6
Biomechanical Characteristics of an Implant Used to Secure Semitendinosus–Gracilis Tendon Grafts in a Canine Model of Extra-Articular Anterior Cruciate Ligament Reconstruction MANDI J. LOPEZ, DVM, MS, PhD, Diplomate ACVS, NAKIA SPENCER, BS, JOHN P. CASEY, and WILLIAM TODD MONROE, PhD Objective—To compare initial femoral fixation properties of a new implant, Graftgrab (GG), with 2 established methods of extra-articular femoral graft fixation, spiked washers (SW) and bone staples (ST). Study Design—Experimental in vitro cohort study. Methods—Canine semitendinosus–gracilis tendon grafts were passed through bone tunnels and fixed to the lateral surface of femoral condyles with spiked washers, bone staples, or new implant prototypes. The fixations were tested to failure with a single-cycle load at a rate of 50 mm/min. Failure and yield strength, stiffness, energy, and elongation were determined from load–displace- ment curves and failure modes were recorded. Results—The graft failed midsubstance in 4 SW, 4 ST, and 1 GG fixations. In 3 SW, 3 ST, and 1 GG specimens, the graft slipped from the fixation. The graft ruptured at the clip (3) and the intra- articular (2) surface of the bone tunnel in the remaining GG specimens. There were no significant differences between fixation groups in femoral tunnel length, femoral width, or the mechanical properties evaluated. Conclusions—The initial in vitro mechanical properties of the new fixation implant are comparable with those of spiked washers and bone staples. Clinical Relevance—The initial mechanical performance of the new implant tested in this study was similar to those of comparable, established implants. The new implant is novel and may be useful for human anterior and veterinary cranial cruciate ligament graft reconstruction fixation. r Copyright 2007 by The American College of Veterinary Surgeons INTRODUCTION T HE BONE-PATELLAR bone and semitendinosus– gracilis grafts are 2 of the most widely used anterior cruciate ligament (ACL) reconstruction grafts in human patients. 1,2 Various implants are available for graft fix- ation to the lateral aspect of the femoral condyle. Endo- Button (Acufex Microsurgical Inc., Mansfield, MA), spiked washers, and bone staples are 3 implants used to secure ACL reconstruction grafts to bone around the femoral bone tunnel. 2 Optimal fixation of an ACL graft requires sufficient initial strength to avoid fixation failure, sufficient stiffness to restore knee stability, anatomic fix- ation to minimize graft motion within the tunnel, and resistance against slippage to avoid loosening in the im- mediate postoperative period. 3 Graft elongation behavior during implantation con- tributes to anteroposterior knee translation and is detri- This study was funded in part by the American College of Veterinary Surgeons Research Foundation and National Institutes of Health grants K01AR002174-04 and K01AR002174-06. Address reprint requests to Dr. Mandi J. Lopez, DVM, MS, PhD, Diplomate ACVS, Department of Veterinary Clinical Sciences, Louisiana State University, Baton Rouge, LA 70803-8410. E-mail: [email protected]. Submitted January 2007; Accepted May 2007 From the Laboratory for Equine and Comparative Orthopedic Research, School of Veterinary Medicine, and the Department of Biological and Agricultural Engineering, Louisiana State University and AgCenter, Baton Rouge, LA. r Copyright 2007 by The American College of Veterinary Surgeons 0161-3499/07 doi:10.1111/j.1532-950X.2007.00310.x 599 Veterinary Surgery 36:599–604, 2007

Biomechanical Characteristics of an Implant Used to Secure Semitendinosus–Gracilis Tendon Grafts in a Canine Model of Extra-Articular Anterior Cruciate Ligament Reconstruction

Embed Size (px)

