2
AIF 503S1 Flicker Impedance 3-phase 32A for Harmonics & Flicker testing AIF 503S1 is a three-phase flicker impedance designed for testing Harmonics and Flicker. The specifications of the two flicker impedances Zref and Ztest are as per EN / IEC 61000-3-3 / -11. The AIF 503 offers space to accommodate the 3-phase Harmonic and Flicker analyser DPA 503. Specification EN/IEC 61000-3-3 EN/IEC 61000-3-11 IEC 60725 emc test equipment

AIF 503S1 Flicker Impedance - EM Test · AIF 503S1 Flicker Impedance ... The AIF 503 offers space to accommodate the 3-phase Harmonic and ... figuration (Bypass for

  • Upload
    vuongtu

  • View
    272

  • Download
    1

Embed Size (px)

Citation preview

Page 1: AIF 503S1 Flicker Impedance - EM Test · AIF 503S1 Flicker Impedance ... The AIF 503 offers space to accommodate the 3-phase Harmonic and ... figuration (Bypass for

AIF 503S1 Flicker Impedance 3-phase 32A

for Harmonics & Flicker testing

AIF 503S1 is a three-phase flicker impedance designed for testing

Harmonics and Flicker.

The specifications of the two flicker impedances Zref and Ztest are as

per EN / IEC 61000-3-3 / -11.

The AIF 503 offers space to accommodate the 3-phase Harmonic and

Flicker analyser DPA 503.

S p e c i f i c a t i o n

• EN/IEC 61000-3-3

• EN/IEC 61000-3-11

• IEC 60725

e m c t e s t e q u i p m e n t

Page 2: AIF 503S1 Flicker Impedance - EM Test · AIF 503S1 Flicker Impedance ... The AIF 503 offers space to accommodate the 3-phase Harmonic and ... figuration (Bypass for

Specifications

EM TEST AG Tel: +41 (0)61 717 91 91Sternenhofstr. 15 Fax: +41 (0)61 717 91 99 data sheet AIF 503S1 V200.docCH-4153 Reinach email: [email protected] 10.12.2007Switzerland URL http://www.emtest.com Page 2/2

The Flicker Impedance AIF 503 S1 is designed for Flicker and Harmonic test according IEC 61000-3-3 and IEC 61000-3-11 with built in Zrefand Ztest.

Reference Impedance ZrefZref is a conventional impedance as specified in IEC 61000-3-3 with the impedance values in accordance with IEC 60725.

Test Impedance ZtestThe test impedance Ztest is lower than Zref and used for 3-phase or single phase equipment intended to be connected to public low-voltagedistribution systems having a service current supply with a capacity ≥ 100A.

Block diagram

The AIF 503S1 is controlled from the DPA 503.

The DPA 503 measures the voltage and current for harmonic andflicker analyzing. The flicker impedance Zref / Ztest is bypassed for theharmonic measurement.

The system is operated from the computer with the windows softwareISMDPA.

FeaturesLED’s indicated the used system con-figuration (Bypass for harmonics, or theused flicker impedance Zref or Ztest).

Dual size 80A banana plug connectorwith 32A for 4mm banana connector.

Rear side view AIF 503S1

Technical dataPhase 3 – phaseLine Voltage 3x 400V ± 10%Line Currrent 32A per phaseLine Frequency 50 / 60Hz Flicker impedance 2 ( Zref & Ztest )

Bypass / Flicker change Switch remote controlledOutput connector CEE connector 5 pole 32A

Safety banana plugInterface 15 pin Sub D from DPA 503

ImpedanceZref at 50Hz RA = 0.24Ω XA = 0.15Ω

RN = 0.16Ω XN = 0.10ΩZtest at 50Hz RA = 0.15Ω XA = 0.15Ω

RN = 0.10Ω XN = 0.10ΩAccuracy better than 3%

Each inductor is designed as a non-saturable air coil and is matched manuallyto the specified value.

Ambient conditionsTemperature operating 0° to 35°Humidity 80% non condensingCooling 6 ventilators controlled

DimensionsHousing 19”, 25HUWeight 100 kgAIF 503 S1 1245 x 553 x 780Position Transportable with castors

Power mains supplySupply voltage 230V or 115V L/N/PE Power 200WLine frequency 50/ 60HzFuse 10A

Technical data subject to change without notice