30
Chapter 74: Biopotentials and Electrophysiology Measurement Teemu Rämö [email protected] m butler.cc.tut.fi/~malmivuo/bem/bembook/

Bio Potentials

Embed Size (px)

DESCRIPTION

Bio Potentials

Citation preview

  • Chapter 74: Biopotentials and Electrophysiology Measurement

    Teemu [email protected]

    butler.cc.tut.fi/~malmivuo/bem/bembook/

  • Agenda

    1st half Introduction to biopotentials Measurement methods

    Traditional: ECG, EEG, EMG, EOG Novell: VCG

    2nd half Measurement considerations

    Electronics Electrodes Practices

    Q&A

  • What are biopotentials

    Biopotential: An electric potential that is measured between points in living cells, tissues, and organisms, and which accompanies all biochemical processes.

    Also describes the transfer of information between and within cells This book focuses strictly on the measurement of potentials

  • Mechanism behind biopotentials 1/2

    Concentration of potassium (K+) ions is 30-50 times higher inside as compared to outside

    Sodium ion (Na+) concentration is 10 times higher outside the membrane than inside

    In resting state the member is permeable only for potassium ions

    Potassium flows outwards leaving an equal number of negative ions inside

    Electrostatic attraction pulls potassium and chloride ions close to the membrane

    Electric field directed inward formsElectrostatic force vs. diffusional force

    Nernst equation:

    Goldman-Hodgkin-Katz equation:

    mVVm 100...70

    ko

    ki

    kk c

    cFz

    RTV,

    ,ln

    CliClNaiNaKiK

    CliClNaiNaKiK

    km cPcPcP

    cPcPcPFz

    RTV,,,

    ,,,ln

    mVVm 100...70

  • Mechanism behind biopotentials 2/2 When membrane stimulation exceeds a threshold

    level of about 20 mV, so called action potential occurs:1. Sodium and potassium ionic permeabilities of the

    membrane change2. Sodium ion permeability increases very rapidly at first,

    allowing sodium ions to flow from outside to inside, making the inside more positive

    3. The more slowly increasing potassium ion permeability allows potassium ions to flow from inside to outside, thus returning membrane potential to its resting value

    4. While at rest, the Na-K pump restores the ion concentrations to their original values

    The number of ions flowing through an open channel >106/sec

    Body is an inhomogeneous volume conductor and these ion fluxes create measurable potentials on body surface

  • Electrocardiography (ECG)

    Measures galvanically the electric activity of the heart Well known and traditional, first measurements by

    Augustus Waller using capillary electrometer (year 1887) Very widely used method in clinical environment Very high diagnostic value

    1. Atrial depolarization

    2. Ventriculardepolarization

    3. Ventricular repolarization

  • ECG basics

    Amplitude: 1-5 mV Bandwidth: 0.05-100 Hz

    Largest measurement error sources: Motion artifacts 50/60 Hz powerline interference

    Typical applications: Diagnosis of ischemia Arrhythmia Conduction defects

  • 12-Lead ECG measurement

    Most widely used ECG measurement setup in clinical environment Signal is measured non-invasively with 9 electrodes Lots of measurement data and international reference databases Well-known measurement and diagnosis practices This particular method was adopted due to historical reasons, now it is already rather

    obsolete

    Einthoven leads: I, II & III Goldberger augmented leads: VR, VL & VF Precordial leads: V1-V6

  • Why is 12-lead system obsolete?

    Over 90% of the hearts electric activity can be explained with a dipole source modelOnly 3 orthogonal components need to be measured,

    which makes 9 of the leads redundant

    The remaining percentage, i.e. nondipolar components, may have some clinical valueThis makes 8 truly independent and 4 redundant leads

    12-lead system does, to some extend, enhance pattern recognition and gives the clinician a few more projections to choose frombut.