Citation preview

Biomechanical Characteristics of an Implant Used to Secure

Semitendinosus–Gracilis Tendon Grafts in a Canine Model of

Extra-Articular Anterior Cruciate Ligament Reconstruction

MANDI J. LOPEZ, DVM, MS, PhD, Diplomate ACVS, NAKIA SPENCER, BS, JOHN P. CASEY, andWILLIAM TODD MONROE, PhD

Objective—To compare initial femoral fixation properties of a new implant, Graftgrab (GG), with 2established methods of extra-articular femoral graft fixation, spiked washers (SW) and bone staples(ST).Study Design—Experimental in vitro cohort study.Methods—Canine semitendinosus–gracilis tendon grafts were passed through bone tunnels andfixed to the lateral surface of femoral condyles with spiked washers, bone staples, or new implantprototypes. The fixations were tested to failure with a single-cycle load at a rate of 50mm/min.Failure and yield strength, stiffness, energy, and elongation were determined from load–displace-ment curves and failure modes were recorded.Results—The graft failed midsubstance in 4 SW, 4 ST, and 1 GG fixations. In 3 SW, 3 ST, and 1GG specimens, the graft slipped from the fixation. The graft ruptured at the clip (3) and the intra-articular (2) surface of the bone tunnel in the remaining GG specimens. There were no significantdifferences between fixation groups in femoral tunnel length, femoral width, or the mechanicalproperties evaluated.Conclusions—The initial in vitro mechanical properties of the new fixation implant are comparablewith those of spiked washers and bone staples.Clinical Relevance—The initial mechanical performance of the new implant tested in this study wassimilar to those of comparable, established implants. The new implant is novel and may be usefulfor human anterior and veterinary cranial cruciate ligament graft reconstruction fixation.r Copyright 2007 by The American College of Veterinary Surgeons

INTRODUCTION

THE BONE-PATELLAR bone and semitendinosus–gracilis grafts are 2 of the most widely used anterior

cruciate ligament (ACL) reconstruction grafts in humanpatients.1,2 Various implants are available for graft fix-ation to the lateral aspect of the femoral condyle. Endo-Button (Acufex Microsurgical Inc., Mansfield, MA),spiked washers, and bone staples are 3 implants used to

secure ACL reconstruction grafts to bone around thefemoral bone tunnel.2 Optimal fixation of an ACL graftrequires sufficient initial strength to avoid fixation failure,sufficient stiffness to restore knee stability, anatomic fix-ation to minimize graft motion within the tunnel, andresistance against slippage to avoid loosening in the im-mediate postoperative period.3

Graft elongation behavior during implantation con-tributes to anteroposterior knee translation and is detri-

This study was funded in part by the American College of Veterinary Surgeons Research Foundation and National Institutes of

Health grants K01AR002174-04 and K01AR002174-06.

Address reprint requests to Dr. Mandi J. Lopez, DVM, MS, PhD, Diplomate ACVS, Department of Veterinary Clinical Sciences,

Louisiana State University, Baton Rouge, LA 70803-8410. E-mail: [email protected].

Submitted January 2007; Accepted May 2007

From the Laboratory for Equine and Comparative Orthopedic Research, School of Veterinary Medicine, and the Department of

Biological and Agricultural Engineering, Louisiana State University and AgCenter, Baton Rouge, LA.

r Copyright 2007 by The American College of Veterinary Surgeons

0161-3499/07

doi:10.1111/j.1532-950X.2007.00310.x

599

Veterinary Surgery

36:599–604, 2007

mental to long-term knee stability and graft tensile prop-erties.4 Loss of initial graft tension is inevitable when us-ing current fixation methods and contributes to problemsassociated with joint instability.5,6 Loss of initial intra-articular graft tension occurs during application of spikedwashers and bone staples for tibial graft fixation5 buteven greater loss occurs during application of femoralfixation implants, which allow more initial lengtheningthan the tibial implants.3,7 Tension loss during implantfixation varies widely even with careful surgical tech-nique.5 An implant that allows graft fixation with min-imal further action after graft tensioning may counteractsome of the factors that contribute to tension loss.

We reasoned that an implant seated in the extra-articular side of the femoral bone tunnel that had a sim-ple mechanism for anchoring the graft immediately aftertensioning would minimize the likelihood of graft tensionloss. Accordingly, we developed an implant that allowsgraft fixation in a single motion after graft tensioning(Fig 1). To evaluate the fixation characteristics of thisimplant we compared it against 2 established methods ofextra-articular graft fixation: spiked washers and bonestaples. Our intent was to determine if initial fixationstrength of this implant was comparable with that ofestablished implants before further testing and designrefinement.

MATERIALS AND METHODS

Implant Description

The implant named Graftgrab (GG) has a stainless-steelbase (diameter, 13.5mm; maximum height, 2.7mm) with acentral hole (4.5mm diameter; Fig 1). A 2.7-mm-long sleeve(5.5mm outer diameter, 4.5mm inner diameter) and taperedspikes (2mm long) extend from the medial surface of the im-plant into the femoral tunnel and bone, respectively. There is achannel from the center to the perimeter of the implant toposition the graft for attachment once it is pulled through thehole. An acetal copolymer clip with a hinge on one side and aclasp on the other crosses this channel to fit securely intonotches in the implant base. The clip has a serrated centralsurface with 2 central serrations that protrude further into thechannel and 2 shorter peripheral serrations that secure thegraft without impeding clip closure. The geometry of the clipwas designed to hold grafts of uniform size and facilitateclasping into the base.

