Upload
gerard
View
46
Download
1
Tags:
Embed Size (px)
DESCRIPTION
An Introduction to Quality Assurance in Analytical Science. Dr Irene Mueller-Harvey Mr Richard Baker Mr Brian Woodget. Part 2 - The Analytical Method. Contents: VAM principles (slides 3&4) Fit for purpose (slide 5) Choice of method (slide 6) Method development and validation (slide 7) - PowerPoint PPT Presentation
Citation preview
An Introduction to Quality Assurance in Analytical
ScienceDr Irene Mueller-Harvey
Mr Richard BakerMr Brian Woodget
University of Reading
Part 2 - The Analytical Method
Contents:• VAM principles (slides 3&4)• Fit for purpose (slide 5)• Choice of method (slide 6)• Method development and validation (slide 7)• Method performance characteristics (slides 8 - 15)• Method validation (slides 16 - 25)
The presentation contains some animation which will be activated automatically (no more than a 2 second delay), by mouse click or by use
of the ‘page down’ key on your keyboard.
VAM Principles (1)
The DTI’s programme on Valid Analytical Measurement (VAM) is an integral part of the UK National Measurement System. The aim of the VAM programme is to help analytical laboratories to demonstrate the validity of their data and to facilitate mutual recognition of the results of analytical measurements. You can find out more about VAM by logging onto the web site [http://www.vam.org.uk/]
The VAM Bulletin is publishedhalf yearly and is sent free to all registered subscribers. Youmay subscribe via the website.
VAM Principles (2) There are six VAM principles:
Analytical measurements should be made to satisfy an agreed requirement
Analytical measurements should be madeusing methods and equipment which have been
tested to ensure they are fit for their purpose
Staff making analytical measurements should be both qualified and competent
to undertake the task
There should be a regular independent assessment of the technical performance of the laboratory
Analytical measurements made in one location should be
consistent with those elsewhere
Organisations making analytical measurementsshould have well definedquality control and quality
assurance procedures
Fit for PurposeYou have seen in part 1 of this presentation that it can be expensiveto generate good quality analytical data. The cost of producing datacan often be reduced by selecting analytical methods and technologiesthat produce data in accordance with the stated objectives forcarrying out the analysis or test. Consider two examples:
A commercial limescale remover states that it contains between 5 to15% w/v of sulphamic acid. Therefore any analysis for quality control
purposes needs only to show that the quantity present is indeed between these two limits and does not need to be accurate to the nearest + 0.1%
The EU current statutory limit for lead in potable water is 25 μg/l. Thus theuse of a method to perform the analysis, which only provides a ‘less than’
value of 100 μg/l, is not suitable for this purpose
Choice of MethodIt is important to appreciate the difference between an ‘analytical method’ (combination of steps illustrated by the ‘analytical process’) and an ‘analytical technique’ (chemical or instrumental procedure by which analyticaldata is eventually obtained). In selecting a method we shall need to consider the following parameters:
• sample type (matrix) and size (lot or a little);• data required (qualitative/quantitative);• expected level(s) of analyte(s);• precision & accuracy expected;• likely interferences;• number and frequency of samples for analysis.
Always use a standard method ifone is available as
this will save ondevelopment time.
However the methodmust be checked to prove that it suitable
for yourlaboratory/situation.
Modification may well be required.
Method developmentand validation‘Never attempt to re-invent the wheel!’
Before embarking on the development of a new method, always research the chemical literature to see if a suitable one already exists. If a suitable one is found, it will still be necessary however to perform some method validation to prove that the method can be successfully adapted to your laboratory, equipment and personnel. More extensive validation is required for a brand new method. Methods in any field of analysis may be defined in terms ‘Method performance characteristics’ and it is these parameters plus a few others, that are quantified during a method validation exercise.
