 Automated External Defibrillators  Defibrillation  Cardioversion  Pacing Electrical Therapies

  • View
    215

  • Download
    1

Embed Size (px)

Transcript

  • Slide 1
  • Automated External Defibrillators Defibrillation Cardioversion Pacing Electrical Therapies
  • Slide 2
  • Early Defibrillation (1) the most frequent initial rhythm in witnessed SCA is VF, (2) the treatment for VF is electrical defibrillation, (3) the probability of successful defibrillation diminishes rapidly over time (4) VF tends to deteriorate to asystole within a few minutes.
  • Slide 3
  • Slide 4
  • Slide 5
  • Slide 6
  • Sudden Cardiac Arrest The survival rate from CPR alone is 0-2% CPR will buy you time, it will not stop a VF Electrical Therapies
  • Slide 7
  • New Recommendations to Integrate CPR and AED Use Two critical questions Whether CPR should be provided before defibrillation is attempted ? The number of shocks to be delivered in a sequence before the rescuer resumes CPR?
  • Slide 8
  • Shock First Versus CPR First Witnessed: an out-of-hospital arrest and an AED is immediately available on-site, the rescuer should use the AED as soon as possible Not witnessed : EMS personnel, they may give about 5 cycles of CPR before checking the ECG rhythm and attempting defibrillation (Class IIb)
  • Slide 9
  • Shock First Versus CPR First HCP witnessed cardiac arrest in hospitals and with AEDs on-site should provide immediate CPR and should use the AED/defibrillator as soon as it is available EMS call-to-arrival intervals were 4 to 5 minutes or longer, victims who received 1 to 3 minutes of CPR before defibrillation
  • Slide 10
  • 1-Shock Protocol Versus 3- Shock Sequence In 2 studies of out-of-hospital and in-hospital CPR by HCP, chest compressions were performed only 51% to 76% of total CPR time 3-shock sequence performed by AEDs resulted in delays of up to 37 seconds the first-shock efficacy of >90% reported by current biphasic defibrillators
  • Slide 11
  • 1-Shock AED Rescuers using monophasic AEDs should give an initial shock of 360 J; if VF persists after the first shock, second and subsequent shocks of 360 J should be given But it is not a mandate to recall monophasic AEDs for reprogramming
  • Slide 12
  • Defibrillation Defibrillation (shock success) is typically defined as termination of VF for at least 5 seconds following the shock. VF frequently recurs after successful shocks, but this recurrence should not be equated with shock failure.
  • Slide 13
  • Defibrillator Monophasic Damped Sinusoidal Wave (MDS) Monophasic truncated Exponential Wave (MTE) Biphasic Truncated Exponential (BTE) Rectilinear Biphasic (RBW)
  • Slide 14
  • Slide 15
  • Damped Sine Wave Unchanged for 30 Years Requires high energy and current. Not highly effective for patients with high transthoracic impedance.
  • Slide 16
  • Biphasic Truncated Exponential The First Generation: Adapted from low impedance ICD applications. Impedance causes waveform to change shape.
  • Slide 17
  • Rectilinear Biphasic Waveform Designed Specifically for External Use: Constant CurrentConstant Current eliminates high peaks Fixed DurationFixed Duration stabilizes waveform in face of varying impedance levels.
  • Slide 18
  • Effect of Patient Impedance on Biphasic Waveforms Rectilinear Biphasic High Impedance First Generation Biphasic Low Impedance -20 0 10 20 30 40 50 04812 -10 -20 0 10 20 30 40 50 04 8 12 -10 -20 0 10 20 30 40 50 04812 -10 -20 0 10 20 30 40 50 04 8 12 -10
  • Slide 19
  • Biphasic Defibrillation with biphasic waveforms of relatively low energy ( 200 J) is safe and has equivalent or higher efficacy for termination of VF than monophasic waveform shocks of equivalent or higher energy (Class IIa). None of the available evidence has shown superiority of either nonescalating or escalating energy biphasic waveform defibrillation for termination of VF
  • Slide 20
  • Biphasic Current research confirms that it is reasonable to use selected energies of 150 J to 200 J with a biphasic truncated exponential waveform or 120 J with a rectilinear biphasic waveform for the initial shock For second and subsequent biphasic shocks, use the same or higher energy (Class IIa).
  • Slide 21
  • AED Survival rate of 41% to 74% from out-of-hospital witnessed VF SCA when immediate bystander CPR is provided and defibrillation occurs within about 3 to 5 minutes of collapse Reviewers found no studies that documented the effectiveness of home AED deployment, so there is no recommendation for or against personal or home deployment of AEDs (Class Indeterminate).
  • Slide 22
  • AED Implantable medical device Transdermal medication patch Lying in water Hairy chest
  • Slide 23
  • AED in Children Biphasic shocks appear to be at least as effective as monophasic shocks and less harmful Dose: 2 J/kg for the first attempt and 4 J/kg for subsequent attempts (Class Indeterminate). If CPR to a child in cardiac arrest and does not have an AED with a pediatric attenuator system, the rescuer should use a standard AED Not used for infants
  • Slide 24
  • In-Hospital Use of AEDs AEDs should be considered for the hospital setting as a way to facilitate early defibrillation (a goal of 3 minutes from collapse)
  • Slide 25
  • Biphasic defibrillator Manufacturers should display the device-specific effective waveform dose range on the face of the device, and providers should use that dose range when attempting defibrillation with that device
  • Slide 26
  • Biphasic defibrillator BTE: 150 J to 200 J for initial shock RBW: 120 J for initial shock For second and subsequent shocks, use the same or higher energy (Class IIa). If unaware dose range, 200 J for the first shock and an equal or higher dose for the second and subsequent shocks
  • Slide 27
  • Synchronized Cardioversion shock delivery that is timed (synchronized) with the QRS complex avoid shock delivery during the relative refractory portion of the cardiac cycle, when a shock could produce VF If impossible to synchronize a shock (eg, the patient s rhythm is irregular), use high-energy unsynchronized shocks.
  • Slide 28
  • Synchronized Cardioversion Recommended : SVT due to reentry, Af, and atrial flutter Not recommended: (automatic focus)MAT. Atrial Tach, Sinus Tachycardia
  • Slide 29
  • Synchronized Cardioversion Af : 100 J to 200 J (Monophasic) AF and other SVT: 50J to 100 J (Monophasic) Cardioversion with biphasic waveforms is now available: 100J to 120J(Biphasic) Monomorphic VT: 100 J, 200 J, 300 J, 360 J). Unstable polymorphic (irregular) VT with or without pulses is treated as VF using unsynchronized high-energy shocks (ie, defibrillation doses).
  • Slide 30
  • Pacing Not recommended: asystolic cardiac arrest Considered: symptomatic bradycardia Immediate pacing is indicated if the patient is severely symptomatic, especially when the block is at or below the His Purkinje level If the patient does not respond to transcutaneous pacing Transvenous
  • Slide 31