Constructs

Hind limbs of 11 hounds (� 2 years old; mean � SEMweight, 25.6 � 0.65 kg) euthanatized for unrelated reasonswere collected immediately after death and frozen at �201C.Specimens were thawed on the day of testing and ran-domly assigned to 1 of 3 fixation groups (n¼ 7/fixation): SW,screw and spiked washer; ST, bone staples; GG, Graftgrab(Fig 1).

Stifles were disarticulated and grafts were prepared fromthe combined tendinous insertion of the gracilis and semite-ndinosus muscles and cranial tibialis fascia on the medial as-pect of each tibia.8 Briefly, the tendons were dissected free oftheir muscular origins maintaining the combined bony inser-tion. The insertion was elevated and the attached dense con-nective tissue and muscle fascia of the cranial tibialis musclewere dissected free of bony and muscular attachments to alength of � 7 cm. Tissue was trimmed to a width of 1.5 cmand twisted from the distal end to a final graft diameter of� 4.5mm and a length of 7 cm. A Chinese finger trap of #1polyglactin 910 was placed around the length of each graft,9

and the combined bony insertion was transected. Grafts werepassed through a 4.4mm diameter stainless-steel tube to en-sure diameter consistency.

All soft tissue was removed from the femurs which weresectioned 6 cm proximal to the distal articular surface andsagittally equidistant between trochlear ridges. An aimingdevice was used to guide a 4.5mm drill bit from the intra-articular origin of the cranial cruciate ligament (CrCL) to thecenter of the lateral aspect of each femoral condyle at an angleof � 451 to the long axis of the femur. The femoral width justproximal to the condyle and the tunnel length were measuredwith a Vernier caliper (McMaster-Carr, Chicago, IL) anddepth gauge (Synthes, Monument, CO), respectively. Con-structs were assembled by 1 investigator. Grafts were pulledthrough the tunnel to a length of 1.5 cm beyond the lateralfemoral surface in all groups (Fig 2).

Fig 1. Schematic of a prototype fixation implant (Graftgrab;

GG). The stainless-steel base of the new implant has a round,

tapered lateral surface with a central hole (H). A sleeve (S) and

tapered spikes (black arrows) extend from the inner surface of

the implant. There is a channel (C) from the center to the

perimeter of the implant to position the graft for attachment

once it is pulled through the sleeve. An acetal copolymer clip

with a hinge (Hi) on one side and a clasp (Cl) on the other

crosses the channel to fit securely into notches in the implant

base (black arrow heads). The clip has a serrated central sur-

face with 2 central prongs (gray arrow heads) that protrude

further into the channel and 2 shorter peripheral prongs (gray

arrows) that secure the graft without impeding clip closure.

600 BIOMECHANICAL CHARACTERISTICS OF A PROTOTYPE IMPLANT FOR LIGAMENT GRAFTS

SW

A 3.2mm pilot hole was drilled 5mm caudal to the bonetunnel in the SW group. The graft was secured to the femur(Fig 3) with a 13mm diameter stainless-steel spiked bonewasher (IMEX Veterinary Inc., Longview, TX) and 4.5mmdiameter, 22mm long, self-tapping cortical screw (Synthes).The graft was wrapped around the screw which was insertedthrough the washer and into the pilot hole. The washer washeld firmly against the graft as the screw was tightened.

ST

The graft was secured to the femur just proximal and cra-nial to the lateral fabella using a belt–buckle technique withtwo 7 � 7mm bone staples (Fig 4) from a powered metaphy-seal stapler (Stapilizer; 3M, St. Paul, MN).8 A staple wasplaced over the graft � 5mm caudal to the bone tunnel, thegraft was folded over the staple in an cranial direction, and

another staple was placed over the graft between the bonetunnel and the first staple.

GG

The base of the implant was tapped into place with anosteotomy mallet. After the graft was passed through thefemoral tunnel and implant sleeve, the hinge side of the clipwas placed in the base, and the graft was secured by closingthe clip over it (Fig 5).

Biomechanical Testing

Femoral sections were mounted on a testing fixture at-tached to the load cell of a materials test system (Model 319.10

Fig 2. Tendinous reconstruction graft (black arrow) extend-

ing from the femoral bone tunnel. The cranial aspect of the

lateral condyle is to the right (C).

