25
10/27/14 Atomic Force Microscope Vecco Dimension 3100 (supported by Bruker) Location White Bldg. Room 613

Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14

Atomic Force Microscope Vecco Dimension 3100

(supported by Bruker)

Location White Bldg.

Room 613

Page 2: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Atomic Force Microscope

AFM Control unit and software Atomic Force Microscope

Page 3: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Atomic Force Microscope Components

(S)canning (P)robe (M)icroscope head Stage holder and (2) types of tips 1. Tapping 2. Contact 3. Magnetic (not shown)

Page 4: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14

Atomic Force Microscope Tips

Tapping tip on holder.

Zoom in on tip.

www.brukerafmprobes.com

Typical Silicon Tips 5-20 nm tip radii 20-100 N/m stiffness

Page 5: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Atomic Force Microscope Tips

Contact tip on holder.

Zoom in on tip.

www.brukerafmprobes.com

Important to match the stiffness of the tip to the sample being measured.

Page 6: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Atomic Force Microscope Simplistic Operation Click to play

Page 7: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Example of AFM measurements – Tapping Mode

Nanoindentation – Optical image

Nitinol material

Page 8: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 AFM scan - 30 um x 30 um

Nitinol material – Limitations of scan 100 um x 4.8 um deep

Page 9: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 AFM scan - 30 um x 30 um – isometric

Nitinol material

Page 10: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 AFM Geometry Analysis

Nitinol material

Page 11: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 AFM Geometry Analysis – Vertical Distance

Nitinol material

Page 12: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 AFM Geometry Analysis – Rmax

Nitinol material

Page 13: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 AFM Geometry Analysis – Depth of indent

Nitinol material

Page 14: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

9/9/14 AFM Geometry Analysis – Angle of indent

Nitinol material

Page 15: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Picking the right AFM Tip – Size matters

Click to play

Page 16: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14

Understanding contamination and its effects Click to play

• In ambient air, surface are covered with contamination • A probe encounters contamination when approaching a surface • Capillary forces pull the contamination up onto the probe

Contact AFM – Force Microscopy

Page 17: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Using contact mode with Force modulation

Consider a simple case of Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

The X & Y of the scanner is fixed. It moves in Z as shown by Extension.

Rubber sample material

A

Van der Waals forces can be seen as the tip approaches the surface (A).

Page 18: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

These attractive forces pull the tip down.

Rubber sample material

B Point of contact (B).

A

Page 19: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

Extension continues and the cantilever bends upward.

Rubber sample material

B

C

Until (C), where the extension stops and so does the deflection.

Point of contact (B).

Page 20: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

C – As the scanner retracts, the deflection decreases (C to D) – red curve. The cantilever relaxes.

Rubber sample material

D

C

There may be surface attraction between the tip and sample which keeps the tip in contact with the surface. A monolayer of water can also provide this attraction.

Page 21: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

E – As this force is overcome, the tip snaps quickly back as shown from by the vertical trajectory from D to E.

Rubber sample material

D

E

Page 22: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14 Tip / Sample interaction

-700

-600

-500

-400

-300

-200

-100

0

100

200

300

0 500 1000 1500 2000 2500

Def

lect

ion

(nm

)

Extension (nm)

Extend

Retract

Force Curve Summary

Rubber sample material

D

E

A

B

C

Page 23: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the Bruker Dimension 3100 atomic force microscope. Two scans are made over the same area.

The first scan determines the topology of the surface.

The second scan is done at a fixed height (200 nm) above the surface. Any changes in the phase of resonance of the tip are due to the magnetic field of the specimen.

10/27/14

Page 24: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

MFM Cross-Sectional Image of a Carburized Duplex 2205 Stainless Steel

surface

g 20 mm

Magnetic Ferrite

Nonmagnetic Austenite: the carbon at the surface prevents polishing-induced martensite

Austenite with some polishing-induced Surface Martensite (which is magnetic)

10/27/14

Page 25: Atomic Force Microscope Vecco Dimension 3100 · 2016-12-27 · Magnetic Force Microscopy Magnetic force microscopy (MFM) scans are performed using a resonating Co-coated tip on the

10/27/14

Thank you