremed Biodeg 3:170. doi:10.4172/2155-6199.1000170

Volume 3 • Issue 12 • 1000170J Bioremed BiodegISSN: 2155-6199 JBRBD, an open access journal

Page 6 of 6

1 1 1 1 12 1 2 2 1 1

2 2 2 2

48 24 48 48

n n n n

n i i n i ii i i i n

b b b b b b b b bb b+ +

= = = = ++ + ⋅ + + ⋅ − ⋅ − ⋅

−= − = =

∑ ∑ ∑ ∑

(16)

where i refers to the time at which the respirometric peak is reached.

This expression (Equation 16) was applied to the data of the respirometric graphs of the analyzed oxygen consumption. In all the cases, the result was the same with the difference between the two RI24 assessed with the two calculation methods presented in this paper.

ConclusionsThe comparison between the two analyzed respirometers

(Costech andAIR-nl)pointedout that thedevelopmentof this kindof instrumentation must be coupled with a careful development of the methodology to be used. Indeed the critical analysis of this approach has demonstrated that some misunderstandings should be solved or corrected.

Forinstance,theAIR-nlrespirometerisconsideredinsomecirclesa system that underestimate the oxygen consumption. On the contrary, this paper demonstrated that the underestimation is a characteristic of the Costech respirometer.

Another aspect is related to the choice of the moisture ofreference (indeed moisture must be corrected before performing the respirometric test); the reference value proposed in the methodology coupled with Costech seems to be too high, resulting out of the range generally considered not limiting.

The role of temperature is another important parameter to be considered: the AIR-nl respirometer is based on the concept ofsteady temperature during the test (temperature is set according to the proposed method). In the approach proposed by Costech, the temperature is not regulated.

Asa consequenceof thesedifferences, it is important todevelopcorrelations recognized at international level in order to avoid a non optimized use of those devices which are more and more adopted in the waste and biomass sector.

References

1. Italian Council Directive (1999) 1999/31/EC of on the landfill of waste. Gazzetta Ufficiale L 182.

2. Italian National Standard Body (2007) CEN/TR 15590: Solid recovered fuels - Determination of potential rate of microbial self-heating using the real dynamic respiration index.

3. Rada EC, Andreottola G (2012) RDF/SRF: which perspective for its future in the EU. Waste Manag 32: 1059-1060.

4. Spanjers H, Vanrolleghem P, Olsson G, Dold P (1996) Respirometry in control of the activated sludge process. Water Sci Technol 34: 117-126.

5. Cossu R, Raga R (2008) Test methods for assessing the biological stability of biodegradable waste. Waste Manag 28: 381-388.

6. Gea T, Barrena R, Artola A, Sánchez A (2004) Monitoring the biological activity of the composting process: Oxygen uptake rate (OUR), respirometric index (RI), and respiratory quotient (RQ). Biotechnol Bioeng 88: 520-527.

7. Komilis DP, Tziouvaras IS (2009) A statistical analysis to assess the maturity and stability of six composts. Waste Manag 29: 1504-1513.

8. Polo AM, Tobajas M, Sanchis S, Mohedano AF, Rodríguez JJ (2011) Comparison of experimental methods for determination of toxicity and biodegradability of xenobiotic compounds. Biodegradation 22: 751-761.

9. Italian National Standard Body (1998) UNI 10780: Compost – Classification, requirements and modality of use.

10. Barrena Gómez R, Vázquez Lima F, Gordillo Bolasell MA, Gea T, Sánchez Ferrer A (2005) Respirometric assays at fixed and process temperatures to monitor composting process. Bioresour Technol 96: 1153-1159.

11. de Guardia A, Mallard P, Teglia C, Marin A, Le Pape C, et al. (2010) Comparison of five organic wastes regarding their behaviour during composting: part 1, biodegradability, stabilization kinetics and temperature rise. Waste Manag 30: 402-414.

12. Italian National Standard Body (2006) UNI/TS 11184: Waste and recovered fuel from waste -Determination of biological stability through Dynamic Respirometric Index.

13. Adani F, Lozzi P, Genevini P (2001) Determination of biological stability by oxygen uptake on municipal solid waste and derived products. Compost Sci Util 9: 163-178.

14. Andreottola G, Dallago L, Ragazzi M (2001) Rating Index RESPIROMETRY of solid matrices using respirometry open and closed. RS SOLID WASTE 15: 81-87.

15. Andreottola G, Dallago L, Ragazzi M (2005) Dynamic respirometric tests for assessing the biological activity of waste, Proceedings of X International Waste Management and Landfill Symposium. Santa Margherita di Pula (Cagliari), Sardinia, Italy.

16. Scaglia B, Acutis M, Adani F (2011) Precision determination for the dynamic respirometric index (DRI) method used for biological stability evaluation on municipal solid waste and derived products. Waste Manag 31: 2-9.

17. Adani F, Cementero M, Favoino E (2002) Il pretrattamento biologico per il conferimento del rifiuto in discarica – Parte 2: definizione e determinazione della stabilità biologica. Rifiuti Solidi 1: 20-28.

18. Jeris JS, Regan RW (1973) Controlling environmental parameters for optimum composting. Compost Science 14: 16-22.

19. Venturi M, Rada EC, Ragazzi M, Silvestri S (2009) Experimental analysis on different measurement systems breathability of organic matrix. RS SOLID WASTE 23: 162-168.