Fig 3. Tendinous reconstruction graft secured to the lateral

aspect of the femoral condyle with a spiked washer and screw.

The cranial aspect of the lateral condyle is to the right (C).

Fig 4. Tendinous reconstruction graft secured to the lateral

aspect of the femoral condyle with 2 bone staples. A staple

(arrow head) was placed over the graft caudal to the bone

tunnel. The graft was folded over the staple in a cranial di-

rection, and another staple (arrow) was placed over the graft

between the bone tunnel and the first staple. The cranial aspect

of the lateral condyle is to the right (C).

Fig 5. Tendinous reconstruction graft secured to the lateral

aspect of the femoral condyle with a prototype implant (Graft-

grab; GG). The cranial aspect of the lateral condyle is to the

right (C).

601LOPEZ ET AL

A/T system, MTS Systems Corporation, Eden Prairie, MN).The intra-articular surface of each specimen faced the direc-tion of the actuator, and the long axis of the bone tunnel wasaligned with the direction of load application. All specimenswere moistened with saline (0.9% NaCl) solution spray duringtesting. The free ends of the grafts were gripped with a spe-cially designed cryoclamp (Fig 6). The clamp was designedwith a dry ice reservoir to freeze the gripped portion of thegrafts. There was a 25mm length of graft between the clampand the intra-articular edge of the bone tunnel. A 1N preloadwas applied, and specimens were evaluated with a load tofailure at 50mm/min after 10 preconditioning cycles of 15%strain at 0.5Hz. Load and displacement data sampled at 60Hzfrom theMTS were used to generate load displacement curves.

Yield and failure load, strain energy, elongation, and thestiffness of each implant/graft construct were determinedgraphically. Failure load was the load at the point of graftrupture. Yield load was the load where the load displacementcurve first deviates from a straight line, indicating plastic de-formation. Yield and failure strain energy were defined as thetotal area under the curve at the yield and failure points, re-spectively. Stiffness was calculated as the change in load overthe change in displacement in the elastic portion of the load–displacement curves. Yield and failure elongation were thegrip to grip displacements measured at yield and failure, re-spectively. Each test was recorded with a digital camcorder,and failure mode was confirmed after testing.

Data Analysis

Comparison of mechanical properties, tunnel length, andcondylar width between groups was determined by 1-wayANOVA (GraphPad Prism v4.0, GraphPad Software Inc.,San Diego, CA). Significance was set at Po.05. When com-parisons were not significant, the difference (d) betweengroups necessary to detect a significant difference was calcu-lated.10 Results are reported as mean � SEM.

RESULTS

Grafts failed midsubstance in 4 SW, 4 ST, and 1 GGfixations. In 3 SW, 3 ST, and 1 GG specimens, the graftslipped from the fixation. The graft ruptured at the clip(3) and the intra-articular (2) surface of the bone tunnelin the remaining GG specimens. There were no signifi-cant differences between fixation groups in femoral tun-nel length, femoral width, or the mechanical propertiesevaluated (Table 1).

DISCUSSION

Implant Design

We designed this implant to achieve graft fixation tothe femur in a single motion after tensioning the graft.

Fig 6. Femoral section mounted on a testing fixture attached

to the load cell of a materials test system (white arrow). The

intra-articular surface of the specimen faces the direction of the

actuator, and the long axis of the bone tunnel is aligned with

the direction of load application. The free end of the graft

(black arrow) is gripped with a cryoclamp.

Table 1. Mean ( � SEM)Mechanical Properties of 3 Implants (Spiked Washer and Screw [SW]; Bone Staple [ST]; Prototype Implant [Graftgrab, GG]) for

Fixation of Semitendinosus–Gracilis Muscle Tendons Grafts to the Lateral Aspect of the Canine Femur for Cranial Cruciate Ligament Reconstruction

Variable SW ST GG P d (%)