Method performance characteristics
A method’s performance is defined by a number of importantindividual characteristics. There are:
Sensitivity Precision
Accuracy Limit of Detection (LoD)
Limit of Determination Bias
Selectivity Linear Range
Dynamic range
Accuracy and precision
The dictionary definition of both ‘accuracy’ and ‘precision’ are roughly the same, indicating that these words may be used synonymously. However in ‘Analytical Science’ they have two separate meanings, the difference between them is best illustrated by using target diagrams
Poor precisionpoor accuracy
Good precisionpoor accuracy
Good mean accuracypoor precision
Good accuracygood precision
Accuracy and precision (2)You saw from the previous slide, a set of results can beeither accurate and/or precise or can be neither accuratenor precise. Thus accuracy may be defined as:
The closeness of the mean value from a replicate set of results to the true or accepted value
Precision may be defined as:
The spread of results from a replicate set of measurements
The difference between the true value and the mean measured value is termed bias. The spread of replicate data ismeasured in terms of standard deviation (s) or variance (s2)
Random and systematic errors
There are 2 types of error for which allowance may be made: Random error Systematic error
Random error arises from variationsin parameters which are outside the
control of the analyst, but which influence the value of the
measurement being made. Becausethese errors are statistically random,
the mean error should be zero if sufficient measurements are made.
Systematic error remains constant or may vary in a predictable way over a
series of measurements and cannot bereduced by making replicate
measurements. In theory, if known, this error can be allowed for.
Eg: subtraction of blank values
mean mean
Bias and variance
A solution containing copper was analysed 10 times using atomic
absorption spectroscopy.The results obtained in ppm were:10.08, 9.80, 10.10, 10.21, 10.14,9.88, 10.02, 10.12, 10.11, 10.09
We can now calculate the precision of the data as standard deviation
If the true value is known to be 10.00 ppm, we can also calculate
the bias
Cu by AAS
9.69.810
10.210.4
0 2 4 6 8 10 12
Replicate sample
Cu in
ppm
Bias = Mean value - true value = 10.06 - 10.00 = 0.06 ppm
Standard deviation (SD) = 0.12(4)
Conclusion - the method gives both good accuracy (low bias) andacceptable precision (RSD of 1.2%)
Relative SD = 100 X SD/10.00 = 1.2%
Sensitivity and selectivity Assessment of method sensitivity
0
200
400
600
800
1000
1200
0 10 20 30
Concentration
Sign
al
Sensitivity is the change in measuredsignal for unit change in concentrationand can be obtained from the calibration graph
Sensitivity = dy/dx
dy
dx
Selectivity is the ability of a methodto discriminate between the targetanalyte and other constituents of thesample. In many instances selectivityis achieved by high performanceseparation using chromatographic orelectrophoretic techniques.
Hplc chromatogram
Limits of detection (LoD) and determination These values refer to the statistical limits below which results of detection or accurate quantitative measurements (determination) should not be reported. The levels of both are dependent upon the variability of the signal when a blank containing none of the analyte is being measured. The signal generated under these conditions is mostly signal noise and is assumed to exhibit a normal distribution pattern. Both the blank signal and the standard deviation of the blank signal need to be measured. From this data we can calculate both limits.
Example: In an analysis of trace Cd by plasma emission spectrometry the following datawere obtained:• mean blank (Bl) signal 4• SD of blank signal 12• 500 ppb Cd 2000
LoD = Bl signal + 3(SD of Bl) = 4 + 3(12) = 40This equates to: [40/2000] X 500 ppb
= 10 ppb CdThe limit of determination uses a similar formula, replacing the 3 SD’s by 10. This gives the limit ofdetermination as 31 ppb Cd
Linear and Dynamic rangesThese terms refer to the extent to which the method maybe used to produce accurate quantitative data
From the graph, it would appear that the data is linear to about 25 ppm and dynamic until about 75 ppm. After 75 ppm there is only minimal increase in signal for increased concentration.
Linear & dynamic ranges
00.10.20.30.40.50.60.70.80.9
0 20 40 60
Concentration/ppm
Sign
al
Expanding thelower section ofthe graph however,shows that non-linearity starts ataround 20 ppm.
Extent of dynamic range
Top of linearrange
Graph showing linear & dynamic ranges
0
0.2
0.4
0.6
0.8
1
0 50 100 150 200
Concentration/ppm
Sign
al
Method validation Validation, is the proof needed to ensure that an analytical method can produce results which are reliable and reproducibleand which are fit for the purpose intended. The parameters that need to be demonstrated are those associated with the ‘Performance characteristics’ together with robustness, repeatability and reproducibility.