    If there was no legacy problem with current systems, 12-lead system wouldve been discarded ages ago

  • Electroencephalography (EEG)

    Measures the brains electric activity from the scalp

    Measured signal results from the activity of billions of neurons

    Amplitude: 0.001-0.01 mV Bandwidth: 0.5-40 Hz

    Errors: Thermal RF noise 50/60 Hz power lines Blink artifacts and similar

    Typical applications: Sleep studies Seizure detection Cortical mapping

  • EEG measurement setup 10-20 Lead system is most

    widely clinically accepted Certain physiological features

    are used as reference points Allow localization of diagnostic

    features in the vicinity of the electrode

    Often a readily available wire or rubber mesh is used

    Brain research utilizes even 256 or 512 channel EEG hats

  • Electromyography (EMG)

    Measures the electric activity of active muscle fibers Electrodes are always connected very close to the muscle

    group being measured Rectified and integrated EMG signal gives rough indication

    of the muscle activity Needle electrodes can be used to measure individual muscle fibers

    Amplitude: 1-10 mV Bandwidth: 20-2000 Hz

    Main sources of errors are 50/60 Hz and RF interference

    Applications: muscle function, neuromuscular disease, prosthesis

  • Electrooculography (EOG)

    Electric potentials are created as a result of the movement of the eyeballs Potential varies in proportion to the amplitude of the movement In many ways a challenging measurement with some clinical value

    Amplitude: 0.01-0.1 mV Bandwidth: DC-10 Hz

    Primary sources of error include skin potential and motion

    Applications: eye position, sleep state, vestibulo-ocular reflex

  • Vectorcardiogram (VCG or EVCG)

    Instead of displaying the scalar amplitude (ECG curve) the electric activation front is measured and displayed as a vector (dipole model, remember?) It has amplitude and direction

    Diagnosis is based on the curve that the point of this vector draws in 2 or 3 dimensions

    The information content of the VCG signal is roughly the same as 12-lead ECG system. The advantage comes from the way how this information is displayed

    A normal, scalar ECG curve can be formed from this vectro representation, although (for practical reasons) transformation can be quite complicated

    Plenty of different types of VCG systems are in use No legacy problem as such

  • Short break,Kahvia ja pullaa!

  • The biopotential amplifier

    Small amplitudes, low frequencies, environmental and biological sources of interference etc.

    Essential requirements for measurement equipment: High amplification

    High differential gain, low common mode gain high CMRR

    High input impedance Low Noise Stability against temperature and voltage fluctuations Electrical safety, isolation and defibrillation protection

  • The Instrumentation Amplifier

    Potentially combines the best features desirable for biopotential measurements High differential gain, low common mode gain, high CMRR, high input resistance

    A key design component to almost all biopotential measurements! Simple and cheap, although high-quality OpAmps with high CMRR should be used

    1

    21 21 R

    RG 3

    42 R

    RG

    CMRR fine tuning

  • Application-specific requirements

    ECG amplifier Lower corner frequency 0.05 Hz, upper 100Hz Safety and protection: leakage current below safety standard limit of 10 uA Electrical isolation from the power line and the earth ground Protection against high defibrillation voltages

    EEG amplifier Gain must deal with microvolt or lower levels of signals Components must have low thermal and electronic noise @ the front end Otherwise similar to ECG

    EMG amplifier Slightly enhanced amplifier BW suffices Post-processing circuits are almost always needed (e.g. rectifier + integrator)

    EOG amplifier High gain with very good low frequency (or even DC) response DC-drifting electrodes should be selected with great care Often active DC or drift cancellation or correction circuit may be necessary

  • Electrical Interference Reduction

    Power line interference (50 or 60 Hz) is always around us Connects capacitively and causes common mode interference The common mode interference would be completely rejected by the instrumentation

    amplifier if the matching would be ideal Often a clever driven right leg circuit is used to further enhance CMRR

    Average of the VCM is inverted and driven back to the body via reference electrode

    0RiV DCM

    1

    2

    0

    21RR

    RiV DCM

  • Filtering Filtering should be included in the front end of the InstrAmp Transmitters, motors etc. cause also RF interference

    Small inductorsor ferrite beadsin the lead wires

    block HF frequencyEM interference

    RF filtering withsmall capacitors

    High-pass filterto reject DC drifting

    Low-pass filteringat several stages

    is recommended toattenuate residualRF interference

  • 50 or 60 Hz notch filter

    Sometimes it may be desirable to remove the power line interference Overlaps with the measurement bandwidth

    May distort the measurement result and have an affect on the diagnosis!