Failure load (N) 110.1 � 9.3 123.8 � 9.1 137.9 � 20.3 0.39 30

Failure elongation (mm) 23.2 � 2.6 25.6 � 2.7 28.9 � 3.1 0.36 21

Yield load (N) 82.8 � 7.2 105.4 � 3.1 94.5 � 18.4 0.40 35

Yield elongation (mm) 15.7 � 2.1 20.0 � 2.2 17.0 � 2.1 0.37 34

Stiffness (N/mm) 8.2 � 1.6 7.8 � 1.0 7.4 � 0.8 0.90 41

Yield strain energy (J) 0.55 � 0.13 0.87 � 0.13 0.76 � 0.25 0.46 65

Failure strain energy (J) 1.38 � 0.36 1.72 � 0.33 1.95 � 0.38 0.53 60

Femoral condyle width (mm) 24.3 � 1.1 24.0 � 0.61 26.9 � 1.1 0.07 10

Femoral tunnel length (mm) 19.7 � 0.07 19.8 � 0.03 19.8 � 0.06 0.25 0.71

Fixation was tested in single cycle uniaxial distraction (50mm/min) to failure.

d¼ differences between populations necessary for significance at power¼ 0.80.

602 BIOMECHANICAL CHARACTERISTICS OF A PROTOTYPE IMPLANT FOR LIGAMENT GRAFTS

The GG implant has a low profile and is free of sharpsurfaces that may potentially cause graft damage. De-signed to facilitate implant alignment, the central sleevecounteracts translational forces preventing implant mo-tion and should also prevent acute or cyclic damage tothe graft at the bone tunnel edge. The spikes extendingfrom the implant base are for rotational stability. Theposition of the implant against the bone is maintained byinherent graft tension, so bone necrosis because of pres-sure is unlikely.

The implant can be created in any size, and multipleclip sizes and configurations can be fabricated to facilitatestable fixation without unnecessary graft compression.Insertion and use of the implant does not require specialequipment and it can be used with arthrotomy orarthroscopy. Incorporation of bioabsorbable compo-nents or a comparable bioabsorbable implant should bepossible.

Bony incorporation of a soft-tissue graft in the bonetunnel is required for survival of the tissue.11 Graftincorporation, healing quality, and strength of the graft–bone interface are augmented by optimizing the fitbetween the graft and the bone tunnel and maximizingthe length of graft within the tunnel.12 The sleeve thatextends from the GG base into the bone tunnel isdesigned to serve a protective and stabilizing function. Inthese femoral specimens, the sleeve covered 12–13% ofthe length of bone tunnel. It is likely that the sleevewill inhibit graft incorporation in that location, so thispotential effect will need to be investigated and the designmodified accordingly.

Study Design Considerations

Canine CrCL grafts were harvested and prepared byestablished methods.8 Because initial graft strength is notaffected by rotation, graft rotation may be used toaddress graft–tunnel mismatch.13 Rotating the free end ofthe graft increases its diameter to achieve a better fitwithin the bone tunnel. We used graft rotation to achievea final diameter of � 4.5mm (confirmed by use of a4.4mm stainless-steel sizing tube) to achieve graft sizeconsistency between groups. We minimized variation intunnel length and femoral condylar width between con-structs (Table 1); however, there was significant variationin mechanical properties between constructs, reducing theoverall power of the study. Wide standard deviations arenot uncommon for investigations of this nature.2

Differences in failure mode may have contributed tolarge variations in failure load and strain energy observedfor GG constructs. It is also possible that clip design wasa factor when failure occurred at that level. Despitethis, mean values for failure, yield, elongation, and strainenergies were relatively similar between groups. Whereas

it is difficult to compare values between studies, the initialmechanical properties of the implant/graft constructswere similar to comparable canine soft tissue fixationstudies,14–16 supporting our observations.

Graft Fixation Considerations

Graft elongation after implantation is one of the crit-ical factors that affect long-term outcome of knee sta-bility. Loss of initial graft tension is seeminglyunavoidable using current graft fixation techniques,4–6

and may result from inability to tension the graft opti-mally because of bone tunnel friction, loss of tensionduring insertion of the device used to secure the graft tothe femur, and subsequent slippage at the fixation site.5

Graft slippage at the fixation site occurred in all 3 con-struct types but less often with GG. The sleeve compo-nent of GG may have caused increased bone tunnelfriction compared with the other 2 fixations. The com-parison techniques (spiked washer and screw insertion,staple insertion) require substantially more manipulationthan clip closure so seemingly there should be less like-lihood of slippage when securing the graft with the clip.However, a comparative assessment of initial graft ten-sion loss is needed to definitively conclude that GGwould have this specific advantage.