Many analytical methods appearing in the literature have not beenthrough a thorough validation exercise and thus should be treated withcaution until full validation has been carried out. Validation of a new
method (new to your laboratory), is a costly and time-consuming exercise,however the result of not carrying out method validation could result in
litigation, failure to get product approval, costly repeat analysis andloss of business and prestige.
You can now consider in more detail how validation iscarried out
Method validation -linearity
LinearityMost analytical methods are of a comparative type and thus require
calibration against accurately known standards to generate quantitative data. Where possible calibration data should show a linear relationship
between analyte concentration and measured signal, however it is acceptable under some circumstances, to use a non-linear relationship
up to the limit of the dynamic range.
Calibration graph
00.1
0.20.30.4
0.50.6
0 5 10 15 20 25
Concentration
Sign
al
Check linearity between50 - 150% of the expectedanalyte concentration
Method validation -specificityLoss of specificity can be due to interferences and matrix
effects.All likely interferences should be investigated and their effects on analyte
response determined over a range of concentrations. Measures can then beput into place to mask, eliminate or separate them from the analyte.
Standard addition procedures can be usedto identify matrix effects
no yes
Method validation -precision
You have seen already that,precision is measured in terms of standard deviation (SD). Assuming that the variability of the measurements is totally random (obeys a normal distribution curve) then a formula derived from this distribution may be usedto calculate standard deviation.
SD = [ Σ(xi - x)2/(n - 1)]1/2
where:xi = individual data pointx = mean value of the datan = the total number of data pointsΣ = the sum of
Normal distribution curve
0
20
40
60
80
100
0 50 100 150
Data points
Freq
uenc
y of
oc
curr
ence
Estimation of true mean
In practice around 8 - 10 data points are used normally to calculate the SD, although statisticians would recommend 50
Method Validation -repeatability & reproducibility
Methods need to be shown to be both repeatable and reproducible. A replicate set of data produced at a particulartime point by an operator working with a particular set of equipment in a given laboratory will verify repeatability. To show reproducibility, the method must produce similar results when any of these parameters are changed. The most likely changes are to time and operator.
Two different operatorsanalysing milk
using different pieces ofequipment at different
times. The laboratory isthe same.
Method validation -reliabilityThe reliability of a method can be tested in a number of ways
Test results from the new method against an existingmethod which is known to
be accurate
Add a known quantity of pure analyte(spike) to a real sample or real samplematrix and check that all of the added
substance can be measured (recovered)
The best way of demonstratingaccuracy is to analyse a reference
material or certified reference material (CRM) if one is available[see part 3 of this presentation]
Selection of reference materials
from LGC
Method validation -detection & quantitation limits
A method is not acceptable for accurate detection orquantitation if the analyte level is likely be fall beneaththe limit(s) calculated based upon the blank signal andits standard deviation (Refer to the formula given in slide 14,in this part of the presentation). Analyte pre-concentration then becomes necessary.
Variation in blank signal
Mean blanksignal
Variation insample signal
Mean samplesignal
Mean sample signalmust be sufficiently
larger than the blankso that positive
detection or accuratequantitation is
possible
Method validation -robustnessRobustness of an analytical method refers to it’s abilityto remain unaffected when subjected to small changesin method parameters.
For exampleIn an hplc analysis the mobile phase is defined in terms of % organic
modifier, pH of the mobile phase, buffer composition, temperatureetc. A perfect mobile phase is one which allows small changes in the
composition without affecting the selectivity or the quantitation of the method.
Alter all major parameters in order to ascertain when the method ceasesto function in accordance with specifications
Method validation -establish stability
In routine analysis where numerous samples and standardsare measured each day, it is essential to assess the stability of the prepared solutions. Stability of these solutions should be tested by repeat analysis over at least a 48 hour period.
Signal
Time
Apparent onset ofsolution instability
Method validation -
additional readingAn article entitled “A Practical Guide to Analytical Method Validation” was published in Analytical Chemistry in 1996 [ Anal. Chem. (68) 305A-309A]
The article may be downloaded free from the acs web site:
http://pubs.acs.org/hotartcl/ac/96/may/may.html