    Option often available with EEG & EOG measuring instruments

    Twin Tnotch filter

    Determinesnotch

    frequency

    Notchtuning

  • Artifact reduction

    Electrode-skin interface is a major source of artifact Changes in the junction potential causes slow changes in the baseline Movement artifacts cause more sudden changes and artifacts

    Drifting in the baseline can be detected by discharging the high-pass capacitor in the amplifier to restore the baseline

  • Electrical isolation

    Electrical isolation limits the possibility of passage of any leakage current from the instrument in use to the patient

    Such passage would be harmful if not fatal!

    1. Transformer Transformers are inherently high frequency

    AC devices Modulation and demodulation needed

    2. Optical isolation Optical signal is modulated in proportion to

    the electric signal and transmitted to the detector

    Typically pulse code modulated to circumvent the inherent nonlinearity of the LED-phototransistor combination

  • Defibrillation Protection

    Measuring instruments can encounter very high voltages E.g. 15005000V shocks from defibrillator Front-end must be designed to withstand these high voltages

    1. Resistors in the inputleads limit the current

    3. Protection against much higher voltages

    is achieved withlow-pressure gasdischarge tubes

    (e.g. neon lamps)

    (note: even isolationcomponents such as

    transformers andoptical isolators need

    these spark gaps)Discharge @ ~100V

    2. Diodes or Zener diodesprotect against high

    voltages

    Discharge @ 0.7-15V

  • Electrodes Basics

    High-quality biopotential measurements require Good amplifier design Use of good electrodes and their proper placement on the patient Good laboratory and clinical practices

    Electrodes should be chosen according to the application Basic electrode structure includes:

    The body and casing Electrode made of high-conductivity material Wire connector Cavity or similar for electrolytic gel Adhesive rim

    The complexity of electrode design often neglected

  • Electrodes - Basics

    Skin preparation by abrasion or cleansing Placement close to the source being measured Placement above bony structures where there is less muscle mass Distinguishing features of different electrodes:

    How secure? The structure and the use of strong but less irritant adhesives How conductive? Use of noble metals vs. cheaper materials How prone to artifact? Use of low-junction-potential materials such as Ag-AgCl If electrolytic gel is used, how is it applied? High conductivity gels can help reduce the junction

    potentials and resistance but tend to be more allergenic or irritating

    Baseline drift due to thechanges in junction

    potential or motion artifactsChoice of electrodes Muscle signal

    interference Placement

    Electromagneticinterference Shielding

  • Ag-AgCl, Silver-Silver Chloride Electrodes

    The most commonly used electrode type Silver is interfaced with its salt silver-chloride Choice of materials helps to reduce junction potentials

    Junction potentials are the result of the dissimilar electrolytic interfaces

    Electrolytic gel enhances conductivity and also reduces junction potentials

    Typically based on sodium or potassium chloride, concentration in the order of 0.1 M weak enough to not irritate the skin

    The gel is typically soaked into a foam pad or applied directly in a pocket produced by electrode housing

    Relatively low-cost and general purpose electrode Particularly suited for ambulatory or long term use

  • Gold Electrodes Very high conductivity suitable for low-noise meas. Inertness suitable for reusable electrodes Body forms cavity which is filled with electrolytic gel Compared to Ag-AgCL: greater expense, higher

    junction potentials and motion artifacts Often used in EEG, sometimes in EMG

    Conductive polymer electrodes

    Made out of material that is simultaneously conductive and adhesive Polymer is made conductive by adding monovalent metallic ions Aluminum foil allows contact to external instrumentation No need for gel or other adhesive substance High resistivity makes unsuitable for low-noise meas. Not as good connection as with traditional electrodes

  • Metal or carbon electrodes

    Other metals are seldom used as high-quality noblemetal electrodes or low-cost carbon or polymericelectrodes are so readily available

    Historical value. Bulky and awkward to use Carbon electrodes have high resistivity and are noisier

    but they are also flexibleand reusable Applications in electrical stimulation and impedance plethysmography

    Needle electrodes Obviously invasive electrodes Used when measurements have to be taken from the organ itself Small signals such as motor unit potentials can be measured Needle is often a steel wire with hooked tip

  • Thats it,Now for Q&A

    SQUID = Superconducting Quantum Interference Device