Fixation is the weakest point of graft reconstruction inthe immediate postoperative period, so knowledge of fix-ation properties is essential for any implant designed forthis purpose.3 Further, the importance of implant/graftconstruct stiffness has been well established, and a highstiffness construct requires substantially less initial tensionto achieve the same joint stability as a low-stiffness con-struct.6 We attempted to control those variables that caninfluence stiffness and found no differences between con-struct types. For this initial evaluation we compared themechanical properties of these graft–femoral fixationtechniques by single cycle uniaxial distraction to failure.More comprehensive evaluation will require assessment ofcyclic loading behavior and in situ mechanical properties.

We found that the initial mechanical performance of anew bone implant designed to secure a tendinous or lig-amentous graft to the canine femur was similar to com-parable, established implants (spiked washer and screw;staples). Further evaluation is needed to determine thefatigue and graft tensioning characteristics of this proto-type implant, which may be a useful for femoral CrCL(or ACL) graft reconstruction fixation after additionaloptimization.

ACKNOWLEDGMENTS

The authors thank Jason Guy for technical expertise and

the Center for Bio-Modular Multi-Scale Systems for access

to fabrication equipment.

603LOPEZ ET AL

REFERENCES

1. Beynnon BD, Meriam CM, Ryder SH, et al: The effect of

screw insertion torque on tendons fixed with spiked wash-

ers. Am J Sports Med 26:536–539, 1998

2. Jagodzinski M, Behfar V, Hurschler C, et al: Femoral press-

fit fixation of the hamstring tendons for anterior cruciate

ligament reconstruction. Am J Sports Med 32:1723–1730,

2004

3. Kousa P, Jarvinen TL, Vihavainen M, et al: The fixation

strength of six hamstring tendon graft fixation devices in

anterior cruciate ligament reconstruction. Part I: femoral

site. Am J Sports Med 31:174–181, 2003

4. Tohyama H, Beynnon BD, Johnson RJ, et al: The effect of

anterior cruciate ligament graft elongation at the time of

implantation on the biomechanical behavior of the graft

and knee. Am J Sports Med 24:608–614, 1996

5. Grover DM, Howell SM, Hull ML: Early tension loss in an

anterior cruciate ligament graft. A cadaver study of four

tibial fixation devices. J Bone Jt Surg Am 87:381–390, 2005

6. Thompson DM, Hull ML, Howell SM: Does a tensioning

device pinned to the tibia improve knee anterior–posterior

load–displacement compared to manual tensioning of the

graft following anterior cruciate ligament reconstruction?

A cadaveric study of two tibial fixation devices. J Orthop

Res 24:1832–1841, 2006

7. To JT, Howell SM, Hull ML: Contributions of femoral fix-

ation methods to the stiffness of anterior cruciate ligament

replacements at implantation. Arthroscopy 15:379–387,

1999

8. Lopez MJ, Markel MD, Kalscheur V, et al: Hamstring graft

technique for stabilization of canine cranial cruciate liga-

ment deficient stifles. Vet Surg 32:390–401, 2003

9. Smeak DD: The Chinese finger trap suture technique for

fastening tubes and catheters. J Am Anim Hosp 26:215–

218, 1990

10. Markel MD: The power of a statistical test. What does in-

significance mean? Vet Surg 20:209–214, 1991

11. Weiler A, Hoffmann RF, Bail HJ, et al: Tendon healing in a

bone tunnel. Part II: histologic analysis after biodegradable

interference fit fixation in a model of anterior cruciate lig-

ament reconstruction in sheep. Arthroscopy 18:124–135,

2002

12. Greis PE, Burks RT, Bachus K, et al: The influence of tendon

length and fit on the strength of a tendon–bone tunnel

complex. A biomechanical and histologic study in the dog.

Am J Sports Med 29:493–497, 2001

13. Verma N, Noerdlinger MA, Hallab N, et al: Effects of graft

rotation on initial biomechanical failure characteristics of

bone–patellar tendon–bone constructs. Am J Sports Med

31:708–713, 2003

14. Cabaud HE, Rodkey WG, Feagin JA: Experimental studies

of acute anterior cruciate ligament injury and repair. Am J

Sports Med 7:18–22, 1979

15. Butler DL, Hulse DA, Kay MD, et al: Biomechanics of cra-

nial cruciate ligament reconstruction in the dog. Vet Surg

12:113–118, 1983

16. Markel MD, Rock MG, Bergenthal DS, et al: A mechanical

comparison of gluteus medius attachment methods in a

canine model. J Orthop Res 11:457–461, 1993

604 BIOMECHANICAL CHARACTERISTICS OF A PROTOTYPE IMPLANT FOR LIGAMENT GRAFTS